18 November 2013 Role of contrast and fractality of laser speckle image in assessing flow velocity and scatterer concentration in phantom body fluids
Author Affiliations +
Abstract
Blood flow velocity and red blood cell concentration are of vital importance in assessing tissue microcirculation. Laser speckle contrast analysis is being considered as a promising tool in the qualitative assessment of flow velocity as well as scatterer concentration in different body fluids, though the quantification part still remains challenging. The fractal-based spatial correlation analysis of speckle flow images along with the corresponding contrast analysis for the quantitative assessment of flow and scatterer concentration is investigated. In this study, phantom body fluid solution (intralipid 20%) of different concentrations is pumped at different flow rates through the designed flow channel using a syringe pump and the corresponding speckle images are acquired. The fractality of the acquired speckle images in response to the changes in concentration of the fluid as well as the variations in fluid flow is analyzed along with the corresponding contrast-based analysis. Following this qualitative analysis, an experimental model is attempted toward quantification of these parameters from a single acquired speckle image by considering the contrast and fractality changes together.

1.

Introduction

Blood supplies nutrients and oxygen to all the parts of our body and also plays a major role in thermal regulation. Monitoring blood flow is, therefore, important and helps clinicians assess tissue health. Various diseases like diabetes, peripheral vascular disease, and Raynaud’s phenomenon impair the tissue blood supply and, therefore, measurement of the same can provide information for the diagnosis/assessment of similar diseases. Changes in microvascular perfusion (tissue blood flow) are found to occur at the onset of a disease and hence, these changes, if detected, can provide early disease diagnosis, which in turn can increase the survival rates.

Laser speckle contrast imaging (LSCI) introduced in 1970s is a nonscanning, noninvasive wide field optical technique, which is being explored widely in microvascular perfusion assessment.1 The principle of LSCI can be found elsewhere.2,3 This technique was a breakthrough in the field of noninvasive monitoring of blood flow. Since then, various researchers have demonstrated a qualitative relationship between motion of scatterers and the resulting spatial or temporal speckle contrast of the integrated speckle pattern. LSCI offers good spatial and temporal resolution and has found applications in real-time blood flow imaging of retina and skin,48 cerebral blood flow imaging,9,10 monitoring of blood flow dynamics during neurosurgeries,11 and also in photodynamic therapy.12 Most of the discussed techniques rely on the variations in the contrast of the obtained speckle image.5,1315 Although LSCI has found tremendous applications in biomedical imaging, the quantification of flow velocities in its absolute units still remains a challenge and continues to be of increased research interest.1619 Although flow quantification has been tried out previously using alternate methods,1,2029 LSCI, due to its inherent advantage of whole field nature and noninvasiveness, holds a potential to be probed in quantitative studies.

It is imperative to keep the scatterer concentration uniform to ensure the relationship between speckle contrast and flow velocities while employing LSCI. As speckle image contrast is also dependent on the concentration of scattering centers, it is also possible to extract concentration related information by analyzing the contrast changes.30 Hence, simultaneous variations of flow and scatterer concentration will be of great challenge to get assessed using LSCI. Extracting additional features of speckle pattern may enable the simultaneous extraction of flow and concentration details. Spatial and temporal speckle statistics, such as autocorrelation and power spectrum, have also been widely used to find the velocity and concentration of the moving scatterers, in addition to the texture analysis of surfaces and images.3134 Recently, fractal nature of the speckle patterns has also been proposed for characterization of rough surfaces, texture analysis, and spatial characterization of multiscattering medium.3537 The fractal analysis of LSCI data has also been carried out.38 Our group has worked previously on the comparison of fractality of laser Doppler flow meter signals and laser speckle images acquired from different parts of the body in healthy subjects previously and has shown that the fractality of speckle images are found to be in agreement with the fractality Doppler signals in characterizing the flow changes.39 From all these reported literature, the fractal nature of speckle pattern could be understood as an alternate tool in assessing flow or scatterer concentration–related parameters. In the present work, an experimental model is attempted analyzing the contrast and fractality features of a captured speckle pattern for extracting the flow velocity as well as scatterer concentration. The initially obtained results are found to be in agreement, suggesting the possible exploration for a more detailed and reliable quantified approach toward the synchronous analysis of multiple parameters related to fluid flow.

2.

Materials and Methods

2.1.

Laser Speckle Imaging Experimental Setup

The schematic of the experimental setup used in this study is shown in Fig. 1. Intralipid 20% solution (Fresenius Kabi, Bad Homburg, Germany) mimicking the body fluid of different concentrations (1%, 2%, 3%, 4%, and 5% v/v) is prepared and pumped at different flow rates through a phantom flow channel ranging from 10 to 190ml/h in steps of 20ml/h using a syringe pump (Model RH-SY10, Ravel Hiteks Pvt. Ltd, India).

Fig. 1

Schematic of laser speckle imaging setup.

JBO_18_11_111419_f001.png

The flow phantom was custom made by grooving channels of 3-mm width and 5-mm depth in perspex sheet (120×40mm2). The flow phantom was illuminated by diode pumped solid state laser (671 nm, 30-mW power) and speckle images were captured by the CCD camera (DCU223M, Thorlabs, Newton, NJ) zoom lens system. In the red and near-infrared region of light, most of the interactions in the tissue are scattering events rather than absorption events4042 and hence the choice of intralipid solution mainly representing the scatterers was made for the experiments. The distance between the CCD camera and the flow channel was maintained constant (40 cm) for all experiments. The laser source to sample distance was 15 cm. The optical setup provided a magnification of 11 and f number of 2.5, thus producing speckles of 8.12μm in diameter, which is approximately twice the pixel size of 4.65 μm, hence being optimal for speckle contrast imaging.8

Exposure time of the CCD camera was set to 20 ms, and subsequent image processing was done using MATLAB (Mathworks, Massachusetts).

2.2.

Laser Speckle Imaging of the Flow Channel—Contrast Analysis

The acquired speckle image for the flow of intralipid 1% solution at a rate of 190ml/h is shown in Fig. 2(a). For the spatial contrast analysis, the flow channel of size 500×21pixels was selected and processed. The speckle contrast was calculated using Eq. (1) by taking 5×5 blocks of pixels over the entire channel image, thus producing a false color contrast map,2 as shown in Fig. 2(b).

Eq. (1)

K=σI,
where K is the speckle contrast, σ is the standard deviation, and I is the mean intensity. The average contrast of three maps corresponding to images taken with an interval of 10 s between them is calculated after enhancement procedures, which were adopted uniformly for different flow conditions.

Fig. 2

(a) Raw speckle image and (b) speckle contrast image for 1% intralipid, 190ml/h.

JBO_18_11_111419_f002.png

2.3.

Laser Speckle Imaging of the Flow Channel: Fractal Analysis

First order and second order statistics of speckle pattern have been used to analyze properties of the scattering media. First-order statistics describe the properties of speckle fields at each point. This includes intensity probability distribution function and contrast. Statistics of second order show how fast the intensity changes from point to point in the speckle pattern. Second-order statistics are described by autocorrelation function (g2) of intensity fluctuations given by Eq. (2):43

Eq. (2)

g2(x)=I(x+Δx)I(x),
where x is the spatial or temporal variable, Δx denotes the change in the variable, and I(x) is the intensity of speckle at x.

Small intensity fluctuations are described by autocorrelation function (g2˜) of the form Eq. (3) and structure function (D ) in Eq. (4) is used:43

Eq. (3)

g2˜(Δx)=[I(x+Δx)I][I(x)I]

Eq. (4)

D(Δx)=[I(x+Δx)I(x)]2.

The structure function is preferred over autocorrelation function while considering small scale intensity fluctuations. In the case of speckle image processing, structure function is preferred37 and in the present work, we have analyzed the intensity fluctuations of the speckle pattern with variations in both flow and scatterer concentrations based on the structure function approach.

The power law behavior of power spectral density of speckle intensity fluctuations characterizes self-similar processes, as shown in Fig. 3 and hence speckle statistics can be modeled by fractional Brownian motion as per the fractal theory,35 and structure function of the speckle intensity fluctuations can be redefined by using a Brownian motion approach, wherein the variance of the process is proportional to the spatial or temporal increments as given by Eq. (5).

Eq. (5)

D(Δx)=[I(x+Δx)I(x)]2Δx2H,
where H is the Hurst coefficient (0<H<1).35

Fig. 3

Power law behavior of speckle intensity fluctuations (|β|>1, hence the fractal Brownian motion).

JBO_18_11_111419_f003.png

In our analysis, the structure function was calculated over each column of the channel image (500×21pixels) considering the same images acquired as mentioned in Secs. 2.1 and 2.2 after subtracting the mean from the image, as given by Eqs. (6) and (7), and was averaged over all the columns.

Eq. (6)

I(x,y)=I(x,y)I(x,y),

Eq. (7)

D(x,y+Δy)=[I(x,y+Δy)I(x,y)]2Δy2H,
where I(x,y) represents the intensity at location (x,y) of the original image, I(x,y) denotes the mean subtracted image, and D(x,y+Δy) represents the variance of the intensity fluctuations along each column of the image.

The Hurst coefficient was calculated by fitting a regression line to the linear part of log–log plot of the increments in variance versus proximity of pixels in the image. The plot of the structure function versus pixel lag on a log–log scale is shown in Fig. 4 for intralipid concentrations of 1% and 5%, respectively, at flow rate of 190ml/h.

Fig. 4

Plot of the structure function with different concentrations of intralipid at flow rate 190ml/h. (a) 1% intralipid and (b) 5% intralipid.

JBO_18_11_111419_f004.png

H indicates the extent of correlation between successive elements of the signal.44,45 H>0.5 indicates that the stochastic process is positively correlated, i.e., an increasing trend in the past follows an increasing trend in the future and vice versa. H<0.5 indicates negative correlation or antipersistence indicating successive elements in the data following alternating increments and decrements.

3.

Results and Discussions

The speckle images obtained for different flows and concentrations of intralipid solution, as mentioned in Sec. 2, were analyzed for the fractality and contrast. Table 1 shows the mean H and contrast values for different flow rates/flow velocities and intralipid concentrations along with the standard deviation measurements. The range of flow velocities is so chosen in order to represent the normal and varied range of tissue perfusion.46

Table 1

Hurst coefficient (H) and speckle contrast (K) values for different velocities of flow (FLV) and scatterer (intralipid) concentrations in % V/V.

Intralipid concentration
SC1%2%3%4%5%
Flow rate (ml/h)FLV (mm/s)HKHKHKHKHK
100.1850.252±0.0200.153±0.00040.381±0.0040.148±0.00050.470±0.0080.144±0.00040.567±0.0070.140±0.00020.8±0.0150.135±0.0002
300.5550.297±0.0130.150±0.00090.470±0.0310.145±0.00040.554±0.0130.142±0.00010.653±0.0010.138±0.00010.830±0.0130.135±0.0003
500.9250.310±0.0040.148±0.00050.525±0.0040.143±0.00060.564±0.0160.141±0.00020.674±0.0010.135±0.00010.834±0.0020.133±0.0005
701.2960.336±0.0070.146±0.00010.533±0.0090.141±0.00050.597±0.0070.139±0.00020.688±0.00050.133±0.00020.839±0.0040.132±0.0003
901.6660.364±0.0070.145±0.00050.561±0.0050.139±0.00010.616±0.0080.136±0.00080.693±0.0080.132±0.00010.865±0.0040.130±0.0004
1102.0730.384±0.0060.143±0.00070.602±0.0020.138±0.00030.642±0.0020.135±0.00050.720±0.0020.130±0.00020.872±0.00130.127±0.0002
1302.4070.394±0.0040.140±0.00010.624±0.010.136±0.00030.653±0.0010.133±0.00020.757±0.0030.128±0.00010.873±0.0020.126±0.0007
1502.7770.408±0.0050.138±0.00060.651±0.0010.135±0.00050.703±0.0070.131±0.00010.816±0.0040.127±0.00020.890±0.0020.124±0.0007
1703.1480.441±0.0270.136±0.00050.679±0.0090.133±0.00020.831±0.0020.129±0.00040.860±0.0080.125±0.00060.891±0.0010.123±0.0009
1903.5180.515±0.0130.135±0.00030.710±0.010.132±0.00030.852±0.00080.128±0.00010.877±0.0010.124±0.00010.904±0.0020.122±0.0002

3.1.

Variation of H with Flow Velocities

An increasing trend was observed for H with an increase in the flow velocity. The fractional Brownian motion analyzed spatially in the intralipid fluid speckle image is found to be more positively correlated with increase in the flow velocity, as is evident from the increased H. The fractional Brownian motion, which could also be considered as a diffusion process exhibited in the speckle image spatial analysis, has previously been investigated for blood.47 This could be seen as an effect to streamline or positively correlate the existing fractional Brownian motion with the effect of increased flow velocity. We hypothesize that the randomness in a fractional Brownian motion reduces with increased flow and this may be attributed to the cause of positive correlation seen with increased flow. An increase in H indicates that the variance of the intensity fluctuations as a function of the pixel location within the speckle images is increasing with increasing flow rates. A plot showing the relationship of H with flow velocity is shown in Fig. 5(a).

Fig. 5

(a) Plot of the experimental data and linear fit showing dependence of H on scatterer concentration and flow velocity. (b) Plot of the experimental data and linear fit showing dependence of contrast on scatterer concentration and flow velocity.

JBO_18_11_111419_f005.png

3.2.

Variation of H with Scatterer Concentration

An increasing trend for spatial H is observed with increase in the concentration of the scatterers similar to that of the response for flow velocity. When the particle concentration is increased, the single particle chaotic movement may be restricted, facilitating particle agglomerations within the medium, which can be considered as the cause of increasing positive correlation for the fractional Brownian movement represented by the increased value of H corresponding to the spatial speckle image analysis. Also, as the scatterer concentration increases, the backscattered image intensity is increased as a result of more scattering centers in the medium leading to an increase in pixel intensity fluctuations, thereby increasing H. The relationship of H with scatterer concentration for constant flow velocities can be observed from Fig. 5(a).

3.3.

Variation of Speckle Contrast with Flow Velocity and Scatterer Concentration

Analyzing average speckle contrast in the region of interest for flow analysis has been established by many researchers and is referenced widely in Sec. 1 of this article. As flow velocity increases, the image becomes blurred and the average contrast decreases. A plot showing the relationship of speckle contrast and flow velocity is shown in Fig. 5(b). It has also been established previously that the average speckle image contrast increases with increase in the scatterer concentration in static fluids30 by our group. The dependence of speckle contrast on scatterer concentration and flow velocity can also be seen in reports.48 We have observed a similar trend for different flows, with measurements being taken for a specific flow at a time. The relationship of scatterer concentration with speckle image contrast can be observed from Fig. 5(b).

3.4.

Dual Parametric Model for Simultaneous Scatterer Concentration and Flow Velocity Measurements

From Figs. 5(a) and 5(b), it is observed that H and contrast follow an inverse relationship in the case of flow velocity. For a given concentration of intralipid and for increasing flow rates, H is increasing while the speckle contrast decreases. On the other hand, for a given flow velocity, increased H and contrast were observed with the increase in scatterer concentration. Also, it is observed that at higher scatterer concentrations, the change in H with varying flow velocities is less compared to that at lower concentrations.

As both Hurst coefficient and speckle contrast are found to be dependent on scatterer concentration and flow velocity, we have used multiple linear regression with our experimental data in order to model the dependence of scatterer concentration and flow rate on H and contrast. Below equations [Eqs. (8) and (9)] obtained represent the fitted plot, and corresponding plots are given along with the experimental results in Figs. 5(a) and 5(b). The parametric dependence (flow and scatterer concentration) of H as well as contrast was approximated to a linear fit here, which can offer the possibility of assessing the parameters on a quantitative scale. This pilot study emphasizes that by taking a single-speckle pattern image of the region of interest and calculating its contrast as well as Hurst coefficient spatially can yield important information regarding the velocity of flow and concentration of scatterers present in the fluid.

Eq. (8)

H=0.0741*(Flow velocity)+0.1136*(Scatterer concentration)+0.1588,

Eq. (9)

K=0.0048*(Flow velocity)0.0037*(Scatterer concentration)+0.1557.

4.

Conclusion

The flow and scatterer concentration dependent variables of the speckle pattern image, such as contrast and Hurst coefficient, have been explored in this paper and could yield clear distinction between various flow values and scatterer concentrations. This is, in turn, intended to provide a quantitative assessment of flow velocities and scatterer concentrations, where the results provided an error within 5% limit. It is to be noted that the methodology discussed here yield a quantitative tool for assessing flow velocity and scatterer concentration simultaneously from a single acquired speckle pattern using simple instrumentation. The possible expression of these changes in a more analytical way will open up new directions in the field of LSCI. On the application side, this methodology can serve in the long run as a helping tool in characterizing normal and malignant tissues with associated variation in blood flow complexities and scatterer concentration based on the values of obtained Hurst coefficient as well as speckle contrast.

Acknowledgments

The authors acknowledge financial support from ICSR & IITM for carrying out the experimental research.

References

1. 

A. F. FercherJ. D. Briers, “Flow visualization by means of single-exposure speckle photography,” Opt. Commun., 37 (5), 326 –330 (1981). http://dx.doi.org/10.1016/0030-4018(81)90428-4 OPCOB8 0030-4018 Google Scholar

2. 

J. D. BriersS. Webster, “Laser speckle contrast analysis (LASCA): a no scanning, full- field technique for monitoring capillary blood flow,” J. Biomed. Opt., 1 (2), 174 –179 (1996). http://dx.doi.org/10.1117/12.231359 JBOPFO 1083-3668 Google Scholar

3. 

J. D. Briers, “Laser speckle contrast imaging for measuring blood flow,” Opt. Appl., 37 (1–2), 139 –151 (2007). OPAPBZ 0078-5466 Google Scholar

4. 

J. D. Briers, “Laser Doppler, speckle and related techniques for blood perfusion mapping and imaging,” Physiol. Meas., 22 (4), R35 –R66 (2001). http://dx.doi.org/10.1088/0967-3334/22/4/201 PMEAE3 0967-3334 Google Scholar

5. 

H. Y. Chenget al., “Modified laser speckle imaging method with improved spatial resolution,” J. Biomed. Opt, 8 (3), 559 –564 (2003). http://dx.doi.org/10.1117/1.1578089 JBOPFO 1083-3668 Google Scholar

6. 

H. ChengT. Q. Duong;, “Simplified laser-speckle-imaging analysis method and its application to retinal blood flow imaging,” Opt. Lett., 32 (15), 2188 –2190 (2007). http://dx.doi.org/10.1364/OL.32.002188 OPLEDP 0146-9592 Google Scholar

7. 

H. Chenget al., “Laser speckle imaging of blood flow in microcirculation,” Phys. Med. Biol., 49 (7), 1347 –1357 (2004). http://dx.doi.org/10.1088/0031-9155/49/7/020 PHMBA7 0031-9155 Google Scholar

8. 

D. A. BoasA. K. Dunn, “Laser speckle contrast imaging in biomedical optics,” J. Biomed. Opt., 15 (1), 011109 (2010). http://dx.doi.org/10.1117/1.3285504 JBOPFO 1083-3668 Google Scholar

9. 

H. Y. Chenget al., “Hyperosmotic chemical agent’s effect on in vivo cerebral blood flow revealed by laser speckle,” Appl. Opt, 43 (31), 5772 –5777 (2004). http://dx.doi.org/10.1364/AO.43.005772 APOPAI 0003-6935 Google Scholar

10. 

S. Yuanet al., “Determination of optimal exposure time for imaging of blood flow changes with laser speckle contrast imaging,” Appl. Opt., 44 (10), 1823 –1830 (2005). http://dx.doi.org/10.1364/AO.44.001823 APOPAI 0003-6935 Google Scholar

11. 

A. B. Parthasarathyet al., “Laser speckle contrast imaging of cerebral blood flow in humans during neurosurgery: a pilot clinical study,” J. Biomed. Opt., 15 (6), 066030 (2010). http://dx.doi.org/10.1117/1.3526368 JBOPFO 1083-3668 Google Scholar

12. 

T. K. Smithet al., “Microvascular blood flow dynamics associated with photodynamic therapy and pulsed dye laser irradiation,” Lasers Surg. Med., 38 (5), 532 –539 (2006). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

13. 

F. Domokiet al., “Evaluation of laser-speckle contrast image analysis techniques in the cortical microcirculation of piglets,” Microvasc. Res., 83 (3), 11 –317 (2012). http://dx.doi.org/10.1016/j.mvr.2012.01.003 MIVRA6 0026-2862 Google Scholar

14. 

O. YangB. Choi, “Laser speckle imaging using a consumer-grade color camera,” Opt. Lett., 37 (19), 3957 –3959 (2012). http://dx.doi.org/10.1364/OL.37.003957 OPLEDP 0146-9592 Google Scholar

15. 

R. LiuJ. QinR.K. Wang, “Motion-contrast laser speckle imaging of microcirculation within tissue beds in vivo,” J. Biomed. Opt., 18 (6), 060508 (2013). http://dx.doi.org/10.1117/1.JBO.18.6.060508 JBOPFO 1083-3668 Google Scholar

16. 

D. D. DuncanS. J. Kirkpatrick, “Can laser speckle flowmetry be made a quantitative tool?,” J. Opt. Soc. Am. A Opt. Image Sci. Vis., 25 (8), 2088 –2094 (2008). http://dx.doi.org/10.1364/JOSAA.25.002088 JOAOD6 0740-3232 Google Scholar

17. 

D. Brierset al., “Laser speckle contrast imaging: theoretical and practical limitations,” J. Biomed. Opt., 18 (6), 066018 (2013). http://dx.doi.org/10.1117/1.JBO.18.6.066018 JBOPFO 1083-3668 Google Scholar

18. 

B. Choiet al., “Linear response range characterization and in vivo application of laser speckle imaging of blood flow dynamics,” J. Biomed. Opt., 11 (4), 041129 (2006). http://dx.doi.org/10.1117/1.2341196 JBOPFO 1083-3668 Google Scholar

19. 

A. K. Dunn, “Laser speckle contrast imaging of cerebral blood flow,” Ann. Biomed. Eng., 40 (2), 367 –377 (2012). http://dx.doi.org/10.1007/s10439-011-0469-0 ABMECF 0090-6964 Google Scholar

20. 

M. D. Stern, “In vivo evaluation of microcirculation by coherent light scattering,” Nature, 254 (5495), 56 –58 (1975). http://dx.doi.org/10.1038/254056a0 NATUAS 0028-0836 Google Scholar

21. 

R. BonnerR. Nossal, “Model for laser Doppler measurements of blood flow in tissue,” Appl. Opt., 20 (12), 2097 –2107 (1981). http://dx.doi.org/10.1364/AO.20.002097 APOPAI 0003-6935 Google Scholar

22. 

A. Serovet al., “Laser Doppler perfusion imaging with a complimentary metal oxide semiconductor image sensor,” Opt. Lett., 27 (5), 300 –302 (2002). http://dx.doi.org/10.1364/OL.27.000300 OPLEDP 0146-9592 Google Scholar

23. 

M. N. Kimet al., “Noninvasive measurement of cerebral blood flow and blood oxygenation using near-infrared and diffuse correlation spectroscopies in critically brain-injured adults,” Neurocrit. Care, 12 (2), 173 –180 (2010). http://dx.doi.org/10.1007/s12028-009-9305-x NCEACB 1541-6933 Google Scholar

24. 

J. Hagbladet al., “A technique based on laser Doppler flowmetry and photoplethysmography for simultaneously monitoring blood flow at different tissue depths,” Med. Biol. Eng. Comput., 48 (5), 415 –422 (2010). http://dx.doi.org/10.1007/s11517-010-0577-2 MBECDY 0140-0118 Google Scholar

25. 

S. Margaretaet al., “Non-invasive monitoring of muscle blood perfusion by photoplethysmography: evaluation of a new application,” Acta Physiol. Scand., 183 (4), 335 –343 (2005). http://dx.doi.org/10.1111/aps.2005.183.issue-4 APSCAX 0001-6772 Google Scholar

26. 

V. J. Srinivasanet al., “Optical coherence tomography for the quantitative study of cerebrovascular physiology,” J. Cereb. Blood Flow Metab., 31 (6), 1339 –1345 (2011). http://dx.doi.org/10.1038/jcbfm.2011.19 JCBMDN 0271-678X Google Scholar

27. 

J. D. BriersS. Webster, “Laser speckle contrast analysis (LASCA): a no scanning, full- field technique for monitoring capillary blood flow,” J. Biomed. Opt., 1 (2), 174 –179 (1996). http://dx.doi.org/10.1117/12.231359 JBOPFO 1083-3668 Google Scholar

28. 

D. FixlerZ. Zalevsky, “Estimation of flow rate and direction of medium with low scattering coefficient via linear polarization measurement,” Opt. Lasers Eng., 51 (2), 91 –95 (2013). http://dx.doi.org/10.1016/j.optlaseng.2012.09.007 OLENDN 0143-8166 Google Scholar

29. 

D. Fixleret al., “Depolarization of light in biological tissues,” Opt. Lasers Eng., 50 (6), 850 –854 (2012). http://dx.doi.org/10.1016/j.optlaseng.2012.01.011 OLENDN 0143-8166 Google Scholar

30. 

A. K. JayanthyN. SujathaM. Ramasubba Reddy, “Non-invasive assessment of static scatterer concentration in phantom body fluids using laser speckle contrast analysis technique,” Opt. Lasers Eng., 49 (4), 553 –556 (2011). http://dx.doi.org/10.1016/j.optlaseng.2010.12.003 OLENDN 0143-8166 Google Scholar

31. 

J. ChurnsideH. Yura, “Velocity measurement using laser speckle statistics,” Appl. Opt., 20 (2), 3539 –3541 (1981). http://dx.doi.org/10.1364/AO.20.003539 APOPAI 0003-6935 Google Scholar

32. 

N. TakaiT. IwaiT. Asakura, “Real-time velocity measurement for a diffuse object using zero-crossings of laser speckle,” J. Opt. Soc. Am., 70 (4), 450 –455 (1980). http://dx.doi.org/10.1364/JOSA.70.000450 JOSAAH 0030-3941 Google Scholar

33. 

A. OhtsuboJ. A. Asakura, “Velocity measurement of a diffuse object by using time-varying speckles,” J. Opt. Quant. Electron., 8 (6), 523 –529 (1976). http://dx.doi.org/10.1007/BF00620143 OQELDI 0306-8919 Google Scholar

34. 

A. OulamaraG. TribillonJ. Duvernoy, “Biological activity measurement on botanical specimen surfaces using a temporal decorrelation effect of laser speckle,” J. Modern Opt., 36 (2), 165 –179 (1989). http://dx.doi.org/10.1080/09500348914550221 JMOPEW 0950-0340 Google Scholar

35. 

S. GuyotM. C. PéronE. Deléchelle, “Spatial speckle characterization by Brownian motion analysis,” Phys. Rev. E, 70 (4), 046618 (2004). http://dx.doi.org/10.1103/PhysRevE.70.046618 PLEEE8 1063-651X Google Scholar

36. 

O. Carvalhoet al., “Statistical speckle study to characterize scattering media: use of two complementary approaches,” Opt. Express, 15 (21), 13817 –13831 (2007). http://dx.doi.org/10.1364/OE.15.013817 OPEXFF 1094-4087 Google Scholar

37. 

O. CarvalhoM. BenderitterL. Roy, “Noninvasive radiation burn diagnosis using speckle phenomenon with a fractal approach to processing,” J. Biomed. Opt., 15 (2), 027013 –027013 (2010). http://dx.doi.org/10.1117/1.3394261 JBOPFO 1083-3668 Google Scholar

38. 

A. H. Heurtieret al., “Laser speckle contrast imaging: multifractal analysis of data recorded in healthy subjects,” Med. Phys., 39 (10), 5847 –5849 (2012). Google Scholar

39. 

N. SujathaL. Cerine, “Laser based signal and image fractal analysis for assessment of blood flow,” Proc. SPIE, 8427 84272M (2012). http://dx.doi.org/10.1117/12.921743 PSISDG 0277-786X Google Scholar

40. 

B. C. WilsonS. L. Jacques, “Optical reflectance and transmittance of tissues: principles and applications,” IEEE J. Quantum Electron., 26 (12), 2186 –2199 (1990). http://dx.doi.org/10.1109/3.64355 IEJQA7 0018-9197 Google Scholar

41. 

C. Simpsonet al., “Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the Monte Carlo inversion technique,” Phys. Med. Biol., 43 (9), 2465 –2478 (1998). http://dx.doi.org/10.1088/0031-9155/43/9/003 PHMBA7 0031-9155 Google Scholar

42. 

R. Graaffet al., “Optical properties of human dermis in vitro and in vivo,” Appl. Opt., 32 (4), 435 –447 (1993). http://dx.doi.org/10.1364/AO.32.000435 APOPAI 0003-6935 Google Scholar

43. 

V. V. Tuchin, Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis, SPIE Press, Bellingham, WA (2007). Google Scholar

44. 

J. J. CollinsC. J. De Luca, “Open-loop and closed-loop control of posture: a random-walk analysis of center-of-pressure trajectories,” Exp. Brain Res., 95 (2), 308 –318 (1993). http://dx.doi.org/10.1007/BF00229788 EXBRAP 0014-4819 Google Scholar

45. 

A. Ekeet al., “Physiological time series: distinguishing fractal noises from motions,” Eur. J. Physiol., 439 (4), 403 –415 (2000). http://dx.doi.org/10.1007/s004240050957 0031-6768 Google Scholar

46. 

M. Stückeret al., “Capillary blood cell velocity in human skin capillaries located perpendicularly to the skin surface: measured by a new laser Doppler anemometer,” Microvasc. Res., 52 (2), 188 –192 (1996). http://dx.doi.org/10.1006/mvre.1996.0054 MIVRA6 0026-2862 Google Scholar

47. 

E. C. Eckstein, “Fractal Brownian motion and particle motions in blood flow,” in Annual Int. Conf. IEEE Engineering in Medicine and Biology Society, 2240 –2241 (1991). Google Scholar

48. 

A. B. Parthasarathyet al., “Laser speckle contrast imaging of flow in a microfluidic device,” Proc. SPIE, 6446 644604 (2007). http://dx.doi.org/10.1117/12.700572 PSISDG 0277-786X Google Scholar
© 2013 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2013/$25.00 © 2013 SPIE
Cerine Lal, Arnab Banerjee, and Narayanan Unni Sujatha "Role of contrast and fractality of laser speckle image in assessing flow velocity and scatterer concentration in phantom body fluids," Journal of Biomedical Optics 18(11), 111419 (18 November 2013). https://doi.org/10.1117/1.JBO.18.11.111419
Published: 18 November 2013
Lens.org Logo
CITATIONS
Cited by 18 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Speckle

Laser scattering

Laser speckle contrast imaging

Blood circulation

Speckle pattern

Tissues

Fractal analysis

Back to Top