27 October 2017 Multifocal spectral-domain optical coherence tomography based on Bessel beam for extended imaging depth
Luying Yi, Liqun Sun, Wuwen Ding
Author Affiliations +
Abstract
To advance the practical application of optical coherence tomography (OCT) in the field of biomedical imaging, the imaging depth must be extended without sacrificing resolution while maintaining sufficient sensitivity. However, there is an inherent trade-off between lateral resolution and depth of field (DOF) in OCT. To address this shortcoming, this article proposes a multifocal Bessel beam spectral-domain optical coherence tomography (MBSDOCT) capable of increasing the DOF with unchanged lateral resolution and a high signal-to-noise ratio. The proposed technique is demonstrated by simulation and experiment. A three-focal MBSDOCT with an axicon lens theoretically achieved a DOF of ∼6  mm with a lateral resolution of ∼13  μm. In imaging experiments performed on the acinar cells of orange tissue, a measured DOF of ∼4  mm was demonstrated with a sensitivity penalty of ∼18.1  dB, relative to the Gaussian beam spectral-domain OCT, with a 9-mW light source.

1.

Introduction

Optical coherence tomography (OCT), a noninvasive three-dimensional (3-D) imaging modality based on low coherence interferometry,1,2 has demonstrated effectiveness for in vivo imaging,35 and in particular for biomedical applications.69 To advance OCT technology for practical applications, the imaging depth must be extended without sacrificing resolution or sensitivity. Lenses with low numerical aperture are usually used in conventional Gaussian beam OCT to maintain a long depth of field (DOF); nevertheless, this approach incurs low lateral resolution. Different approaches have been taken to overcome this inherent trade-off between lateral resolution and DOF in OCT; those approaches can be classified into two categories in terms of the focusing beam: Gaussian beam OCT (GOCT) and Bessel beam OCT (BOCT).

There are two kinds of GOCT with extended DOF: multifocal GOCT (MGOCT)10,11 and interferometric synthetic aperture microscopy (ISAM).1216 In MGOCT, multiple apertures are used simultaneously in the scanning process. Each aperture is focused at a different depth during the scan. Compared with each set of OCT data, the MGOCT shows an improved resolution over the whole depth; however, the DOF of a Gaussian beam with high resolution is very small. In contrast, ISAM takes the OCT data and corrects for the spreading of the beam by applying a specific algorithm, resulting in an infinite extension of the DOF and invariant spatial resolution along the depth axis. However, in ISAM, the effective DOF is limited because the signal-to-noise ratio (SNR) decreases with 1/z, where z is the depth coordinate. Xu et al.17 proposed the multifocal ISAM (MISAM), which can further extend the DOF.

BOCT18,19 extends the DOF using a Bessel beam to illuminate the sample. The Bessel beam is generated using an axicon lens. The scattered signal is collected using either the axicon lens or a conventional objective lens. Lee and Rolland19 reported a doublepass BOCT that achieved a measured invariant resolution of 8  μm across a 4-mm DOF measured using the microoptic axicon while achieving an invariant sensitivity of 80  dB with a 25-mW input power. Leitgeb et al.20 proposed a nondual path imaging scheme to image biological samples and reported a 1.5-μm resolution across a 200-μm DOF. Curatolo et al.21 reported that superior contrast (by up to 40%) may be obtained over an extended depth range by MGOCT compared with BOCT. However, the GOCT has a very small DOF when the lateral resolution is high. Even with the MGOCT, it is difficult to achieve the same DOF as is achieved by the BOCT for a given lateral resolution when the SNR penalty of the BOCT can be accepted.

Recent developments in detector technology have enabled high acquisition speeds; therefore, spectral-domain optical coherence tomography (SDOCT), which is capable of high-resolution, real-time, and 3-D imaging of the internal structures of biological tissues, is currently an important tool for biological tissue imaging.2226 To address the trade-off between DOF and lateral resolution in OCT, this article proposes capitalizing on the advantages of SDOCT in a multifocal Bessel beam SDOCT (MBSDOCT) that combines the advantages of both BOCT and MGOCT. The imaging depth of the proposed technique is demonstrated by simulation and experiments. An MBSDOCT system equipped with a superluminescent diode (SLD) operating at a 1301-nm center wavelength (especially apt for depth imaging) is used to image the acinar cells of orange tissue. We obtained a clear imaging depth of 4  mm, with a sensitivity penalty of 18.1  dB at an input power of 9 mW.

2.

Methods

2.1.

Description of Bessel Beam

Figure 1(a) shows an axicon with a cone angle γ illuminated by a Gaussian beam with 1/e2 radius w0. Assuming a Gaussian input field with unit power, we may write the intensity distribution behind the axicon as27

Eq. (1)

I(ρ˜,z˜)=8πw02Nz˜exp(2z˜2)J02(2πNρ˜),
in cylindrical coordinates (ρ˜=ρ/w0, z˜=z/LB), where N is defined as the Fresnel number27 and J0 is the zero-order Bessel function of the first kind. The axial and radial extents of the beam maximum, generally referred to as the DOF LB and the beam radius RB, respectively, are of interest. They can be expressed as28

Eq. (2)

LB=w0×{[tan(β)1]-tan(γ)}
and

Eq. (3)

RB=2.4048k0tanβ2.4048k0β,
where β=arcsin(nasinγ)γ, na is the refractive index of the axicon, and k0 is the wave number. The peak intensity of the central beam can be expressed as27

Eq. (4)

Ibessel=I(0,12)=4πNew02.

Fig. 1

(a) Schematic diagram of generating Bessel beam and (b) relationships between Ibessel, DOF, and w0.

JBO_22_10_106016_f001.png

Equation (2) shows that there are two methods of extending the LB: decreasing the γ and increasing the w0. However, decreasing γ will increase RB, as shown in Eq. (3). Although increasing w0 can increase LB with unchanged RB, increasing w0 also causes a sharp decline in the peak intensity of the central beam Ibessel, as shown in Eq. (4) and Fig. 1(b). Decreased intensity of light incident on the sample will decrease the sensitivity of the system.

2.2.

Experimental Setup

To benchmark the performance of the proposed MBSDOCT in terms of DOF and SNR, a 1301-nm SDOCT system was built, as shown in Fig. 2(a). The system used an SLD with a 9-mW output power and 85-nm wavelength bandwidth, yielding an axial resolution of 8.8  μm in air. A Bessel beam generating system (L2, L3, axicon, L4, and L5) generated the needed Bessel beam. The γ of the axicon was 10 deg, and the refractive index was 1.43, resulting in a lateral resolution of 13  μm. Interference signals between the reference light and backscattered light, which carry structural information regarding the sample, were collected by a spectrometer with a spectral resolution of 0.05 nm, yielding an imaging depth of 6  mm. Depth structure information was obtained by inverting the spectral signal. For example, Fig. 2(b) shows the spectral signals from two reflectors at depths of 200 and 400  μm, respectively, and the structural signal inverted from the spectral signals.

Fig. 2

(a) Bessel beam SDOCT imaging system; (b) top, spectral signals from two reflectors, and bottom, structural signal inverted from the spectral signals.

JBO_22_10_106016_f002.png

The proposed MBSDOCT is similar to conventional BSDOCT, except that multiple channels are used. Because biological samples move (especially live samples), the samples can easily become displaced. Therefore, the scans were achieved by shifting axicons, L4 and L5, simultaneously to change the focus of the beam. It needs to be noted that the axicons, L4 and L5, were united together in the experimental setup.

The sensitivity of the proposed SDOCT configuration was measured. For Gaussian beam imaging using a conventional lens, the peak sensitivity has conventionally been measured at the focal plane of the lens (L5), where a mirror was used in place of a biological sample;19 an 101-dB peak sensitivity was measured with an input power of 9 mW. This measurement was made with a difference of 1.6 mm in the optical path lengths between the sample mirror and the reference mirror. The sensitivity of the Bessel beam setup (N=9.9, w0=170  μm) was found to be 83  dB based on a comparison with the SNR of the Gaussian beam SDOCT (GSDOCT) setup using the same system parameters and sample.27

3.

Simulation Results

3.1.

Comparative Analysis of Different Categories of SDOCT

In Sec. 1, we introduced several categories of OCT with extended DOF, and this classification method is also applicable to SDOCT. Numerical simulations of methods in each of these categories are useful for providing a theoretical corroboration of the method proposed in this paper. Table 1 describes the parameters of the GSDOCT, ISAM, and BSDOCT, and D is the spot size of Gaussian and Bessel beam, which is considered the lateral resolution here. The spectral resolution was 0.1 nm, and the bandwidth was 80 nm. The BSDOCT simulation in this paper followed published methods for the GSDOCT and ISAM,16 except that the Bessel beam replaced the Gauss beam. The numerical simulation consists of the forward model and the reconstruction. The forward model is the interaction between the sample and the incident light, and the spectral datum is collected in this process. The susceptibility of the sample is obtained by inverse Fourier transform of the spectral datum in the reconstruction process.

Table 1

Imaging parameters of GSDOCT, ISAM, and BSDOCT.

MethodTheoretical DOF (μm)Theoretical D (μm)
GSDOCTπD22λ0=204, λ0=1.3  μm2πk0NA=13, NA=0.1
ISAMπD22λ0(SNR0SNRL)212πk0NA=13, NA=0.1
BSDOCTw0×{[tan(β)1]tan(γ)}=1599, w0=120  μm, γ=10  deg2×2.4048k0tanβ13

Figures 3(a), 3(b), and 3(c) show simulation results for the imaging of isotropic point scatterers with diameters of 2  μm located randomly, from which we can see that the BSDOCT extended the DOF over those of the two other methods, all with the same lateral resolution.

Fig. 3

Simulations of (a) GSDOCT, (b) ISAM, and (c) BSDOCT applied to a scattering particles sample.

JBO_22_10_106016_f003.png

To illustrate quantitatively the advantages of MBSDOCT in terms of DOF, the relationships between the lateral resolution and the DOF in the methods mentioned in Sec. 1 were analyzed. The DOF of multifocal GSDOCT (MGSDOCT) can be approximated as the sum of the DOFs of all channels, which can be expressed as

Eq. (5)

DOFG=nc2πnw02λ,
where nc is the number of channels, n is the refractive index of the tissue sample, and w0 is the waist radius of the Gaussian beam. The relationships between the lateral resolution and DOF of GSDOCT and MGSDOCT (nc=2) are plotted by the black and brown curves as shown in Fig. 4, respectively. ISAM suffers from an SNR that decreases with 1/z, which limits its effective DOF, expressed as17

Eq. (6)

DOFI=2πnw02λ0(SNR0SNRL)21,
where SNR0 is the SNR at the focal plane and SNRL is the lowest tolerable SNR. Similarly, the DOF of MISAM can be approximated as the sum of the DOFs of all channels. The relationships between the lateral resolution and DOF of ISAM and MISAM are plotted by the green and blue curves as shown in Fig. 4, respectively, where SNR0=10SNRL and nc=2. In Sec. 2.1, we explained that in BSDOCT, when γ is decreased or w0 is increased to extend the DOF, either the lateral resolution or the SNR worsens. Therefore, this paper proposes MBSDOCT, which can extend the imaging depth under an unchanged lateral resolution and high SNR. The relationships between the lateral resolution and the DOF of BSDOCT and MBSDOCT are plotted by the pink and red curves as shown in Fig. 4, respectively, where w0=500  μm and nc=2.

Fig. 4

Relationships between lateral resolution and DOF of GSDOCT, MGSDOCT, ISAM, MISAM, BSDOCT, and MBSDOCT.

JBO_22_10_106016_f004.png

Figure 4 shows that MISAM and MBSDOCT have prominent advantages in extending the DOF and MBSDOCT has a larger DOF in the range of high lateral resolution (small D).

3.2.

MBSDOCT Simulation

Figure 5 shows the results of a computer simulation to verify the feasibility of MBSDOCT. The sample used in the simulation consisted of a series of point scatterers located randomly. The system parameters were consistent with the experimental system introduced in Sec. 2.2.

Fig. 5

(a)–(d) Simulation results of MBSDOCT (w0=170  μm) and BSDOCT (w0=340  μm) and (e) signal intensity (dB) of (c) and (d).

JBO_22_10_106016_f005.png

Figures 5(a) and 5(b) show the BSDOCT images acquired with focal planes at different depths of 0 and 2.2 mm, in which the w0 was 170  μm, yielding a theoretical DOF of 2.2 mm. Figure 5(c) shows the stitched MBSDOCT image synthesized from the data in Figs. 5(a) and 5(b), the depth of which is 4.4  mm, equivalent to the BSDOCT case, where the w0 was 340  μm, shown in Fig. 5(d). Although Figs. 5(c) and 5(d) present results from the same imaging depth and lateral resolutions, Fig. 5(c) shows a larger signal intensity than Fig. 5(d), as shown in Fig. 5(e). We deduced that the SNR of MBSDOCT would be higher in the actual experiment, which was later verified.

Based on these analyses, the MBSDOCT image with a smaller w0 and more channels was predicted to have a higher SNR than the MBSDOCT image with a larger w0 and fewer channels, which was later verified experimentally.

4.

Experimental Results

Two datasets were acquired separately using a BSDOCT setup. The focal planes of the scanning beams were adjusted before each scan. As shown in Fig. 6, two datasets (a) and (b) were acquired with focal planes at 0 and 1.5 mm, respectively. Figure 6(c) shows the MBSDOCT image of the two datasets. The SDOCT system had a w0 of 207  μm, yielding a DOF of 2.7 mm in air. The sample was orange tissue, and we assumed a refractive index of 1.33 for water. The image from each channel becomes unclear far away from the focal plane. Figure 6(a) shows an estimated effective DOF of 1.5  mm (from the blue dotted frame); Fig. 6(b) shows an estimated effective DOF of 1.5  mm (from the red dotted frame); and Fig. 6(c) shows an estimated effective DOF of 3  mm, 2 times greater than either single channel. Figures 6(d) and 6(e) indicate that the image obtained by MBSDOCT exhibited high signal intensity throughout the entire depth.

Fig. 6

(a)–(c) Two channels of BSDOCT images and the resulting MBSDOCT image, respectively; (d) signal intensity of one transverse point selected from the two BSDOCT figures; and (e) signal intensity of one transverse point selected from the MBSDOCT image.

JBO_22_10_106016_f006.png

Table 2 shows the SNRs of the images outlined by the blue and red dashed boxes in the three graphs in Fig. 6. Because there is a transverse offset between the two BSDOCT images, the registration image is not the sum of the two images; rather, the registration image is spliced by the clear part of the two BSDOCT images. The table shows that the MBSDOCT image exhibited a high SNR throughout the entire depth.

Table 2

SNRs of the three images in Fig. 6.

ImageSNR1 (0.5 to 2 mm)SNR2 (2 to 3.5 mm)
Channel 120.617
Channel 216.420.1
MBSDOCT20.620.1

The cell structure size of the two channel images in Fig. 6 is inconsistent. There may be two reasons for this problem. First, the measuring arm had a lateral swing while moving in the longitudinal direction, and the lengthened optical path difference resulting from this swing, considered the vertical depth of the structure, inevitably causes the different sizes of cells imaged in the two channels, both in the longitudinal and lateral directions. Second, the scan speed of the spectrometer used in this experiment was slow, and the imaging of one channel takes several minutes, during which the orange tissue changes. Therefore, a stable mobile station, precise adjustment of the optical system, and a fast spectrometer are needed to avoid inconsistent cell sizes in different channels images.

In Fig. 7, three datasets (a), (b), and (c) were acquired with focal planes at 0, 1.5, and 3 mm, respectively. Figure 7(d) shows the MBSDOCT image of the three datasets, with an effective DOF of 4  mm. The SDOCT system had a w0 of 170  μm, yielding a 2.2-mm DOF in air. From Figs. 7(e) and 7(f), we can see that the MBSDOCT image exhibited a high SNR throughout the entire depth. The SNR in the red dashed frame of Fig. 7(a) was 22.1, which was higher than that shown in Fig. 6(a). Therefore, the MBSDOCT image with a smaller w0 and more channels exhibited a higher SNR than the MBSDOCT image with a larger w0 and fewer channels. However, when the number of channels increases, the system becomes more complex and the acquisition time becomes longer.

Fig. 7

(a)–(d) Three channels of BSDOCT images and the resulting MBSDOCT image, respectively; (e) signal intensity of one transverse point selected from the three BSDOCT figures; and (f) signal intensity of one transverse point selected from the MBSDOCT image.

JBO_22_10_106016_f007.png

In Sec. 3.2, we demonstrated by simulation that increasing w0 can also extend the DOF. However, the image taken with w0=340  μm exhibited lower signal intensity compared with the MBSDOCT image taken with w0=170  μm when their DOFs were the same. Experiments with the same parameters as were used in the simulation were implemented to demonstrate the predicted result; however, the focal plane of each channel was shifted by 1 mm.

Figures 8(a) and 8(b) show the BSDOCT images with a w0 of 170  μm, and Fig. 8(c) shows the MBSDOCT image resulting from these two datasets. Figure 8(d) shows the BSDOCT image with a w0 of 340  μm. Figures 8(c) and 8(d) show approximately the same image depth; however, the MBSDOCT image exhibited a larger signal intensity, as shown in Fig. 8(e). Similarly, the SNRs of the images outlined by the blue dashed boxes in the two pictures were calculated. The results indicate that the image shown in Fig. 8(d) suffered from a lower SNR of 20.33, whereas the MBSDOCT image exhibited a larger SNR of 22.09. Therefore, the MBSDOCT setup with w0 exhibited a larger SNR than the BSDOCT setup with 2w0, although both setups had the same lateral resolution and theoretical DOF.

Fig. 8

(a)–(c) Two channels of BSDOCT images and the resulting MBSDOCT image with w0 of 170  μm; (d) BSDOCT image with w0 of 340  μm; and (e) signal intensity of one transverse point selected from the two figures.

JBO_22_10_106016_f008.png

Table 3 gives the results of different BSDOCT experimental sets under different w0. These results are consistent with the theory: the larger a w0 the system has, the lower its SNR. The theoretical SNR penalty relative to GSDOCT can be expressed as a function of the Fresnel number N27

Eq. (7)

SNRpenalty(dB)=20log(2eN)=20log(1.213N).

Table 3

Comparison of BSDOCT with different w0.

Bessel 1Bessel 2Bessel 3
Wrist radius w0 (μm)170207340
Fresnel number N9.91219.8
Theoretical DOF (mm)2.22.74.5
Measured DOF (mm)22.53
Theoretical SNR penalty (dB)18.219.924
Measured SNR penalty (dB)18.121.929.8

The reason for which the imaging depth does not reach the theoretical value and for which the number of channels cannot be further increased is the lower power of the light source. Larger input power has greater imaging depth and SNR; however, the biological tissue’s damage threshold must be considered.

5.

Conclusion

In this paper, we demonstrated that the proposed MBSDOCT can improve imaging depth and SNR. The simulation results showed that the MBSDOCT can obtain a larger DOF and signal intensity, and these results were verified by experiments on acinar cells of orange tissue. When three foci were used, we obtained an imaging depth of 4  mm with a lateral resolution of 13  μm, and this imaging depth can increase when a larger input power is used. Furthermore, MBSDOCT images with a smaller w0 and more channels were demonstrated to achieve a higher SNR than MBSDOCT images with a larger w0 and fewer channels, although they have the same imaging depth and lateral resolution. Therefore, the proposed MBSDOCT method is offers an important advance for the deep imaging of biological tissue and diagnosis of deep lesions in medicine. In addition, the multifocal Bessel beam scheme is suitable for a variety of OCT modes, compared with SDOCT.

Disclosures

The authors have no relevant financial interests in the article and no other potential conflicts of interest to disclose.

Acknowledgments

This research was funded by the State Key Lab Precision Measurement Technology and Instruments, Tsinghua University, and the National Major Scientific Instrument and Equipment Development Project of China (Grant Nos. 2012YQ200182 and 2012YQ0901670602).

References

1. 

D. Huang et al., “Optical coherence tomography,” Science, 254 (5035), 1178 –1181 (1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

2. 

A. F. Fercher et al., “Optical coherence tomography—principles and applications,” Rep. Prog. Phys., 66 (2), 239 –303 (2003). http://dx.doi.org/10.1088/0034-4885/66/2/204 RPPHAG 0034-4885 Google Scholar

3. 

G. J. Tearney et al., “In vivo endoscopic optical biopsy with optical coherence tomography,” Science, 276 (5321), 2037 –2039 (1997). http://dx.doi.org/10.1126/science.276.5321.2037 SCIEAS 0036-8075 Google Scholar

4. 

X. Y. Zhang et al., “Dual-band spectral-domain optical coherence tomography for in vivo imaging the spectral contrasts of the retinal nerve fiber layer,” Opt. Express, 19 (20), 19653 –19667 (2011). http://dx.doi.org/10.1364/OE.19.019653 OPEXFF 1094-4087 Google Scholar

5. 

Y. Zhao et al., “Evaluation of burn severity in vivo in a mouse model using spectroscopic optical coherence tomography,” Biomed. Opt. Express, 6 (9), 3339 –3345 (2015). http://dx.doi.org/10.1364/BOE.6.003339 BOEICL 2156-7085 Google Scholar

6. 

M. R. Hee et al., “Optical coherence tomography of the human retina,” Arch. Ophthalmol., 113 (3), 325 (1995). http://dx.doi.org/10.1001/archopht.1995.01100030081025 AROPAW 0003-9950 Google Scholar

7. 

M. R. Hee et al., “Quantitative assessment of macular edema with optical coherence tomography,” Arch. Ophthalmol., 113 (8), 1019 (1995). http://dx.doi.org/10.1001/archopht.1995.01100080071031 AROPAW 0003-9950 Google Scholar

8. 

Y. K. K. Tao et al., “Intraoperative spectral domain optical coherence tomography for vitreoretinal surgery,” Opt. Lett., 35 (20), 3315 –3317 (2010). http://dx.doi.org/10.1364/OL.35.003315 OPLEDP 0146-9592 Google Scholar

9. 

S. P. Chong et al., “Quantitative microvascular hemoglobin mapping using visible light spectroscopic optical coherence tomography,” Biomed. Opt. Express, 6 (4), 1429 –1450 (2015). http://dx.doi.org/10.1364/BOE.6.001429 BOEICL 2156-7085 Google Scholar

10. 

J. Holmes et al., “Multi-channel Fourier domain OCT system with superior lateral resolution for biomedical applications,” Proc. SPIE, 6847 68470O (2008). http://dx.doi.org/10.1117/12.761655 PSISDG 0277-786X Google Scholar

11. 

J. Holmes, “Theory and applications of multi-beam OCT,” Proc. SPIE, 7139 713908 (2008). http://dx.doi.org/10.1117/12.821006 PSISDG 0277-786X Google Scholar

12. 

Y. Xu et al., “Automated interferometric synthetic aperture microscopy and computational adaptive optics for improved optical coherence tomography,” Appl. Opt., 55 (8), 2034 –2041 (2016). http://dx.doi.org/10.1364/AO.55.002034 APOPAI 0003-6935 Google Scholar

13. 

T. S. Ralston et al., “Interferometric synthetic aperture microscopy,” Nat. Phys., 3 (2), 129 –134 (2007). http://dx.doi.org/10.1038/nphys514 NPAHAX 1745-2473 Google Scholar

14. 

P. S. Carney et al., “Automation of interferometric synthetic aperture microscopy,” in Frontiers in Optics, (2015). Google Scholar

15. 

B. J. Davis et al., “Interferometric synthetic aperture microscopy: computed imaging for scanned coherent microscopy,” Sens. Basel, 8 (6), 3903 –3931 (2008). http://dx.doi.org/10.3390/s8063903 Google Scholar

16. 

B. J. Davis et al., “Nonparaxial vector-field modeling of optical coherence tomography and interferometric synthetic aperture microscopy,” J. Opt. Soc. Am. A, 24 (9), 2527 –2542 (2007). http://dx.doi.org/10.1364/JOSAA.24.002527 JOAOD6 0740-3232 Google Scholar

17. 

Y. Xu et al., “Multifocal interferometric synthetic aperture microscopy,” Opt. Express, 22 (13), 16606 –16618 (2014). http://dx.doi.org/10.1364/OE.22.016606 OPEXFF 1094-4087 Google Scholar

18. 

C. Blatter et al., “High-speed functional OCT with self-reconstructive Bessel illumination at 1300 nm,” Proc. SPIE, 8091 809104 (2011). http://dx.doi.org/10.1117/12.889669 Google Scholar

19. 

K. S. Lee and L. P. Rolland, “Bessel beam spectral-domain high-resolution optical coherence tomography with micro-optic axicon providing extended focusing range,” Opt. Lett., 33 (15), 1696 –1698 (2008). http://dx.doi.org/10.1364/OL.33.001696 OPLEDP 0146-9592 Google Scholar

20. 

R. A. Leitgeb et al., “Extended focus depth for Fourier domain optical coherence microscopy,” Opt. Lett., 31 (16), 2450 –2452 (2006). http://dx.doi.org/10.1364/OL.31.002450 OPLEDP 0146-9592 Google Scholar

21. 

A. Curatolo et al., “Quantifying the influence of Bessel beams on image quality in optical coherence tomography,” Sci. Rep., 6 23483 (2016). http://dx.doi.org/10.1038/srep23483 SRCEC3 2045-2322 Google Scholar

22. 

J. F. de Boer et al., “Improved signal-to-noise ratio in spectral-domain compared with time-domain optical coherence tomography,” Opt. Lett., 28 (21), 2067 –2069 (2003). http://dx.doi.org/10.1364/OL.28.002067 OPLEDP 0146-9592 Google Scholar

23. 

M. Wojtkowski et al., “Full range complex spectral optical coherence tomography technique in eye imaging,” Opt. Lett., 27 (16), 1415 –1417 (2002). http://dx.doi.org/10.1364/OL.27.001415 OPLEDP 0146-9592 Google Scholar

24. 

M. Wojtkowski et al., “Real-time in vivo imaging by high-speed spectral optical coherence tomography,” Opt. Lett., 28 (19), 1745 –1747 (2003). http://dx.doi.org/10.1364/OL.28.001745 OPLEDP 0146-9592 Google Scholar

25. 

Z. H. Hu et al., “Automated segmentation of neural canal opening and optic cup in 3D spectral optical coherence tomography volumes of the optic nerve head,” Invest. Ophthalmol. Visual Sci., 51 (11), 5708 –5717 (2010). http://dx.doi.org/10.1167/iovs.09-4838 IOVSDA 0146-0404 Google Scholar

26. 

B. J. Kaluzny, “Spectral optical coherence tomography: a novel technique for cornea imaging,” Cornea, 25 (8), 960 (2006). http://dx.doi.org/10.1097/01.ico.0000224644.81719.59 CORNDB 0277-3740 Google Scholar

27. 

D. Lorenser et al., “Energy-efficient low-Fresnel-number Bessel beams and their application in optical coherence tomography,” Opt. Lett., 39 (3), 548 –551 (2014). http://dx.doi.org/10.1364/OL.39.000548 OPLEDP 0146-9592 Google Scholar

28. 

N. Weber et al., “Highly compact imaging using Bessel beams generated by ultraminiaturized multi-micro-axicon systems,” J. Opt. Soc. Am. A, 29 (5), 808 –816 (2012). http://dx.doi.org/10.1364/JOSAA.29.000808 JOAOD6 0740-3232 Google Scholar

Biography

Luying Yi is a PhD student in optical engineering at the Department of Precision Instruments, Tsinghua University. She received her bachelor’s degree in optoelectronics and information engineering at Tianjin University in 2015. She is currently studying the spectral-domain optical coherence tomography under the guidance of Professor Sun in State Key Laboratory of Precision Measurement Technology & Instruments. Her current works include the brain imaging and deep imaging.

Liqun Sun received his BS degree in applied optics from Beijing Institute of Technology in 1988, and his PhD in physical electronics in 2000 from Xi’an Jiao Tong University, China. He had been an Engineer in Xi’an Institute of Applied Optics since 1988. In 2000, he joined the Department of Precision Instruments, Tsinghua university. He is now a professor of optical engineering. His research interests include spectrum measurement technology, optical sensing and bio-optics.

Wuwen Ding is a PhD student in optical engineering at the Department of Precision Instruments, Tsinghua University. He received his bachelor’s degree in Measurement and Control Technology and Instrumentation Program at Tsinghua University in 2012. His works focus on gas absorption spectroscopy. He currently studying tunable diode laser absorption spectroscopy (TDLAS) and chirped laser dispersion spectroscopy (CLADS) under the guidance of Professor Sun in State Key Laboratory of Precision Measurement Technology & Instruments.

© 2017 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2017/$25.00 © 2017 SPIE
Luying Yi, Liqun Sun, and Wuwen Ding "Multifocal spectral-domain optical coherence tomography based on Bessel beam for extended imaging depth," Journal of Biomedical Optics 22(10), 106016 (27 October 2017). https://doi.org/10.1117/1.JBO.22.10.106016
Received: 14 July 2017; Accepted: 11 October 2017; Published: 27 October 2017
Lens.org Logo
CITATIONS
Cited by 21 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical coherence tomography

Signal to noise ratio

Bessel beams

Axicons

Image resolution

Sun

Gaussian beams

Back to Top