1 September 2004 Accurate measurement of blood vessel depth in port wine stained human skin in vivo using pulsed photothermal radiometry
Author Affiliations +
Abstract
We report on application of pulsed photothermal radiometry (PPTR) to determine the depth of port wine stain (PWS) blood vessels in human skin. When blood vessels are deep in the PWS skin (<100 μm), conventional PPTR depth profiling can be used to determine PWS depth with sufficient accuracy. When blood vessels are close or partially overlap the epidermal melanin layer, a modified PPTR technique using two-wavelength (585 and 600 nm) excitation is a superior method to determine PWS depth. A direct difference approach in which PWS depth is determined from a weighted difference of temperature profiles reconstructed independently from two-wavelength excitation is demonstrated to be appropriate for a wider range of PWS patients with various blood volume fractions, blood vessel sizes, and depth distribution. The most superficial PWS depths determined in vivo by PPTR are in good agreement with those measured using optical Doppler tomography (ODT).
© (2004) Society of Photo-Optical Instrumentation Engineers (SPIE)
Bincheng Li, Bincheng Li, Boris Majaron, Boris Majaron, John A. Viator, John A. Viator, Thomas E. Milner, Thomas E. Milner, Zhongping Chen, Zhongping Chen, Yonghua Zhao, Yonghua Zhao, Hongwu Ren, Hongwu Ren, John Stuart Nelson, John Stuart Nelson, } "Accurate measurement of blood vessel depth in port wine stained human skin in vivo using pulsed photothermal radiometry," Journal of Biomedical Optics 9(5), (1 September 2004). https://doi.org/10.1117/1.1784470 . Submission: