17 February 2015 Noninvasive glucose sensing by transcutaneous Raman spectroscopy
Author Affiliations +
J. of Biomedical Optics, 20(5), 051036 (2015). doi:10.1117/1.JBO.20.5.051036
Abstract
We present the development of a transcutaneous Raman spectroscopy system and analysis algorithm for noninvasive glucose sensing. The instrument and algorithm were tested in a preclinical study in which a dog model was used. To achieve a robust glucose test system, the blood levels were clamped for periods of up to 45 min. Glucose clamping and rise/fall patterns have been achieved by injecting glucose and insulin into the ear veins of the dog. Venous blood samples were drawn every 5 min and a plasma glucose concentration was obtained and used to maintain the clamps, to build the calibration model, and to evaluate the performance of the system. We evaluated the utility of the simultaneously acquired Raman spectra to be used to determine the plasma glucose values during the 8-h experiment. We obtained prediction errors in the range of ∼1.5−2  mM. These were in-line with a best-case theoretical estimate considering the limitations of the signal-to-noise ratio estimates. As expected, the transition regions of the clamp study produced larger predictive errors than the stable regions. This is related to the divergence of the interstitial fluid (ISF) and plasma glucose values during those periods. Two key contributors to error beside the ISF/plasma difference were photobleaching and detector drift. The study demonstrated the potential of Raman spectroscopy in noninvasive applications and provides areas where the technology can be improved in future studies.
© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE)
Wei-Chuan Shih, Kate L. Bechtel, Mihailo V. Rebec, "Noninvasive glucose sensing by transcutaneous Raman spectroscopy," Journal of Biomedical Optics 20(5), 051036 (17 February 2015). http://dx.doi.org/10.1117/1.JBO.20.5.051036
JOURNAL ARTICLE
5 PAGES


SHARE
KEYWORDS
Glucose

Raman spectroscopy

Calibration

Ear

Blood

Error analysis

Statistical analysis

RELATED CONTENT

Study and implementation of a soft x ray 100 eV...
Proceedings of SPIE (June 27 2006)
Repetitively pulsed plasma illumination sources
Proceedings of SPIE (December 12 1997)
Design of a multisensor optical surface scanner
Proceedings of SPIE (October 06 1994)
Knomes: Knowledge Monitoring Expert System
Proceedings of SPIE (October 31 1987)

Back to Top