Open Access
1 July 2003 Fiber dispersion in time domain measurements compromising the accuracy of determination of optical properties of strongly scattering media
Author Affiliations +

1.

Introduction

The determination of optical properties, in particular the absorption coefficient μa (measured at different wavelengths) and the reduced scattering coefficient μs , of strongly scattering tissue is one of the main tasks in biomedical optics.1 2 3 4 5 Various methods for measuring these optical properties of tissues have been reported. Integrating spheres were used for in vitro investigation of diffusely reflected and transmitted light with isolated tissue samples.6 Noninvasive determination of in vivo optical properties was achieved by analysis of diffusely reflected or transmitted continuous wave (c.w.) light using several source–detector separations.7 Improved in vivo measurements of optical properties of tissues were reported using intensity-modulated light.8 The most advanced method for noninvasive in vivo determination of optical properties is based on time-resolved measurement of picosecond light pulses transmitted through or diffusely reflected from the tissue.9 10 The temporal spreading of the light pulses contains information about the optical coefficients to be determined.

In practice, time-resolved measurements are often performed by time-correlated single-photon counting (TCSPC), measuring distributions of times of flight of photons (DTOF) through the tissue. Optical coefficients are determined by fitting a theoretical DTOF to experimental data. It is well known, however, that generally the instrumental response of the measuring system cannot be neglected; i.e., the experimental DTOF does not represent the true impulse response of the tissue. Thus deconvolution of the measured DTOF with the overall instrumental response is required to obtain the true optical parameters. Alternatively, the theoretical curve can be convolved with the instrumental response before fitting the data. In either case, it is crucial for the accuracy of the analysis that the instrumental response function must be known as accurately as possible.11

It is the purpose of this paper to analyze the influence of temporal dispersion effects in the detection fibers or fiber bundles—as usually employed—on the instrumental response, and to investigate the consequences of these effects on the accuracy of the optical properties evaluated. In particular, a method for the proper determination of the response function as suggested recently by Schmidt et al.;12 is investigated in more detail. Furthermore, we will discuss how fiber dispersion may affect the results of frequency-domain experiments as well. It is worth noting that the temporal dispersion effects discussed in the following pages are some of the most dominant contributions to the overall instrumental response, in particular if longer (⩾2 m) fibers with a high numerical aperture (e.g., NA=0.54) are used, as is often the case.

2.

Experimental

The optical setup is illustrated in Fig. 1. A diode laser (PDL 800 system from PicoQuant GmbH, Germany) emitted pulses of 100-ps duration at a repetition frequency of 64 MHz and at a wavelength of 687 nm. The mean power of the laser light was 0.75 mW and a beam 1 mm in diameter was directly illuminating the phantom.

Figure 1

Experimental setup.

026303j.1.jpg

The light transmitted through the phantom was collected opposite the source point and routed into the detection optics by a fiber bundle. Four commercially available fiber bundles with different properties were investigated. The bundles were constructed from multimode glass fibers with a typical diameter of 70 μm. Their lengths and numerical apertures are summarized in Table 1. The overall transmission of light by the bundles varied between 50 and 60.

Table 1

Properties of fiber bundles investigated.
Bundle # Manufacturer Diameter
mm
Length
m
NA
1 Schoelly 5 1.5 0.22
2 Loptek 4 1.5 0.54
3 Loptek 4 3 0.54
4 Schott 5 3.6 0.54

To study the influence of the numerical aperture on the accuracy of the optical properties determined in more detail, the effective numerical aperture of the detecting setup was varied with the help of a diaphragm located in front of the detection optics. A metal package photomultiplier tube (PMT) (R7400U-02, Hamamatsu Photonics GmbH, Germany) was used as a photon counting detector. The exit face of the fiber bundle was imaged onto the multiplier photocathode surface by two aspherical lenses (f 1=37 mm, d 1=52 mm; f 2=38.5 mm, d 2=50 mm). The output pulses of the photodetector were amplified and distributions of the times of flight of photons were measured by time-correlated single-photon counting (SPC 300, Becker & Hickl GmbH, Germany). Distributions of times of flight were accumulated in 1024 channels, each 20.8 ps in width. The data acquisition times were typically 10 s for phantom measurements and 1 s for response measurements, respectively. Neutral-density filters were inserted to obtain count rates on the level of 800 kHz for a fully opened diaphragm.

The phantom consisted of a polymer host containing monodispersed quartz glass spheres with a diameter of 0.5 μm as scatterers and an IR absorbing dye as absorber. The shape of the phantom was close to that of a slightly compressed female breast, i.e., a slab of 4 cm thickness with rounded edges, as used for optical mammography.13 The measurements were performed on a spot where lateral distances to the edges of the phantom were not less than 3 cm. The refractive index of the phantom was 1.55.

Several methods to determine the response function of the setup were employed. First, the so-called direct response was measured by directly illuminating the tip of the detecting bundle with the laser beam in a setup similar to that proposed by Ntziachristos et al.;14 Second, the response was measured after placing thin diffusers in front of the detection fiber bundle. Two kinds of diffusers were employed for this purpose: a sheet of white paper (100 μm thickness, 80 g/m2) and a milky white opal coating (450 μm thick) flashed onto one surface of a glass plate (Edmund Optics).

The optical properties of the phantom were evaluated by fitting the photon flux density obtained from the solution of the diffusion equation15 16 for an infinite slab to the measured distributions of the times of flight of photons. The extrapolated boundary conditions were applied.17 The instrumental response was taken into account by performing convolution of the theoretical curve with the response in each step of the fitting procedure. In practice, only that part of the measured DTOF between the 5 levels of the maximum value was fitted. Time zero was fixed according to response measurements. The DTOFs were corrected for the differential nonlinearity of the TCSPC electronics.

3.

Results

The inherent instrumental response function as limited solely by the laser pulse duration, the photodetector, and the TCSPC electronics was determined as 170 ps FWHM. The influence of the effective numerical aperture of the detector setup, including the fiber bundle on the response, is shown in Fig. 2. It is important to note that a thin, strongly scattering medium (white paper) was placed in front of the detection fiber bundle to ensure excitation and propagation of all higher transverse modes within the NA of the fibers. The paper was thin enough so that pulse broadening by scattering in the paper itself could be neglected. In measurements reported by other authors,18 the mean time of flight was about 5 ps at a wavelength of 700 nm using a similar sheet of paper (with a basis weight of 90 g/m2). The use of a thin diffuser leads to much stronger dispersion effects than the direct illumination of the entrance face of the fiber or fiber bundle, as will be discussed later. As can be seen from Fig. 2, the response is almost unaffected if a low effective NA is used, whereas a broadening of the response up to an FWHM of 1 ns was observed with an increase of the effective NA.

Figure 2

Instrumental response measurements using various effective numerical apertures obtained by changing the diaphragm diameter. The measurements were performed with the Schott bundle (NA=0.54, l=3.6 m), illumination with short (≈100 ps) input laser pulses, and a thin paper in front of the entrance face.

026303j.2.jpg

The observed time spread inside the fibers is well known to result from different effective propagation speeds of photons for various transverse modes within the (effective) NA. It can be derived from the group delay19 as

Eq. (1)

Δt=nlc0 (11NA2n21),
where n is the refractive index of the core material, l the length of the fiber bundle, and c0 the speed of light in vacuum. Assuming a glass fiber (bundle) with l=3.6 m, NA=0.54, and n=1.5, the highest-order mode is delayed by 1.3 ns in comparison with the lowest-order mode. Assuming the same number of photons in all modes within NA, this leads to a rectangular pulse of the width 2σ=0.75 ns (where σ2 is the variance of the pulse), which is in good agreement with experimental observations. The temporal dispersion caused by imperfect imaging with the aspherical lens setup was estimated to be less than 30 ps, and the dispersion caused by different lengths of individual fibers in the fiber bundle (assuming <0.1) is less than 20 ps for l=3.6 m; i.e., both can be neglected in comparison with mode dispersion.

The reduction of intensity with decreasing diameter of the diaphragm (i.e., decreasing effective NA) is shown in Fig. 3. For these measurements the Schott fiber bundle (l=3.6 m) with a high NA (0.54) was used. If a scattering medium (phantom, paper, or opal glass) is placed in front of the bundle, the whole NA is filled by light and the diaphragm starts to cut off intensity at a diameter of about d=26 mm, which corresponds to an effective NA of 0.54 (which is identical with the NA of the bundle as provided by the manufacturer). Without a scattering medium (i.e., direct illumination), mainly low-order modes propagate and the cutoff in intensity appears at a diaphragm diameter of d=14 mm, which corresponds to an effectively illuminated NA of about 0.32.

Figure 3

Normalized intensity and width of the pulses (expressed as double the standard deviation 2σ) versus the diameter of the diaphragm measured using various scatterers in front of the entrance face of the Schott bundle (NA=0.54, l=3.6 m). Note that for small diameters of the diaphragm (d<7 mm), the width of the response is not affected by temporal dispersion of the fiber bundle, i.e., is given by the inherent response (laser pulse duration and detector).

026303j.3.jpg

The results concerning the influence of the effective numerical aperture on the width 2σ of the measured pulses without a scattering medium (direct response) and using a sheet of paper, opal glass, or a thick phantom as scatterers, respectively, are summarized in Fig. 3 as well. Whereas for the direct response the pulse width is almost unaffected, the measurements with paper and opal glass resemble an increase in width from opening the NA. Apart from the thick, scattering breast phantom, the influence of the instrumental response on the width of the DTOF is clearly visible.

This influence has consequences for the calculation of optical properties, as can be seen in Fig. 4. If the correct instrumental response—measured with the help of a thin diffuser—is used for convolution of the theoretical curve to fit the experimental data, the calculated absorption as well as the reduced scattering coefficients almost do not depend on the fiber bundle employed or on the effective NA, etc. If, however, the narrow instrumental response from the direct measurement is used instead, the optical properties are strongly overestimated. These problems obviously do not occur with short bundles of low NA.

Figure 4

Deviation of optical properties from their correct values a=0.014 cm −1, μs =8.5 cm −1) versus the diameter of the diaphragm measured with different bundles for direct (closed symbols) and diffused (open symbols) responses. The following symbols refer to measurements obtained with the use of different fiber bundles ▪, □— Scholly bundle, NA=0.22, l=1.5 m; •, ○—Loptek, NA=0.54, l=1.5 m; ▴, ▵—Loptek, NA=0.54, l=3 m; ▾, ▿—Schott, NA=0.54, l=3.6 m.

026303j.4.jpg

In principle, the discussed dispersion effects can be reduced by using graded-index instead of step-index fibers. However, the cladding in such fibers is relatively thick and the optical filling factor of bundles built from graded-index fibers is inferior to that of step-index fiber bundles, causing considerable reduction in the light collection efficiency.

The observed dispersion effects are not only present in time-resolved measurements but are also important for frequency-domain techniques because the phase shift of the modulated light intensity depends on temporal dispersion within the detection bundle as well. This was verified by performing a zero-phase measurement using a frequency-domain spectrometer (ISS Inc., Champaign, Illinois) at a modulation frequency of 110 MHz, with and without a thin diffuser in front of the detection bundle. The phase shift was measured and the mean time of flight was evaluated for a 3-m-long detection bundle with an NA=0.46 at four wavelengths: 750, 780, 810, and 830 nm. The differences between mean times of flight for diffuse and direct zero-phase measurements ranged between 610 and 700 ps. Consequently, proper zero-phase measurements with the help of thin diffusers are rather important for absolute measurements with frequency-domain devices as well.

4.

Conclusions

Limitations in the accuracy of optical properties determined by time-resolved transmission measurements of short laser pulses through strongly scattering media have been investigated with respect to the numerical aperture and length of the optical detection fibers or fiber bundles employed. It was found that a remarkable broadening of the instrumental response as a result of modal dispersion [even for short l=1.5-m bundles of high (0.54) NA] may influence considerably the optical properties determined. Consequently, the overall time resolution must be measured properly using a thin diffusing medium (e.g., white paper or opal glass) in front of the detecting bundle to fill the whole NA of the fiber in the same manner as during tissue measurements. The results obtained reveal that the use of such a thin diffuser allows all modes to be filled uniformly and provides an excellent way of measuring the response function properly. In contrast, (incorrect) response measurements without such a diffuser resulted in an overestimation of optical properties up to 70 (90) for μs a) for a 3.6-m-long fiber bundle with an NA=0.54.

Acknowledgments

The study was partially supported by the project Optical Mammography: Imaging and characterisation of breast lesions by pulsed near-infrared laser light (OPTIMAMM) funded by the European Union (contract QLG1-2000-00690). A.L. was financially supported by a Marie-Curie Fellowship from the European Community (Quality of Life Program, contract QLGA-CT-2000-52128) and gratefully acknowledges the hospitality of the Division of Medical Physics and Metrological Information Technology of the Physikalisch-Technische Bundesanstalt Berlin-Charlottenburg. The authors would like to thank Mr. Kamil Uludag (Department of Neurology, Charite´ Hospital, Berlin) for collaboration in the frequency-domain experiments.

REFERENCES

1. 

E. M. Sevick , B. Chance , J. Leigh , S. Nioka , and M. Maris , “Quantitation of time- and frequency-resolved optical spectra for the determination of tissue oxygenation,” Anal. Biochem. , 195 330 –351 (1991). Google Scholar

2. 

B. C. Wilson , M. S. Patterson , and D. M. Burns , “Effect of photosensitizer concentration in tissue on the penetration depth of photoactivating light,” Lasers Med. Sci. , 1 235 –244 (1986). Google Scholar

3. 

V. Quaresima, R. Sfareni, A. Pizzi, and M. Ferrari, “Measurement of the muscle optical properties on muscular dystrophy patients by a frequency-domain photometer,” in Biomedical Optical Spectroscopy and Diagnostics, E. Sevick-Muraca and D. Benaron, Eds., pp. 123–125, Vol. 3 of OSA Trends in Optics and Photonics Series, Optical Society of America, Washington, DC (1996).

4. 

G. Gratton , P. M. Corballis , E. Cho , M. Fabiani , and D. C. Hood , “Shades of grey matter: noninvasive optical images of human brain responses during visual stimulation,” Psychophysiology , 32 505 –509 (1995). Google Scholar

5. 

M. Kohl , M. Essenpreis , and M. Cope , “The influence of glucose concentration upon the transport of light in tissue-simulating phantoms,” Phys. Med. Biol. , 40 1267 –1287 (1995). Google Scholar

6. 

T. L. Troy and S. N. Thennadil , “Optical properties of human skin in the near infrared wavelength range of 1000 to 2200 nm,” J. Biomed. Opt. , 6 167 –176 (2001). Google Scholar

7. 

T. J. Farrell , M. S. Patterson , and B. Wilson , “A diffusion theory model of spatially resolved, steady-state diffuse reflectance for the noninvasive determination of tissue optical properties in vivo,” Med. Phys. , 19 879 –888 (1992). Google Scholar

8. 

S. Fantini , M. A. Franceschini-Fantini , J. S. Maier , S. A. Walker , B. Barbieri , and E. Gratton , “Frequency-domain multichannel optical detector for noninvasive tissue spectroscopy and oximetry,” Opt. Eng. , 32 34 –42 (1995). Google Scholar

9. 

R. Cubeddu , A. Pifferi , P. Taroni , A. Torricelli , and G. Valentini , “A solid tissue phantom for photon migration studies,” Phys. Med. Biol. , 42 1971 –1979 (1997). Google Scholar

10. 

R. Cubeddu , A. Pifferi , P. Taroni , A. Torricelli , and G. Valentini , “Compact tissue oximeter based on dual-wavelength multichannel time-resolved reflectance,” Appl. Opt. , 38 3670 –3680 (1999). Google Scholar

11. 

V. Ntziachristos, “Concurrent diffuse optical tomography, spectroscopy and magnetic resonance imaging of breast cancer,” PhD Thesis, Univ. of Pennsylvania (2000).

12. 

F. E. W. Schmidt , M. E. Fry , E. M. C. Hillman , J. C. Hebden , and D. T. Delpy , “A 32-channel time-resolved instrument for medical optical tomography,” Rev. Sci. Instrum. , 71 256 –265 (2000). Google Scholar

13. 

D. Grosenick , H. Wabnitz , and H. Rinneberg , “Time-resolved imaging of solid phantoms for optical mammography,” Appl. Opt. , 36 221 –231 (1997). Google Scholar

14. 

V. Ntziachristos , X. H. Ma , and B. Chance , “Time-correlated single photon counting imager for simultaneous magnetic resonance and near-infrared mammography,” Rev. Sci. Instrum. , 69 4221 –4233 (1998). Google Scholar

15. 

M. S. Patterson , B. Chance , and B. C. Wilson , “Time resolved reflectance and transmittance for the noninvasive measurement of tissue optical properties,” Appl. Opt. , 28 2331 –2336 (1989). Google Scholar

16. 

S. R. Arridge , M. Cope , and D. T. Delpy , “The theoretical basis for the determination of optical path lengths in tissues: temporal and frequency analysis,” Phys. Med. Biol. , 37 1531 –1560 (1992). Google Scholar

17. 

R. C. Haskell , L. O. Svaasand , T.-T. Tsay , T.-C. Feng , M. S. McAdams , and B. J. Tromberg , “Boundary conditions for the diffusion equation in radiative transfer,” J. Opt. Soc. Am. A , 11 2727 –2741 (1994). Google Scholar

18. 

J. Carlsson , P. Hellenyin , L. Malmqvist , A. Persson , W. Persson , and C-G. Wahlstrom , “Time resolved studies of light propagation in paper,” Appl. Opt. , 34 1528 –1535 (1995). Google Scholar

19. 

M. Young, Optics and Lasers: Including Fibers and Optical Waveguides, Springer-Verlag, Berlin (1992).
©(2003) Society of Photo-Optical Instrumentation Engineers (SPIE)
Adam Liebert, Heidrun Wabnitz, Dirk Grosenick, and Rainer Macdonald "Fiber dispersion in time domain measurements compromising the accuracy of determination of optical properties of strongly scattering media," Journal of Biomedical Optics 8(3), (1 July 2003). https://doi.org/10.1117/1.1578088
Published: 1 July 2003
Lens.org Logo
CITATIONS
Cited by 84 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical properties

Scattering

Dispersion

Diffusers

Glasses

Picosecond phenomena

Tissue optics

Back to Top