19 March 2012 Methodological considerations for global analysis of cellular FLIM/FRET measurements
Author Affiliations +
Abstract
Global algorithms can improve the analysis of fluorescence energy transfer (FRET) measurement based on fluorescence lifetime microscopy. However, global analysis of FRET data is also susceptible to experimental artifacts. This work examines several common artifacts and suggests remedial experimental protocols. Specifically, we examined the accuracy of different methods for instrument response extraction and propose an adaptive method based on the mean lifetime of fluorescent proteins. We further examined the effects of image segmentation and a priori constraints on the accuracy of lifetime extraction. Methods to test the applicability of global analysis on cellular data are proposed and demonstrated. The accuracy of global fitting degrades with lower photon count. By systematically tracking the effect of the minimum photon count on lifetime and FRET prefactors when carrying out global analysis, we demonstrate a correction procedure to recover the correct FRET parameters, allowing us to obtain protein interaction information even in dim cellular regions with photon counts as low as 100 per decay curve.
© 2012 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2012/$25.00 © 2012 SPIE
Nur Aida Abdul Rahim, Roger D. Kamm, Peter T. C. So, and Serge B. Pelet "Methodological considerations for global analysis of cellular FLIM/FRET measurements," Journal of Biomedical Optics 17(2), 026013 (19 March 2012). https://doi.org/10.1117/1.JBO.17.2.026013
Published: 19 March 2012
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Fluorescence resonance energy transfer

Infrared imaging

Photon counting

Proteins

Luminescence

Fluorescence lifetime imaging

Image segmentation

Back to Top