27 May 2015 Errata: Lateral and axial measurement differences between spectral-domain optical coherence tomography systems
Author Affiliations +
J. of Biomedical Optics, 20(5), 059802 (2015). doi:10.1117/1.JBO.20.5.059802
Abstract
This PDF file contains the errata for “JBO Vol. 20 Issue 05 Paper JBO-2015-0515-ERR” for JBO Vol. 20 Issue 05
Folgar, Yuan, Farsiu, and Toth: Errata: Lateral and axial measurement differences between spectral-domain optical coherence tomography systems

This article [J. Biomed. Opt. 19(1), 016014 (2014)] was originally published online on 17 January 2014 with errors in the row and column labels of Tables 4 and 5 on p. 5. The corrected tables are reprinted below.

Table 4

Conversion factors for lateral measurements across platforms.

Convert from this platform
Lateral scalingZeiss Cirrus™Heidelberg Spectralis™Bioptigen Envisu™Bioptigen SDOIS
Convert to this platformZeiss Cirrus™0.9800.9910.994
Heidelberg Spectralis™1.0201.0111.014
Bioptigen Envisu™1.0090.9891.003
Bioptigen SDOIS1.0060.9860.997

Table 5

Conversion factors for axial measurements across platforms.

Convert from this platform
Axial scalingZeiss Cirrus™Heidelberg Spectralis™Bioptigen Envisu™Bioptigen SDOIS
Convert to this platformZeiss Cirrus™1.0111.0371.174
Heidelberg Spectralis™0.9901.0251.160
Bioptigen Envisu™0.9640.9751.132
Bioptigen SDOIS0.8520.8620.884

This article was corrected online on 18 May 2015.