12 May 2014 Improved light collection and wavelet de-noising enable quantification of cerebral blood flow and oxygen metabolism by a low-cost, off-the-shelf spectrometer
Author Affiliations +
Abstract
Broadband continuous-wave near-infrared spectroscopy (CW-NIRS) is an attractive alternative to time-resolved and frequency-domain techniques for quantifying cerebral blood flow (CBF) and oxygen metabolism in newborns. However, efficient light collection is critical to broadband CW-NIRS since only a small fraction of the injected light emerges from any given area of the scalp. Light collection is typically improved by optimizing the contact area between the detection system and the skin by means of light guides with large detection surface. Since the form-factor of these light guides do not match the entrance of commercial spectrometers, which are usually equipped with a narrow slit to improve their spectral resolution, broadband NIRS spectrometers are typically custom-built. Nonetheless, off-the-shelf spectrometers have attractive advantages compared to custom-made units, such as low cost, small footprint, and wide availability. We demonstrate that off-the-shelf spectrometers can be easily converted into suitable instruments for deep tissue spectroscopy by improving light collection, while maintaining good spectral resolution, and reducing measurement noise. The ability of this approach to provide reliable cerebral hemodynamics was illustrated in a piglet by measuring CBF and oxygen metabolism under different anesthetic regimens.

1.

Introduction

Prevention of brain injury during the perinatal period remains a primary concern in neonatal intensive care practice due to potentially devastating and life-long neurological consequences.1 Near-infrared spectroscopy (NIRS) has been considered an ideal tool for bedside brain monitoring in newborns due to its portability, safety, and sensitivity to deep tissue oxygenation and hemodynamics.2 As well, the relatively thin layer of superficial tissues on the newborn’s head circumvents problems associated with extra-cerebral signal contamination, which is a major obstacle to adult studies.35 However, quantification of cerebral oxygenation and hemodynamics by NIRS requires separating the effects of light scatter from absorption.

Time and frequency domain NIRS methods have been developed to separate scattering and absorption effects, but the instrumentations are relatively expensive and complex.6,7 A cheaper and technically simpler approach is continuous-wave (CW) broadband NIRS.8 Acquisition of broadband spectra has the added advantage of allowing simultaneous quantification of multiple chromophores.9,10 With this technique, the concentrations of tissue chromophores, both endogenous and exogenous, are quantified by first or second differential analyses, which exploit the hyper-spectral content of the measurements. This approach has been shown to provide accurate measurements of cerebral blood flow (CBF) and oxygen metabolism in a neonatal animal model (piglets).1113 More recently, it has been adapted to clinical studies to investigate potential pharmacological effects on cerebral hemodynamics and metabolism in preterm infants.14

Critical to broadband CW-NIRS monitoring is efficient light collection since only a small fraction of the injected light emerges from the probed tissue due to absorption and strong scattering. The most common method for improving light collection is by optimizing the contact area between the detection system and the probed tissue by means of light guides with large detection surface.15 Since the size of these light guides is not compatible with commercial spectrometers, which are typically equipped with an entrance slit to improve their spectral resolution, deep tissue spectrometers are usually custom-built.10,1416 Nevertheless, off-the-shelf spectrometers have attractive advantages compared to custom-made units, such as low cost, small footprint, and wide availability.17 Some spectrometer manufacturers have attempted to tackle the issue of the compatibility with large detection light guides by removing the entrance slit and using a single fiber to guide light to the spectrometer.18 This approach is, however, impeded by its inability to concomitantly provide good spectral resolution and large contact area with the probed tissue. That is, either the spectrometer is equipped with a large-diameter fiber to optimize light collection, at the expense of spectral resolution, or a small-diameter fiber is used for good spectral resolution, at the expense of sensitivity.

In this paper, we propose an approach that provides high sensitivity, by improving light collection and reducing measurement noise, while maintaining good spectral resolution. Light collection was increased by removing the slit and using the output of the detection light guide as entrance to the spectrometer. To maintain good spectral resolution, the output of the fiber bundle was designed to mimic the typical form-factor of a spectrometer’s entrance slit. The quality of the data was further improved by de-noising the measurements using a wavelet-based algorithm. The spectral resolution, wavelength accuracy, and noise characteristics of the modified instrument were quantified and compared to the specifications of a custom-built spectrometer that has been previously validated for deep tissue spectroscopy.11 The ability of the modified spectrometer, augmented by the wavelet de-noising approach and differential spectroscopy data analysis, to quantify cerebral hemodynamics was illustrated in a piglet by measuring CBF and oxygen metabolism under different anesthetic regimens.

2.

Materials and Methods

2.1.

Instrument Modifications

The approach described here is of general purpose and can be applied to most off-the-shelf spectrometers, although the principle was demonstrated using the QE65000 unit from Ocean Optics, Dunedin, Florida. The spectrometer’s slit was removed, to increase light throughput to the detector, so that the output of the detection fiber bundle became the entrance to the spectrometer. To maintain good spectral resolution, the output of the fiber bundle was designed to mimic the typical shape of an entrance slit; the proximal end of the fiber bundle was made of two columns of seven fibers (Fig. 1), forming a rectangular strip of 0.215mm×1mm (width×height). The tip of the fiber bundle with the rectangular output was positioned at the entrance of the spectrometer, with the two columns of fibers parallel to the vertical dimension of the CCD detector, and locked in place using an SMA905 connector. At the distal end of the bundle, the 125-μm core diameter silica fibers were closely packed (left of Fig. 1) to reduce the footprint of the probe. Note that the detection fiber bundle was custom built by Fiberoptics Technology, Inc. (Pomfret, Connecticut) according to the specific design requirements outlined above.

Fig. 1

Schematic of the detection fiber bundle; the 14 fibers (core diameter=125μm) were closely packed at one end (left) and arranged into two columns of seven fibers at the other end (right) to form a rectangular output acting as the spectrometer’s entrance slit. The end of the fiber bundle with the rectangular strip was secured in front of the spectrometer with the SMA905 connector.

JBO_19_5_057007_f001.png

2.2.

Characterization of the Modified Spectrometer

The spectral resolution and wavelength accuracy of the modified spectrometer were characterized and compared to the specifications of a custom-built unit that has been validated for deep tissue spectroscopy.11,12 The latter was equipped with a front-illuminated CCD camera (S7010-1007, Hamamatsu Photonics, Bridgewater, New Jersey and Japan) cooled to 73°C by a four-stage Peltier cooling system (Wright Instruments Ltd, Enfield, United Kingdom).13 At this temperature, the dark current was 7.7×103electron/pixel/second and the mean readout noise was 6.9 counts. The CCD was operated as an array detector by binning the 124 pixels along the vertical direction.13

The detector of the off-the-shelf spectrometer was also a two-dimensional CCD array (Hamamatsu S7031-1006; 1044 columns by 64 rows of pixels, pixel size of 24μm2, detection range between 200–1100 nm, 90% quantum efficiency at 700-nm peak). Likewise, the pixels along the vertical dimension were also binned to operate the CCD as an array detector.18 The major sources of noise (readout, shot, and dark noise) were characterized using an approach similar to the method described by Zonios.19

2.2.1.

Readout noise

To measure the readout noise, the integration time was set to 8 ms—the fastest operating rate of the spectrometer—and the CCD detector was cooled to its lowest temperature (20°C) to minimize dark noise. The photoelectron (shot) noise was also minimized by covering the entrance of the spectrometer so that no light reached the CCD detector. One hundred spectra were acquired successively in this configuration. The readout noise, defined as the standard deviation from the mean intensity of 100 measurements, was calculated for each pixel.

2.2.2.

Shot noise

To evaluate the shot noise, the integration time was again set to 8 ms and the CCD temperature to 20°C. The output of a broadband light source (Ocean Optics HL-2000-HP) was adjusted by combinations of neutral density filters to obtain 13 different levels of light intensity. One hundred spectra were acquired at each intensity level. For every pixel, the mean and standard deviation (i.e., the noise) were calculated for each set of one hundred measurements. From these data, the detector gain and readout noise were calculated by fitting the measurements with the following equation:19

Eq. (1)

N=NR2+S/G,
where N is the measured noise which is composed of the readout noise (NR), the mean signal intensity (S), and pixel gain (G).

2.2.3.

Dark noise

Dark noise was evaluated by covering the entrance of the spectrometer, to minimize shot noise, and acquiring one hundred spectra at different integration times (8 ms to 20 s). In this configuration, the measured cumulative noise (readout noise and dark noise) can be modeled as19

Eq. (2)

N=NR2+T×nd/G2,
where N is the measured noise, NR is the readout noise, T is the integration time, G is the detector gain, and nd is the dark electron accumulation rate. By measuring the standard deviation (at one pixel) for different integration times, the dark electron accumulation rate can be determined by fitting the measurement with Eq. (2). The dependence of dark noise on temperature was also evaluated by repeating the procedure with the CCD cooled from 10 to 20°C in step of 2°C.

2.2.4.

Stability test

The stability of the modified spectrometer was also tested and compared to that of the custom-built unit. Spectra from a broadband the light source (Ocean Optics HL-2000-HP) were acquired continuously over an hour by each spectrometer using an integration time of 400 ms, the typical duration used in our in vivo measurements. The normalized standard deviation of the area under the spectra was calculated for each spectrometer to evaluate its stability.

3.

Data Analysis

3.1.

Wavelet De-Noising

The de-noising algorithm was adopted from our previous works.20,21 First, the noise’s variance was stabilized by applying the Anscombe transformation to the measured spectra.22 This step efficiently transformed the Poisson noise in the data into white Gaussian noise. Second, the white noise was filtered out by wavelet de-noising using the MATLAB® function wden with Daubechies wavelets of order 10.20 Finally, an inverse transformation (Anscombe inverse exact unbiased transformation) was applied to the de-noised data to obtain an estimate of the original noise-free spectra.22

3.2.

Quantification of Cerebral Chromophore Concentrations

Near-infrared light propagation in turbid media can be described by the radiative transfer equation which, in the case of the present application, can be reduced to its diffusion approximation.23 Since the extra-cerebral tissue layer in newborn piglets is very thin (1 to 2 mm), the head can be approximated as a semi-infinite homogeneous medium.10 As such, the results presented in this paper were obtained by analyzing the piglet’s head measurements using the following solution to the diffusion equation for a semi-infinite homogenous medium10,24

Eq. (3)

ψ=2S(4π)2Dexp[ρ(μaD)1/2]ρ3[1+ρ(μaD)1/2](z0+zb)×{z+3D[1(z0+zb)2+3z22ρ2(3+ρ2μaD1+ρ(μaD)1/2)]},
where D=1/[3*(μa+μs)] is the diffusion coefficient, ρ is the radial distance between the emission and detection probes, S is the source strength in photon per second, μa and μs are the absorption and reduced scattering coefficients, respectively, and z0 and zb are the boundary approximation parameters and are defined by the following expressions:

Eq. (4)

z0=(μa+μs)1,

Eq. (5)

zb=1+Reff1Reff2D,
where Reff represents the fraction of photons that is internally reflected at the boundary, and zb was computed using the approach described by Haskell et al.25

Brain chromophore concentrations were quantified using a modified version of the method described in Ref. 10. First, the reflectance R(λ) was calculated from the spectrum measured on the piglet’s head, the dark spectrum of the spectrometer, and the reference signal obtained with the light source by

Eq. (6)

R(λ)=log10(spectrumλdarkλreferenceλdarkλ).

Next, the second spectral derivative of Eq. (3) was fit to the second spectral derivative of R(λ) between 825 and 850 nm to extract the water concentration. The choice of this wavelength range was guided by previous findings showing that the second spectral derivative of the scattering coefficient and the other tissue chromophores do not change significantly in this spectral window.26 The recovered water concentration was then used as a fixed parameter in the fitting of the first spectral derivative of R(λ) with the first derivative of Eq. (3), to obtain the scattering coefficient and the remaining tissue chromophores. The objective function of the fitting was the norm of the difference between the measured and theoretical derivative spectra. However, instead of an exhaustive search, as in Ref. 10, a least-squares minimization routine based on fminsearchbnd27 was developed to find the fitting parameters that minimized the objective function. This adaptation provided 20-fold increase in the speed of the algorithm compared to the previous approach.10 The fitting parameters included the concentration of oxy-hemoglobin (HbO2), deoxy-hemoglobin (Hb), water (H2O), and fat as well as the reduced scattering coefficient. The upper and lower boundaries of these parameters were set using a range that spans the published literature values for brain tissue and are shown in Table 1.28 Note that both first and second spectral derivatives of the reflectance measurements and theoretical model were computed with a subroutine built in MATLAB® using the diff function. Furthermore, the goodness-of-fit was evaluated by computing the deviation of the measurements from the model, using the summed square of the residuals:

Eq. (7)

SSE=n=1N(RiR^i)2,
where Ri is the measured quantity (first or second spectral derivative) and R^i is the corresponding theoretical prediction.

Table 1

Lower and upper limits of the fitting parameters, i.e., reduced scattering coefficient (μs′) and concentrations of water (H2O), deoxy-hemoglobin (Hb), and oxy-hemoglobin (HbO2).

H2O (%)Hb (μM)HbO2 (μM)Fat (%)μs (mm1)
Lower limit7051000.4
Upper limit904080501.5

For the dynamic contrast-enhanced measurements (described in Sec. 4), the spectra acquired during the indocyanine green (ICG) injection were analyzed by fixing the scattering coefficient and endogenous chromophore concentrations to their preinjection values (i.e., assuming that they did not change during the 2 min duration of the bolus-tracking experiment) so that the ICG concentration was the sole fitting parameter.

4.

In Vivo Measurement

The animal experiments were conducted to illustrate the ability of the modified spectrometer, augmented by the wavelet de-noising and the spectral derivative approach, to measure CBF and oxygen metabolism in newborns. The study was approved by the Animal Use Subcommittee of the Canadian Council on Animal Care at Western University and was conducted according to its guidelines. A Duroc-cross piglet was delivered from a local supplier on the morning of the experiment. The animal was anaesthetized with 1.75% isoflurane (3% during surgery), tracheotomized, and mechanically ventilated on an oxygen/medical air mixture. A femoral artery was catheterized to monitor blood pressure and intermittently collect blood samples for gas and glucose analysis. An ear vein was catheterized to inject the light-absorbing intravascular tracer, ICG (Sigma-Aldrich, St. Louis, Missouri). The piglet was placed in the prone position and a probe holder was strapped to its head to hold the emission and detection probes at a separation of 3 cm. This separation has been shown to provide good sensitivity to the brain with probes positioned on the scalp.29 The piglet was allowed to stabilize for 1 h after surgery before any measurements were acquired.

To investigate the effects of the noise reduction, one hundred spectra were acquired (while the animal was in stable condition) and de-noised. Thereafter, the two sets of data (raw and de-noised spectra) were analyzed with the method described in Sec. 3.2. The mean and standard deviation of the reduced scattering coefficient and cerebral chromophore concentrations, obtained from the two sets of spectra, were compared to reveal the effects of the de-noising.

Dynamic contrast-enhanced measurements were acquired while the piglet was subject to three different anesthetic regimens: isoflurane (2%), light isoflurane (0.75%) plus nitrous (2L/min), and propofol infusion (3.0mL/h). Hemodynamics measurements were acquired at each anesthetic level and required an intravenous bolus injection of ICG (0.05 mg dissolved in 0.5 mL sterile water). Tissue and arterial ICG concentration curves were concomitantly measured for 2 min by the modified spectrometer and a dye densitometer (DDG 2001, Nihon Kohden, Japan), respectively. CBF was calculated from these curves and combined with the tissue and arterial oxygen saturations to extract the cerebral metabolic rate of oxygen (CMRO2), as previously described.30

5.

Results

Figure 2 displays two spectra from the same neon light source measured by the modified spectrometer (QE65000) and the custom-built unit equipped with the S7010-1007 Hamamatsu CCD detector. This graph shows that subtle spectral features in the neon spectrum were better resolved by the modified spectrometer than the custom-built unit. For example, the neon spectrum measured by the modified spectrometer clearly showed one small peak at 718 nm and a higher one at 724 nm while these two peaks were not resolvable in the spectrum acquired by the custom-built unit. As well, the full-width at-half maximum of the highest peak (at 703 nm) and those of two other representative peaks (one on either side of the 703-nm peak) of the neon light source are displayed in Table 2. These results show that although the modified spectrometer has a relatively lower sensitivity than the custom-built unit in the 725- to 900-nm wavelength range, the spectral lines measured by the former were consistently narrower than those acquired with the latter. Since the spectral dispersion of both spectrometers was 0.37nm/pixel and they also have similar pixel size (24μm), the better resolution of the modified spectrometer is likely due to the slightly smaller width of the detection fiber bundle (215μm) compared to that of the slit (250μm) of the custom-built unit. Furthermore, the wavelength accuracy of the modified instrument was determined by comparing the positions of the spectral lines of a low-pressure mercury–argon calibration light source measured by the spectrometer to the values provided by the manufacturer. The results revealed that the modified spectrometer could measure the positions of the spectral lines with an accuracy of 1nm over the entire wavelength range.

Fig. 2

Comparison of spectra from the same neon light source measured by the modified spectrometer and the custom-built unit.

JBO_19_5_057007_f002.png

Table 2

Full-width at half-maximum (FWHM) of the highest peak (at 703 nm) and those of two other representative peaks (one on either side of the 703-nm peak) of the neon light source measured by each spectrometer.

Peak position (nm)640703811
Custom built spectrometer (FWHM in nm)7.85.55.4
Modified spectrometer (FWHM in nm)5.74.74.8

Figure 3 displays the readout noise measured with the CCD cooled to 20°C; the average value was 1.2 counts, in good agreement with the 1.5 counts provided by the manufacturer. Figure 4 shows the dependence of the shot noise on signal intensity; the solid curve is the theoretical prediction of Eq. (1) and the dots are the measured noise. The model fit yielded a detector gain (G) of 23 and a readout noise (NR) of 1.2 counts, which is in excellent agreement with the value obtained from the data presented in Fig. 3.

Fig. 3

Readout noise acquired at the shortest integration time (8 ms) with the CCD cooled to 20°C to minimize dark noise. The entrance to the spectrometer was also blocked to eliminate shot noise.

JBO_19_5_057007_f003.png

Fig. 4

Experimental noise (blue dots) plotted as a function of the mean signal intensity. The solid curve is the best fit of Eq. (1).

JBO_19_5_057007_f004.png

Figure 5 depicts the expected dependency of dark noise on integration time and detector temperature. The dark electron accumulation rate (nd) obtained by fitting the noise measurements with Eq. (2) was 251 and 84counts2/pixel/s at 10 and 18°C, respectively. Table 3 displays the readout noise and nd at CCD temperatures from 10 to 20°C. As expected, the readout noise remained constant while there was a positive correlation between nd and temperature. Figure 5(a) shows that at 10°C the measured noise was dominated by readout noise up to integration time of 0.6 s, while the dark noise quickly dominated for longer integration times. Cooling the CCD to 18°C extended the integration time up to 3 s without substantial increase in dark noise [Fig. 5(b)].

Fig. 5

The cumulative noise (i.e., readout and dark) in one pixel (blue dots) plotted as a function of integration time at (a) 10°C and (b) 18°C. The solid curves are the best fit to Eq. (2).

JBO_19_5_057007_f005.png

Table 3

Dark electron accumulation rate and readout noise measured at different temperatures of the CCD detector.

Temperature (°C)NR (counts)nd (counts2/pixel/s)
101.25251
121.24181
141.27160
161.26121
181.2584
201.2466

Figure 6(a) illustrates the noise in the raw measurements [same data as in Fig. 5(a)] and after de-noising. This figure demonstrates that de-noising reduced the readout noise from 1.3 to <0.5 counts, as evident by comparing the noise values for integration times <0.6s. In addition, the total noise remained <1 count for higher integration times, showing that de-noising also reduced the dark electron accumulation rate. Figure 6(b) displays a reflectance spectrum from raw and de-noised data measured on the piglet’s head; de-noising reduced the variability in the data without altering the shape of the reflectance curve. These graphs collectively show that the de-noising can efficiently reduce the three main sources of noise without altering the integrity of the data. Furthermore, the SSE of the residuals for the first spectral derivatives was 13×104 for the raw data and 8×104 for the de-noised spectrum, and for the second derivatives they were 7×105 and 3.5×105 for the raw and de-noised data, respectively. The consistent improved goodness-of-fit of the de-noised data for both the first and second spectral derivatives further confirms the ability of the de-noising algorithm to reduce measurement noise.

Fig. 6

(a) Effects of de-noising at different integration times with the CCD cooled to 10°C. The dots and diamonds are the noise in the raw data and after de-noising, respectively. (b) Reflectance spectrum obtained from the piglet’s head before and after de-noising. The inset is the residual between the two reflectance spectra.

JBO_19_5_057007_f006.png

As a final note on the instrument characteristics, the stability test revealed that the normalized standard deviation from the mean intensity, over 1 h of measurements, was 1.4×104 and 7.9×103 for the custom-built spectrometer and the modified unit, respectively. This result indicates that the two instruments have similar stability.

The first and second spectral derivatives of a typical reflectance spectrum acquired on the piglet’s head and the corresponding de-noised data are depicted in Fig. 7. Included in the graphs are the best fits to the spectral derivatives of the theoretical model. Table 4 shows the mean and standard deviation of the reduced scattering coefficient (at 800 nm) and cerebral concentrations of H2O, Hb, and HbO2, obtained by fitting the two sets of one hundred reflectance spectra (raw and de-noised). The mean values of the fitting parameters obtained from both sets of spectra were the same, but less variability was evident after de-noising. Note that we do not report fat concentration values since they were close to zero for both data sets, likely due to a lack of distinctive spectral features of this chromophore in the wavelength range covered by the spectrometer (see Sec. 6).

Fig. 7

First (a) and second (c) spectral derivatives of a typical reflectance spectrum measured on the piglet’s head and the best fit to the model first and second spectral derivatives of the diffusion approximation. (b and d) Same data after de-noising with their corresponding fit to the model.

JBO_19_5_057007_f007.png

Table 4

Mean and standard deviation (SD) of μs′ (at 800 nm), and concentrations of H2O, Hb, and HbO2 obtained by fitting each reflectance spectrum of the two data sets (raw and de-noised).

Raw reflectanceDe-noised reflectance
MeanSDMeanSD
H2O (%)855852
Hb (μM)130.5130.2
HbO2 (μM)265.3252.1
μs (mm1)0.790.090.790.05

Figure 8 shows the ICG concentration curves measured under the three anesthetics. Since the amount of ICG injected was approximately the same for all measurements, these curves suggest that CBF was higher under isoflurane and lower under propofol. These qualitative observations are in agreement with the absolute CBF recovered from the ICG curves and displayed in Table 5. Furthermore, the CMRO2 was higher under nitrous, the lightest anesthetic, and a lower CMRO2 value was measured under propofol, which was the strongest anesthetic.

Fig. 8

Brain indocyanine green concentration curves measured under different anesthetic regimens.

JBO_19_5_057007_f008.png

Table 5

Cerebral blood flow (CBF), cerebral metabolic rate of oxygen (CMRO2), oxy- and deoxy-hemoglobin concentrations, and tissue oxygen saturation (ScO2) under different anesthetics.

CBF (mL/min/100g)CMRO2 (mLO2/min/100g)Hb (μM)HbO2 (μM)ScO2 (%)
Isoflurane291.5142565
Nitrous251.8151651
Propofol211.1112064

6.

Discussion

In contrast to commercially available CW-NIRS units that are typically equipped with few wavelengths to monitor relative oxygenation, broadband CW-NIRS instruments have the ability to quantify the concentrations of tissue chromophores. As such, broadband CW-NIRS is an attractive alternative to other quantitative approaches, such as time-resolved and frequency-domain techniques because it is technically simpler and less expensive. The interest in quantitative techniques is increasing since they provide the ability to measure key parameters of cerebral physiology—CBF and cerebral oxidative metabolism—that are potentially more sensitive markers of brain health than relative oxygenation.3133 Broadband CW-NIRS has been successfully used to quantify cerebral oxygenation and hemodynamics in both experimental and clinical studies.10,11,14,31,3335 However, to date, all broadband CW-NIRS applications have involved custom-built spectrometers. Given that off-the-shelf spectrometers with good noise characteristics are becoming increasingly affordable, the focus of this work was to explore the feasibility of using such a spectrometer to quantify CBF and oxygen metabolism.

A general approach for improving the sensitivity of these spectrometers, while maintaining good spectral resolution and reducing the instrumental noise was introduced. The principle was illustrated with a compact low-cost spectrometer in which the entrance slit was removed to increase the light throughput to the detector. Removing the slit could potentially degrade spectral resolution and wavelength accuracy. To avoid these pitfalls while maintaining efficient light collection, which is critical in broadband NIRS, the modified spectrometer was equipped with a detection light guide made of flexible optical fibers arranged in a configuration that mimics the typical form-factor of an entrance slit. This simple approach enabled to transform a low-cost spectrometer designed for dilute solutions into a suitable instrument for deep tissue spectroscopy as shown in this paper.

The modified spectrometer was characterized and the results showed that its spectral characteristics were comparable to those of a high quality custom-built unit, specifically designed for deep tissue spectroscopy (Fig. 2). Spectral accuracy was accessed by comparing the positions of the spectral lines of a calibration lamp measured by the modified spectrometer to the known values provided by the manufacturer of the light source. The wavelength accuracy of the instrument was found to be 1nm which is adequate for tissue spectroscopy, given the wide spectral features of HbO2, Hb, H2O, fat, and ICG in the near-infrared range.36 As well, the similarity of the positions of the spectral lines of the neon light source measured by the two spectrometers (Fig. 2) is another illustration of the wavelength accuracy of the new instrument.

Since the modified spectrometer was equipped with a CCD array, the major contributors to detector noise were readout noise, photoelectron noise, and dark noise.19,37,38 These noise components were quantified and the results showed that photoelectron noise was the main contributor. Note that in the photoelectron noise characterization, it was not possible to distinguish noise due to instability in the output power of the light source from pure shot noise. As such, the values of photoelectron noise reported here are, in fact, the combined contributions of both shot noise and light source instability. To further improve the noise characteristics of the system, all spectra were subsequently de-noised by a wavelet algorithm that was previously developed for time-resolved NIRS.20 In this study, it was evident that de-noising spectral data improved the precision of the chromophore estimate (Table 4) with negligible effects on the shape of the reflectance spectrum (Fig. 6). Another important feature of the de-noising is its ability to significantly reduce dark noise even for integration times up to 20 s [Fig. 6(a)], which is appealing in low-light conditions such as at large source-detector separations in reflectance measurements or in the case of trans-cranial illumination. The de-nosing approach could also be adopted in other fields of optics dealing with weak spectrum measurements.37

One approach for quantifying cerebral chromophore concentrations from broadband CW-NIRS spectra is by second differential analysis.39 Taking the second derivative of a spectrum removes the effects of scatter since it varies approximately linearly with wavelength. The differential pathlength and concentrations of chromophores can then be determined by assuming a known water concentration in brain, which is typically 80% to 90% depending on age and species.35 However, a limitation with this approach is the inability to quantify the HbO2 concentration due to the lack of any distinctive features in its second derivative spectrum.8,39 A similar method that can quantify tissue HbO2 concentration was proposed by Myers et al.40 Their approach was based on an algebraic simplification of the second spectral derivative of tissue reflectance acquired at four wavelengths and required an empirical calibration with known tissue oxygen saturation. An alternative method that does not require any internal or empirical calibration is to model the wavelength-dependency of scatter with a power law and fit the broadband data with an analytical solution to the diffusion approximation.10,41,42 In this paper, the fitting of the second spectral derivative was used to obtain the water concentration. The recovered H2O concentration was subsequently used to constrain the fitting of the first derivative to obtain estimates of the other chromophores.

When applied to in vivo spectra, the second differential approach resulted in an estimated brain H2O concentration of 85%, in agreement with the expected value.39 As well, the reduced scattering coefficient and concentrations of Hb and HbO2 recovered from the first derivative analysis were also in agreement with previously reported values.43 Furthermore, the hemodynamics results were consistent with the relative strengths of the different anesthetics; the lowest CMRO2 was measured under propofol, the strongest anesthetic. Although their physiological interpretation is beyond the scope of this paper, the CBF and CMRO2 results illustrate the ability of this approach to quantify cerebral hemodynamics.

A limitation of this study is the lack of a direct comparison of the specs of the spectrometer before and after the modifications. Such information would certainly provide a quantitative assessment of the signal-to-noise ratio improvement provided by the modifications and confirm our qualitative observations. However, when compared to a similar (nonmodified QE65000) spectrometer, the detection efficiency of the modified spectrometer was approximately 45% higher. Another potential limitation is that in the data analysis, the piglet’s head was treated as a semi-infinite homogenous medium. Although this is a rough approximation, such approach is widely used in the biomedical optics community to analyze NIRS measurements of neonates and piglets, and has been shown to produce blood flow estimates that strongly correlate with measurements obtained by state-of-the-art medical imaging modalities.11,13

A disadvantage with the low-cost spectrometer used in this paper was that it does not cover the 900 to 1000 nm spectral range. Since the absorption peak of fat is within this wavelength range, this spectrometer was unable to provide reliable estimates of tissue fat content. Nor can it be used to monitor tissue bound-water content or temperature by using the large water peak in this spectral window.4446 This limitation can, however, be mitigated by choosing an off-the-shelf unit that covers the 900 to 1000 nm spectral range or by slightly rotating the grating of the present spectrometer to encompass this wavelength range.

7.

Conclusion

This study demonstrated that an off-the-shelf spectrometer can be easily modified to make it suitable for quantifying deep tissue chromophore concentrations. The noise and spectral characteristics of the modified instrument were characterized and were found comparable to those of a custom-built spectrometer optimized for deep tissue spectroscopy. Combining the data analysis algorithms with the modified spectrometer provided a low-cost technique for measuring cerebral oxygenation and hemodynamics. Given the small footprint of these spectrometers, they could easily be adapted to intensive care units and also used as field-unit neuromonitors.

Acknowledgments

The authors would like to thank Jennifer Hadway and Laura Morrison for their assistance with the animal experiments and Anna Diop for her help in the instrument characterization. This research was supported by the Canadian Institutes of Health Research.

References

1. 

A. Solaet al., “Potential for protection and repair following injury to the developing brain: a role for erythropoietin?,” Pediatr. Res., 57 (5), 110R –117R (2005). http://dx.doi.org/10.1203/01.PDR.0000159571.50758.39 PEREBL 0031-3998 Google Scholar

2. 

G. GreisenT. LeungM. Wolf, “Has the time come to use near-infrared spectroscopy as a routine clinical tool in preterm infants undergoing intensive care?,” Philos. Trans. R. Soc., A, 369 (1955), 4440 –4451 (2011). http://dx.doi.org/10.1098/rsta.2011.0261 PTRMAD 1364-503X Google Scholar

3. 

F. Goraet al., “Noninvasive measurement of cerebral blood flow in adults using near-infrared spectroscopy and indocyanine green: a pilot study,” J. Neurosurg. Anesthesiol., 14 (3), 218 –222 (2002). Google Scholar

4. 

H. Owen-Reeceet al., “Use of near infrared spectroscopy to estimate cerebral blood flow in conscious and anaesthetized adult subjects,” Br. J. Anaesth., 76 (1), 43 –48 (1996). http://dx.doi.org/10.1093/bja/76.1.43 BJANAD 0007-0912 Google Scholar

5. 

J. T. Elliottet al., “Quantitative measurement of cerebral blood flow in a juvenile porcine model by depth-resolved near-infrared spectroscopy,” J. Biomed. Opt., 15 (3), 037014 (2010). http://dx.doi.org/10.1117/1.3449579 JBOPFO 1083-3668 Google Scholar

6. 

F. Bevilacquaet al., “Broadband absorption spectroscopy in turbid media by combined frequency-domain and steady-state methods,” Appl. Opt., 39 (34), 6498 –6507 (2000). http://dx.doi.org/10.1364/AO.39.006498 APOPAI 0003-6935 Google Scholar

7. 

M. Diopet al., “Comparison of time-resolved and continuous-wave near-infrared techniques for measuring cerebral blood flow in piglets,” J. Biomed. Opt., 15 (5), 057004 (2010). http://dx.doi.org/10.1117/1.3488626 JBOPFO 1083-3668 Google Scholar

8. 

S. J. MatcherC. E. Cooper, “Absolute quantification of deoxyhaemoglobin concentration in tissue near infrared spectroscopy,” Phys. Med. Biol., 39 (8), 1295 –1312 (1994). http://dx.doi.org/10.1088/0031-9155/39/8/008 PHMBA7 0031-9155 Google Scholar

9. 

S. J. MatcherM. CopeD. T. Delpy, “Use of the water absorption spectrum to quantify tissue chromophore concentration changes in near-infrared spectroscopy,” Phys. Med. Biol., 38 177 –196 (1993). http://dx.doi.org/10.1088/0031-9155/39/1/011 PHMBA7 0031-9155 Google Scholar

10. 

H. Z. Yeganehet al., “Broadband continuous-wave technique to measure baseline values and changes in the tissue chromophore concentrations,” Biomed. Opt. Express, 3 (11), 2761 –2770 (2012). http://dx.doi.org/10.1364/BOE.3.002761 BOEICL 2156-7085 Google Scholar

11. 

D. W. Brownet al., “Quantitative near infrared spectroscopy measurement of cerebral hemodynamics in newborn piglets,” Pediatr. Res., 51 (5), 564 –570 (2002). http://dx.doi.org/10.1203/00006450-200205000-00004 PEREBL 0031-3998 Google Scholar

12. 

K. M. Tichaueret al., “Measurement of cerebral oxidative metabolism with near-infrared spectroscopy: a validation study,” J. Cereb. Blood Flow. Metab., 26 (5), 722 –730 (2005). http://dx.doi.org/10.1038/sj.jcbfm.9600230 JCBMDN 0271-678X Google Scholar

13. 

M. Diopet al., “A broadband continuous-wave multichannel near-infrared system for measuring regional cerebral blood flow and oxygen consumption in newborn piglets,” Rev. Sci. Instrum., 80 (5), 054302 (2009). http://dx.doi.org/10.1063/1.3123347 RSINAK 0034-6748 Google Scholar

14. 

R. Aroraet al., “Preservation of the metabolic rate of oxygen in preterm infants during indomethacin therapy for closure of the ductus arteriosus,” Pediatr. Res., 73 (6), 713 –718 (2013). http://dx.doi.org/10.1038/pr.2013.53 PEREBL 0031-3998 Google Scholar

15. 

R. Yanget al., “Monitoring angiogenesis using a human compatible calibration for broadband near-infrared spectroscopy,” J. Biomed. Opt., 18 (1), 016011 (2013). http://dx.doi.org/10.1117/1.JBO.18.1.016011 JBOPFO 1083-3668 Google Scholar

16. 

H. Xuet al., “Magnetic-resonance-imaging-coupled broadband near-infrared tomography system for small animal brain studies,” Appl. Opt., 44 (11), 2177 –2188 (2005). http://dx.doi.org/10.1364/AO.44.002177 APOPAI 0003-6935 Google Scholar

17. 

E. WrightK. St. LawrenceM. Diop, “Conversion of a low cost off-the-shelf spectrometer into a suitable instrument for deep tissue spectroscopy,” Proc. SPIE, 8573 85730V (2013). http://dx.doi.org/10.1117/12.2005624 Google Scholar

18. 

“QE65000 Scientific-grade Spectrometer,” (2010). http://www.oceanoptics.com/products/qe65000.asp Google Scholar

19. 

G. Zonios, “Noise and stray light characterization of a compact CCD spectrophotometer used in biomedical applications,” Appl. Opt., 49 (2), 163 –169 (2010). http://dx.doi.org/10.1364/AO.49.000163 APOPAI 0003-6935 Google Scholar

20. 

M. DiopK. St Lawrence, “Deconvolution method for recovering the photon time-of-flight distribution from time-resolved measurements,” Opt. Lett., 37 (12), 2358 –2360 (2012). http://dx.doi.org/10.1364/OL.37.002358 OPLEDP 0146-9592 Google Scholar

21. 

M. DiopK. St Lawrence, “Improving the depth sensitivity of time-resolved measurements by extracting the distribution of times-of-flight,” Biomed. Opt. Express, 4 (3), 447 –459 (2013). http://dx.doi.org/10.1364/BOE.4.000447 BOEICL 2156-7085 Google Scholar

22. 

M. MäkitaloA. Foi, “Optimal inversion of the Anscombe transformation in low-count Poisson image denoising,” IEEE Trans. Image Process., 20 (1), 99 –109 (2011). http://dx.doi.org/10.1109/TIP.2010.2056693 IIPRE4 1057-7149 Google Scholar

23. 

F. Martelliet al., Light Propagation through Biological Tissue and Other Diffusive Media: Theory, Solutions, and Software, Bellingham, WA (2010). Google Scholar

24. 

S. FantiniM. FranceschiniE. Gratton, “Semi-infinite-geometry boundary problem for light migration in highly scattering media: a frequency-domain study in the diffusion approximation,” J. Opt. Soc. Am. B, 11 (10), 2128 –2138 (1994). http://dx.doi.org/10.1364/JOSAB.11.002128 JOBPDE 0740-3224 Google Scholar

25. 

R. C. Haskellet al., “Boundary conditions for the diffusion equation in radiative transfer,” J. Opt. Soc. Am. A, 11 (10), 2727 –2741 (1994). http://dx.doi.org/10.1364/JOSAA.11.002727 JOAOD6 0740-3232 Google Scholar

26. 

H. Zabihi-Yeganeh, “Non-invasive measurement of tissue chromophore concentrations and cerebral blood flow using broad-band continuous wave near infrared spectroscopy,” (2012). Google Scholar

28. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 R37 –R61 (2013). http://dx.doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 Google Scholar

29. 

E. OkadaD. T. Delpy, “Near-infrared light propagation in an adult head model. II. Effect of superficial tissue thickness on the sensitivity of the near-infrared spectroscopy signal,” Appl. Opt., 42 (16), 2915 –2922 (2003). http://dx.doi.org/10.1364/AO.42.002915 APOPAI 0003-6935 Google Scholar

30. 

K. Verdecchiaet al., “Quantifying the cerebral metabolic rate of oxygen by combining diffuse correlation spectroscopy and time-resolved near-infrared spectroscopy,” J. Biomed. Opt., 18 (2), 027007 (2013). http://dx.doi.org/10.1117/1.JBO.18.2.027007 JBOPFO 1083-3668 Google Scholar

31. 

K. M. Tichaueret al., “Assessing the severity of perinatal hypoxia-ischemia in piglets using near-infrared spectroscopy to measure the cerebral metabolic rate of oxygen,” Pediatr. Res., 65 (3), 301 –306 (2009). http://dx.doi.org/10.1203/PDR.0b013e318194faa6 PEREBL 0031-3998 Google Scholar

32. 

D. A. BoasM. A. Franceschini, “Haemoglobin oxygen saturation as a biomarker: the problem and a solution,” Philos. Trans. R. Soc., A, 369 (1955), 4407 –4424 (2011). http://dx.doi.org/10.1098/rsta.2011.0250 PTRMAD 1364-503X Google Scholar

33. 

K. M. Tichaueret al., “Near-infrared spectroscopy measurements of cerebral blood flow and oxygen consumption following hypoxia-ischemia in newborn piglets,” J. Appl. Physiol., 100 (3), 850 –857 (2006). http://dx.doi.org/10.1152/japplphysiol.00830.2005 JAPYAA 0021-8987 Google Scholar

34. 

T. S. Leunget al., “Theoretical investigation of measuring cerebral blood flow in the adult human head using bolus indocyanine green injection and near-infrared spectroscopy,” Appl. Opt., 46 (10), 1604 –1614 (2007). http://dx.doi.org/10.1364/AO.46.001604 APOPAI 0003-6935 Google Scholar

35. 

C. E. Cooperet al., “The noninvasive measurement of absolute cerebral deoxyhemoglobin concentration and mean optical path length in the neonatal brain by second derivative near infrared spectroscopy,” Pediatr. Res., 39 32 –38 (1996). http://dx.doi.org/10.1203/00006450-199601000-00005 PEREBL 0031-3998 Google Scholar

36. 

S. PrahlS. L. Jacques, “Optical properties spectra,” (2001) http://omlc.ogi.edu/spectra/ Google Scholar

37. 

Z. Caiet al., “Quantification and elimination of the CCD dark current in weak spectrum measurement by modulation and correlation method,” Proc. SPIE, 7850 785005 (2010). http://dx.doi.org/10.1117/12.869369 PSISDG 0277-786X Google Scholar

38. 

J. J. Davenportet al., “Noise analysis of a CCD based ultra-violet spectrometry system,” Proc. SPIE, 8439 84391P (2012). http://dx.doi.org/10.1117/12.922328 PSISDG 0277-786X Google Scholar

39. 

R. SpringettY. SakataD. T. Delpy, “Precise measurement of cerebral blood flow in newborn piglets from the bolus passage of indocyanine green,” Phys. Med. Biol., 46 (8), 2209 –2225 (2001). http://dx.doi.org/10.1088/0031-9155/46/8/312 PHMBA7 0031-9155 Google Scholar

40. 

D. E. Myerset al., “Noninvasive method for measuring local hemoglobin oxygen saturation in tissue using wide gap second derivative near-infrared spectroscopy,” J. Biomed. Opt., 10 (3), 034017 (2005). http://dx.doi.org/10.1117/1.1925250 JBOPFO 1083-3668 Google Scholar

41. 

R. Nachabéet al., “Estimation of lipid and water concentrations in scattering media with diffuse optical spectroscopy from 900 to 1,600 nm,” J. Biomed. Opt., 15 (3), 037015 (2010). http://dx.doi.org/10.1117/1.3454392 JBOPFO 1083-3668 Google Scholar

42. 

O. PucciV. ToronovK. St Lawrence, “Measurement of the optical properties of a two-layer model of the human head using broadband near-infrared spectroscopy,” Appl. Opt., 49 (32), 6324 –6332 (2010). http://dx.doi.org/10.1364/AO.49.006324 APOPAI 0003-6935 Google Scholar

43. 

S. Ijichiet al., “Quantification of cerebral hemoglobin as a function of oxygenation using near-infrared time-resolved spectroscopy in a piglet model of hypoxia,” J. Biomed. Opt., 10 (2), 024026 (2005). http://dx.doi.org/10.1117/1.1899184 JBOPFO 1083-3668 Google Scholar

44. 

S. H. Chunget al., “In vivo water state measurements in breast cancer using broadband diffuse optical spectroscopy,” Phys. Med. Biol., 53 (23), 6713 –6727 (2008). http://dx.doi.org/10.1088/0031-9155/53/23/005 PHMBA7 0031-9155 Google Scholar

45. 

S. H. Chunget al., “Non-invasive tissue temperature measurements based on quantitative diffuse optical spectroscopy (DOS) of water,” Phys. Med. Biol., 55 (13), 3753 –3765 (2010). http://dx.doi.org/10.1088/0031-9155/55/13/012 PHMBA7 0031-9155 Google Scholar

46. 

S. H. Chunget al., “Non-invasive measurement of deep tissue temperature changes caused by apoptosis during breast cancer neoadjuvant chemotherapy: a case study,” J. Innovative Opt. Health Sci., 4 (4), 361 –372 (2011). http://dx.doi.org/10.1142/S1793545811001708 JIOHAA 1793-7205 Google Scholar

Biographies of the authors are not available.

© 2014 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2014/$25.00 © 2014 SPIE
Mamadou Diop, Eric Wright, Vladislav Toronov, Ting-Yim Lee, and Keith St. Lawrence "Improved light collection and wavelet de-noising enable quantification of cerebral blood flow and oxygen metabolism by a low-cost, off-the-shelf spectrometer," Journal of Biomedical Optics 19(5), 057007 (12 May 2014). https://doi.org/10.1117/1.JBO.19.5.057007
Published: 12 May 2014
Lens.org Logo
CITATIONS
Cited by 24 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spectroscopy

Wavelets

Tissue optics

Oxygen

Charge-coupled devices

Spectral resolution

Chromophores

Back to Top