5 October 2016 Design of multichannel functional near-infrared spectroscopy system with application to propofol and sevoflurane anesthesia monitoring
Author Affiliations +
Abstract
Monitoring the changes of cerebral hemodynamics and the state of consciousness during general anesthesia (GA) is clinically important. There is a great need for developing advanced detectors to investigate the physiological processes of the brain during GA. We developed a multichanneled, functional near-infrared spectroscopy (fNIRS) system device and applied it to GA operation monitoring. The cerebral hemodynamic data from the forehead of 11 patients undergoing propofol and sevoflurane anesthesia were analyzed. The concentration changes of oxygenated hemoglobin, deoxygenated hemoglobin, total hemoglobin, and cerebral tissue heart rate were determined from the raw optical information based on the discrete stationary wavelet transform. This custom-made device provides an easy-to-build solution for continuous wave-fNIRS system, with customized specifications. The developed device has a potential value in cerebral monitoring in clinical settings.
© 2016 Society of Photo-Optical Instrumentation Engineers (SPIE)
Zhenhu Liang, Zhenhu Liang, Yue Gu, Yue Gu, Xuejing Duan, Xuejing Duan, Lei Cheng, Lei Cheng, Shujuan Liang, Shujuan Liang, Yunjie Tong, Yunjie Tong, Xiaoli Li, Xiaoli Li, } "Design of multichannel functional near-infrared spectroscopy system with application to propofol and sevoflurane anesthesia monitoring," Neurophotonics 3(4), 045001 (5 October 2016). https://doi.org/10.1117/1.NPh.3.4.045001 . Submission:
JOURNAL ARTICLE
11 PAGES


SHARE
RELATED CONTENT


Back to Top