9 May 2018 Effect of intraocular scattering in macular pigment optical density measurements
Author Affiliations +
Abstract
Fundus reflectometry is a common in-vivo, noninvasive method to estimate the macular pigment optical density (MPOD). The measured density, however, can be affected by the individual’s intraocular scattering. Scattering causes a reduction in the contrast of the fundus image, which in turn leads to an underestimation of the measured density. Intraocular scattering was measured optically in a group of seven young, healthy subjects using the method of optical integration and was subsequently used to correctly estimate the MPOD from fundus images. It was shown that when scattering is not considered, the measured optical density using fundus reflectometry can be underestimated by as high as 16% for our group of subjects.

1.

Introduction

Intraocular scattering has a large spatial range, extending well beyond 2 deg14 and causing an overall contrast decrease in recorded fundus images. This decrease depends on three factors: the wavelength, with the decrease being is stronger at the edges of the visible spectrum,5,6 the quality of the ocular optics, and finally the dynamic range of the fundus image.

Scattering in fundus images is manifested in a similar manner as in vision: light originating from the brighter components of the fundus image creates a veil of light over the darker parts of the image. Scattering increases with age in cuartic manner,4 but elevated scattering can also be the result of specific ocular diseases, such as corneal edemas and swelling7,8 or cataracts affecting the crystalline lens.9,10

Macular pigment (MP) is a collective term for the carotenoids lutein, zeaxanthin and mesozeaxanthin, and, as its name suggests, is located mainly, but not exclusively,-at the macula, with its higher concentration observed at the fovea. Its yellow color and the fact that it is located prior to the outer segments of the photoreceptors suggest that MP acts as a short wavelength filter.11 Moreover, carotenoids are well known for their antioxidant activity12 and, therefore, MP is thought to also act as an antioxidant protective agent in the retina.1315 The lack of it has been correlated to degenerative macular diseases, such as age-related macular degeneration.1618 In other studies, MP has been related to improved visual function, particularly acuity and visibility.19,20

In the recent years, several in-vivo methods for the measurement of the optical density of the MP have been developed, classified in two categories: the optical, or objective, methods and the psychophysical, or subjective, ones. From the optical methods, fundus reflectance21,22 and fundus autofluorescence23 are the most commonly used, with good repeatability and coherent results. Among the psychophysical methods, heterochromatic flicker photometry24,25 is the most commonly used, with several commercial devices available. Each of the above methods may suffer from scattering to a different extent,26 which depends on the specifics of the setup and the nature of the measurement.

The objective of this work was to evaluate the effect of intraocular scattering in macular pigment optical density (MPOD) measurements using reflectance values from fundus images, recorded using a flood illumination system. In addition, a method for compensation of the effect for more accurate MPOD measurements is proposed.

2.

Methods

A complete measurement consisted of essentially three distinct steps: (a) the objective reconstruction of the individual’s wide angle point spread function (PSF), (b) the acquisition of foveal images at 450 and 550 nm and their “correction” using the corresponding PSF (calculated previously), and (c) the calculation of the MPOD, through the appropriate formulae. Each step is described in more detail in Secs. 1, 2.2, and 2.3 below.

The instrument used for the reconstruction of the PSF and to capture the fundus images for the measurement of the MPOD is shown in Fig. 1. The setup consisted of an arc lamp backilluminating a mechanical iris (I), which was conjugated to the retina through lenses L1, L2, and L3. For the imaging, light from the fundus was directed to an EMCCD camera with the use of a half mirror, conjugated to the retinal plane through lenses L2, L3, and L4. The wavelength could rapidly switch between 450 and 550 nm using an interference filter wheel (λ) and the appropriate illumination time was achieved with the use of a mechanical shutter (Sh) controlled by a microcontroller. Two appropriately positioned diaphragms (D1, D2) were used to avoid corneal and lens reflections, ensuring that the captured light originates exclusively from the fundus. The latter is particularly important in scattering measurements. A target located at the retinal plane through beam splitter was used for eye fixation. Finally, a stereo camera system was used for the precise alignment of the pupil. Details on the alignment system can be found in authors’ previous works.27,28 A more thorough description of the optical system can be found in previous work by Christaras et al.1

Fig. 1

Optical system for the reconstruction of the wide angle PSF and the recording of the images used in the MPOD measurement.

JBO_23_5_056001_f001.png

The high sensitivity of the EMCCD camera allowed us to use low light power for the imaging of the fundus; maximum light power at the corneal plane was measured using a Newport 1815-C Power Meter (Newport Corporation, United States) and was found to be 0.135 mW, well below the ISO 15004-2:2007 maximum permissible exposure limits.

2.1.

PSF Calculation

The scatter estimation was based on the method of optical integration, and it is explained in detail elsewhere.2,5 In short, the method consists of the projection of a set of disks of increasing diameter on the retina and the subsequent recording of the double pass image. The double-pass PSF at a specific angle is then calculated by differentiating the intensity value at the center of a recorded disk extending up to that angle.2 Therefore, by projecting an adequate number of disks of different size and measuring their reflectance intensity at the center, one can reconstruct the double-pass PSF to the desired spatial extent and resolution. The single-pass PSF can then be calculated by taking the square-root of the Fourier transform of the double-pass PSF. In practice, for a less noisy PSF reconstruction, a monoparametric PSF based on the CIE glare function4 was used. A family of PSFs and subsequently of intensity values was simulated for the different parameter values, and the curve that best fitted the intensity data was determined. The corresponding modulation transfer function (MTF) could then be calculated by taking the absolute value of the Fourier transform of the PSF.

Several fundus images were taken prior to the actual measurement, to find the best focus and an appropriate vein-free retinal site for the scattering measurement. Intensity analysis of the images showed no significant reflectance fluctuations, suggesting that photoreceptor reflectance did not change throughout the measurement.

The actual reconstruction of the PSFs for the set of subjects was carried out in a previous study.1

2.2.

Acquisition, Registration, and Correction of the Foveal Images

For the measurement of the MPOD, the procedure consisted of the intensity analysis of two fundus foveal images of 12 deg diameter, taken at 550 nm, where the MP absorption is insignificant, and at 450 nm, where the absorption is close to its maximum.29,30 The exposure time was kept below 10 ms to avoid motion blur and ensure that the recorded intensity range is similar for all subjects and below camera saturation.

In total, five fundus images were acquired at each wavelength and were registered and averaged to increase the signal-to-noise ratio. Subsequently, the images at 550 and 450 nm were again spatially registered with respect to one another so that each pixel depicted the exact same retinal region. An appropriate filter function was applied in the Fourier domain to all images in order to eliminate irrelevant structures mostly visible in blue31,32 and blood vessels absorbing at both blue and green.33 The low-pass filter was chosen such that frequencies higher than about 2  cycles/deg would be eliminated, leaving the lower frequencies almost intact. This was done by multiplying the MTF with the function:

G(f)=eaf6,
where a is a parameter related to the frequency cut-off limit and f is f=fx2+fy2. The effect of this filter function on a fundus image is shown in Fig. 2.

Fig. 2

Example of a 12-deg fundus image (a) before and (b) after applying the spatial filtering.

JBO_23_5_056001_f002.png

Subsequently, the images were corrected for intraocular scattering following our compensation method.1 Briefly, for the correction of a fundus image, a deconvolution-based technique was used; according to this technique, the appropriate diffraction-limited MTF was divided by the reconstructed “scattering” MTF and the result was multiplied with the image in the Fourier domain. The resulting image was then compensated for the measured scattering effects.

2.3.

MPOD Estimation

For the estimation of the MPOD, a simplified four-layer fundus model was assumed and the method described by Delori et al.23 was followed. In their work, Delori et al.23 observed a good correlation between this method and other methods for the measurement of the MP.

The fundus layers, following the direction of light, were: inner limiting membrane (ILM), MP, photoreceptor layer (Ph), retinal pigment epithelium (RPE), Bruch’s membrane (BM), and choroid. Assuming that no light is attenuated or reflected before the MP, then the reflectance measured at the camera is related with the MPOD through the relation:

Eq. (1)

R(x,y,λ)=R*(x,y,λ)102D(x,y,λ),
where R* is the reflectance of all layers posterior to the MP, D is the MP optical density, and the factor 2 expresses the double pass of the light through the MP. Let (x,y) and (x,y) be two different retinal sites, then applying Eq. (1) at each site, dividing the two equations and with a little bit of algebra, one obtains

Eq. (2)

logR(x,y,λ)R(x,y,λ)=logR*(x,y,λ)R*(x,y,λ)+2[D(x,y,λ)D(x,y,λ)].

Replacing the optical density difference between the two sites DMP(λ)=D(x,y,λ)D(x,y,λ) and, subsequently, expressing that density difference as a function of the extinction coefficient DMP(λ)=DMP(460)KMP(λ) from Brown and Wald,29 Eq. (2) becomes

Eq. (3)

logRP(λ)RF(λ)=logRP*(λ)RF*(λ)+2DMP(460)KMP(λ),
where site (x,y) lies at the perifovea (P) and site (x,y) at the fovea (F). Applying it for two different wavelengths λ1 and λ2 and subtracting one equation from the other, one obtains

Eq. (4)

logRP(λ1)RF(λ1)logRP(λ2)RF(λ2)=logRP*(λ1)RF*(λ1)logRP*(λ2)RF*(λ2)+2DMP(460)[KMP(λ1)KMP(λ2)].

Assuming now that the reflection spectrum at the perifovea is proportional to the reflection spectrum at the fovea, then that the ratio RP*(λ)RF*(λ) is constant for all wavelengths, and therefore, Eq. (4) is simplified to

Eq. (5)

logRP(λ1)RF(λ1)logRP(λ2)RF(λ2)=2DMP(460)[KMP(λ1)KMP(λ2)].

Finally, solving Eq. (5) for DMP(460), one obtains

Eq. (6)

DMP(460)=0.5KMP(λ1)KMP(λ2){logRP(λ1)RP(λ2)+logRF(λ2)RF(λ1)},
where in our experiment λ1=550  nm and λ2=450  nm.

The values for the extinction coefficients were extracted from Brown’s data29 after integrating over the spectrum of each interference filter and they were calculated to be KMP(450)=0.9038 and KMP(550)=0.0089.

There are a number of different metrics used for the measurement of the MPOD like the median, the peak value, the weighted average, etc. However, no golden standard exists and different devices use different metrics. We used the average value for an area of 1 deg around the center of the fovea as the value for the MPOD. This is similar to the field size used in some commercial psychophysical devices.34

We performed the series of measurements in seven healthy volunteers, two females and five males, all Caucasians, with normal vision and no known ocular diseases. The younger subject was 23 years old and the older 34 years old and the average age was 29 years old. Four subjects had light iris pigmentation (blue or green) and three subjects had heavy iris pigmentation (brown). All the research and measurements followed the tenets of the Declaration of Helsinki and were approved by the local ethics committee. Informed consent was obtained from the subjects after they were fully informed about the nature of the measurements.

In order to see the effect of scattering in the MPOD measurement, the scattering compensation technique1 was applied for both the original images without any enhancement and the images after the scattering compensation. The resulted MPOD was then compared, and the relative effect of the correction was estimated.

3.

Results

For each subject, two PSFs, one at each wavelength, were calculated in our previous work.1 The computed scatter parameter is shown in Table 1. The reconstructed PSFs at 450 and 550 nm for two subjects are shown in Fig. 3.

Table 1

MPOD values before and after scattering compensation as well as the relative change between the two measurements for seven subjects.

SubjectAgeScattering parameters (blue)Scattering parameters (green)MPOD (no comp.; D.U.)MPOD (comp.; D.U.)Relative increase (%)
S1311.210.980.1510.1606.0
S2311.350.960.2190.25416.0
S3341.291.010.1230.13711.4
S4231.321.320.2000.2168.0
S5291.151.050.2110.2225.2
S6251.200.960.2430.2638.2
S7301.311.150.0820.09111.0

Fig. 3

(a) and (b) Reconstructed PSFs at 450 nm blue/dashed and 550 nm (green) for two different subjects.

JBO_23_5_056001_f003.png

The profile of the MPOD spatial distributions for all seven subjects is presented graphically in Fig. 4 for both cases, i.e., with scattering compensation (red line) and without scattering compensation (blue line). Table 1 shows the values for the average MPOD for 1 deg before (fifth column) and after (sixth column) scattering compensation, as well as the relative change (last column). As seen in the table, the underestimation of the MP when scattering is not considered ranges between 5.2% and 16%, with an average of 9.4%. The average MPOD before the compensation was found to be 0.176±0.058 and 0.192±0.064 after compensation. The importance of the effect depends on the amount of intraocular scattering at blue and green, their difference, specular reflections in the image, as well as the spatial characteristics of the MP itself.

Fig. 4

MPOD nasal–temporal profile before blue/dashed and after (red) scattering compensation for the seven subjects. All profiles were centered at the fovea.

JBO_23_5_056001_f004.png

4.

Discussion

In this work, the effect of intraocular scattering on MPOD measurements was quantified and, subsequently, compensated using an image processing technique. When the corrected images were used for the measurement of the MPOD, a mean increase of 9.4% in the value was observed for a set of seven young and healthy subjects. Regardless of the possible errors in the reflectometric estimation of MPOD, the incorporation of straylight in the calculation will result in an increase of the estimated value by a factor on the order of 16%. This increase may be higher in eyes with elevated straylight such as in eyes with cataract.

Several assumptions were made for the MPOD extraction from the fundus images. The most important was treating the fundus posterior to the MP as a uniform reflecting surface. For short and middle wavelengths below 580 nm, reflections occur mostly at the RPE and at BM35,36 and, therefore, melanin density at the RPE affects the amount of reflected light. RPE melanin is not constant throughout the fundus but exhibits a higher density at the fovea and, therefore, it has some effect in MPOD from reflectometry measurements. Nevertheless, Delori et al.23 found a contribution to the MPOD measurement of 0.03 D.U., concluding thus that the melanin difference across the fundus has a limited effect on the estimation of the MPOD using the reflectometry method. Reflections occurring prior to the RPE at the Ph, on the other hand, would not be affected by melanin density changes. Another possible scenario is that part of reflected light, however, originates from scattering by the Henle fibers, i.e., where the MP is believed to lie,37 and hence leading to a reduced double-pass through the MP and consequently an additional cause to further underestimate the MPOD.

On the reconstruction of the PSF, it is important to note that in our PSF manipulations throughout the method of optical integration, the phase transfer function is assumed to be 0. This is attributed to the fact that the PSF resulting from scattering is considered as rotationally symmetric, and therefore its Fourier transform will be a real function.

Regarding the directionality of the photoreceptors, because of the pupil configuration used in the system, illuminating and imaging angles are relatively large (about 5 deg) and wave-guiding is relatively low at those angles and, therefore, the SC effect is negligible.38 Tests prior to the experiment supported this fact.

There is yet another important factor that could lead to a wrong estimation of the MPOD using the reflectometry method: specular reflections, mainly taking place at the interface between the aqueous humor and the ILM.35 Those specular reflections depend on the parameters of the system, such as the incident angle of the rays and the curvature of the fundus. Moreover, they are known to be consistently higher in younger subjects.23,39 They are observed in several foveal images in flood illumination systems with unpolarized light, especially at middle and short wavelengths at about 3 deg at the perifovea (see Fig. 2). Those reflections may affect the measurement to some extent, since they act as a glare source, significantly increasing the dynamic range of the image. This was dealt with by ensuring that the image pixels at the sites where the specular reflections occur, were not saturated, though they were kept close to saturation in order to achieve the best possible dynamic range for the rest of the image. The use of cross polarizers could efficiently deal with specular reflections, but it was avoided because it would significantly reduce the available light at the camera and it could potentially have an effect on the measurement of the MP since MP is known to have a polarization preference.40

Finally, it needs to be noted that not all MPOD measurement techniques suffer at the same degree from scattering. In this study, only the effect of scattering for a flood illumination system is examined. For a rough estimate on how scattering affects different methods see Table 2 in the work of Leung.26 As a rule of thumb, one can expect the higher impact to be observed in flood illumination systems, a medium to low impact in psychophysical methods, and no significant impact in systems with confocality.

5.

Conclusions

Intraocular scattering was found to affect the estimations of MP optical density when using fundus reflectometry. We showed that, even for a group of young subjects with low amounts of straylight, the effect of scattering in the measurement of the MPOD using this method could be up to 16%. Our method can compensate for this effect and provide scatter-free values for the MPOD.

Disclosures

The authors declare that there are no conflicts of interest related to this article.

Acknowledgments

This research has been supported by the European Research Council Advanced Grant ERC-2013-AdG-339228 (SEECAT), Secretaría de Estado e Investigación, Desarrollo e Innovación (SEIDI) FIS2016-76163-R; Fundación Séneca-Agencia de Ciencia y Tecnología de la Región de Murcia (grant 19897/GERM/15), and the European Regional Development Fund (EU-FEDER).

References

1. 

D. Christaras et al., “Intraocular scattering compensation in retinal imaging,” Biomed. Opt. Express, 7 (10), 3996 –4006 (2016). https://doi.org/10.1364/BOE.7.003996 BOEICL 2156-7085 Google Scholar

2. 

H. Ginis et al., “The wide-angle point spread function of the human eye reconstructed by a new optical method,” J. Vis., 12 (3), 20 (2012). https://doi.org/10.1167/12.3.20 Google Scholar

3. 

D. Christaras et al., “Scattering contribution to the double-pass PSF using Monte Carlo simulations,” Ophthalmic Physiol. Opt., 37 (3), 342 –346 (2017). https://doi.org/10.1111/opo.12375 Google Scholar

4. 

J. J. Vos and T. J. T. P. Van Den Berg, “The large angle course of the disability glare function and its attribution to components of ocular scatter,” CIE Collect. Color Vision, 124 (2), 11 –29 (1997). Google Scholar

5. 

H. Ginis et al., “Wavelength dependence of the ocular straylight,” Invest. Ophthalmol. Visual Sci., 54 (5), 3702 –3708 (2013). https://doi.org/10.1167/iovs.13-11697 IOVSDA 0146-0404 Google Scholar

6. 

J. E. Coppens, L. Franssen and T. J. T. P. Van Den Berg, “Wavelength dependence of intraocular straylight,” Exp. Eye Res., 82 (4), 688 –692 (2006). https://doi.org/10.1016/j.exer.2005.09.007 EXERA6 0014-4835 Google Scholar

7. 

D. B. Elliott et al., “Relative sensitivity of clinical tests to hydrophilic lens-induced corneal thickness changes,” Optom. Vision Sci., 70 (12), 1044 –1048 (1993). https://doi.org/10.1097/00006324-199312000-00009 OVSCET 1040-5488 Google Scholar

8. 

K. M. Meek et al., “Transparency, swelling and scarring in the corneal stroma,” Eye (Lond.), 17 (8), 927 –936 (2003). https://doi.org/10.1038/sj.eye.6700574 12ZYAS 0950-222X Google Scholar

9. 

J. J. Vos, “Disability glare - a state of the art report,” CIE, 3 (2), 39 –53 (1984). Google Scholar

10. 

D. B. Elliott and M. A. Bullimore, “Assessing the reliability, discriminative ability, and validity of disability glare tests,” Invest. Ophthalmol. Visual Sci., 34 (1), 108 –119 (1993). IOVSDA 0146-0404 Google Scholar

11. 

D. M. Snodderly, J. D. Auran and F. C. Delori, “The macular pigment II. Spatial distribution in primate retinas,” Invest. Ophthalmol. Visual Sci., 25 (6), 674 –685 (1984). IOVSDA 0146-0404 Google Scholar

12. 

W. Stahl and H. Sies, “Antioxidant activity of carotenoids,” Mol. Aspects Med., 24 (6), 345 –351 (2003). https://doi.org/10.1016/S0098-2997(03)00030-X Google Scholar

13. 

E. Loane et al., “The rationale and evidence base for a protective role of macular pigment in age-related maculopathy,” Br. J. Ophthalmol., 92 1163 –1168 (2008). https://doi.org/10.1136/bjo.2007.135566 BJOPAL 0007-1161 Google Scholar

14. 

F. Khachik, P. S. Bernstein and D. L. Garland, “Identification of lutein and zeaxanthin oxidation products in human and monkey retinas,” Invest. Ophthalmol. Visual Sci., 38 (9), 1802 –1811 (1997). IOVSDA 0146-0404 Google Scholar

15. 

T. Berendschot, “Fundus reflectance—historical and present ideas,” Prog. Retinal Eye Res., 22 (2), 171 –200 (2003). https://doi.org/10.1016/S1350-9462(02)00060-5 Google Scholar

16. 

S. Beatty et al., “Macular pigment and age related macular degeneration,” Br. J. Ophthalmol., 83 (83), 867 –877 (1999). https://doi.org/10.1136/bjo.83.7.867 BJOPAL 0007-1161 Google Scholar

17. 

S. Beatty et al., “Macular pigment and risk for age-related macular degeneration in subjects from a Northern European population,” Invest. Ophthalmol. Visual Sci., 42 (2), 439 –446 (2001). IOVSDA 0146-0404 Google Scholar

18. 

J. S. L. Tan et al., “Dietary antioxidants and the long-term incidence of age-related macular degeneration. The blue mountains eye study,” Ophthalmology, 115 (2), 334 –341 (2008). https://doi.org/10.1016/j.ophtha.2007.03.083 OPANEW 0743-751X Google Scholar

19. 

B. R. Wooten and B. R. Hammond, “Macular pigment: influences on visual acuity and visibility,” Prog. Retinal Eye Res., 21 (2), 225 –240 (2002). https://doi.org/10.1016/S1350-9462(02)00003-4 Google Scholar

20. 

J. Loughman et al., “Macular pigment and its contribution to visual performance and experience,” J. Optom., 3 (2), 74 –90 (2010). https://doi.org/10.1016/S1888-4296(10)70011-X Google Scholar

21. 

T. T. J. M. Berendschot, “Imaging the macular pigment,” Medical Retina, 51 –68 Springer-Verlag, Berlin, Heidelberg, Germany (2010). Google Scholar

22. 

T. T. J. M. Berendschot and D. Van Norren, “Objective determination of the macular pigment optical density using fundus reflectance spectroscopy,” Arch. Biochem. Biophys., 430 149 –155 (2004). https://doi.org/10.1016/j.abb.2004.04.029 Google Scholar

23. 

F. C. Delori et al., “Macular pigment density measured by autofluorescence spectrometry : comparison with reflectometry and heterochromatic flicker photometry,” J. Opt. Soc. Am. A, 18 (6), 1212 –1230 (2001). https://doi.org/10.1364/JOSAA.18.001212 JOAOD6 0740-3232 Google Scholar

24. 

B. R. Wooten et al., “A practical method for measuring macular pigment optical density,” Invest. Ophthalmol. Visual Sci., 40 (11), 2481 –2489 (1999). IOVSDA 0146-0404 Google Scholar

25. 

D. M. Snodderly, “Macular pigment measurement by heterochromatic flicker photometry in older subjects: the carotenoids and age-related eye disease study,” Invest. Ophthalmol. Visual Sci., 45 (2), 531 –538 (2004). https://doi.org/10.1167/iovs.03-0762 IOVSDA 0146-0404 Google Scholar

26. 

I. F. Leung, “Macular pigment: new clinical methods of detection and the role of carotenoids in age-related macular degeneration,” Optometry, 79 266 –272 (2008). https://doi.org/10.1016/j.optm.2007.03.017 OVSCET 1040-5488 Google Scholar

27. 

H. Ginis et al., “Compact optical integration instrument to measure intraocular straylight,” Biomed. Opt. Express, 5 (9), 3036 (2014). https://doi.org/10.1364/BOE.5.003036 BOEICL 2156-7085 Google Scholar

28. 

D. Christaras, Light Scattering Phenomena in Vision and in Imaging, Universidad de Murcia, Murcia, Spain (2016). Google Scholar

29. 

P. K. Brown and G. Wald, “Visual pigments in human and monkey retinas,” Nature, 200 (4901), 37 –43 (1963). https://doi.org/10.1038/200037a0 Google Scholar

30. 

D. M. Snodderly et al., “The macular pigment. I. Absorbance spectra, localization, and discrimination from other yellow pigments in primate retinas,” Invest. Ophthalmol. Visual Sci., 25 (6), 660 –673 (1984). IOVSDA 0146-0404 Google Scholar

31. 

R. W. Knighton, S. G. Jacobson and C. M. Kemp, “The spectral reflectance of the nerve fiber layer of the macaque retina,” Invest. Ophthalmol. Visual Sci., 30 (11), 2393 –2402 (1989). IOVSDA 0146-0404 Google Scholar

32. 

R. W. Knighton and X. R. Huang, “Directional and spectral reflectance of the rat retinal nerve fiber layer,” Invest. Ophthalmol. Visual Sci., 40 (3), 639 –647 (1999). IOVSDA 0146-0404 Google Scholar

33. 

E. J. van Kampen and W. G. Zijstra, “Spectrophotometry of hemoglobin and hemoglobin derivatives,” Adv. Clin. Chem., 23 199 –257 (1983). https://doi.org/10.1016/S0065-2423(08)60401-1 Google Scholar

34. 

J. Loughman et al., “An evaluation of a novel instrument for measuring macular pigment optical density: the MPS 9000,” Acta Ophthalmol., 90 (2), e90 –e97 (2012). https://doi.org/10.1111/aos.2012.90.issue-2 Google Scholar

35. 

F. C. Delori and K. P. Pflibsen, “Spectral reflectance of the human ocular fundus,” Appl. Opt., 28 (6), 1061 –1077 (1989). https://doi.org/10.1364/AO.28.001061 APOPAI 0003-6935 Google Scholar

36. 

J. M. Gorrand, R. Alfieri and J. Y. Boire, “Diffusion of the retinal layers of the living human eye,” Vision Res, 24 (9), 1097 –1106 (1984). https://doi.org/10.1016/0042-6989(84)90088-9 VISRAM 0042-6989 Google Scholar

37. 

R. A. Bone and J. T. Landrum, “Macular pigment in henle fiber membranes: a model for Haidinger’s brushes,” Vision Res., 24 (2), 103 –108 (1984). https://doi.org/10.1016/0042-6989(84)90094-4 VISRAM 0042-6989 Google Scholar

38. 

P. J. Delint, T. T. Berendschot and D. van Norren, “Local photoreceptor alignment measured with a scanning laser ophthalmoscope,” Vision Res., 37 (2), 243 –248 (1997). https://doi.org/10.1016/S0042-6989(96)00118-6 VISRAM 0042-6989 Google Scholar

39. 

J. M. Gorrand and F. C. Delori, “Reflectance and curvature of the inner limiting membrane at the foveola,” J. Opt. Soc. Am. A, 16 (6), 1229 –1237 (1999). https://doi.org/10.1364/JOSAA.16.001229 JOAOD6 0740-3232 Google Scholar

40. 

R. A. Bone, “The role of the macular pigment in the detection of polarized light,” Vision Res., 20 (3), 213 –220 (1980). https://doi.org/10.1016/0042-6989(80)90105-4 VISRAM 0042-6989 Google Scholar

Biographies for the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Dimitrios Christaras, Alexandros Pennos, Harilaos Ginis, and Pablo Artal "Effect of intraocular scattering in macular pigment optical density measurements," Journal of Biomedical Optics 23(5), 056001 (9 May 2018). https://doi.org/10.1117/1.JBO.23.5.056001
Received: 23 January 2018; Accepted: 19 April 2018; Published: 9 May 2018