2 July 2015 Analysis of the change in peak corneal temperature during excimer laser ablation in porcine eyes
Samuel Arba Mosquera, Shwetabh Verma
Author Affiliations +
Abstract
The objective is to characterize the impact of different ablation parameters on the thermal load during corneal refractive surgery by means of excimer laser ablation on porcine eyes. One hundred eleven ablations were performed in 105 porcine eyes. Each ablation was recorded using infrared thermography and analyzed mainly based on the two tested local frequencies (40 Hz, clinical local frequency; 1000 Hz, no local frequency). The change in peak corneal temperature was analyzed with respect to varying ablation parameters [local frequency, system repetition rate, pulse energy, optical zone (OZ) size, and refractive correction]. Transepithelial ablations were also compared to intrastromal ablations. The average of the baseline temperature across all eyes was 20.5°C±1.1 (17.7°C to 22.2°C). Average of the change in peak corneal temperature for all clinical local frequency ablations was 5.8°C±0.8 (p=3.3E53 to baseline), whereas the average was 9.0°C±1.5 for all no local frequency ablations (p=1.8E−35 to baseline, 1.6E−16 to clinical local frequency ablations). A logarithmic relationship was observed between the changes in peak corneal temperature with increasing local frequency. For clinical local frequency, change in peak corneal temperature was comparatively flat (r2=0.68 with a range of 1.5°C) with increasing system repetition rate and increased linearly with increasing OZ size (r2=0.95 with a range of 2.4°C). Local frequency controls help maintain safe corneal temperature increase during excimer laser ablations. Transepithelial ablations induce higher thermal load compared to intrastromal ablations, indicating a need for stronger thermal controls in transepithelial refractive procedures.
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Samuel Arba Mosquera and Shwetabh Verma "Analysis of the change in peak corneal temperature during excimer laser ablation in porcine eyes," Journal of Biomedical Optics 20(7), 078001 (2 July 2015). https://doi.org/10.1117/1.JBO.20.7.078001
Published: 2 July 2015
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Laser ablation

Control systems

Excimer lasers

Cornea

Surgery

Temperature metrology

Cameras

Back to Top