1 July 2001 Refractive-index-mismatch induced aberrations in single-photon and two-photon microscopy and the use of aberration correction
Author Affiliations +
J. of Biomedical Optics, 6(3), (2001). doi:10.1117/1.1382808
Abstract
We examine the effects of aberrations induced by a refractive index mismatch on the signal level and resolution of singlephoton (1–p) and two-photon (2–p), conventional and confocal scanning microscopes. In particular, we consider the aberrations introduced by an interface between oil/glass and water. Resolution is defined in terms of enclosed fluorescence, rather than full-width halfmaximum, revealing more useful information for heavily aberrated point spread functions (PSFs). It is shown that, at large focusing depths, the resolution of 2–p conventional and 1–p confocal microscopes are almost identical. The benefits of aberration correction are examined by removing Zernike aberration modes. With aberration correction, the best resolution is found for 1–p confocal and 2–p confocal modes. An approximation based upon geometrical optics is also introduced which shows that the axial resolution of heavily aberrated PSFs is roughly proportional to focusing depth.
Martin J. Booth, Tony Wilson, "Refractive-index-mismatch induced aberrations in single-photon and two-photon microscopy and the use of aberration correction," Journal of Biomedical Optics 6(3), (1 July 2001). http://dx.doi.org/10.1117/1.1382808
JOURNAL ARTICLE
7 PAGES


SHARE
KEYWORDS
Confocal microscopy

Point spread functions

Luminescence

Microscopes

Aberration correction

Refractive index

Geometrical optics

Back to Top