1 October 1998 Biocompatibility of Laser-deposited Hydroxyapatite Coatings on Titanium and Polymer Implant Materials
Author Affiliations +
Abstract
We have investigated the biocompatibility of calcium phosphate coatings deposited by pulsed laser ablation from hydroxyapatite (HA) targets onto polyethylene and Teflon substrates. It was found that the cell density, attachment, and morphology of primary rat calvaria osteoblasts were influenced by both the original polymer and by the nature of the apatite coatings. HA coatings on Teflon were found to have higher biocompatibility in terms of cell adhesion and spreading. In vivo studies of bone response to coatings deposited by KrF excimer and CO2 lasers on commercial Ti6A14V alloy implants show that both deposition techniques suppress fibrous tissue formation and promote osteogenesis.
Evgueni N. Antonov, Evgueni N. Antonov, Victor N. Bagratashvili, Victor N. Bagratashvili, Vladimir K. Popov, Vladimir K. Popov, Emil N. Sobol, Emil N. Sobol, Steven Howdle, Steven Howdle, Christine Joiner, Christine Joiner, Kate G. Parker, Kate G. Parker, Terry L. Parker, Terry L. Parker, Alexander A. Doktorov, Alexander A. Doktorov, Valeri B. Likhanov, Valeri B. Likhanov, Alexander I. Volozhin, Alexander I. Volozhin, Sergey Sergeivich Alimpiev, Sergey Sergeivich Alimpiev, Sergey M. Nikiforov, Sergey M. Nikiforov, } "Biocompatibility of Laser-deposited Hydroxyapatite Coatings on Titanium and Polymer Implant Materials," Journal of Biomedical Optics 3(4), (1 October 1998). https://doi.org/10.1117/1.429852 . Submission:
JOURNAL ARTICLE
6 PAGES


SHARE
RELATED CONTENT

Laser synthesis of Si Mo C O polymer for membrane...
Proceedings of SPIE (September 06 1994)
Characteristic changes of lased dentin
Proceedings of SPIE (April 16 1995)
Excimer laser drilling of polymers
Proceedings of SPIE (August 17 1997)
Laser micromachining at the Danish Technological Institute
Proceedings of SPIE (December 09 2005)

Back to Top