1 July 2005 Comparison of histometric data obtained by optical coherence tomography and routine histology
Author Affiliations +
J. of Biomedical Optics, 10(4), 044008 (2005). doi:10.1117/1.2039086
Abstract
There is a lack of systematic investigations comparing optical coherence tomography (OCT) with histology. OCT assessments were performed on the upper back of 16 healthy subjects. Epidermis thickness (ET) was assessed using three methods: first, peak-to-valley analysis of the A-scan (ET-OCT-V); second, manual measurements in the OCT images (ET-OCT-M); third, light microscopic determination using routine histology (ET-Histo). The relationship between the different methods was assessed by means of the Pearson correlation procedure and Bland and Altman plots. We observed a strong correlation between ET-Histo (79.4±21.9 µm) and ET-OCT-V (79.2±15.5 µm, r=0.77) and ET-OCT-M (82.9±15.8 µm, r=0.75), respectively. Bland and Altman plots revealed a bias of –0.19 µm (95% limits of agreement: –27.94 µm to 27.56 µm) for ET-OCT-V versus ET-Histo and a bias of 3.44 µm (95% limits of agreement: –24.9 µm to 31.78 µm) for ET-OCT-M versus ET-Histo. Despite the strong correlation and low bias observed, the 95% limits of agreement demonstrated an unsatisfactory numerical agreement between the two OCT methods and routine histology indicating that these methods cannot be employed interchangeably. Regarding practicability, precision, and indication spectrum, ET-OCT-V and ET-OCT-M are of different clinical value.
Thilo Gambichler, Stefanie Boms, Markus Stuecker, Alexander Kreuter, Michael Sand, Georg Moussa, Peter Altmeyer, Klaus Hoffmann, "Comparison of histometric data obtained by optical coherence tomography and routine histology," Journal of Biomedical Optics 10(4), 044008 (1 July 2005). http://dx.doi.org/10.1117/1.2039086
JOURNAL ARTICLE
6 PAGES


SHARE
KEYWORDS
Optical coherence tomography

Skin

In vivo imaging

Tissue optics

Statistical analysis

Ultrasonography

Image resolution

Back to Top