9 May 2016 Cerenkov and radioluminescence imaging of brain tumor specimens during neurosurgery
Antonello Enrico Spinelli, Marco Paolo Schiariti, Chiara Maria Grana, Mahila Ferrari, Marta Cremonesi, Federico Boschi
Author Affiliations +
Abstract
We presented the first example of Cerenkov luminescence imaging (CLI) and radioluminescence imaging (RLI) of human tumor specimens. A patient with a brain meningioma localized in the left parietal region was injected with 166 MBq of 90Y-DOTATOC the day before neurosurgery. The specimens of the tumor removed during surgery were imaged using both CLI and RLI using an optical imager prototype developed in our laboratory. The system is based on a cooled electron multiplied charge coupled device coupled with an f/0.95 17-mm C-mount lens. We showed for the first time the possibility of obtaining CLI and RLI images of fresh human brain tumor specimens removed during neurosurgery.

Cerenkov luminescence imaging (CLI) and radioluminescence imaging (RLI) are becoming well-established methods for preclinical in vivo small animal optical imaging, as described in recent reviews.13 CLI and RLI are based, respectively, on the detection of Cerenkov radiation or scintillation light in the visible range. CLI was mainly applied to image the biodistribution of radiopharmaceuticals labeled with beta plus46 and beta minus7,8 emitters. RLI was developed to obtain the biodistribution of alpha9 and gamma emitters10 and can also be used in combination with CLI to enhance the light output.11

Recent examples of CLI applications to humans were presented in Refs. 12 and 13. In Ref. 12, Cerenkov radiation from the thyroid of a patient treated with I131 for hyperthyroidism was imaged, while Ref. 13 presented an endoscopic Cerenkov luminescence imaging approach to detect gastric cancers in patients injected with 2-[18F]fluoro-2-deoxy-D-glucose.

In this work, we present an application of CLI and RLI for the analysis of ex vivo fresh tumor specimens removed during neurosurgery. The main goal of such an approach is to provide a fast molecular imaging estimation of tumor uptake (almost in real time) during surgical procedures. CLI and RLI images were obtained by placing the tumor specimens in an optical imager prototype developed in our laboratory.14

The detector used in the prototype is a cooled (80°C) electron multiplied charge coupled device (EMCCD; Andor iXon Ultra) coupled with an f/0.95 17-mm C-mount lens (Schneider Optics). The EMCCD has a back-illuminated 512×512 sensor with pixel size equal to 16  μm, resulting in a detector with total dimensions equal to 8.2×8.2  mm2.

The smallest achievable field of view (FOV) is 6.1×6.1  cm2, resulting in an image pixel size of 120  μm. To exclude ambient light, the system is mounted on a black light-tight enclosure, in which an adjustable stage is placed for positioning the specimen.

All the CLI and RLI images were acquired using binning = 2 and 5 min of exposure time. A photographic black-and-white image (binning = 2 and exposure time = 0.01 s) was always acquired before any emission image to provide a reference for CLI and RLI. A color image of the sample was also acquired using a conventional camera to show the amount of blood present in the sample.

RLI data were acquired by placing a slab of 1-cm plastic scintillator on top of the specimen. To avoid any contamination, the scintillator was placed 1  mm from the sample. Considering that the plastic scintillator slab is almost transparent, the collected light is thus the sum of Cerenkov and scintillator light produced in the slab. This acquisition scheme (CLI+RLI) was adopted mainly to maximize the light collected by the CCD.

A patient with a grade 1 brain meningioma localized in the left parietal region was injected with 166 MBq of Y90-DOTATOC the day before surgery. The patient participated in an experimental protocol of a radioguided surgery technique for cerebral tumors with the use of a beta minus radiation probe,15 which uses Y90-DOTATOC for diagnostic purposes—contrary to its traditional use with therapeutical quantities for peptide receptor radionuclide therapy. In the new technique in Ref. 15, the Y90-DOTATOC injected activity is calculated based on a diagnostic positron emission tomography–computed tomography (PET-CT) imaging with the analogue Ga68-DOTATOC radiopharmaceutical that allows evaluation of the activity concentrations in tumors and thus can establish the activity to be administered for detection with the beta minus probe. In our case, the tumor to nontumor standard uptake value ratio was 25. The required Y90-DOTATOC activities are of a few hundreds of MBq (typically 185 MBq, 166 MBq in our patient), and the ensuing tumor activity concentrations are also suitable for the study of CLI and RLI.

As shown in Ref. 16, meningiomas express somatostatin receptors; thus, the Ga68-DOTATOC and Y90-DOTATOC uptake in the tumor is specific.

Figure 1(a) shows the Ga68-DOTATOC image of our patient, acquired two weeks before surgery. This time is sufficient to guarantee the absence of any interaction between the two radiotracers. A similar time interval was previously considered in a study aimed at comparing the biodistribution of Y90-DOTATOC in the same patients under two renal protection conditions.17

Fig. 1

(a) The coronal, sagittal, and axial PET-CT images of the patient injected with Ga68-DOTATOC. The PET images were used to estimate the activity concentration of Y90-DOTATOC in the tumor during surgery. (b) The T2 MR images of the patient acquired without contrast medium.

JBO_21_5_050502_f001.png

Ga68 and Y90 have very different physical half-lives, but the labeling of the same peptide—leading to Ga68-DOTATOC and Y90-DOTATOC—allows us to assume similar, if not identical, kinetic processes in tumors and normal organs as well, as applies for In111-DOTATOC used as a surrogate for Y90-DOTATOC, as sustained by the joint IAEA, EANM, and SNMMI practical guidance on peptide receptor radionuclide therapy (PRRT).18 The use of Ga68-DOTATOC to foresee the outcome of therapy with Lu177- or Y90-DOTATOC has been ascertained in the community of PRRT and has been discussed in several papers.1922 Moreover, in the protocol of Ref. 15, the presumed activity of Y90 derived from the Ga68 estimates has also been compared with the Monte Carlo simulations of the beta probe sensitivity, showing results in good agreement (data not shown from a paper submitted by Solfaroli Camillocci et al.).

The diagnostic T2 magnetic resonance images of the patient acquired without contrast medium before surgery are shown in Fig. 1(b). The study was approved by the ethical committee of the Carlo Besta Neurological Institute, and informed consent was signed by the patient.

The tumor diameter was 3.5  cm and the volume was 18  cm3. From the DOTATOC uptake derived from PET images, we could estimate the activity concentration of Y90-DOTATOC in the tumor during surgery to be equal to 4.8  kBq/ml.

Y90 (T1/2=2.66  days) decays mainly by beta minus emission with end point energy equal to 2.28 MeV and a small fraction of pair production. Because of the direct interaction with the CCD of bremsstrahlung and gamma rays emitted by Y90, a small sparse spike noise pattern is generated during the emission image acquisition. The spikes have been removed by applying a box median filter with a size equal to 3×3  pixels on the emission data.

The specimens of the tumor removed by the surgeon were imaged using both CLI and RLI as described before. The sample activity could not be directly measured simply because surgery was carried out in a hospital that did not have a nuclear medicine division with appropriate instrumentation. The sample has a dimension of 5×11×9  mm3; thus, the estimated activity was equal to 2.4 kBq.

Figure 2(a) shows the photographic picture of the tumor specimen, Fig. 2(b) the Cerenkov image, and Fig. 2(c) the RLI signal. The images confirm an appropriate localization of the visible Cerenkov and radioluminescence light signal within the tumor region.

Fig. 2

(a) A color image evincing the amount of blood present in the tumor specimen, and (b) and (c) the CLI and RLI signals. As can be seen from the images in (b) and (c), there is a good agreement between the localization of the visible Cerenkov and radioluminescence light signal within the tumor region. The color scale is in arbitrary units.

JBO_21_5_050502_f002.png

To our knowledge, this is the first published experimental evidence of the possibility of detecting Cerenkov and radioluminescence light from a fresh human tumor specimen removed during surgery.

As expected, the detected light signal is higher in RLI with respect to CLI; we used the same color scale in order to provide a better comparison between CLI and RLI signals. A circular (3 mm diameter) region of interest was drawn at the center of the tumor specimen, and the mean value was calculated for both CLI and RLI. We found that the average values of the CLI and RLI signal were, respectively, equal to 582 and 652 (arbitrary units).

The higher background signal in RLI data around the specimen is due to direct interaction of the high-energy beta particles with the scintillator slab.

In summary, we showed, for the first time, examples of CLI and RLI images of a fresh human brain tumor specimen removed during neurosurgery of a patient previously administered, for investigational purposes, with 166 MBq of Y90-DOTATOC.

The FOV of 6.1×6.1  cm allows the acquisition of few surgical samples simultaneously, reducing the total time needed for the imaging process.

As shown by recent Monte Carlo simulation,23 the use of Y90 is ideal considering the higher number of Cerenkov photons emitted in the tissues with respect, for example, to F18.

The CLI signal of the specimen is well confined in the tumor region, while the RLI signal is higher but also contains higher background outside the sample. We can conclude that CLI is the best option when the Cerenkov signal is high enough to be clearly detected, while RLI can be a good alternative option when the Cerenkov light is low. More precisely, RLI can be the best imaging approach when the radiotracer uptake is low, when using isotopes with low Cerenkov light yield, or with end point energy below the Cerenkov threshold.

The approach presented in this work could provide a fast molecular imaging method for the estimation of tumor radiopharmaceutical uptake during surgical procedures using Cerenkov and radioluminescence light.

Acknowledgments

The authors would like to acknowledge Professor Riccardo Faccini and his collaborators for their cooperation.

References

1. 

A. E. Spinelli et al., “Optical imaging using radioisotopes: a novel multimodal approach to molecular imaging,” Q J Nucl Med Mol Imaging, 56 279 –289 (2012). Google Scholar

2. 

A. E. Spinelli and F. Boschi, “Novel biomedical applications of Cerenkov radiation and radioluminescence imaging,” Phys. Med., 31 (2), 120 –129 (2015). http://dx.doi.org/10.1016/j.ejmp.2014.12.003 Google Scholar

3. 

F. Boschi and A. E. Spinelli, “Cerenkov luminescence imaging at a glance,” Curr. Mol. Imaging, 3 106 –117 (2014). http://dx.doi.org/10.2174/2211555203666141128002406 Google Scholar

4. 

R. Robertson et al., “Optical imaging of Cerenkov light generation from positron-emitting radiotracers,” Phys. Med. Biol., 54 N355 –N365 (2009). http://dx.doi.org/10.1088/0031-9155/54/16/N01 PHMBA7 0031-9155 Google Scholar

5. 

A. E. Spinelli et al., “Cerenkov radiation allows in vivo optical imaging of positron emitting radiotracers,” Phys. Med. Biol., 55 483 –495 (2010). http://dx.doi.org/10.1088/0031-9155/55/2/010 PHMBA7 0031-9155 Google Scholar

6. 

F. Boschi et al., “In vivo (18)F-FDG tumour uptake measurements in small animals using Cerenkov radiation,” Eur. J. Nucl. Med., 38 120 –127 (2011). http://dx.doi.org/10.1007/s00259-010-1630-y EJNMD9 0340-6997 Google Scholar

7. 

A. E. Spinelli et al., “Cerenkov radiation imaging of beta emitters: in vitro and in vivo results,” Nucl. Instrum. Methods Phys. Res. A, 648 S310 –S312 (2011). http://dx.doi.org/10.1016/j.nima.2010.11.038 NIMAER 0168-9002 Google Scholar

8. 

A. E. Spinelli et al., “Multispectral Cerenkov luminescence tomography for small animal optical imaging,” Opt. Express, 19 12605 –12618 (2011). http://dx.doi.org/10.1364/OE.19.012605 OPEXFF 1094-4087 Google Scholar

9. 

F. Boschi et al., “Optical imaging of alpha emitters: simulations, phantoms and in vivo results,” J. Biomed. Opt., 16 (12), 126011 (2011). http://dx.doi.org/10.1117/1.3663441 JBOPFO 1083-3668 Google Scholar

10. 

A. E. Spinelli et al., “Optical imaging of Tc-99m based tracers, in vitro and in vivo results,” J. Biomed. Opt., 16 (11), 116023 (2011). http://dx.doi.org/10.1117/1.3653963 JBOPFO 1083-3668 Google Scholar

11. 

M. Shimamoto et al., “Hybrid light imaging using Cerenkov luminescence and liquid scintillation for preclinical optical imaging in vivo,” Mol. Imaging Biol., 1 –10 (2016). http://dx.doi.org/10.1007/s11307-016-0928-y Google Scholar

12. 

A. E. Spinelli et al., “First human Cerenkography,” J. Biomed. Opt., 18 (2), 020502 (2013). http://dx.doi.org/10.1117/1.JBO.18.2.020502 JBOPFO 1083-3668 Google Scholar

13. 

H. Hu et al., “Feasibility study of novel endoscopic Cerenkov luminescence imaging system in detecting and quantifying gastrointestinal disease: first human results,” Eur. Radiol., 25 (6), 1814 –1822 (2015). http://dx.doi.org/10.1007/s00330-014-3574-2 Google Scholar

14. 

A. E. Spinelli, C. R. Gigliotti and F. Boschi, “Unified approach for bioluminescence, Cerenkov, β, X and γ rays imaging,” Biomed. Opt. Express, 6 2168 (2015). http://dx.doi.org/10.1364/BOE.6.002168 BOEICL 2156-7085 Google Scholar

15. 

F. Collamati et al., “Toward radioguided surgery with β- decays: uptake of a somatostatin analogue, DOTATOC, in meningioma and high-grade glioma,” J. Nucl. Med., 56 (1), 3 –8 (2015). http://dx.doi.org/10.2967/jnumed.114.145995 Google Scholar

16. 

M. Bartolomei et al., “Peptide receptor radionuclide therapy with (90)Y-DOTATOC in recurrent meningioma,” Eur. J. Nucl. Med. Mol. Imaging, 36 (9), 1407 –1416 (2009). http://dx.doi.org/10.1007/s00259-009-1115-z Google Scholar

17. 

L. Bodei et al., “Receptor-mediated radionuclide therapy with Y90-DOTATOC in association with amino acid infusion: a phase I study,” Eur. J. Nucl. Med. Mol. Imaging, 30 (2), 207 –216 (2003). http://dx.doi.org/10.1007/s00259-002-1023-y Google Scholar

18. 

L. Bodei et al., “The joint IAEA, EANM, and SNMMI practical guidance on peptide receptor radionuclide therapy (PRRNT) in neuroendocrine tumours,” Eur. J. Nucl. Med. Mol. Imaging, 40 (5), 800 –816 (2013). http://dx.doi.org/10.1007/s00259-012-2330-6 Google Scholar

19. 

L. Bodei et al., “Peptide receptor radionuclide therapy for advanced neuroendocrine tumors,” Thorac. Surg. Clin., 24 (3), 333 –349 (2014). http://dx.doi.org/10.1016/j.thorsurg.2014.04.005 Google Scholar

20. 

A. Filice et al., “Radiolabeled somatostatin analogues therapy in advanced neuroendocrine tumors: a single centre experience,” J. Oncol., 2012 320198 (2012). http://dx.doi.org/10.1155/2012/320198 Google Scholar

21. 

S. Koukouraki et al., “Evaluation of the pharmacokinetics of 68Ga-DOTATOC in patients with metastatic neuroendocrine tumours scheduled for 90Y-DOTATOC therapy,” Eur. J. Nucl. Med. Mol. Imaging, 33 (4), 460 –466 (2006). http://dx.doi.org/10.1007/s00259-005-0006-1 Google Scholar

22. 

M. Ö. Öksüz et al., “Peptide receptor radionuclide therapy of neuroendocrine tumors with (90)Y-DOTATOC: is treatment response predictable by pre-therapeutic uptake of (68)Ga-DOTATOC?,” Diagn. Interv. Imaging, 95 (3), 289 –300 (2014). http://dx.doi.org/10.1016/j.diii.2013.07.006 Google Scholar

23. 

F. Boschi, M. Pagliazzi and A. E. Spinelli, “Cerenkov luminescence imaging of human breast cancer: a Monte Carlo simulations study,” J. Instrum., 11 (03), C03032 (2016). http://dx.doi.org/10.1088/1748-0221/11/03/C03032 Google Scholar
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2016/$25.00 © 2016 SPIE
Antonello Enrico Spinelli, Marco Paolo Schiariti, Chiara Maria Grana, Mahila Ferrari, Marta Cremonesi, and Federico Boschi "Cerenkov and radioluminescence imaging of brain tumor specimens during neurosurgery," Journal of Biomedical Optics 21(5), 050502 (9 May 2016). https://doi.org/10.1117/1.JBO.21.5.050502
Published: 9 May 2016
Lens.org Logo
CITATIONS
Cited by 30 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tumors

Brain

Surgery

Neuroimaging

Cerenkov radiation imaging

Charge-coupled devices

Luminescence

Back to Top