|
1.IntroductionRaman spectroscopy is a powerful analytical method due to its chemical bond specificity. The molecular vibrations of each substance produce a characteristic fingerprint spectrum that can be used to determine the chemical and structural composition of the sample. Applications in biomedical research are wide-ranging, including the diagnosis of cancer,1, 2, 3 atherosclerosis,4 and blood analysis.5 The major advantages of Raman spectroscopy compared to fluorescence applications are that Raman signals do not bleach, and that Raman bands are spectrally narrow, which improves the separability of the signals. A basic challenge for Raman spectroscopic measurements in turbid media is the signal quantification. The spectral line shape and overall intensity of the measured spectra depend on the absorption and scattering properties of the sample. In the case of biological tissue, absorption arises from chromophores such as hemoglobin and melanin. Light scattering in tissue occurs due to the refractive index inhomogeneities of the cell nuclei, cell organelles, or collagen fibers. Due to different biochemical tissue compositions and microstructures, the optical properties of tissues may vary strongly between tissues, tissue sites, and individuals. Since a contrast in optical properties can, on its own, indicate pathological changes such as tumor angiogenesis or dysplasia, extensive research has been carried out on the measurement of optical properties for diagnostic purposes.6, 7, 8, 9 Nevertheless, most Raman spectroscopic tissue measurements are not corrected for these influences, which hinders the quantitative comparison of the measured Raman spectra. Since the measured Raman scattered light and purely elastically scattered light are both influenced by variations in the optical properties, correction methods have been suggested based on diffuse reflectance measurements. Some approaches have been developed for applications where either absorption or scattering is assumed to change10, 11 and are therefore not suitable for in vivo applications, where independent and simultaneous variations of the optical properties are common. The first theoretical and experimental studies on the correction of Raman signals in biological tissue were published recently12, 13, 14 for a parameter range of absorption coefficient and scattering coefficient , which are typical for measurements in the near infrared wavelength range (NIR). The correction functions utilize the elastic diffuse reflectance, which is measured in the same excitation-detection geometry as the Raman spectrum. However, knowledge of the total attenuation coefficient given by is required,12, 13 or the reduced scattering coefficient is used.14 However, the determination of optical properties from a single elastic reflectance signal requires an elastic diffuse reflectance spectrum, together with sample specific assumptions regarding the wavelength dependency of the absorption and scattering coefficients,15 which reduces the method’s flexibility. The aim of our investigation was to develop a method that could correct Raman signals for the influence of the optical properties in a large parameter range without a priori assumptions of the absorption and scattering properties. Therefore, we combined Raman spectroscopy with spatially resolved reflectance measurements.16 By performing measurements on various homogenous tissue phantoms as well as Monte Carlo simulations, we investigated how the Raman signal depends on the optical properties. To include parameters typical for measurements in the NIR as well as in the visible wavelength range (VIS), was varied from and from . These investigations exceed the parameter range of previous studies.12, 13, 14 To correct the Raman signals for the influence of the optical properties, we investigated possible correction functions that are dependent on the elastic diffuse reflectance. We also implemented a correction function based directly on the Monte Carlo simulation of the Raman signals. 2.Experimental Setup and Monte Carlo SimulationIn this section, we explain the Monte Carlo code that was used to simulate the Raman signal and describe our combined Raman and spatially resolved reflectance setup as well as the production of the tissue phantoms. 2.1.Monte Carlo Model for the Simulation of Raman and Reflectance SignalsIn accordance with our experiments, we assumed that the absorption coefficient and the coefficient for Raman scattering were constant throughout the sample. Because was very small, multiple Raman scattering was neglected and the Raman signal was assumed to be proportional to . We further assumed a Henyey-Greenstein phase function12 and an anisotropy factor for the elastic scattering processes. We defined a flux density of photons that were not Raman scattered and a flux density of Raman-scattered photons. Since multiple Raman scattering was neglected, both densities could be calculated with the conventional Monte Carlo method described in Ref. 17, which takes into account elastic scattering and absorption. Only the source term was different for both flux densities. For , the source term was a boundary condition at the sample surface that corresponded to the excitation light entering the sample. For , the source term was the number density of Raman scattering events given by , which was proportional to the number density of the absorbed excitation photons . Therefore, was used as the source density to calculate and scale the results with . In the Monte Carlo calculation for , the density was given by the ensemble of absorption process positions. Thus, the following scheme was used to calculate : At each position where an excitation photon was absorbed, a Raman scattered photon was launched with an isotropic initial direction. The photon propagated with optical parameters that corresponded to the Stokes shifted wavelength. A fraction of these photons was remitted at the sample surface and reached a Raman detector, which was defined as a circular area on the sample surface. Normalization with the number of incident excitation photons and scaling led to the detected Raman signal . In this paper, corresponds to the calculated or measured relative Raman signal of a Raman scatterer embedded in a turbid matrix. The elastic diffuse reflectance of the sample was determined by the number of excitation photons detected at the sample surface. In the results shown below, the relative elastic diffuse reflectance signal is given by . The size and position of the excitation and detector area corresponds to the images of the fiber endfaces on the sample surface in Fig. 1 . For simplicity, the fiber bundle for the Raman detection in Fig. 2 was assumed to correspond to one fiber with a radius, and the slant of excitation and receiving optics were neglected. The propagation direction of the remitted photons were analyzed by taking into account the finite numerical aperture (NA) of the imaging optics. For our application, we used and scaled the results by a constant factor accordingly. We used photons for , leading to a computation time of a few seconds for each set of parameters but photons for . 2.2.Combined Raman and Spatially Resolved Reflectance SetupA schematic overview of our setup is shown in Fig. 2. For the alternating measurement of Raman spectra and spatially resolved reflectance, we combined a Raman excitation channel, a Raman detection channel, and one channel for the spatially resolved reflectance measurement. The three channels were imaged onto the sample with the optics described in Fig. 2. The laser power for Raman and spatially resolved reflectance measurements was . 2.2.1.Raman spectroscopyThe resonance Raman scattering of carotenoids was excited at using an argon laser with a cleanup filter (max line, AHF Analysentechnik). For the detection, we used a CCD-based spectrometer (Triax 320, Jobin Yvon) with the CCD chip cooled to . We used a grating and an exposure time of for each spectrum. For the following analysis, we evaluated the Raman peak at , which originates from the stretch vibrations of -carotene. In addition, we evaluated the fraction of the elastically scattered laser light at , which passed the long pass filter (edge basic, AHF Analysentechnik). To obtain the Raman peak values from the fluorescence background, we performed a nonlinear least square fit of a Lorentzian-shaped curve to the spectra (gnuplot). 2.2.2.Spatially resolved reflectance measurementThe spatially resolved reflectance measurement was based on a customized fiber bundle, which consisted of 58 multimode silica fibers, each in diameter with (see Fig. 2, Fiberguide). One central illumination fiber was surrounded by three concentric rings of detection fibers ( , 19, and 30). The fiber bundle was imaged onto the sample surface at an angle of to avoid detection of specular reflection and to permit the arrangement of the three channels in one plane. The resulting source-detector distances were 0.4, 0.8, and , which were the average distances of the slightly distorted image. The fibers of each detection ring were bundled and fixed in close proximity to photodiodes (PDB-C160SM, Laser Components). The amplification was varied automatically for each readout to optimize analog-to-digital (A/D) conversion. The background signal was acquired automatically before each measurement and subtracted. The reflectance signals were normalized to the illumination power, which was measured simultaneously (see Fig. 2). The optical properties of the sample were measured at the Raman excitation wavelength with the argon laser at and close to the Raman peak with a frequency-doubled Nd:YAG laser at . Using lasers instead of white light was motivated by the fact that simple photodiodes could be used for the measurement without spectral filtering of the reflected light. 2.2.3.Lookup table-based determination of and from spatially resolved reflectanceIn order to infer the optical properties of the samples, we used a lookup table (LUT),18 which is a table with forward simulations of the reflectance for various combinations of and , for each of the source-detector distances given in Fig. 2. The data were produced with the Monte Carlo code described in Sec. 2.1 for the purely elastically scattered photons using a detector at the source-detector distances given above. The sample geometry, anisotropy factor, and refractive index were constant within the LUT. The calibration factors for the measured reflectance values, which accounted for the transmission properties of the optics, differences in alignment, and the detector response, were obtained by measuring a set of three calibration phantoms with known optical properties. The measured reflectance values were then compared to the LUT by a customized program.19 The program output was and of the best-fitting LUT-data. When the method was evaluated for 15 phantoms with and , the optical properties differed from the values obtained by integrating sphere measurements (Sec. 2.3).The average deviation was and , and the average absolute deviation was 11% and 10% . This range of optical properties corresponds to the results of preliminary in vivo measurements of skin. Since the setup was designed for in vivo skin measurements, the illumination power was limited to . Phantoms with a higher than could not be measured with the current implementation of our setup because there was no reflectance signal detectable at the second distance using . 2.3.Phantom Preparation and CharacterizationTissue phantoms with to and to were produced based on silicone (Wacker, RT 601) using red silopren color paste (GE Bayer Silicones) as an absorber and titanium dioxide as a scatterer. To ensure that the highly lipophilic -carotene (Fluka) was completely dissolved, it was first dissolved in methylene chloride and paraffin (Vaseline, Unilever) and then mixed with the silicone. The final -carotene concentration was in all phantoms. The contribution of -carotene to the of the silicone phantom was at , as measured in a transparent phantom. The Raman spectra were measured on the day of preparation. A thin layer from each phantom mixture was used to determine the optical properties with an integrating sphere spectrometer (Lambda 900, Perkin Elmer). The total transmission, diffuse transmission, and diffuse reflectance were measured and used as the input parameters for an inverse Monte Carlo algorithm20 to yield , , and . For comparison with the spatially resolved reflectance measurements, was calculated from and . 3.Dependency of the Raman Signal on the Macroscopic Optical PropertiesFirst we investigated how the Raman peak at [ bonds, see Fig. 3 ] depends on the optical properties of the silicone phantom, which correspond to typical values for skin in the VIS.7 The Raman signal dropped roughly proportional to , while the influence of was fairly small, see Fig. 3. The optical properties of the phantoms are the mean optical coefficients at excitation and Stokes wavelength. These coefficients were measured with the integrating sphere spectrometer, since the whole parameter range was not accessible using our implementation of the spatially resolved reflectance setup. Since the silicone color paste contributed to the scattering coefficient of the phantom, of the phantoms was not constant within one series of increasing absorber concentration. Next we compared our phantom measurements with simulations [Fig. 3]. The simulation results of the Raman signal were all scaled with the same factor to fit the experimental results. The factor includes the unknown in vivo Raman cross section as well as setup-specific constants. The trends of the experimental values corresponded well to the simulations. However, the experimental and simulated values did not match for all phantoms within the given errors, which could be explained by errors incurred by variations in the sample preparation. We further performed Monte Carlo simulations to investigate the dependency of the Raman signal on and in a larger parameter range, as with the phantom measurements (Fig. 4 ). For our measurement geometry, was roughly proportional to in the VIS ( and ). For smaller (typical for the NIR), the curves deviated significantly from this behavior. When the same simulation was performed for a pencil-beam excitation and a semi-infinite detection geometry, which is a typical imaging geometry, the behavior continued into the NIR parameter range (data not shown). When the excitation and detection geometry corresponded to a fiber, the dependency was weaker, as for the measurement geometry used here (data not shown). This can be explained by the influence of the measurement geometry on the pathlength distribution of the detected Raman photons, which determines the influence of the optical properties on the Raman scattered light. For our measurement geometry, the influence of in the NIR is larger than the VIS. In addition to the Raman signal, we also analyzed the elastic diffuse reflectance of the laser light, which was extracted from the same spectra as the Raman signals (Fig. 5 ). The elastic signal was much more sensitive to variations of and less sensitive to . As a consequence, a simple ratio of Raman and elastic diffuse reflectance signal still depended on and and therefore was not sufficient for correction in this parameter range, which covered tissue optical properties in the VIS. 4.Determination of by Correction FunctionsTo link Raman signals from turbid samples directly to Raman scattering coefficients, we had to correct the Raman signal for the influence of the macroscopic optical properties. The measured or calculated relative Raman signal, which depended on , , and but also on the sample geometry, measurement geometry, and sensitivity of the setup, is called . We assumed that the application of a dimensionless correction function , which was sample and measurement geometry-dependent, led to The right-hand side of Eq. 1 is independent from and of the sample and can be interpreted as the signal that could be measured if the sample were transparent, similar to that defined in Refs. 12, 13, 14. The factor , in units of , depends only on the geometry and the sensitivity of the setup. The constant is the average pathlength of excitation photons in the transparent sample for the given sample and measurement geometry. In the case of a simulation (data not shown) where a transparent layer of Raman scatterers with refractive index was measured in backscattering geometry using an ideal detector , was obtained and corresponded to the layer thickness. The factor can be obtained by measuring a calibration sample of the same geometry and known and . For many applications, it might be sufficient to correct relative to a reference sample of the same geometry with and byNext we will describe two types of the correction function . Motivated by the existing literature, we investigated correction functions that utilize the elastic diffuse reflectance, which is measured in the same geometry as the Raman signal (Sec. 4.1). We also suggest an alternative correction function that is based only on the simulations of the Raman signal (Sec. 4.2).4.1.Correction Utilizing the Elastic Diffuse Reflectance SignalIt would be very convenient if a simple ratio of and could completely remove the influence of the optical properties. However, as shown for our measurement geometry in Fig. 6 , the ratio only decreases the influence of . The influence of increases except for the parameter range and , where the influence of decreases as well. This explains why the ratio improved the results in Ref. 14, where a measurement spot of similar size was used . In Fig. 7 , is plotted versus for various combinations of and and for three different measurement geometries. Obviously, cannot be described as a function of only without knowledge of the other variables. However, it was shown recently for measurements in the NIR that , with , can be described by a function of only; see Refs. 12, 13. This leads to the correction function , where is a geometry-dependent function of the diffuse reflectance , and is a length to make dimensionless as defined in Eq. 1. Using our Monte Carlo model, we searched for similar empirical correction functions and evaluated their validity with respect to the range of optical properties and the measurement geometry. While we found some correction functions that were only valid in the semi-infinite measurement geometry or the VIS parameter range, only and could be used for our measurement geometry in the whole parameter range investigated here. Since and performed very similarly, we present only the application of in Fig. 7. It is obvious that works in the semi-infinite as well as in our measurement geometry, but not for a measurement spot with a diameter of . In order to find a function for the whole parameter range, we applied a 4th-order polynomial to the log of the data. We accounted for different optical properties at the excitation and Stokes wavelength by using the mean optical properties. A wavelength-dependent variation of can be accounted for by using the geometric mean instead. When was applied to the simulated data in Fig. 4, the variation of the data decreased from 109% to 6%. This means that the accuracy of the to be determined is strongly improved. When was applied to our silicone phantom data (Sec. 2.3), the variation of the data decreased from 88% to 21%. The remaining variability of the signal could result from variations in the sample preparation and from the fact that the diffuse reflectance was only measured at the excitation wavelength. Therefore, an experimental investigation of the correction effect was limited with this setup. 4.2.Correction Based on Raman Signal SimulationsSince our Monte Carlo model for Raman measurements in turbid media corresponded sufficiently well with our experiments, we also used directly the forward simulation of to obtain the correction function according to Eq. 1. For a relative correction according to Eq. 2, we simulated To minimize the computational effort during the measurement, the correction factors were simulated once for various combinations of and and stored in a table. To account for the different optical properties at the excitation and Stokes wavelength, can be replaced by the geometric mean of at both wavelengths. However, if the values are simulated a priori, using four variables would drastically increase the size of this table. Therefore, it is more convenient to use the Raman signal at the mean optical properties of excitation and Stokes wavelength, which we found to be equal to the letter approach in the case of .The effect of this correction procedure on the simulated data led to a straight line (not shown). When applied to the phantom data, the variation of the data decreased from 88% to 16%. The remaining variability of the signal could be due to variations in the sample preparation. 5.Determination of Molecular Raman Cross Section or Raman Scatterer ConcentrationIn Sec. 4 we described how could be determined once the calibration factors were measured. Another way to determine independent from a calibration measurement is by comparing the simulated ratio of the Raman to elastic signal (see Fig. 6) to the measured ratio given in Fig. 5. We estimated the magnitude of to be for our phantoms with . For known Raman scatterer concentration c and , the differential molecular Raman scattering cross-section can be calculated by We obtained for our silicone phantoms. Compared to the results of Berdyugin 21 ( , measured in cyclohexane at excitation and concentration of ), our result was one order of magnitude larger. This can be at least partially explained by the enhanced resonance excitation at . Surprisingly, the results of Tian 22 varied up to 3 orders of magnitude depending on the -carotene concentration , which was dissolved with pyridine and diluted with water. Thus, for the absolute quantification of -carotene in biomedical applications, more investigations of the molecular resonance Raman cross-section in different biological matrices and for different concentrations should be performed.Assuming the molecular Raman scattering cross-section can be found in the literature, the concentration of the Raman scatterer can be calculated by Eq. 3 using the measured . 6.Summary and DiscussionWe introduced a method to combine Raman and spatially resolved reflectance spectroscopy that allows the measurement of the optical properties together with the Raman spectrum. To determine the absolute Raman scatterer coefficients independent from the optical properties, we compared to different methods to correct for the turbidity-induced signal variation. One correction method utilized empirical functions that are dependent on the elastic diffuse reflectance but also require knowledge of the optical properties. An alternative method used a Monte Carlo model to generate and propagate Raman signals in turbid media to directly calculate the correction factors, regardless of any functional relationship. Both correction methods reduced the turbidity-induced signal variation by 95% or more for our simulated data, and by 76% or more for our phantom data. Both correction methods need to be adjusted once to the excitation and the detection using phantom measurements or Monte Carlo simulations. Since the empirical equations still require knowledge of the optical properties, we see no advantage to this method. The correction method based on the Monte Carlo simulation, however, does not require the measurement of elastic reflectance in the Raman measurement geometry and can even be used in geometries such as a measurement spot, where an empirical correction function has not been described yet. For the data presented here, we used correction functions obtained for samples that were homogenous within the sampling volume. However, this simplification may be critical if a sampling volume includes different layers, like the epidermis and the dermis in skin measurements. The influence of the sample structure on the correction function is the subject of a current investigation. From the determined , we calculated the molecular Raman scattering cross-section of a given -carotene concentration in our silicone phantoms. For a known , the absolute Raman scatterer concentration can be determined from . For unknown molecular Raman cross-sections, which is a possible scenario especially in complex biological matrices, the methods can also correct Raman signal variations induced by sample-to-sample turbidity variation, which allows the quantitative comparison of results obtained from different individuals, tissues, or tissue sites. Although we took advantage of the outstanding high-resonance Raman scattering cross-section of -carotene, the correction methods are transferable to any other Raman scatterer. For the correction of multiple Raman peaks at different wavelengths, the optical property measurement can be easily performed using white light instead of lasers. In addition to biomedical applications, the correction method might also be beneficial in the food industry or environmental research, where quantitative measurements of analytes independent of turbidity variations are necessary. Furthermore, we suggest that the correction procedures may be applied similarly to fluorescence measurements. AcknowledgmentsWe thank M. Meinke and M. Darvin (Center of Experimental and Applied Cutaneous Physiology, Charité-Universitätsmedizin Berlin) for useful discussions. The project was funded by the Senate of Berlin and by the European Union (EFRE, FKZ 10138595). ReferencesA. S. Haka, K. E. Shafer-Peltier, M. Fitzmaurice, J. Crowe, R. R. Dasari, and M. S. Feld,
“Diagnosing breast cancer by using Raman spectroscopy,”
Proc. Natl. Acad. Sci. U.S.A., 102
(35), 12371
–12376
(2005). https://doi.org/10.1073/pnas.0501390102 0027-8424 Google Scholar
M. D. Keller, E. M. Kanter, C. A. Lieber, S. K. Majumder, J. Hutchings, D. L. Ellis, R. B. Beaven, N. Stone, and A. Mahadevan-Jansen,
“Detecting temporal and spatial effects of epithelial cancers with Raman spectroscopy,”
Dis. Markers, 25 323
–337
(2008). 0278-0240 Google Scholar
A. Nijssen, S. Koljenovic, T. C. Bakker Schut, P. J. Caspers, and G. J. Puppels,
“Towards oncological application of Raman spectroscopy,”
J. Biophotonics, 2 29
–36
(2009). https://doi.org/10.1002/jbio.200810055 Google Scholar
S. W. E. van de Poll, K. Kastelijn, T. C. Bakker Schut, C. Strijder, G. Pasterkamp, G. J. Puppels, and A. van der Laarse,
“On-line detection of cholesterol and calcification by catheter based Raman spectroscopy in human atherosclerotic plaque ex vivo,”
Heart, 89
(9), 1078
–1082
(2003). https://doi.org/10.1136/heart.89.9.1078 1355-6037 Google Scholar
A. J. Berger, T. W. Koo, I. Itzkan, G. Horowitz, and M. S. Feld,
“Multicomponent blood analysis by near-infrared Raman spectroscopy,”
Appl. Opt., 38 2916
–2926
(1999). https://doi.org/10.1364/AO.38.002916 0003-6935 Google Scholar
D. Arifler, C. MacAulay, M. Follen, and R. Richards-Kortum,
“Spatially resolved reflectance spectroscopy for diagnosis of cervical precancer: Monte Carlo modeling and comparison to clinical measurements,”
J. Biomed. Opt., 11
(6), 064027
(2006). https://doi.org/10.1117/1.2398932 1083-3668 Google Scholar
E. Salomatina, B. Jiang, J. Novak, and A. N. Yaroslavsky,
“Optical properties of normal and cancerous human skin in the visible and near-infrared spectral range,”
J. Biomed. Opt., 11 064026
(2006). https://doi.org/10.1117/1.2398928 1083-3668 Google Scholar
R. Reif, O. A’Amar, and I. J. Bigio,
“Analytical model of light reflectance for extraction of the optical properties in small volumes of turbid media,”
Appl. Opt., 46
(29), 7317
–7328
(2007). https://doi.org/10.1364/AO.46.007317 0003-6935 Google Scholar
M. Pilz, S. Honold, and A. Kienle,
“Determination of the optical properties of turbid media by measurements of the spatially resolved reflectance considering the point-spread function of the camera system,”
J. Biomed. Opt., 13
(5), 054047
(2008). https://doi.org/10.1117/1.2983675 1083-3668 Google Scholar
S. J. Tinnemans, M. H. F. Kox, T. A. Nijhuis, T. Visser, and B. M. Weckhuysen,
“Real time quantitative Raman spectroscopy of supported metal oxide catalysts without the need of an internal standard,”
Phys. Chem. Chem. Phys., 7 211
–216
(2005). https://doi.org/10.1039/b414427a 1463-9076 Google Scholar
D. N. Waters,
“Raman spectroscopy of powders—Effects of light absorption and scattering,”
Spectrochim. Acta Mol. Spectros., 50 1833
–1840
(1994). https://doi.org/10.1016/0584-8539(94)80196-7 Google Scholar
W. C. Shih, K. L. Bechtel, and M. S. Feld,
“Intrinsic Raman spectroscopy for quantitative biological spectroscopy Part 1: Theory and simulations,”
Opt. Express, 16
(17), 12726
–12736
(2008). 1094-4087 Google Scholar
K. L. Bechtel, W. C. Shih, and M. S. Feld,
“Intrinsic Raman spectroscopy for quantitative biological spectroscopy Part 2: Experimental applications,”
Opt. Express, 16
(17), 12737
–12745
(2008). 1094-4087 Google Scholar
I. Barman, G. P. Singh, R. R. Dasari, and M. S. Feld,
“Turbidity-corrected Raman spectroscopy for blood analyte detection,”
Anal. Chem., 81 4233
–4240
(2009). https://doi.org/10.1021/ac8025509 0003-2700 Google Scholar
G. Zonios and A. Dimou,
“Light scattering spectroscopy of human skin in vivo,”
Opt. Express, 17
(3), 1256
–1267
(2009). https://doi.org/10.1364/OE.17.001256 1094-4087 Google Scholar
C. Reble, I. Gersonde, S. Andree, J. Helfmann, and G. Illing,
“Correction of Raman signals for tissue optical properties,”
Proc. SPIE, 7368 73680C
(2009). https://doi.org/10.1117/12.831595 0277-786X Google Scholar
L. H. Wang, S. L. Jacques, and L. Q. Zheng,
“MCML Monte Carlo modeling of photon transport in multi-layered tissues,”
Comput. Methods Programs Biomed., 47 131
–146
(1995). https://doi.org/10.1016/0169-2607(95)01640-F 0169-2607 Google Scholar
N. Rajaram, T. H. Nguyen, and J. W. Tunnell,
“Lookup table-based inverse model for determining optical properties of turbid media,”
J. Biomed. Opt., 13
(5), 050501
(2008). https://doi.org/10.1117/1.2981797 1083-3668 Google Scholar
S. Andree, C. Reble, J. Helfmann, I. Gersonde, and G. Illing,
“Spatially resolved reflectance to deduce absorption and reduced scattering coefficients,”
Proc. SPIE, 7368 73680I
(2009). https://doi.org/10.1117/12.831601 0277-786X Google Scholar
M. Friebel, A. Roggan, G. Müller, and M. Meinke,
“Determination of optical properties of human blood in the spectral range using Monte Carlo simulations with hematocrit-dependent effective scattering phase functions,”
J. Biomed. Opt., 11
(3), 034021
(2006). https://doi.org/10.1117/1.2203659 1083-3668 Google Scholar
V. V. Berdyugin, K. Y. Burshtein, and P. P. Shorygin,
“Cross section of resonance Raman scattering of light by polyenes,”
Opt. Spectrosc., 61
(2), 192
–194
(1986). 0030-400X Google Scholar
Y.-J. Tian, J. Zuo, L.-Y. Zhang, Z.-W. Li, S.-Q. Gao, and G.-H. Lu.,
“Study of resonance Raman cross section of aqueous -carotene at low concentrations,”
Appl. Phys. B, 87
(4), 727
–730
(2007). https://doi.org/10.1007/s00340-007-2633-7 0946-2171 Google Scholar
|