You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
22 August 2017Assessment of organ and effective dose when using region-of-interest attenuators in cone-beam CT and interventional fluoroscopy
In some medical-imaging procedures using cone-beam CT (CBCT) and fluoroscopy, only the center of the field of view (FOV) may be needed to be visualized with optimal image quality. To reduce the dose to the patient while maintaining visualization of the entire FOV, a Cu attenuator with a circular aperture for the region of interest (ROI) is used. The potential organ and effective dose reductions of ROI imaging when applied to CBCT and interventional fluoroscopic procedures were determined using EGSnrc Monte Carlo code. The Monte Carlo model was first validated by comparing the surface dose distribution in a solid-water block phantom with measurement by Gafchromic film. The dependence of dose reduction on the ROI attenuator thickness, the opening size of the ROI, the axial beam position, and the location of the different organs for both neuro and thoracic imaging was evaluated. The results showed a reduction in most organ doses of 45% to 70% and in effective dose of 46% to 66% compared to the dose in a CBCT scan and in an interventional procedure without the ROI attenuator. This work provides evidence of a substantial reduction of organ and effective doses when using an ROI attenuator during CBCT and fluoroscopic procedures.
The alert did not successfully save. Please try again later.
Zhenyu Xiong, Sarath Vijayan, Stephen Rudin, Daniel R. Bednarek, "Assessment of organ and effective dose when using region-of-interest attenuators in cone-beam CT and interventional fluoroscopy," J. Med. Imag. 4(3) 031210 (22 August 2017) https://doi.org/10.1117/1.JMI.4.3.031210