The Origins Space Telescope is one of four flagship missions under study for the 2020 Astrophysics Decadal Survey. With a 5.9-m cold (4.5 K) telescope deployed from space, Origins promises unprecedented sensitivity in the near-, mid-, and far-infrared from 2.8 to 588 μm. This mandates the use of ultrasensitive and stable detectors in all of the Origins instruments. At the present, no known detectors can meet Origins’ stability requirements in the near- to mid-infrared or its sensitivity requirements in the far-infrared. We discuss the applicability of transition-edge sensors, as both calorimeters and bolometers, to meet these requirements, and lay out a path toward improving the present state-of-the-art. |
1.IntroductionThe Origins Space Telescope (Origins) traces our cosmic history, from the formation of the first galaxies and the rise of metals to the development of habitable worlds and present-day life. Origins does this through exquisite sensitivity to infrared radiation from ions, atoms, molecules, dust, water vapor and ice, and observations of extrasolar planetary atmospheres, protoplanetary disks, and large-area extragalactic fields. Origins operates in the wavelength range of 2.8 to and is more than 1000 times more sensitive than its predecessors due to its large, cold (4.5 K) telescope and advanced instruments. A complete description of Origins, its science goals, and its instrumentation is provided by the Origins Space Telescope Mission Concept Study Report.1 We reproduce some key details here. Origins has three instruments, each with a variety of observing modes. The Origins Survey Spectrometer (OSS) is designed to perform extragalactic surveys of far-infrared (FIR) line emission from 25 to , probing galaxy evolution out to . OSS uses a series of grating modules to disperse incident light into six logarithmically spaced bands, each with spectral resolving power . A Fourier transform spectrometer (FTS) can be inserted into the optical path to achieve , and the FTS can be used in conjunction with a scanning etalon to achieve in the 100 to band. In order to make background-limited observations of FAR spectral lines at , OSS requires instrument sensitivity of in an hour of observing at .2 This corresponds to a required detector noise equivalent power (NEP) of and per-pixel saturation power of to 0.2 fW, a level of sensitivity that has yet to be achieved in an FIR detector. OSS has six focal planes, each with . The Far-Infrared Imager Polarimeter (FIP) instrument will perform wide-field photometric surveys of astrophysical objects, bridging the gap between observations by the James Webb Space Telescope and the Atacama Large Millimeter/submillimeter Array. FIP has polarized imaging capability at both 50 and . It requires , a per-pixel NEP of , and per-pixel saturation power of to 20 fW. The Mid-Infrared Spectrometer Camera Transit spectrometer (MISC-T) will make spectral measurements of transiting exoplanets in the 2.8 to band with exquisite photometric precision, looking for biosignatures in the atmospheres of Earth-like exoplanets that orbit M stars. MISC-T uses a series of grisms to disperse light into three bands. The short-wavelength bands (2.8 to and 5.5 to ) will have to 100 and require stability over 60 transits, and the long-wavelength band (11 to ) will have to 295 and requires ppm stability over 60 transits. Bright targets of K magnitude will be observed, leading to photon arrival rates of up to a few hundred thousand photons per pixel per second. Detector systems that meet these stringent stability requirements in the MISC-T band do not exist today. The choice of detector used for each instrument will have a tremendous impact on the development, construction, and eventual operation of Origins. The sensitivity requirements of the FIR instruments mandate the use of sub-Kelvin detectors. There are several detector technologies that have the potential to meet the requirements of the FIR instruments. Among them are microwave kinetic inductance detectors (MKIDs), quantum capacitance detectors (QCDs), and transition-edge sensors (TESs), the subject of this paper. MKIDs and QCDs will be discussed elsewhere in this issue. The MISC-T instrument’s stability requirement does not mandate the use of a low-temperature detector. As a result, the baseline detectors under consideration for MISC-T (HgCdTe arrays for wavelengths shorter than and Si:As arrays for wavelengths longer than ) do not require sub-Kelvin coolers. They operate at and , respectively. Neither detector type has shown that it can meet the stability requirements of MISC-T, so, in this paper, we also present a TES calorimeter option for MISC-T. TES bolometers are also considered as candidates for MISC-T, but we argue that operation in a calorimetric mode provides an ideal choice. This paper is organized as follows. In Sec. 2, we introduce the TES and its most common implementations: calorimeters and bolometers that use a superconductor as a resistive thermometer. Then, we present a path toward meeting the requirements of each Origins instrument using TESs. In Sec. 3, we present the use of a photon-counting TES calorimeter for MISC-T. In Sec. 4, we discuss how TES bolometers can be designed to meet the requirements of OSS and FIP. Since no detector systems currently satisfy Origins’ requirements, we describe how TESs can. Our conclusions are in Sec. 5. 2.Transition-Edge SensorsTES detectors measure incident power or energy using the temperature dependence of a superconductor’s resistance. First realized by Andrews et al.,3 TESs are now the most sensitive detectors to radiation from the gamma-ray through the millimeter wave. TESs can operate as calorimeters or bolometers. A calorimeter is designed to measure discrete depositions of energy, and a bolometer is designed to measure quasistatic power dissipated by a flux of photons. Both implementations share the same basic architecture. Figure 1 shows a cartoon of a TES and its basic readout circuit. The detector consists of a thermal mass of heat capacity at a temperature , isolated from a thermal bath at temperature across a thermal link with conductance . The thermometer—a superconductor biased in its transition—measures the temperature of the isolated thermal mass, which can consist of electrons or electrons and phonons. The device is operated at near-constant voltage bias to stabilize the device in the superconducting transition via negative electrothermal feedback.4 A shunt resistor whose resistance is much smaller than the TESs at its operating point is typically used to achieve voltage bias. The shunt resistance value, in conjunction with the device and bias circuit inductance, sets the electrical time constant. The TES is readout/multiplexed out using one or more superconducting quantum interference device (SQUID) amplifiers. A range of SQUID multiplexing options currently exists for TES readout [e.g., time-division multiplexing (TDM),5 frequency-division multiplexing,5 microwave SQUID multiplexing (),6 and code-division multiplexing7]. Many TDM schemes are deployed in the field for reading out kilopixel-scale arrays of TES bolometers. Providing more bandwidth per pixel, microwave multiplexing shows the most promise for reading out large arrays of TES calorimeters; the Lynx mission is baselining arrays of TES calorimeters that are read out with . For simplicity, the thermal conductance and heat capacity are commonly assumed to be independent of temperature over the range of operation, however, extension of this limiting case is necessary to model the full dynamic range and temporal response of the device in practical settings.8 TESs are amendable to production entirely by available microfabrication techniques and are suitable for realization as large-format detector arrays (e.g., see, for example, Refs. 910.11.–12). Both TES bolometers and TES calorimeters have achieved measured performance consistent with a linear near-equilibrium thermodynamics model of the system, with no hidden variables or unaccounted-for noise sources. This TES model is a powerful tool used to design detectors that meet the combined requirements for an application. It gives confidence in the ability to predict performance, help identify sources of unwanted characteristics, and make necessary changes that accelerate development programs. 3.TES CalorimetersTES calorimeters are designed to measure discrete depositions of energy. In this case, when an incident energetic particle of energy is absorbed by the TES, the temperature of the detector rises by . This sudden temperature increase yields a corresponding increase of the superconductor’s resistance, shunting current through the shunt resistor, and reducing the current flowing through the SQUID input coil. The device then relaxes to its steady-state temperature with an exponential decay time constant , where accounts for electrothermal feedback gain.4 This process yields a pulse in the time domain (Fig. 5). In the x-ray band, where TES calorimeters are most commonly deployed, typical time constants are . TES calorimeters designed for the near- to mid-infrared have time constants faster than .13 TES calorimeters are most commonly operated linearly, where the pulse height is proportional to the deposited energy. In this way, each TES pixel is a spectrometer, and an array of TES calorimeters is an imaging spectrometer on a chip (i.e., an integral field spectrograph that does not need dispersive optics). The energy scale of linear device operation is set by the heat capacity; the saturation energy—the maximum deposited energy in the linear regime—is defined as , where is the logarithmic temperature sensitivity of the device. With nonlinear data analysis techniques, TES calorimeters can also be effectively operated nonlinearly with little negative impact on device performance.14,15 TES calorimeter development for astronomy applications has concentrated in the development of x-ray calorimeter arrays. The x-ray work, along with complementary development of bolometers for microwave applications, has led to significant advancement in TES understanding and performance. In this band, TESs have achieved the highest resolving power of any nondispersive spectrometer, measuring the energy of photons to better than a part in 3400.16 Some notable advancements in both TES understanding and performance include: (1) identification of sources of excess noise in TESs17–22; (2) identification of the resistive mechanisms in TESs and ways to control the shape of the resistive transition surface23–29; (3) improved understanding of energy losses from athermal phonons and quasiparticle excitations30; (4) improved fabrication methods and understanding of the thermal conductance of membranes and leg structures31–34; (5) improved coupling to radiation at longer wavelengths using tuned optical stacks13,35; and (6) improved signal processing including methods for nonlinear signals.14,15,36,37 Fundamental thermodynamic noise limits a TES calorimeter’s achievable energy resolution . For a TES, the known thermodynamic fluctuations are associated with electrical resistance (Johnson noise in the TES resistance and in the bias shunt resistor ) and thermal impedance (phonon noise across the thermal link that couples the sensor to the bath). The expression for simplifies to a compact form18 under the assumptions of negligible amplifier noise, negligible shunt resistor Johnson noise, and large loop gain: Here, and are the temperature of the TES and bath, respectively, is a thermal exponent, describing the power through the thermal link , is the total heat capacity, and and are both dimensionless parameters, characterizing the sensitivity of the resistive transition to changes in temperature and current, respectively. More precisely, and are defined as the logarithmic derivative of the resistance with respect to temperature and current, respectively: and . The spectral resolving power is the ratio of the photon energy to the energy resolution: (for of a few or more, ). A detector with smaller is a higher resolution detector. The work by Bandler et al.38 and Iyomoto et al.21 are early examples that demonstrate a TES calorimeter’s measured noise spectrum and energy resolution agrees with hose calculated from the model thermodynamic system [Eq. (1)]. Assuming near-equilibrium thermodynamics and modeling the calorimeter system as one heat capacity connected through a single thermal conductance to a heat bath at temperature , the calculated energy resolution departs from the measured value by less than 30%. If rather than the TES resistance being ohmic, it is assumed to be near-equilibrium with a linear current dependence, then the first order nonequilibrium correction term to the Johnson noise is shown to account for all the measured excess noise at the operating point. Measurements providing inputs to Eq. (1) include impedance measurements and current–voltage () curves at different bath temperatures and bias currents, and measurements of the natural decay time of a pulse with no electrothermal feedback (). Together with Eq. (1), these measurements predicted an energy resolution . The measured energy resolution fitting to the Mn kα complex at 5.9 keV agrees by giving . TES calorimeters have been designed to give high detection efficiency and constant energy resolution over a large spectral range. In the visible, TES calorimeters exhibited uniform energy resolution from 0.3 eV to .39 TES calorimeters developed for quantum information science applications have achieved detection efficiencies of 99%, the highest of any detector in the optical.13,35 3.1.Photon-Counting TES for the Mid-InfraredIn this section, we describe how a TES operated as a calorimeter, instead of a bolometer, can be used for the MISC-T instrument, reducing the requirements on instrument stability while maximizing effective throughput and overall observing time. A bolometer is unable to distinguish various external factors from the optical signal and therefore puts more stringent requirements on control of instrumental systematics, stability of the system, and the length of time over which such stability must be maintained. Compared to a calorimeter, a bolometer is more negatively impacted by system drift, stray power coupling into the detector, changes in ambient operating conditions, including the local magnetic field environment, low frequency () noise, and array nonuniformities. These effects manifest as a shift or drift in the baseline signal, the measured current flowing in the TES. Because the drift can be confused with an optical signal (Fig. 2), the burden to account for any drifts rests on the instrument as a whole. Especially as bolometers become more sensitive and therefore slower, commonly employed techniques to push the optical signal into a higher frequency band (e.g., by chopping) are more difficult to execute. By contrast, as long as the intrinsic resolution of the detector is high enough, this shift or drift in the baseline signal has a negligible impact on a calorimeter’s ability to detect a photon event. Moreover, photon-counting observations are more efficient. No external modulation is required and no observation time is spent observing a calibrator. From a sensitivity perspective, it has been shown in several previous studies that TES calorimeters can be realized with sufficient intrinsic energy sensitivity to count THz photons.40,41 A TES dark count analog occurs when no photon is incident upon the detector and an anomalous noise trace triggers an event acquisition and application of the optimal filter to this noise record results in the energy of an in-band photon. One finds that if you have a TES in its steady-state operating bias collecting noise records at 50 MHz continuously in the dark with an energy-resolving power of 4 for the lowest photon energy of the band of interest, the dark count event rate is of order one per age of the universe. We use this criterion for a “noiseless” detector (effectively no detector false positives). In Fig. 3, we show an example for conditions that are just barely satisfying our criterion for noiseless photon counting with . In red and black are TES current time series plotted for one photon event with (red) and a corresponding record for the TES at its operating point with a no photon record (black). To the right is the distribution of many such records after application of an optimal filter to extract the energy of each record. We see even for only the population of 0 and 1 photon records are well separated. The likelihood of mistaking a noise record for a photon record is extremely low. This can also be seen as the 0 photon distribution at an energy of 1 is very small. As increases beyond , the dark count rate remains negligible. Since the energy resolution is constant, as the photon energy increases, the height of the pulse signal proportionally increases, leading to a proportionally higher measurement. The energy sensitivity in the X-ray is used to determine the energy of an incident photon with high accuracy. In this application, the high energy sensitivity of a TES can be used to push noiseless single-photon detection down to lower photon energies and into the MIR for Origins. The combination of high speed, noiseless single-photon detection, and sensitivity is used to unambiguously distinguish in-band photon events from noise and identify and remove out-of-band events caused by cosmic rays or background. In grating-based dispersive spectrometers, the energy sensitivity also enables rejection of higher-order photons that correspond to a different spectral channel, a capability unique to energy-resolving single-photon detectors and not found in other single-photon technologies [e.g., electron-multiplying charge-coupled devices,42 superconducting nanowire single-photon detectors,43 avalanche photodiodes,44 quantum dot detectors,45 etc.]. For the MISC-T instrument, the TES calorimeter design for the lowest photon energy of the band () is most challenging. It requires the lowest for noiseless photon counting while also needing the largest absorbing area for an absorber-coupled strategy [see Sec. 4.3; is sufficient46,47]. We therefore focus our discussion on the detector, which uses a resistive Bi absorber coupled to a TES. We find applying the TES model to our MISC-T TES design achieves the required photon count rates, sensitivity, and absorption efficiency with known materials parameters. In Fig. 4, we plot resolving power for versus over a range of typical values. We find that internal thermal fluctuation noise (ITFN) in the absorber is not negligible and departs from the simple analytic single body expression in Eq. (1). The ITFN limits sensitivity and shows that increasing much above 20 provides a limited improvement in sensitivity. The key take away is that even accounting for ITFN, the sensitivity is greater than the photon-counting criterion (over six times larger). This gives significant sensitivity margin even at the lowest photon energy of the MISC-T band. Noiseless photon counting becomes easier (even greater margin) for the rest of the MISC-T band. The thermal recovery time of TES calorimeters designed for the optical is typically of order , accommodating count rates of million counts per second.35,49–51 During the thermal recovery time, the TES is still able to receive and detect a photon and therefore has no dead time, a feature unique among single-photon detectors. If an additional absorption event occurs during the detector, recovery signal pile-up occurs. In such a case, the ability to extract the energy of each event is degraded below the maximum resolving power of the detector. For spectroscopy applications with no dispersive optics, such records are tagged as lower resolution events and different processing is employed to extract the event energies. For MISC-T, where the TESs are distributed over a spectral channel, pulse pile-up can be tolerable and the single-photon events are identified by the fast rise of the leading edge provided the detector stays below the saturation energy. Even for , we see in Fig. 4 that the sensitivity is over six times larger than required to photon count and an increase in rate will not impact the ability to count photons. The energy sensitivity margin increases further as the wavelength decreases. As the rate increases further, the calorimeter operation begins to become more bolometer-like and the advantages of photon counting start to diminish. Increasing the number of spectral channels increases the spectral information of the measurement and decreases the photon rate per channel. Increasing the event rate per spectral channel much above 1 million counts per second is achievable by: (1) decreasing the pulse recovery time, (2) increase the number of TES pixels per spectral channel (oversampling the point spread function), and (3) implementing photon counting algorithm for high count rates. The faster pulse recovery time is achievable by operating the TES at higher temperatures (see Fig. 5(b)) or by engineering a TES’s thermal coupling. Calkins et al.51 have demonstrated engineering of the thermal coupling of the TES to achieve a thermal recovery time of less than half a microsecond. In Fig. 5, we show the MISC-T TES response to a (60 meV) photon. In this TES model simulation, 60 meV of energy is deposited at nine different locations in a high-efficiency absorber. The simulation on the left is at 70 mK and on the right 100 mK. By raising the temperature, we increase the speed of the detector. The faster pulses at higher temperatures show positional dependence of absorption, which impacts the energy resolution. But even here, the sensitivity is sufficient for noiseless photon counting with margin. The maximum count rate expected for MISC-T is photons per second incident on the detector,1 thus, we do not expect as-demonstrated device speeds to be limiting. 4.TES BolometersTES bolometers operate similarly to TES calorimeters, but as quasi-steady-state devices, where power balance is achieved between sources of power dissipation (Joule heating of the sensor and incident radiation) and heat flow out the thermal link. Instead of being sensitive to single-photon events, bolometers are sensitive to a quasi-steady flux of photons absorbed by the detector and manifested as deposited power. Bolometric operation of the TES has many of the traditional advantages of a transducer with large negative feedback, including improved linearity, device speed, and immunity to parameter nonuniformity. When operated under negative electrothermal feedback,52 TES bolometers are highly linear devices. Thermal conductance to the bath sets the device saturation power and upper range of linear operation as a detector. TES bolometers are the most sensitive detectors used to measure millimeter and submillimeter radiation. For applications requiring background-limited sensitivity, they are also the most widely deployed, with arrays of thousands of pixels operating from ground (e.g., BICEP/Keck53), aircraft (e.g., SOFIA/HAWC+12), and balloon (e.g., SPIDER54 and EBEX55) platforms. Due to both their high sensitivity and the comparative ease of reading-out large arrays, TES bolometers have largely replaced bolometers that use high-impedance thermistors as thermometers. At the present, however, no bolometers have demonstrated the ultimate sensitivity required for Origins. In the following, we discuss what dictates a bolometer’s sensitivity, the state-of-the-art achieved, and some of the strategies being employed and natural tradeoffs that occur when designing a bolometer to meet Origins requirements. These include thermal isolation techniques, radiation coupling techniques, and strategies to mitigate the impact of cosmic ray events. 4.1.Bolometer SensitivityA bolometer’s sensitivity is parametrized by its NEP. In the dark, a bolometer’s NEP is ideally limited by phonon noise (), the fundamental thermodynamic noise that arises due to the random exchange of energy across the thermal link coupling the bolometer pixel to the thermal bath (see Fig. 1). In this limit, NEP . Under optical bias, a background-limited detector is limited by photon noise, with as the leading subdominant noise term. Thus, is the figure of merit for a bolometer’s sensitivity (see Mather56 for a complete treatment of bolometer noise). The power spectral density of the phonon noise is given by where is Boltzmann’s constant, is the temperature of the bolometer pixel, is the thermal conductance between the pixel and the thermal bath, and is a constant that accounts for potential nonequilibrium effects. In the equilibrium case, where the pixel temperature is equal to the temperature of the thermal bath , . In the extreme nonequilibrium case, where , . Bolometers are usually designed to operate somewhere between these extrema; for example, bolometers designed for cosmic microwave background studies typically operate with . Thus, to make a more sensitive detector, one must reduce the operating temperature and/or reduce the thermal conductance. The former must meet practical constraints of available cryostats (space qualified cryostats can achieve a minimum operating temperature of ), so once a minimum operating temperature is reached, reductions in NEP are accomplished by reducing .NEP can be expressed in several ways that represent different measurements and device characteristics. It is important to note the distinctions between them. First is the thermal fluctuation noise NEP (). is calculated from Eq. (4) using a value for extracted from curve measurements. It represents a theoretical limit, where there is no significant contribution to the noise from other known or unknown noise sources (e.g., Johnson noise, amplifier noise, or excess noise). Next is the electrical NEP (). is calculated from the measured current noise spectral density and frequency-dependent electrical responsivity , where .57 Here, is the DC electrical responsivity to dissipated Joule power, extracted from the curve, and is the time constant of the device. Finally, there is the optical NEP (). In contrast to and , is necessarily derived from optical measurements and is calculated from the measured current noise spectral density and the device’s responsivity to incident optical power . In a background-limited detector, would be dominated by photon noise. Both and account for all noise sources, but only accounts for potential inefficiencies and loss mechanisms associated with photon absorption. Reductions in operating temperature and thermal conductance to achieve higher sensitivity impact the bandwidth of a bolometer. Recall that the response time of a bolometer is proportional to . At low temperatures, both and scale with temperature: and . Bolometers are engineered to achieve a specific value depending on the physics of the heat transfer. Typical values include , 4, or 5, corresponding to ballistic phonon conduction, Kapitza conduction, or electron–phonon conduction, respectively. In a realized bolometer, the measured value may take on intermediate values if there are multiple series/parallel conduction paths through differing means or if there are few propagating modes.58,59 Figure 6 illustrates the natural tradeoffs that occur between operating temperature, time constant, and NEP. In short, a more sensitive bolometer is also slower, thus introducing more stringent requirements on overall instrument stability. Engineering to achieve a certain value also impacts the saturation power of a bolometer , defined as the power required to heat the bolometer to a temperature above which the responsivity of the device falls below some critical level. The relationship between and can be understood from the power balance requirement of steady-state operation, where the power flowing out the thermal link equals the sum of absorbed optical power and Joule power dissipated by the TES resistance. Note that stray power can manifest itself as either optical power or Joule power. For , where is a constant and is the exponent of the thermal conductance, is given by Thus, for a given bath temperature and sensor transition temperature (), the saturation power of the device scales with the thermal conductance. As is reduced to achieve higher sensitivity, the stray power requirements of the experimental platform—electrical and optical—become proportionally more stringent. In particular, implementation of techniques to mitigate stray power like filtered connectors at the cryostat vacuum feedthroughs, thermal blocking and powder filters in close proximity to the detectors, and single-point grounding is necessary to achieve a sufficiently quiet experimental space. In the dark, the most sensitive bolometer demonstrated to date was developed by SRON and is described by Ridder et al.,60 Khosropanah et al.,61 and Suzuki et al.62 It uses a TiAu TES on a leg-isolated SiN membrane, achieving of and of , with a phonon noise-limited bandwidth of and a saturation power below a few femtowatts. Figure 6 illustrates the impact on this detector’s time constant if were reduced by the necessary factor of to meet Origins sensitivity requirements. Under optical bias, Karasik and Cantor measured of using a Ti hot electron bolometer, the lowest optical NEP demonstrated to date in a TES bolometer.63 4.2.Thermal Isolation TechniquesThere are several thermal isolation techniques that show promise toward enabling Origins-type sensitivities in bolometers. In no particular order, these fall into four basic categories (Fig. 7): isolation via long (diffusive) legs, short (ballistic) legs, phononic filter/bandgap legs, and weak electron–phonon (e-p) coupling in the sensor material (a “hot electron” bolometer). “Long (diffusive) leg bolometers” generate thermal isolation via diffuse phonon transport across the legs that support a membrane-isolated bolometer. The legs and the membrane are typically micromachined from Si or SiN. Conduction via diffuse phonon transport follows Fourier’s law of thermal conduction: , where is the material’s bulk thermal conductivity, is the cross-sectional area of the leg, and is the length of the leg. As has been pointed out in the literature,64,65 the Fourier limit is an oversimplification of physics. is not a bulk property, but rather a property that depends on the details of the leg’s fabrication process, which can lead to nonuniform conductance across an array or between fabrication runs. In addition, the behavior only holds for legs longer than a certain threshold; has been reported.65 Bolometers that use long-leg isolation are exemplified in practice by the SPICA/SAFARI bolometers developed by SRON60–62 (see Sec. 4.1). A representative optical micrograph of a long-leg bolometer is shown in Fig. 7. Despite the excellent achieved sensitivity of long-legged bolometers, an obvious limitation is the low fill fraction achieved. By area, the SPICA/SARARI bolometers have a fill fraction of . SPICA/SAFARI has three bands between 600 and 2000 pixels per band, yielding far fewer pixels than the required for OSS. With , reducing the conductance by 100× to meet OSS requirements would necessarily reduce the fill fraction. For FIP, however, these bolometers already meet the sensitivity requirement and the array’s size is similar to SPICA/SAFARI. An added complication to longer legs is that as the leg length increases, the heat capacity of the legs can become non-negligible, leading to measurable ITFN degrading device performance. Long-legged bolometers lend themselves most naturally to resistive absorber coupling. Antenna coupling would likely have to incorporate the legs themselves, and further investigation is required to determine whether the electromagnetic requirements of an antenna could be consistent with the thermal requirements of the legs. “Short (ballistic) leg bolometers” achieve thermal isolation by minimizing the solid angle between the radiating sensor and legs that support a membrane. The ballistic limit occurs when the length of the leg is shorter than the phonon mean free path in the material, so it is assumed that phonons that pass through the leg aperture escape to the bath. For very small cross-sectional areas, ballistic legs have been shown to approach the so-called “quantum limit” of thermal conductance, where only four phonon modes can propagate.66 The bolometers developed for the HIRMES instrument58 used this technique to achieve kilopixel arrays with ,67 in good agreement with the targeted value of NEP for background-limited operation. Compared to long-legged bolometers, ballistic leg bolometers can achieve much higher array fill fractions, and the achieved thermal conductance of realized legs is both calculable from real bulk material properties and is much more tolerant of fabrication nonidealities and nonuniformities.66,68,69 “Phononic filter bolometers” use nanomachined legs to generate coherent phonon scattering that suppresses propagated phonon modes to below the quantum limit.34,68–70 Like ballistic leg bolometers, phononic filter bolometers promise short legs that enable high array filling fractions but with the additional advantage of reduced thermal conductivity relative to ballistic leg designs. Theoretical work by Rostem et al.68 indicates that a factor of 5 improvement over the quantum limit is possible with phononic filter legs that are just long and are machinable using known techniques. Devices presented by Williams et al.69 achieved a factor of reduction in thermal conductance below the quantum limit using phononic filter legs. Fabrication of phononic filters is challenging as submicrometer lithography techniques are required, but recent work by Denis et al.71 demonstrates the reliable and robust fabrication of phononic leg-isolated TESs. “Hot electron bolometers (HEBs)” use the inherent decoupling of electrons and phonons in certain metals at low temperatures for thermal isolation. The conductance between electrons and phonons at low temperature is given by where is the electron–phonon coupling constant (material property), is the volume of the sensor, is temperature, and is the thermal exponent (typically for e-p coupling). A table of measured coupling constants for superconducting metals that would act as the hot-electron bolometer is provided by Karasik et al.72 Among superconductors, W has the weakest known e-p coupling and therefore the greatest potential as a hot-electron bolometer sensor material. Karasik et al.63,72,73 have achieved below and of using Ti as the hot electron sensor; we have performed simulations that indicate W-based HEBs can exceed OSS requirements at 70 mK. A significant advantage of HEBs over the other bolometer types described above is that HEBs do not require any membranes, thus fabrication is simplified, large fill fractions can be realized, and the final devices are mechanically robust. The fabrication problems that can occur in HEBs (e.g., inconsistent transition temperatures or interface problems) are shared among all TES bolometer types and ultimately are not limiting. Note that while there are comparatively few examples of realized hot-electron bolometers in the literature, calorimeters are frequently designed to operate in the electron–phonon decoupled limit using materials that would also yield exceptional sensitivity in a bolometer (e.g., see Calkins et al.51).4.3.Radiation Coupling TechniquesAntenna coupling can be used to define the sensor’s angular acceptance and coupling to the electromagnetic radiation field. It converts the incident fields (photons) into electronic excitations in the absorber media, which are dissipated and subsequently detected as heat. To achieve high antenna coupling, efficiency requires transforming the modal symmetry and impedance scales encountered by the wave in free space to that present in the circuit elements employed to absorb the radiation. A wide variety of techniques exist to carry out these functions; however, due to the need to realize large arrays of sensors, in particular, methods amendable to implementation as planar lithographic structures are primarily of interest. Consideration of the overlap in the angular response of the individual antenna elements in a multibeam array can provide insights into the interpixel isolation and maximum achievable power coupling.74 Similarly, care is warranted in providing isolation between the detector bias/readout and the optical signal paths. Rejection of out-of-band radiation sources, which can potentially load a bolometric sensor and degrade the achievable response, also needs to be realized. These functions are typically achieved through the use of choke circuits on the sensor wafer and cooled thermal blocking filters at appropriate locations in the instrument system. A qualitative survey of commonly encountered planar antennas, such as resonant and traveling-wave line- and slot-structures, is provided in Fig. 8. The angular response (or “antenna pattern”) and the impedance scales of complementary line- and slot-like aerial structures are linked by Babinet’s principle in its vector form.75 Extension of these symmetry concepts to antenna geometries that are only a function of the angle enables the realization of frequency-independent antenna structures.76 Examples of frequency-independent traveling-wave structures include planar logarithmic spiral and log-periodic antennas, which have continuous and discrete scaling symmetries, respectively. In practice, such antennas exhibit deviations from frequency independence from truncation at the inner and outer length scales (i.e., set by the feed network and the maximum extent of the aerials). While this tends to simplify impedance matching to the detector’s absorber circuit, these structures tend to have a low filling fraction when used in arrays over a wide spectral range. The operation of planar antenna arrays in an immersion lens configuration77–79 has enabled greater control over electromagnetic substrate losses and has allowed a natural separation of the radiation and readout functions of the focal plane. While antenna-coupled arrays have found utility in instrument applications, where high spatial-sampling presents a driving consideration,80 an alternative approach to achieve full sampling is presented by “absorber-coupled” sensor arrays.81 In this limit, the “antenna response” of the sensor is essentially uniform over the range of interest and the telescope optics in concert with a cold Lyot stop is used to specify and limit radiation presented to the focal plane.82 In this configuration, a homogeneous thin film for each pixel’s absorber has resonant (quarter-wave) or frequency-independent back-termination83,84 depending on the desired spectral range and coupling. Incorporation of frequency-selective surfaces (FSSs) in the absorber structure85,86 provides an alternative path to achieving the desired impedance scales for high in-band-coupling and improved out-of-band rejection in bolometric sensor settings. Prior discussions of antenna coupling for bolometer implementations can be found in the literature.72,87,88 It is important to note that for a bolometric sensor, the detailed thermal mode of operation can place practical constraints on how the device is electromagnetically coupled to the radiation field. For example, in hot-electron bolometers, the antenna has direct electronic coupling and appropriate material selection enables operation of the device. Similarly, in phononic filter, ballistic leg, and diffusive leg bolometers, direct electrical contact would thermally short the detector and reactively electromagnetically coupling the antenna to a resistive absorber element on a thermally isolated element is typically employed. Either approach can be employed for calorimeter applications, though the device we present in Sec. 3 uses a resistive absorber element. In all cases, the choice of absorber strategy should reflect that array fill fraction is a parameter which should be maximized in large array applications like Origins. For completeness, related antenna-coupled sensors at infrared wavelengths89 employing other physical detection mechanisms are also conveyed. These noncryogenic implementations have included integrated dipole antennas,90 bowtie antennas,91,92 log-periodic/spiral antennas,91 microstrip patch antennas,93 microstrip dipole antennas,94 and Yagi-Uda arrays.95 4.4.Susceptibility to Cosmic RaysThe interaction between energetic charged particles and the materials used in the detector system leads to a stochastic background of energetic events observed by the sensors in the focal plane over the course of a space mission. Consider, for example, the Planck Surveyor mission, which reported a rate of 80 cosmic ray-related events per minute for the High Frequency Instrument array during operations in its L2 orbit. Even with template fitting of its radiometric data, a science data loss of was experienced due to these “glitch”-like temporal events.96,97 Beyond impacting observational efficiency, nonideal detector responses associated with these energetic events can lead to instrumental stability and calibration issues, which reduce imaging fidelity if unmitigated. Origins will require higher sensitivity and, by extension, a lower focal plane operating temperature than used in previously deployed systems and thus increase the relative importance of the sensor’s thermal bus implementation on minimizing the impact of cosmic rays. Extension of the detector design techniques employed for calorimetry98–101 provides a viable path to address this instrumentation need. Mitigating cosmic rays is one particular advantage of a device known as the ideal integrating bolometer,102–105 which combines leg isolation (diffusive, ballistic, or phononic) with a switchable thermal short. Upon an upset (e.g., cosmic ray hit), the device can be reset quickly to mitigate the data losses that Planck experienced. 5.ConclusionsThe unprecedented sensitivity enabled by Origins’ large, cold, and space-based telescope places stringent requirements on its instruments. In particular, the detector systems employed by each instrument must be designed to enable background-limited observations, contributing negligible noise to the overall instrumental budget. Detectors that enable background-limited observations in each channel do not exist today. For MISC-T, the most difficult specification the detector system must meet is stability. While bolometers and semiconducting detectors have many sources of instability (e.g., environmental drifts for bolometers, dark current, and read noise for semiconductors), all these terms go to zero if a TES calorimeter is employed. The TES calorimeter we designed using well-established physical models and measured material parameters can overcome many of the other challenges associated with the MISC-T detector system, such as high photon flux and the large required bandwidth, delivering all the advantages of photon-counting using an energy-resolving detector. For OSS and FIP, the most challenging requirement the detector subsystems must meet is sensitivity, with radiation coupling and cosmic ray immunity also leading design drivers. There are multiple promising paths toward achieving the required sensitivity using TES bolometers, with demonstrated sensitivities now approaching the OSS requirement and improving. For all the instruments, the TES detector options benefit from strong heritage in astrophysics instrumentation and well-understood physics that enables the design and implementation of devices that operate at the thermodynamic limit across the Origins band. AcknowledgmentsThe research was supported by the National Aeronautics and Space Administration (NASA) under the Goddard Space Flight Center (GSFC) Internal Research and Development (IRAD) program. The authors are grateful for figures and technical feedback provided by Ari Brown, Kevin Denis, and Karwan Rostem of the GSFC. We are also grateful for the feedback provided by the referees, which improved the quality of this paper. The authors declare no conflicts of interest. ReferencesM. Meixner et al.,
“Origins space telescope mission concept study report,”
(2019). Google Scholar
C. M. Bradford et al.,
“The Origins Survey Spectrometer (OSS): a far-IR discovery machine for the Origins Space Telescope,”
Proc. SPIE, 10698 1069818
(2018). https://doi.org/10.1117/12.2314049 Google Scholar
D. H. Andrews et al.,
“Attenuated superconductors i. for measuring infra-red radiation,”
Rev. Sci. Instrum., 13
(7), 281
–292
(1942). https://doi.org/10.1063/1.1770037 RSINAK 0034-6748 Google Scholar
K. D. Irwin,
“An application of electrothermal feedback for high resolution cryogenic particle detection,”
Appl. Phys. Lett., 66
(15), 1998
–2000
(1995). https://doi.org/10.1063/1.113674 APPLAB 0003-6951 Google Scholar
K. Irwin,
“Squid multiplexers for transition-edge sensors,”
Physica C, 368
(1), 203
–210
(2002). https://doi.org/10.1016/S0921-4534(01)01167-4 Google Scholar
K. D. Irwin and K. W. Lehnert,
“Microwave squid multiplexer,”
Appl. Phys. Lett., 85
(11), 2107
–2109
(2004). https://doi.org/10.1063/1.1791733 APPLAB 0003-6951 Google Scholar
K. M. Morgan et al.,
“Code-division-multiplexed readout of large arrays of tes microcalorimeters,”
Appl. Phys. Lett., 109
(11), 112604
(2016). https://doi.org/10.1063/1.4962636 APPLAB 0003-6951 Google Scholar
K. D. Irwin and G. C. Hilton, Transition-Edge Sensors, 99 63
–150 Springer-Verlag, Berlin, Heidelberg
(2005). Google Scholar
J. N. Ullom and D. A. Bennett,
“Review of superconducting transition-edge sensors for x-ray and gamma-ray spectroscopy,”
Supercond. Sci. Technol., 28 084003
(2015). https://doi.org/10.1088/0953-2048/28/8/084003 SUSTEF 0953-2048 Google Scholar
C. M. Posada et al.,
“Fabrication of large dual-polarized multichroic TES bolometer arrays for CMB measurements with the SPT-3g camera,”
Supercond. Sci. Technol., 28 094002
(2015). https://doi.org/10.1088/0953-2048/28/9/094002 SUSTEF 0953-2048 Google Scholar
P. A. R. Ade et al.,
“Antenna-coupled TES bolometers used in BICEP2, Keck array, and spider,”
Astrophys. J., 812 176
(2015). https://doi.org/10.1088/0004-637X/812/2/176 ASJOAB 0004-637X Google Scholar
D. A. Harper et al.,
“Hawc+, the far-infrared camera and polarimeter for sofia,”
J. Astron. Instrum., 07
(4), 1840008
(2018). https://doi.org/10.1142/S2251171718400081 Google Scholar
A. E. Lita, A. J. Miller and S. W. Nam,
“Counting near-infrared single-photons with 95% efficiency,”
Opt. Express, 16 3032
(2008). https://doi.org/10.1364/OE.16.003032 OPEXFF 1094-4087 Google Scholar
D. J. Fixsen et al.,
“Optimal energy measurement in nonlinear systems: an application of differential geometry,”
J. Low Temp. Phys., 176 16
–26
(2014). https://doi.org/10.1007/s10909-014-1149-x JLTPAC 0022-2291 Google Scholar
S. E. Busch et al.,
“Progress towards improved analysis of TES x-ray data using principal component analysis,”
J. Low Temp. Phys., 184 382
–388
(2016). https://doi.org/10.1007/s10909-015-1357-z JLTPAC 0022-2291 Google Scholar
A. R. Miniussi et al.,
“Performance of an x-ray microcalorimeter with a 240 μm absorber and a TES bilayer,”
J. Low Temp. Phys., 193 337
–343
(2018). https://doi.org/10.1007/s10909-018-1974-4 JLTPAC 0022-2291 Google Scholar
A. Luukanen et al.,
“Fluctuation-Limited noise in a superconducting transition-edge sensor,”
Phys. Rev. Lett., 90 238306
(2003). https://doi.org/10.1103/PhysRevLett.90.238306 PRLTAO 0031-9007 Google Scholar
K. Irwin and G. Hilton, Transition-Edge Sensors, 63
–150 Springer, Berlin, Heidelberg
(2005). Google Scholar
J. N. Ullom et al.,
“Characterization and reduction of unexplained noise in superconducting transition-edge sensors,”
Appl. Phys. Lett., 84 4206
(2004). https://doi.org/10.1063/1.1753058 APPLAB 0003-6951 Google Scholar
K. M. Kinnunen, M. R. J. Palosaari and I. J. Maasilta,
“Normal metal-superconductor decoupling as a source of thermal fluctuation noise in transition-edge sensors,”
J. Appl. Phys., 112 034515
(2012). https://doi.org/10.1063/1.4745908 JAPIAU 0021-8979 Google Scholar
N. Iyomoto et al.,
“Modeling of TES x-ray microcalorimeters with a novel absorber design,”
J. Low Temp. Phys., 151 406
–412
(2008). https://doi.org/10.1007/s10909-007-9668-3 JLTPAC 0022-2291 Google Scholar
K. Irwin,
“Thermodynamics of nonlinear bolometers near equilibrium,”
Nucl. Instrum. Methods Phys. Res. Sect. A, 559
(2), 718
–720
(2006). https://doi.org/10.1016/j.nima.2005.12.115 NIMAER 0168-9002 Google Scholar
J. E. Sadleir et al.,
“Longitudinal proximity effects in superconducting transition-edge sensors,”
Phys. Rev. Lett., 104 047003
(2010). https://doi.org/10.1103/PhysRevLett.104.047003 PRLTAO 0031-9007 Google Scholar
J. E. Sadleir et al.,
“Proximity effects and nonequilibrium superconductivity in transition-edge sensors,”
Phys. Rev. B, 84 184502
(2011). https://doi.org/10.1103/PhysRevB.84.184502 Google Scholar
J. E. Sadleir et al.,
“Magnetically tuned superconducting transition-edge sensors,”
IEEE Trans. Appl. Supercond., 23 2101405
(2013). https://doi.org/10.1109/TASC.2013.2251391 ITASE9 1051-8223 Google Scholar
J. E. Sadleir et al.,
“The magnetically-tuned transition-edge sensor,”
J. Low Temp. Phys., 176 392
–399
(2014). https://doi.org/10.1007/s10909-014-1194-5 JLTPAC 0022-2291 Google Scholar
S. J. Smith et al.,
“Implications of weak-link behavior on the performance of Mo/Au bilayer transition-edge sensors,”
J. Appl. Phys., 114 074513
(2013). https://doi.org/10.1063/1.4818917 JAPIAU 0021-8979 Google Scholar
A. G. Kozorezov et al.,
“Microscopic model of a transition edge sensor as a weak link,”
IEEE Trans. Appl. Supercond., 21 250
–253
(2011). https://doi.org/10.1109/TASC.2010.2094595 ITASE9 1051-8223 Google Scholar
A. Kozorezov et al.,
“Modelling the resistive state in a transition edge sensor,”
Appl. Phys. Lett., 99 063503
(2011). https://doi.org/10.1063/1.3621829 APPLAB 0003-6951 Google Scholar
A. G. Kozorezov et al.,
“Athermal energy loss from x-rays deposited in thin superconducting films on solid substrates,”
Phys. Rev. B, 87 104504
(2013). https://doi.org/10.1103/PhysRevB.87.104504 Google Scholar
W. Holmes et al.,
“Measurements of thermal transport in low stress silicon nitride films,”
Appl. Phys. Lett., 72 2250
(1998). https://doi.org/10.1063/1.121269 APPLAB 0003-6951 Google Scholar
H. F. C. Hoevers et al.,
“Radiative ballistic phonon transport in silicon-nitride membranes at low temperatures,”
Appl. Phys. Lett., 86 251903
(2005). https://doi.org/10.1063/1.1949269 APPLAB 0003-6951 Google Scholar
J. T. Karvonen, T. Kühn and I. J. Maasilta,
“Temperature profile for ballistic and diffusive phonon transport in a suspended membrane with a radial symmetric heat source,”
Chin. J. Phys., 49 435
(2011). Google Scholar
N. Zen et al.,
“Engineering thermal conductance using a two-dimensional phononic crystal,”
Nat. Commun., 5 3435
(2014). https://doi.org/10.1038/ncomms4435 NCAOBW 2041-1723 Google Scholar
D. Fukuda et al.,
“Titanium-based transition-edge photon number resolving detector with 98% detection efficiency with index-matched small-gap fiber coupling,”
Opt. Express, 19 870
–875
(2011). https://doi.org/10.1364/OE.19.000870 OPEXFF 1094-4087 Google Scholar
D. J. Fixsen et al.,
“Optimal fitting of non-linear detector pulses with nonstationary noise,”
AIP Conf. Proc., 605 339
–342
(2002). https://doi.org/10.1063/1.1457659 APCPCS 0094-243X Google Scholar
D. J. Fixsen et al.,
“Pulse estimation in nonlinear detectors with nonstationary noise,”
Nucl. Instrum. Methods Phys. Res. Sect. A, 520 555
–558
(2004). https://doi.org/10.1016/j.nima.2003.11.313 Google Scholar
S. R. Bandler et al.,
“Performance of TES x-ray microcalorimeters with a novel absorber design,”
J. Low Temp. Phys., 151 400
–405
(2008). https://doi.org/10.1007/s10909-007-9673-6 JLTPAC 0022-2291 Google Scholar
R. W. Romani et al.,
“First astronomical application of a cryogenic transition edge sensor spectrophotometer,”
Astrophys. J. Lett., 521 L153
–L156
(1999). https://doi.org/10.1086/312202 AJLEEY 0004-637X Google Scholar
B. S. Karasik and A. V. Sergeev,
“THz hot-electron photon counter,”
IEEE Trans. Appl. Supercond., 15 618
–621
(2005). https://doi.org/10.1109/TASC.2005.849963 ITASE9 1051-8223 Google Scholar
J. Wei et al.,
“Ultrasensitive hot-electron nanobolometers for terahertz astrophysics,”
Nat. Nanotechnol., 3 496
–500
(2008). https://doi.org/10.1038/nnano.2008.173 NNAABX 1748-3387 Google Scholar
L. Zhang et al.,
“A characterization of the single-photon sensitivity of an electron multiplying charge-coupled device,”
J. Phys. B, 42 114011
(2009). https://doi.org/10.1088/0953-4075/42/11/114011 Google Scholar
V. B. Verma et al.,
“Towards single-photon spectroscopy in the mid-infrared using superconducting nanowire single-photon detectors,”
Proc. SPIE, 10978 109780N
(2019). https://doi.org/10.1117/12.2519474 PSISDG 0277-786X Google Scholar
M. D. Petroff, M. G. Stapelbroek and W. A. Kleinhans,
“Detection of individual 0.4- wavelength photons via impurity-impact ionization in a solid-state photomultiplier,”
Appl. Phys. Lett., 51 406
–408
(1987). https://doi.org/10.1063/1.98404 APPLAB 0003-6951 Google Scholar
S. Komiyama,
“Single-photon detectors in the terahertz range,”
IEEE J. Sel. Top. Quantum Electron., 17 54
–66
(2011). https://doi.org/10.1109/JSTQE.2010.2048893 IJSQEN 1077-260X Google Scholar
D. T. Chuss et al.,
“Diffraction considerations for planar detectors in the few-mode limit,”
Pub. Astron. Soc. Pac., 120 430
(2008). https://doi.org/10.1086/587494 PASPAU 0004-6280 Google Scholar
C. N. Thomas et al.,
“Modeling the intensity and polarization response of planar bolometric detectors,”
J. Opt. Soc. Am. A, 27 1219
(2010). https://doi.org/10.1364/JOSAA.27.001219 JOAOD6 0740-3232 Google Scholar
M. P. Chiao et al.,
“Parametric characterization of TES detectors under DC bias,”
IEEE Trans. Appl. Supercond., 27 1
–5
(2017). https://doi.org/10.1109/TASC.2016.2645164 ITASE9 1051-8223 Google Scholar
D. Fukuda et al.,
“Photon number resolving detection with high speed and high quantum efficiency,”
Metrologia, 46 S288
–S292
(2009). https://doi.org/10.1088/0026-1394/46/4/S29 MTRGAU 0026-1394 Google Scholar
D. Fukuda et al.,
“Titanium superconducting photon-number-resolving detector,”
IEEE Trans. Appl. Supercond., 21
(3), 241
–245
(2011). https://doi.org/10.1109/TASC.2010.2089953 ITASE9 1051-8223 Google Scholar
B. Calkins et al.,
“Faster recovery time of a hot-electron transition-edge sensor by use of normal metal heat-sinks,”
Appl. Phys. Lett., 99 241114
(2011). https://doi.org/10.1063/1.3659686 APPLAB 0003-6951 Google Scholar
A. T. Lee et al.,
“A superconducting bolometer with strong electrothermal feedback,”
Appl. Phys. Lett., 69 1801
–1803
(1996). https://doi.org/10.1063/1.117491 APPLAB 0003-6951 Google Scholar
H. Hui et al.,
“BICEP3 focal plane design and detector performance,”
Proc. SPIE, 9914 99140T
(2016). https://doi.org/10.1117/12.2232986 PSISDG 0277-786X Google Scholar
A. S. Bergman et al.,
“280 GHz focal plane unit design and characterization for the SPIDER-2 suborbital polarimeter,”
J. Low Temp. Phys., 193 1075
–1084
(2018). https://doi.org/10.1007/s10909-018-2065-2 JLTPAC 0022-2291 Google Scholar
“The EBEX balloon borne experiment—detectors and readout,”
Astrophys. J. Suppl. Ser., 239
(1), 8
(2018). https://doi.org/10.3847/1538-4365/aae436 APJSA2 0067-0049 Google Scholar
J. C. Mather,
“Bolometer noise: nonequilibrium thoery,”
Appl. Opt., 21 1125
–1129
(1982). https://doi.org/10.1364/AO.21.001125 APOPAI 0003-6935 Google Scholar
R. C. Jones,
“The general theory of bolometer performance,”
J. Opt. Soc. Am., 43 1
–10
(1953). https://doi.org/10.1364/JOSA.43.000001 JOSAAH 0030-3941 Google Scholar
E. M. Barrentine et al.,
“Characterization of Si membrane TES bolometer arrays for the HIRMES instrument,”
J. Low Temp. Phys., 193 241
–248
(2018). https://doi.org/10.1007/s10909-018-1966-4 JLTPAC 0022-2291 Google Scholar
J. Bartlett, K. Rostem and E. J. Wollack,
“Robust modeling of acoustic phonon transmission in nanomechanical structures,”
Appl. Phys. Lett., 114 113101
(2019). https://doi.org/10.1063/1.5068685 APPLAB 0003-6951 Google Scholar
M. L. Ridder et al.,
“Fabrication of low-noise TES arrays for the SAFARI instrument on SPICA,”
J. Low Temp. Phys., 184 60
–65
(2016). https://doi.org/10.1007/s10909-015-1381-z JLTPAC 0022-2291 Google Scholar
P. Khosropanah et al.,
“Ultra-Low noise TES bolometer arrays for SAFARI instrument on SPICA,”
Proc. SPIE, 9914 99140B
(2016). https://doi.org/10.1117/12.2233472 PSISDG 0277-786X Google Scholar
T. Suzuki et al.,
“Development of ultra-low-noise TES bolometer arrays,”
J. Low Temp. Phys., 184 52
–59
(2016). https://doi.org/10.1007/s10909-015-1401-z JLTPAC 0022-2291 Google Scholar
B. S. Karasik and R. Cantor,
“Demonstration of high optical sensitivity in far-infrared hot-electron bolometer,”
Appl. Phys. Lett., 98 193503
(2011). https://doi.org/10.1063/1.3589367 APPLAB 0003-6951 Google Scholar
K. Rostem et al.,
“Precision control of thermal transport in cryogenic single-crystal silicon devices,”
J. Appl. Phys., 115 124508
(2014). https://doi.org/10.1063/1.4869737 JAPIAU 0021-8979 Google Scholar
S. Withington et al.,
“Thermal elastic-wave attenuation in low-dimensional bars at low temperatures,”
J. Appl. Phys., 122 054504
(2017). https://doi.org/10.1063/1.4997466 JAPIAU 0021-8979 Google Scholar
D. Osman et al.,
“Transition edge sensors with few-mode ballistic thermal isolation,”
J. Appl. Phys., 116
(6), 064506
(2014). https://doi.org/10.1063/1.4893019 JAPIAU 0021-8979 Google Scholar
A. Brown,
(2020). Google Scholar
K. Rostem et al.,
“Wide-stopband aperiodic phononic filters,”
J. Phys. D Appl. Phys., 49 255301
(2016). https://doi.org/10.1088/0022-3727/49/25/255301 Google Scholar
E. A. Williams et al.,
“Superconducting transition edge sensors with phononic thermal isolation,”
J. Appl. Phys., 124 144501
(2018). https://doi.org/10.1063/1.5041348 JAPIAU 0021-8979 Google Scholar
A. N. Cleland, D. R. Schmidt and C. S. Yung,
“Thermal conductance of nanostructured phononic crystals,”
Phys. Rev. B, 64 172301
(2001). https://doi.org/10.1103/PhysRevB.64.172301 Google Scholar
K. L. Denis et al.,
“Fabrication of phononic filter structures for far-IR/sub-mm detector applications,”
Proc. SPIE, 10927 1092720
(2019). https://doi.org/10.1117/12.2506387 PSISDG 0277-786X Google Scholar
B. S. Karasik, A. V. Sergeev and D. E. Prober,
“Nanobolometers for THz photon detection,”
IEEE Trans. Terahertz Sci. Technol., 1 97
–111
(2011). https://doi.org/10.1109/TTHZ.2011.2159560 Google Scholar
B. S. Karasik et al.,
“Development of the nano-HEB array for low-background far-IR applications,”
Proc. SPIE, 7741 774119
(2010). https://doi.org/10.1117/12.856682 PSISDG 0277-786X Google Scholar
S. Stein,
“On cross coupling in multiple-beam antennas,”
IRE Trans. Anntenas Propag., 10
(5), 548
–557
(1962). https://doi.org/10.1109/TAP.1962.1137917 Google Scholar
H. G. Booker,
“Slot aerials and their relation to complementary wire aerials (Babinet’s principle),”
J. Inst. Electr. Eng. Part IIIA, 93
(4), 620
–626
(1946). https://doi.org/10.1049/ji-3a-1.1946.0150 Google Scholar
V. Rumsey,
“Frequency independent antennas,”
in 1958 IRE Int. Convention Record,
114
–118
(1966). https://doi.org/10.1109/IRECON.1957.1150565 Google Scholar
D. Rutledge and M. Muha,
“Imaging antenna arrays,”
IEEE Trans. Antennas Propag., 30
(4), 535
–540
(1982). https://doi.org/10.1109/TAP.1982.1142856 IETPAK 0018-926X Google Scholar
D. Rutledge, D. Neikirk, D. Kasilingam,
“Integrated-circuit antennas,”
Infrared and Millimeter Waves, 10 1
–90 Academic Press, New York
(1983). Google Scholar
R. Compton et al.,
“Bow-Tie antennas on a dielectric half-space: theory and experiment,”
IEEE Trans. Antennas Propag., 35
(6), 622
–631
(1987). https://doi.org/10.1109/TAP.1987.1144162 IETPAK 0018-926X Google Scholar
M. J. Griffin, J. J. Bock and W. K. Gear,
“Relative performance of filled and feedhorn-coupled focal-plane architectures,”
Appl. Opt., 41 6543
–6554
(2002). https://doi.org/10.1364/AO.41.006543 APOPAI 0003-6935 Google Scholar
A. Kusaka, E. J. Wollack and T. R. Stevenson,
“Angular and polarization response of multimode sensors with resistive-grid absorbers,”
J. Opt. Soc. Am. A, 31 1557
(2014). https://doi.org/10.1364/JOSAA.31.001557 JOAOD6 0740-3232 Google Scholar
E. J. Wollack, D. T. Chuss and S. H. Moseley,
“Electromagnetic considerations for pixellated planar bolometer arrays in the single-mode limit,”
Proc. SPIE, 6275 62750V
(2006). https://doi.org/10.1117/12.672600 PSISDG 0277-786X Google Scholar
J. Clarke et al.,
“Superconductive bolometers for submillimeter wavelengths,”
J. Appl. Phys., 48
(12), 4865
–4879
(1977). https://doi.org/10.1063/1.323612 JAPIAU 0021-8979 Google Scholar
B. Carli and D. Iorio-Fili,
“Absorption of composite bolometers,”
J. Opt. Soc. Am., 71 1020
–1025
(1981). https://doi.org/10.1364/JOSA.71.001020 JOSAAH 0030-3941 Google Scholar
M. S. Kowitt et al.,
“Frequency selective bolometers,”
Appl. Opt., 35 5630
–5635
(1996). https://doi.org/10.1364/AO.35.005630 APOPAI 0003-6935 Google Scholar
B. Monacelli et al.,
“Infrared frequency selective surface based on circuit-analog square loop design,”
IEEE Trans. Antennas Propag., 53
(2), 745
–752
(2005). https://doi.org/10.1109/TAP.2004.841290 IETPAK 0018-926X Google Scholar
S. E. Schwarz and B. T. Ulrich,
“Antenna-coupled infrared detectors,”
J. Appl. Phys., 48
(5), 1870
–1873
(1977). https://doi.org/10.1063/1.323940 JAPIAU 0021-8979 Google Scholar
J. Mees, M. Nahum and P. L. Richards,
“New designs for antenna-coupled superconducting bolometers,”
Appl. Phys. Lett., 59
(18), 2329
–2331
(1991). https://doi.org/10.1063/1.106059 APPLAB 0003-6951 Google Scholar
S. Y. Wang, T. Izawa and T. K. Gustafson,
“Coupling characteristics of thin-film metal-oxide-metal diodes at 10.6 microns,”
Appl. Phys. Lett., 27
(9), 481
–483
(1975). https://doi.org/10.1063/1.88541 APPLAB 0003-6951 Google Scholar
C. Fumeaux et al.,
“Measurement of the resonant lengths of infrared dipole antennas,”
Infrared Phys. Technol., 41
(5), 271
–281
(2000). https://doi.org/10.1016/S1350-4495(00)00047-5 IPTEEY 1350-4495 Google Scholar
N. Chong and H. Ahmed,
“Antenna-coupled polycrystalline silicon air-bridge thermal detector for mid-infrared radiation,”
Appl. Phys. Lett., 71
(12), 1607
–1609
(1997). https://doi.org/10.1063/1.119992 APPLAB 0003-6951 Google Scholar
F. J. González et al.,
“Antenna-coupled infrared detectors for imaging applications,”
IEEE J. Sel. Top. Quantum Electron., 11
(1), 117
–120
(2005). https://doi.org/10.1109/JSTQE.2004.841474 IJSQEN 1077-260X Google Scholar
I. Codreanu et al.,
“Microstrip antenna-coupled infrared detector,”
Electron. Lett., 35
(25), 2166
–2167
(1999). https://doi.org/10.1049/el:19991486 ELLEAK 0013-5194 Google Scholar
I. Codreanu and G. D. Boreman,
“Infrared microstrip dipole antennas—FDTD predictions versus experiment,”
Microw. Opt. Technol. Lett., 29
(6), 381
–383
(2001). https://doi.org/10.1002/mop.1184 MOTLEO 0895-2477 Google Scholar
E. Mohammadi, M. A. Ghaffari and N. Behdad,
“Wide dynamic range, angle-sensing, long-wave infrared detector using nano-antenna arrays,”
Sci. Rep., 10 2488
(2020). https://doi.org/10.1038/s41598-020-59440-2 SRCEC3 2045-2322 Google Scholar
“Planck early results. IV. First assessment of the high frequency instrument in-flight performance,”
Astron. Astrophys., 536 A4
(2011). https://doi.org/10.1051/0004-6361/201116487 AAEJAF 0004-6361 Google Scholar
“Planck early results. VI. The high frequency instrument data processing,”
Astron. Astrophys., 536 A6
(2011). https://doi.org/10.1051/0004-6361/201116462 AAEJAF 0004-6361 Google Scholar
T. Saab et al.,
“GEANT modeling of the low-earth-orbit cosmic-ray background for the Astro-E2 XRS instrument,”
Proc. SPIE, 5501 320
–327
(2004). https://doi.org/10.1117/12.551945 PSISDG 0277-786X Google Scholar
C. A. Kilbourne et al.,
“Analysis of the Suzaku/XRS background,”
Nucl. Instrum. Methods Phys. Res. Sect. A, 559 620
–622
(2006). https://doi.org/10.1016/j.nima.2005.12.116 Google Scholar
N. Iyomoto et al.,
“Heat sinking, crosstalk, and temperature uniformity for large close-packed microcalorimeter arrays,”
IEEE Trans. Appl. Supercond., 19
(3), 557
–560
(2009). https://doi.org/10.1109/TASC.2009.2017704 ITASE9 1051-8223 Google Scholar
F. M. Finkbeiner et al.,
“Development of embedded heatsinking layers for compact arrays of x-ray TES microcalorimeters,”
IEEE Trans. Appl. Supercond., 21 223
–226
(2011). https://doi.org/10.1109/TASC.2010.2091237 ITASE9 1051-8223 Google Scholar
A. Kogut,
“Ideal integrating bolometer,”
in Far-IR, Sub-Mm & MM Detector Technol. Workshop,
51
(2002). Google Scholar
A. Kogut, M. Dipirro, S. H. Moseley,
“Ideal integrating bolometer,”
New Concepts for Far-Infrared and Submillimeter Space Astronomy, 342 2004). Google Scholar
P. C. Nagler et al.,
“Development of superconducting magnetic heat switches for an ideal integrating bolometer,”
J. Low Temp. Phys., 184 280
–285
(2016). https://doi.org/10.1007/s10909-015-1365-z JLTPAC 0022-2291 Google Scholar
E. R. Canavan et al.,
“Demonstration of a pixel-scale superconducting heat switch for an ideal integrating bolometer,”
IEEE Trans. Appl. Supercond., 27 1
–5
(2017). https://doi.org/10.1109/TASC.2017.2664830 ITASE9 1051-8223 Google Scholar
BiographyPeter C. Nagler is an experimental physicist in the Instrument Systems and Technology Division at NASA’s Goddard Space Flight Center. His research focuses on low-temperature detectors and custom instrumentation for astrophysics applications. |