1 November 2005 Evaluation of enamel dental restoration interface by optical coherence tomography
Author Affiliations +
Abstract
Evaluation of molar dental restorations on enamel is performed using optical coherence tomography (OCT) with 10 µm resolution. Images of ~50 µm failure gaps in the restorations are demonstrated and the OCT images are compared with x-ray and optical microscopy pictures. The results demonstrate the potential of the technique for clinical evaluation of dental restorations.
© (2005) Society of Photo-Optical Instrumentation Engineers (SPIE)
L. S.A. de Melo, L. S.A. de Melo, Renato E. de Araujo, Renato E. de Araujo, Anderson Zanardi Freitas, Anderson Zanardi Freitas, Denise Maria Zezell, Denise Maria Zezell, Nilson Dias Vieira, Nilson Dias Vieira, John M. Girkin, John M. Girkin, Andrew F. Hall, Andrew F. Hall, Mariana Torres Carvalho, Mariana Torres Carvalho, Anderson Stevens Leônidas Gomes, Anderson Stevens Leônidas Gomes, } "Evaluation of enamel dental restoration interface by optical coherence tomography," Journal of Biomedical Optics 10(6), 064027 (1 November 2005). https://doi.org/10.1117/1.2141617 . Submission: