6 April 2012 Errata: Localization capability and limitation of electron-multiplying charge-coupled, scientific complementary metal-oxide semiconductor, and charge-coupled devices for superresolution imaging
Author Affiliations +
J. of Biomedical Optics, 17(4), 049801 (2012). doi:10.1117/1.JBO.17.4.049801
Abstract
Abstract unavailable.
Quan, Zeng, and Huang: Errata: Localization capability and limitation of electron-multiplying charge-coupled, scientific complementary metal-oxide semiconductor, and charge-coupled devices for superresolution imaging

This article [J. Biomed. Opt. 15(6), 066005 (2010)] was originally published online on Nov. 19, 2010 with an error in Eqs. (12) and (20). These equations contain errors in the factor of the second term. The error was due to the reason that electron multiplication process to background noise was not being fully included into the equation. The corrected form of these identical equations is

Δx2=2s2+a2/12ϕN+16πs4(ϕIb)a2(ϕN)2.

The correction has a negligible effect on Fig. 6, but has no effect on the remainder of the manuscript.

Tingwei Quan, Shaoqun Zeng, Zhen-Li Huang, "Errata: Localization capability and limitation of electron-multiplying charge-coupled, scientific complementary metal-oxide semiconductor, and charge-coupled devices for superresolution imaging," Journal of Biomedical Optics 17(4), 049801 (6 April 2012). http://dx.doi.org/10.1117/1.JBO.17.4.049801