1 October 2003 Small-volume frequency-domain oximetry: phantom experiments and first in vivo results
Author Affiliations +
J. of Biomedical Optics, 8(4), (2003). doi:10.1117/1.1608892
Abstract
We describe a new method to determine the oxygen saturation and the total hemoglobin content of tissue in vivo absolutely at small source-detector separations (<10 mm). Phase and mean intensity of modulated laser light of various wavelengths was measured at several predetermined source-detector separations in the frequency domain. From these measured quantities, the absorption coefficient was derived using the modified time-integrated microscopic Beer-Lambert law (MBL). In addition, the interaction volume of the photons was determined using a multi-layer Monte-Carlo model of human skin. To evaluate the method, we employed homogenous solid phantoms (consisting of TiO2 particles embedded in resin) with mean scattering and absorbing properties comparable to those of human skin. Furthermore, in vivo measurements were performed in a healthy volunteer to demonstrate that the technique is applicable for the determination of the oxygen saturation and the total hemoglobin content in the skin in vivo. The proposed technique is especially suited for the on-line determination of the oxygen saturation and total hemoglobin content in applications where small applicators are required (e.g., fetal oxygen monitoring sub partu).
Stefan Willmann, Albert Terenji, Jens Osterholtz, Joerg Meister, Peter Hering, Hans-Joachim Schwarzmaier, "Small-volume frequency-domain oximetry: phantom experiments and first in vivo results," Journal of Biomedical Optics 8(4), (1 October 2003). https://doi.org/10.1117/1.1608892
JOURNAL ARTICLE
11 PAGES


SHARE
Back to Top