15 March 2018 Measurement of refractive index of hemoglobin in the visible/NIR spectral range
Author Affiliations +
Abstract
This study is focused on the measurements of the refractive index of hemoglobin solutions in the visible/near-infrared (NIR) spectral range at room temperature for characteristic laser wavelengths: 480, 486, 546, 589, 644, 656, 680, 930, 1100, 1300, and 1550 nm. Measurements were performed using the multiwavelength Abbe refractometer. Aqua hemoglobin solutions of different concentrations obtained from human whole blood were investigated. The specific increment of refractive index on hemoglobin concentration and the Sellmeier coefficients were calculated.
© 2018 Society of Photo-Optical Instrumentation Engineers (SPIE)
Ekaterina N. Lazareva, Valery V. Tuchin, "Measurement of refractive index of hemoglobin in the visible/NIR spectral range," Journal of Biomedical Optics 23(3), 035004 (15 March 2018). https://doi.org/10.1117/1.JBO.23.3.035004 . Submission: Received: 21 November 2017; Accepted: 20 February 2018
Received: 21 November 2017; Accepted: 20 February 2018; Published: 15 March 2018