Author Affiliations +
Abstract
Polarization-sensitive optical coherence tomography (PS-OCT) is a light-based, high-resolution, real-time, noninvasive, and nondestructive imaging modality yielding quasimicroscopic cross-sectional images of cartilage. As yet, comprehensive parameterization and quantification of birefringence and tissue properties have not been performed on human cartilage. PS-OCT and algorithm-based image analysis were used to objectively grade human cartilage degeneration in terms of surface irregularity, tissue homogeneity, signal attenuation, as well as birefringence coefficient and band width, height, depth, and number. Degeneration-dependent changes were noted for the former three parameters exclusively, thereby questioning the diagnostic value of PS-OCT in the assessment of human cartilage degeneration.
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
Nicolai Brill, Nicolai Brill, Mathias Wirtz, Mathias Wirtz, Dorit Merhof, Dorit Merhof, Markus Tingart, Markus Tingart, Holger Jahr, Holger Jahr, Daniel Truhn, Daniel Truhn, Robert Schmitt, Robert Schmitt, Sven Nebelung, Sven Nebelung, } "Polarization-sensitive optical coherence tomography-based imaging, parameterization, and quantification of human cartilage degeneration," Journal of Biomedical Optics 21(7), 076013 (22 July 2016). https://doi.org/10.1117/1.JBO.21.7.076013 . Submission: