Open Access
22 August 2017 Simultaneous en-face imaging of multiple layers with multiple reference optical coherence tomography
Author Affiliations +
Abstract
A technique based on multiple reference optical coherence tomography (MR-OCT) is proposed for simultaneous imaging at multiple depths. The technique has been validated by imaging a reference sample and a fingerprint in-vivo. The principle of scanning multiple selected layers is shown by imaging a partial fingerprint with 200×200×200 voxels of 3×3×0.5  mm size and obtaining an arbitrary amount of layers merely by digital processing. The spacing among the layers can be adjusted arbitrarily, and the SNR roll-off is shown for three different spacings. At a mirror scan frequency of 1 kHz and an A-line rate of 2 kHz, the acquisition time was 20 s for one volume. The results show the feasibility of the application of layer scanning MR-OCT that uses a partial mirror in the reference arm of the Michelson interferometer. The reduced scan range required for layer scanning allows even higher scan rates that are limited only by the voice coil design and the mass-spring system, e.g., mirror mass, spring constant, and damping.
© 2017 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2017/$25.00 © 2017 SPIE
Kai Neuhaus, Seán O’Gorman, Paul M. McNamara, Sergey A. Alexandrov, Josh Hogan, Carol Wilson, and Martin J. Leahy "Simultaneous en-face imaging of multiple layers with multiple reference optical coherence tomography," Journal of Biomedical Optics 22(8), 086006 (22 August 2017). https://doi.org/10.1117/1.JBO.22.8.086006
Received: 25 May 2017; Accepted: 24 July 2017; Published: 22 August 2017
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Phase modulation

Imaging systems

Mirrors

Optical coherence tomography

In vivo imaging

Multilayers

Biomedical optics

Back to Top