26 January 2018 Prefrontal hemodynamic changes measured using near-infrared spectroscopy during the Valsalva maneuver in patients with orthostatic intolerance
Author Affiliations +
Abstract
The Valsalva maneuver (VM) with beat-to-beat blood pressure and heart rate monitoring are used to evaluate orthostatic intolerance (OI). However, they lack the ability to detect cerebral hemodynamic changes, which may be a cause of OI symptoms. Therefore, we utilized near-infrared spectroscopy during VM. Patients with OI symptoms and normal healthy subjects were recruited. Patients were subgrouped according to VM results: patients with normal VM (NVM) and abnormal VM (AbVM). Oxyhemoglobin (HbO), deoxyhemoglobin, and total hemoglobin changes were measured at four different source–detector distances (SD) (15, 30, 36, and 45 mm), and latency, amplitude, duration, and integrated total signal were calculated. Those parameters were compared between a normal healthy control (HC) group and the two OI patient subgroups. We found that HbO increment latency at 30-mm SD in the HC, NVM, and AbVM groups was as follows: 0.39±0.23  s, 2.79±0.36  s, and 8.14±0.55  s, respectively (p<0.05). Among the four parameters we evaluated, latency of HbO increment was the best marker for differentiating OI.

1.

Introduction

Various autonomic function tests (AFTs) have been used to evaluate autonomic dysfunction; however, only a few tests have been validated as clinically quantitative methods to evaluate major autonomic domains, including sudomotor, cardiovagal, and adrenergic.1 The Valsalva maneuver (VM) is one of the validated quantifying AFTs. The VM is easy to perform and is a useful method for simultaneously evaluating both sympathetic adrenergic functions using blood pressure (BP) responses and parasympathetic cardiovagal functions using heart rate (HR) responses.2 Moreover, Valsalva activity evokes various symptoms related to dysregulation of the autonomic nervous system (ANS), such as dizziness, light headedness, blurred vision, and palpitation. The maneuver can be performed with prolonged forced exhalation through a closed airway and can occur naturally during routine activities, such as heavy lifting, straining on the toilet, or vomiting. Overall hemodynamic autoregulation during VM is a key factor in preventing the development of significant clinical symptoms such as orthostatic intolerance (OI), which is one of the representative manifestations of autonomic dysfunction or syncope.3,4

However, BP and HR monitoring alone is not enough to detect pathologic dysfunction of the ANS. In addition, the mechanisms behind VM that determine the symptoms caused by autonomic dysfunction are still unclear. Many previous studies have suggested a decrease in cerebral perfusion evoked by autonomic reflexes during Valsalva activity as a possible mechanism.59

To fully explore the hemodynamic changes during VM, we chose to use near-infrared spectroscopy (NIRS) to monitor the relative concentration changes in oxyhemoglobin (HbO), deoxyhemoglobin (Hb), and total hemoglobin (HbT). NIRS has been shown to be a useful tool for evaluating cerebral hemodynamics during various dynamic activities. In particular, NIRS has successfully been used to observe cerebral hemodynamics regulated by AFTs.1013 Various studies have been conducted to explore cerebral hemodynamics during VM for healthy subjects as monitored by NIRS. Perry et al.9 reported that greater VM intensity affects cerebral hemodynamics with reduced oxygenation during each VM phase. The main finding of Tsubaki et al.14 was that HbO measured by NIRS was strongly correlated with mean arterial pressure (MAP) during VM. Various other relationships between VM and cerebral autoregulation were reported in healthy subjects using NIRS.1517 However, the effects of autonomic dysfunction on cerebral hemodynamics during VM have yet to be thoroughly investigated. In normal healthy people, there is a close relationship between cerebral hemodynamic autoregulation and VM, but we hypothesized that, since autonomic dysfunction can have a profound effect on the delivery and regulation of cerebral hemodynamics,18 the VM hemodynamics of patients with autonomic dysfunction might differ from those of healthy subjects.

In this study, we explore the effectiveness of NIRS for proper assessment of patients with OI through various metrics that quantify cerebral hemodynamic activity during VM. In this way, we hoped to present NIRS as a possible method for understanding the mechanisms that govern significant clinical symptoms involving dysregulation of the ANS.

2.

Materials and Methods

2.1.

Ethics Statement

All participants gave written informed consent before study inclusion. All procedures were performed in accordance with the Declaration of Helsinki and were approved by the Korea University Medical Center Institutional Review Board (IRB No. ED15363).

2.2.

Subjects

Patients with symptoms of OI were recruited from January 2016 to June 2016. For study accuracy, we rigorously adjusted the inclusion criteria. Inclusion criteria were (1) symptoms suggesting OI (memory loss, visual difficulties, light headedness, headache, fatigue, increases or decreases in BP, weakness, nausea and abdominal pain, sweating, tremulousness, and exercise intolerance),19 (2) no history of central nervous system diseases, including stroke, Parkinson’s disease, and Alzheimer’s dementia, and (3) no history of significant head injury, alcohol, or psychotropic drug abuse. Demographic and clinical data, including age, gender, and comorbid chronic diseases, were obtained. Normal healthy control (HC) volunteers who did not have any medical history that might affect ANS were recruited. Healthy subjects who had any abnormal findings in the AFTs of this study protocol and had any OI symptoms according to the OI questionnaires were excluded from final analysis. In addition, subjects with arrhythmia were excluded because of the possibility of VM measurement error. A total of 31 patients with OI and 9 healthy subjects who completed all study protocols were recruited. However, four patients and three healthy subjects who showed arrhythmia, which made parameters unreliable, were excluded from final analysis. Two patients who had NIRS signals that were not sufficient to be analyzed due to excessive motion artifacts were excluded. Another three healthy subjects were excluded due to abnormal VM (AbVM) parameter findings: two subjects showed abnormal Valsalva ratios (VRs) and one subject showed abnormal cardiovagal barosensitivity (BRSv). Finally, data from 25 patients and 3 normal HCs were analyzed.

2.3.

Study Design

All participants taking medications affecting autonomic function were asked to discontinue their medicine for at least 24 h prior to VM. To minimize the effects of confounds, VM was controlled and regulated by the standard electrodiagnostic laboratory environment.20 Tests were performed in the following sequence: (1) autonomic dysfunction self-questionnaires were completed by the patient and (2) VMs were conducted while attaching the NIRS probe on the forehead to investigate cerebral hemodynamics.

2.4.

Autonomic Dysfunction Questionnaires

Self-report questionnaires were administered. Autonomic dysfunction was measured using the Korean version of the Composite Autonomic Symptom Score 31 (K-COMPASS 31) consisting of 31 items that assess the following categories: 4 items for OI, 3 items for vasomotor, 4 items for secretomotor, 12 items for gastrointestinal, 3 items for bladder, and 5 items for pupillomotor. The sum of items from each category and the sum of all 31 items were used for analysis. Of note, higher scores indicate more severe autonomic dysfunction.21 We also only scored the four items related to OI separately among the many items of the K-COMPASS 31 questionnaire.

2.5.

Valsalva Maneuver

The VM was performed with Finometer equipment (Finapres Medical Systems BV, Amsterdam, The Netherlands) with a cuff placed on the middle finger. Subjects blew through a mouthpiece attached to a manometer and maintained a pressure of 40 mm Hg for 15 s at rest and in the recumbent position. VM is a difficult procedure to perform consecutively, so each trial was followed by a rest period of 2 min, with testing repeated until two responses with similar beat-to-beat BP and HR recordings were obtained. The VM trial that showed the correct four phases of VM in the BP recordings was chosen for further NIRS analysis. Therefore, there was only one VM trial per subject in which BP and NIRS data were used for analysis. The following BP metrics were determined for each patient. (1) VR was calculated as the maximum HR during the maneuver in phase II divided by the lowest HR in phase IV. The VR of each subject was compared with age-specific reference values.2,22 (2) Pressure recovery time (PRT) was defined as the time from the lowest systolic BP (SBP) during the valley of phase III (start of phase IV) to the return of SBP to baseline before overshooting of SBP during phase IV.1,23 (3) Alternative adrenergic barosensitivity (BRSa1) was defined as the sum of the SBP decrement during early phase II and two-thirds of the SBP decrement in phase III divided by PRT.24,25 (4) BRSv was determined as the regression slope of the RR interval (milliseconds) over SBP during early phase II.2427 Using the cutoff value for each parameter, patients with an abnormal result in any of the four parameters were placed into the AbVM group.

2.6.

Near-Infrared Spectroscopy

A custom-built 108-channel NIRS system was administered to the subject pool to monitor cerebral hemodynamics during VM (Fig. 1). The NIRS probe was built to span the subject’s entire forehead and was secured in place by a medical band. The center bottom of the NIRS probe was aligned with the FpZ point, and the left and right corners were aligned with the F7 and F8 points, according to the International 10–10 system. This alignment has been shown to be appropriate for monitoring cerebral oxygenation changes in the prefrontal area.28,29 The experiment administrator visually ensured good contact for all sources and detectors before the experiment started. In addition, the NIRS software included a module that identified saturated or low-signal channels based on an empirically tested range of light intensities. This was confirmed by the administrator before the tests began. The subjects were instructed to avoid large head motions that might induce motion artifacts in the NIRS signal. The NIRS probe remained on the subject as they performed multiple VM trials.

Fig. 1

A 108-channel matrix of NIRS. (a) Shows the four different source–detector separations (15, 30, 36, and 45 mm) with the bottom of the probe aligned with the F7-Fpz-F8 line on the subject’s head. (b) The actual NIRS probe consisted of 15 photodetectors and 12 LED sources.

NPH_5_1_015002_f001.png

The NIRS probe was attached to the forehead area because of the lack of hair that could significantly degrade signal quality. The probe consisted of 15 photodetectors and 12 light-emitting diodes (LED) sources. Each of the LED sources alternated between wavelengths of 760 and 830 nm at a temporal frequency of 5 Hz for scanning the entire forehead. The power emitted from the LED sources was 4  mW. The 108 channels came from the 108-unique source–detector pairings with different source–detector distances (SD) of 15 mm (40 channels), 30 mm (20 channels), 36 mm (32 channels), and 45 mm (16 channels). The channel matrix is shown in Fig. 1; the green line connecting a source and detector signifies one channel. The separation between source and detector determines the penetration depth, with the greatest sensitivity to gray matter changes being shown at a separation of 30 mm or greater.30 Before the calculation of the NIRS metrics, the data were low-passed filtered (cutoff frequency of 0.2 Hz) and smoothed with a 5-point moving average filter, and the first point of the calculated HbO, Hb, and HbT signals was realigned to zero. In this study, we analyzed changes in HbO, Hb, and HbT to determine whether the observed hemodynamic response varied according to SD.

Resolving for changes in optical density or the log ratio of baseline intensity, Io and transient intensity I changes at the two wavelengths (760 and 830 nm) allowed accurate measurement of relative changes in HbO, Hb, and HbT,31 with HbT being proportional to blood volume.32 In short, the relative concentration change in HbO and Hb can be resolved using the modified Beer–Lambert’s law33

Eq. (1)

ΔHbO=ϵHbλ2log(IoI)λ1ϵHbλ1log(IoI)λ2L(ϵHbλ2ϵHbOλ1  ϵHbλ1ϵHbOλ2),

Eq. (2)

ΔHb=ϵHbOλ1log(IoI)λ2ϵHbOλ2log(IoI)λ1L(ϵHbλ2ϵHbOλ1ϵHbλ1ϵHbOλ2),
where ΔHbO and ΔHb are the change in HbO and Hb, respectively, from the baseline state. The baseline intensity was set to the average intensity of the first 100 acquired points before the VM started. L is the optical pathlength, and ϵHbO,  Hbλ1 and ϵHbO,  Hbλ2 are the extinction coefficients for HbO and Hb for the two wavelengths λ1=760  nm and λ2=830  nm, respectively, taken from Ref. 34. A full review of NIRS methodology and the terms above can be found elsewhere.35

2.7.

NIRS Parameters

Upon preliminary review of the hemodynamic changes during VM, NIRS parameters were determined to compare the hemodynamic changes between subjects. These parameters were latency, duration, and amplitude calculated for HbO and Hb, and the integrated signal calculated for HbT. Latency was defined as the time between reaching the appropriate pressure to perform the VM (40-mm Hg exhalation pressure) and the beginning of the NIRS hemodynamic response. The amplitude was defined as the maximum concentration change in the calculated NIRS response. Duration was defined as the width of the NIRS response or the time between the start of the NIRS response and when the hemodynamics returned to the baseline state. For HbT, the area under the curve was calculated during the VM exhalation period using the trapz function in MATLAB®. Because HbT is proportional to blood volume, the integration of this signal can be thought of as sustained blood volume during the VM period. The NIRS parameters were only calculated through the NIRS response obtained from the VM trials suited best for analysis (Fig. 2). The VM trial was selected based on appropriate phase data in the BP measurements. We also used a VM trial that showed an NIRS response with well-defined latency, duration, and amplitude parameters for NIRS analysis.

Fig. 2

NIRS response during VM. Comparison of BP (blue) and HbO (red) changes for (a) an HC and (b) an AbVM patient with OI. In addition, a visual depiction of calculated metrics is shown. (a) The NIRS parameters (latency, amplitude, and duration) and (b) four phases of the VM (according to BP measurements) are shown.

NPH_5_1_015002_f002.png

2.8.

Statistical Analysis

Patients were divided into two groups: an NVM group and an AbVM group according to their calculated clinical parameters based on VM results. In addition, data for VM and NIRS in healthy subjects were evaluated. Differences in demographic characteristics, questionnaire scores, and proportion of subjects with abnormal results from VM among the patient groups and the normal healthy group were analyzed using either the Kruskal–Wallis test or Fisher’s exact test. The NIRS parameters were calculated and compared between the two patient groups with OI and normal healthy subjects using a custom-built script written in MATLAB® (2010A, The MathWorks Inc., Natick, Massachusetts). A significant difference for metrics between the groups was tested using the two-tailed t-test. A p-value <0.05 was considered statistically significant. Pearson’s correlation analyses were performed to identify relationships between VM parameters and the NIRS signal.

3.

Results

3.1.

Subject Characteristics

Data from 25 patients and 3 normal HCs were analyzed. The patients with OI were subgrouped according to VM findings as follows: (1) NVM (n=9; median age 66.0; 2 males) and (2) AbVM group (n=16; median age 49.0; 9 males). Table 1 presents demographic and clinical characteristics, including VM results, for all subjects.

Table 1

Subject demographic and clinical characteristics.

NVM (n=9)AbVM (n=16)Normal healthy (n=3)p-value
Clinical features
Male (%)22.25066.70.27
Age (year)66.00 (31, 80)49.00 (22, 79)66.00 (66, 74)0.11
Height (m)1.61 (1.54, 1.79)1.62 (1.49, 1.75)1.61 (1.56, 1.65)0.86
Weight (kg)61.00 (53, 94)60.00 (45, 84)62.00 (55.5, 68.0)0.64
BMI (
kg/m2
)
23.05 (20.20, 28.06)23.64 (17.93, 30.48)24.98 (21.33, 25.48)0.97
VM
PRT1.41 (0.58, 2.90)2.03 (0.62, 25.2)1.28 (0.60, 1.68)0.18
BRSa148.58 (30.34, 78.82)34.05 (6.18, 128.1)40.82 (40.33, 117.61)0.47
BRSv4.20 (3.50, 6.70)2.60 (1.2, 14.3)3.60 (3.30, 6.20)0.14
VR abnormality (%)029.400.12
Questionnaire
K-COMPASS 319.00 (0, 20)19.00 (2, 48)0.00 (0, 0)0.00
OI score1.00 (0, 6)4.50 (0, 10)0.00 (0, 0)0.02
Note: Statistically significant values (p < 0.05) are highlighted in bold.Note: Data are presented as median (range; minimum, maximum).VM, Valsalva maneuver; BMI, body mass index; PRT, pressure recovery time; BRSa, adrenergic barosensitivity; BRSv, cardiovagal barosensitivity; VR, Valsalva ratio; and OI, orthostatic intolerance.

3.2.

Comparison of NIRS Parameters Between Patients and Healthy Subjects

NIRS parameters (latency, amplitude, duration, and integration) were calculated for the three subject groups: HC, NVM, and AbVM. Figure 2 shows a sample BP and HbO response by (a) HC and (b) AbVM. In this example of an HC and AbVM patient, it can be seen that the AbVM had a longer latency and lower amplitude than the HC; while the HbO peak of the HC occurred during the phase II period of VM, the peak for the AbVM HbO signal came later, and the slope to peak was also lower compared with the HC HbO.

We first investigated whether there was evidence of localized hemodynamic changes across the prefrontal area during VM. The sectioning of the NIRS probe according to its location on the forehead is shown in Fig. 3. The location on the prefrontal area is denoted by the International 10–10 coordinates. Figure 4 shows the box plot for the HbO metrics [(a) latency, (b) amplitude, (c) duration, and (d) HbT integration] for the three groups at a 30-mm SD. It can be seen from the figure that between the six areas, all metrics were measured similarly for the three subject groups. The biggest differences between groups can be seen in the HC group; however, this can be attributed to the low number of subjects. Thus, for simplicity, the proceeding analysis between groups will be shown for all channels averaged across the prefrontal area.

Fig. 3

Probe sectioning used for channel averaging. The NIRS channels were averaged according to six different groups across the forehead. The International 10–10 coordinates are noted on the diagram for orientation.

NPH_5_1_015002_f003.png

Fig. 4

NIRS metrics for the six regions: (a) HbO latency for 30-mm SD divided by the six regions (as shown in Fig. 3), (b) HbO amplitude for 30-mm SD divided by the six regions, (c) HbO duration for 30-mm SD divided by the six regions, and (d) HbT integration for 30-mm SD divided by the six regions.

NPH_5_1_015002_f004.png

The average HbO and Hb latency for the three groups is shown in Fig. 5. The latencies of the HC and NVM groups were significantly shorter than that of the AbVM group (p<0.05). For the HC group, the HbO latency at 30 mm was 0.39±0.23  s. The NVM group had an HbO latency of 2.79±0.36  s, and the AbVM group had an HbO latency of 8.14±0.55  s. For Hb, the latency at 30 mm for HC, NVM, and AbVM was 1.64±0.51  s, 3.33±0.41  s, and 5.23±0.34  s, respectively. Figure 5 shows the statistically significant difference between Hb latency at the 45-mm distance and shorter channels (i.e., 15 and 30 mm). However, the 45-mm distance difference in latency may be attributed to increased scattering of the light reaching that detector. In addition, there were no significant differences between SDs by latency. Overall, our results showed that OI patients had a longer latency for onset of the hemodynamic response than the HC group, with the AbVM group having the longest latency.

Fig. 5

Latency for all averaged channels: (a) HbO latency and (b) Hb latency for all channels are averaged according to source–detector separation. The bars are ordered from left to right from shortest separation (15 mm) to farthest separation (45 mm). Healthy subjects showed the smallest latency at the shorter distance source–detector separations, especially for the HbO signal. Statistically significant values (p<0.5) compared with the healthy group or patient NVM group are depicted by * or +, respectively. Statistically significant values (p<0.5) between distances within a group are connected via arrows.

NPH_5_1_015002_f005.png

While latency clearly distinguished between the groups, amplitude and duration were not significantly different between groups for the NIRS response. The full list of p-values between groups for NIRS metrics obtained with a two-tailed t-test is presented in Table 2. Although there was no statistical significance, certain patterns can be observed in Figs. 6 and 7. The amplitude of the HbO and Hb response tended to peak at the shorter separations for all groups and decreased as channel distance became longer, as shown in Fig. 6. The drop-off in amplitude is especially noticeable in the Hb response. For amplitude, there was a statistical difference at the 15-mm distance and all other distances for all groups in their HbO and Hb responses. For farther channels, the number of distances that were statistically different was reduced, showing that amplitude was drastically reduced with longer channels.

Table 2

p-values from a two-tailed t-test for groups of patients (healthy, patient normal, and patient abnormal) for each metric (latency, amplitude, duration, and HbT integration) at different hemodynamic measurements (HbO, Hb, and HbT).

15 mm30 mm36 mm45 mm
HbO
Latency
Healthy/patient normal0.1820.1180.1080.176
Healthy/patient abnormal0.0110.0130.0120.018
Patient normal/patient abnormal0.0030.0040.0080.023
Amplitude
Healthy/patient normal0.4130.4810.1490.372
Healthy/patient abnormal0.2720.2340.1800.572
Patient normal/patient abnormal0.7910.4580.6860.703
Duration
Healthy/patient normal0.6690.2500.0710.810
Healthy/patient abnormal0.5150.8910.6840.787
Patient normal/patient abnormal0.1020.0910.1050.462
Hb
Latency
Healthy/patient normal0.04160.14310.99760.4696
Healthy/patient abnormal0.02160.01600.57900.5914
Patient normal/patient abnormal0.02020.03790.40250.7288
Amplitude
Healthy/patient normal0.4940.7460.6150.827
Healthy/patient abnormal0.3200.7020.6980.593
Patient normal/patient abnormal0.5430.8400.7710.274
Duration
Healthy/patient normal0.1770.0830.8890.236
Healthy/patient abnormal0.1450.0820.2350.469
Patient normal/patient abnormal0.4290.3920.0630.499
HbT
Integration
Healthy/patient normal0.2430.2390.1810.305
Healthy/patient abnormal0.0300.0260.0260.102
Patient normal/patient abnormal0.2300.1290.1510.422
Note: Statistically significant values (p<0.05) are highlighted in bold.

Fig. 6

Amplitude for all channels averaged according to source–detector separation. The amplitude of (a) HbO and (b) Hb changes peaked at shorter channels for all patient groups and dropped for the longer channels. Statistically significant values (p<0.5) compared with the healthy group or patient NVM group are depicted by * or +, respectively. Statistically significant values (p<0.5) between distances within a group are connected via arrows.

NPH_5_1_015002_f006.png

Fig. 7

Duration for all channels averaged according to source–detector separation. The bars are ordered from left to right from shortest separation (15 mm) to farthest separation (45 mm) for the recorded NIRS duration of (a) HbO and (b) Hb. The duration parameter shows less difference among the groups. Statistically significant values (p<0.5) compared with the healthy group or patient NVM group are depicted by * or +, respectively. Statistically significant values (p<0.5) between distances within a group are connected via arrows.

NPH_5_1_015002_f007.png

In terms of duration, Fig. 7 shows that HbO and Hb duration did not show any significant pattern between the groups. The difference between channel distances within a group did not demonstrate any recurring trend; however, the 45-mm channels did show the longest duration in the Hb signal for the healthy group.

As a measure of sustained blood volume in the prefrontal area during VM, HbT integration was compared among the three groups (Fig. 8). The HbT integration was significantly different between the HC and AbVM groups. The HC group showed a higher sustained blood volume change in the prefrontal superficial layer during VM compared with the OI groups. Similar to the amplitude metric, the shortest channels were significantly different from other distances for all groups, and the number of statistically significant distances was reduced as the channels became farther apart.

Fig. 8

HbT integration for all channels averaged according to source–detector separation. The bars are ordered from left to right from shortest separation (15 mm) to farthest separation (45 mm) for the integration of the HbT signal. Statistically significant values (p<0.5) compared with the healthy group or patient NVM group are depicted by * or +, respectively. Statistically significant values (p<0.5) between distances within a group are connected via arrows.

NPH_5_1_015002_f008.png

To observe the effect of superficial channels, the shorter channels (15 mm) were regressed from 30, 36, and 45 mm data, using previously described procedures.15 Figure 9 shows the HbO metrics recalculated after short channel removal. At 30 mm, latency was significantly different (p<0.05) between healthy and NVM/AbVM [Fig. 9(a)]. In addition, once the short channels were regressed, HbO duration was significantly different between the healthy and OI patients at 30 mm [Fig. 9(c)]. However, there was no significant difference between NVM and AbVM in the HbO and HbT integration signal when the short channel was regressed.

Fig. 9

Calculated metrics with short channel removed: (a) HbO latency, (b) amplitude, (c) duration, and (d) HbT integration calculated for 30, 36, and 45 mm, with the 15-mm channel regressed from the data. Statistically significant values (p<0.5) compared with the healthy group or patient normal group are depicted as a * or +, respectively.

NPH_5_1_015002_f009.png

3.3.

Correlation of NIRS Parameters to VM Parameters

We also investigated whether there was any relationship between our derived NIRS parameter and the typical VM parameters used to diagnosis clinical symptoms. We performed a linear regression for the NIRS parameters (HbO, Hb: latency, duration, amplitude, and HbT integration) averaged across the entire prefrontal area, per SD, and the VM parameters (VR, PRT, BRSa1, and BRSv). According to our regression analysis, the only combination of NIRS and VM parameters with a statistically significant relationship was HbO duration at 15-mm source–detector separation and VR (R2=0.164, p=0.032). The results of the linear regression are shown in Fig. 10. When HbO duration at 15 mm and VR were tested for each group, there was no significant relationship between the parameters. Results of the linear regression per group were as follows: HC (R2=0.003, p=0.964), NVM (R2=0.159, p=0.288), and AbVM (R2=0.207, p=0.076). The small HC size and similar values in VR between NVM (average: 1.58±0.28) and AbVM (average: 1.62±0.35) resulted in a poor fit for each group individually. However, when all the subjects were grouped together, the fit was statistically significant.

Fig. 10

Linear regression analysis of NIRS and VM parameters. Each patient group is designated by a specific shape. This combination of HbO duration (15-mm source–detector separation) and VR was the only combination that showed a statistically significant difference (p<0.05).

NPH_5_1_015002_f010.png

4.

Discussion

VM performance can be measured by various clinical parameters using BP and HR data. Baroreceptor sensitivity has been represented through parameters such as BRSv and BRSa1.27 In addition, we measured VR and PRT for all subjects. VR is the most commonly used parameter to estimate cardiovagal function. However, VR is often not sensitive enough to accurately detect patient abnormal cardiovagal function;36 therefore, other clinical parameters should also be considered to discriminate abnormal cardiovagal function. The symptom questionnaire showed significantly higher scores in the patient group compared with the HC group and even higher in patients with AbVM compared with patients with NVM. This suggests that discrimination of VM abnormality is best achieved through multiple parameters to be reliable. In addition to clinical parameters, we analyzed VM performance between groups with NIRS to monitor relative HbO, Hb, and HbT concentration changes. We analyzed the latency, amplitude, duration, and HbT integration (i.e., sustained blood volume) of the NIRS response to distinguish between HC, NVM, and AbVM during VM.

Of these parameters, latency showed the highest contrast among the groups. The AbVM group had longer latency for HbO changes for all source–detector separations compared with NVM (p<0.05). Cerebral perfusion pressure is defined as MAP minus intracranial pressure (ICP) or MAP minus central venous pressure (CVP), if CVP>ICP. VM leads to a proportional elevation in CVP, a marked reduction in cardiac output, and characteristic changes in MAP.3739 In particular, the abrupt reduction in MAP challenges the regulation of cerebral perfusion and can result in syncope.40 Our hypothesis is that increased CVP leads to sustained blood volume retention in the superficially distributed venous system of the forehead, which leads to rapid detection of cerebral hemodynamics in a shallow channel and a short latency for the NIRS signal. We believe that, in the AbVM, the latency can be increased due to the interruption of hemodynamic autoregulation compared with NVM. AbVM also showed a pattern of more diminished HbO response amplitude compared with NVM, with a reduced response as the SD increased, except in the 45-mm channel; however, the results between groups were not statistically significant, but the difference between distances was significant. Compared with AbVM, the NVM did not differ statistically in sustained volume, but the HC showed a difference during VM except for the farthest channel (Fig. 8). The lack of significant differences in the 45-mm channel for the NIRS metrics may be attributed to strong light scattering.

The calculated NIRS metrics can provide clinical insight into cerebral perfusion during VM. Delayed HbO latency may represent delayed cerebral perfusion in patients with AbVM compared with patients with NVM. Although statistical significance was not shown, even OI patients with NVM results showed slightly delayed HbO compared with HC. According to the 2017 ACC/AHA/HRS guideline for the evaluation and management of patients with syncope, the mechanism of syncope, one of the symptoms of OI, is explained by cerebral hypoperfusion.41 Lankford et al. showed that hypotension associated with OI leads to decreased cerebral perfusion. This decreased perfusion is working as a cause for neurologic clinical manifestations.13 Gulli et al. stated that patients suffering from poor OI fail to maintain an adequate brain perfusion pressure. This study showed that prolonged delay in the baroreflex response was associated with the poor OI.42 These findings underlie suspicion that OI symptoms might develop due to delayed cerebral perfusion, which might cause an oxygen deficit for a few seconds.

To assess the overall recruitment of cerebral hemodynamics during the VM, we evaluated the amplitude of the NIRS signal; there was no difference in HbO amplitude by group, but there was a difference according to NIRS channel distance. While HbO amplitude showed a pattern among the groups, Hb amplitude did not show any pattern except for the 15-mm SD distance. In general, neuronal activities in normal subjects cause a decrease in Hb associated with increases in HbO and HbT in the activated cortical area caused by O2 delivery in excess of O2 consumption.43,44 However, the brain dysfunction may present various forms of Hb metabolism during neuronal activation and the NIRS signal in a short source–detector channel (e.g., 15 mm) may be affected by sinus and/or forehead skin vasculature, so there may not be any difference in Hb response.32,45

The HbT integration, which is related to cerebral blood volume,46 was not significantly different between OI groups. The HC group showed higher sustained blood volume changes in the prefrontal superficial layer during VM compared with the OI groups. VM leads to a proportional elevation in intrathoracic and CVP. The increase in CVP causes retention of sustained blood volume in the venous system of the forehead and causes superficial blood volume changes. We observed an increased HbT in the mainly HC group and the patients with NVM group compared with patients with AbVM group. The reason for the increased HbT in mainly shallow channels over other channels is believed to reflect the venous and/or sinus system and skin vascular pooling occurring at the measurement sites. VM produces a marked increase in blood volume in extracranial compartments. The increased extracranial blood volume heavily affects HbT parameters. This may reflect skin vascular pooling, since extracranial signals may be present in these NIRS signals.9,47 A study showed that the average scalp thickness was 6.9±3.6  mm, while the average skull thickness was 6.0±1.9  mm.48 Considering the important effects of overlying scalp and skull layers on the measurement of brain hemodynamics, the reason that the short SD (15 mm) is consistently showing a significant difference in NIRS analysis during VM may be due to general superficial hemodynamics influenced during VM, rather than cerebral hemodynamics. Moreover, once the short channels were regressed, HbO latency was significantly different between the groups.

Consequently, HbT integration, along with duration and amplitude, did not show as large a contrast between the NVM and AbVM groups as latency did; thus, we concluded that latency had the most potential as a quality metric for identifying an AbVM response in patients with OI.

This study had some limitations. First, there were no data related to cerebrovascular status in our subjects. We tried to exclude any possibility of CNS abnormality, including cerebrovascular abnormality, via a thorough medical interview and neurologic examination. Further studies with cerebral vascular evaluation, such as MR angiography, would provide more reliable scientific results. The other limitation is the small HC sample size. There was no statistically significant different response for latency, but this may be attributed to the limited HC group sample size. Although the number of HCs may be seen as a limitation of this study, we believe that the dataset is reliable. We rigorously enforced inclusion criteria for the 15 HC subjects. The criteria excluded subjects with any problem that showed up in the OI questionnaires and subjects who showed AbVM data or potential signs of disorders of the ANS or CNS. After the screening, we were left with three HCs who were able to pass multiple OI tests [i.e., questionnaires, VM, etc. (Table 1)]. Because of this strict screening, we believe that our HC subject data represent subjects with no signs of OI.

An additional limitation may result from the distances of the NIRS probe sources and detectors. With multiple source–detector analysis, shorter channel regressions are typically performed to isolate the hemodynamic response from the superficial area and the cerebral response. Further investigation in this area would be worthwhile, but it has been shown that VM is largely a systemic maneuver whose signal comes predominantly from superficial hemodynamic changes rather than cerebral changes.15 Previous studies have shown that the ideal distance to separate the superficial changes is 8  mm.49 However, considering the thickness of the overlying scalp and skull layers, the suitable distance to measure cerebral activity is between 20- and 45-mm SD separations.48 For this reason, although the shortest SD distance was 15 mm due to the design of our NIRS probe, it may be a reasonable distance to observe superficial changes.

Although Fig. 10 shows a correlation between NIRS parameters (HbO duration) and the clinical VM parameter (VR), it is worth investigating the intertwined relationships between our derived NIRS parameters and the clinically used VM parameters. Our current study focused on the prefrontal hemodynamic changes rather than the physiological significance and relationship between the two parameter groups (NIRS and VM). By closely understanding how cerebral perfusion and autoregulatory systems affect these metrics during VM, we can better quantify and analyze the cerebral autoregulatory system.

NIRS offers an alternate solution for understanding the effects of a dysregulated ANS on the delivery of blood to the brain. In addition, NIRS has the potential to monitor patients with inconclusive BP or HR readings. Our results confirm the presence of a delay in hemodynamic response according to the VM procedure. Although the delay in the recruitment of hemodynamics to the frontal area during VM is still largely unknown,50 differentiating between healthy subjects and OI patients according to VM performance is important for understanding the autoregulatory nature of the nervous system. For further study of cerebral hemodynamics during VM, larger sample sizes are required, including normal healthy participants, and detailed studies should be conducted at deeper layers during NIRS.

Disclosures

All of the authors had no relevant financial or competing interests and no other potential conflicts of interest.

Acknowledgments

This research was supported by the Original Technology Research Program for Brain Science through the National Research Foundation (NRF) of Korea funded by the Ministry of Science ICT and Future Planning (MSIP) (No. 2017M3C7A1048566) and the NRF of Korea Grant funded by the Korean Government (MSIP) (No. NRF-2015R1A5A7037674). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the paper.

References

1. 

P. Novak, “Quantitative autonomic testing,” J. Visualized Exp., (53), e2502 (2011). http://doi.org/10.3791/2502 Google Scholar

2. 

P. A. Low, “Autonomic nervous system function,” J. Clin. Neurophysiol., 10 (1), 14 –27 (1993). http://dx.doi.org/10.1097/00004691-199301000-00003 Google Scholar

3. 

P. A. Low and V. A. Tomalia, “Orthostatic hypotension: mechanisms, causes, management,” J. Clin. Neurol., 11 (3), 220 –226 (2015). http://dx.doi.org/10.3988/jcn.2015.11.3.220 Google Scholar

4. 

G. Tian et al., “External counterpulsation reduces beat-to-beat blood pressure variability when augmenting blood pressure and cerebral blood flow in ischemic stroke,” J. Clin. Neurol., 12 (3), 308 –315 (2016). http://dx.doi.org/10.3988/jcn.2016.12.3.308 Google Scholar

5. 

W. Wendling et al., “Cardiovascular and cerebrovascular effects of the applied Valsalva manoeuvre in anaesthetized neurosurgical patients,” Eur. J. Anaesthesiol., 11 (2), 81 –87 (1994). EJANEG 1365-2346 Google Scholar

6. 

G. Nath, G. Korula and M. J. Chandy, “Effect of intrathecal saline injection and Valsalva maneuver on cerebral perfusion pressure during transsphenoidal surgery for pituitary macroadenoma,” J. Neurosurg. Anesthesiol., 7 (1), 1 –6 (1995). http://dx.doi.org/10.1097/00008506-199501000-00001 JNANEV 0898-4921 Google Scholar

7. 

G. Ossard et al., “Cerebral blood flow velocity response induced by a 70-hPa Valsalva manoeuvre associated with normo- and hypergravity in humans,” Eur. J. Appl. Physiol. Occup. Physiol., 72 (5–6), 502 –508 (1996). http://dx.doi.org/10.1007/BF00242282 EJAPCK 1432-1025 Google Scholar

8. 

F. Pott et al., “Middle cerebral artery blood velocity during a Valsalva maneuver in the standing position,” J. Appl. Physiol., 88 (5), 1545 –1550 (2000). http://dx.doi.org/10.1152/jappl.2000.88.5.1545 Google Scholar

9. 

B. G. Perry et al., “Cerebral hemodynamics during graded Valsalva maneuvers,” Front. Physiol., 5 349 (2014). http://dx.doi.org/10.3389/fphys.2014.00349 Google Scholar

10. 

B. Wang et al., “Posture-related changes in brain functional connectivity as assessed by wavelet phase coherence of NIRS signals in elderly subjects,” Behav. Brain Res., 312 238 –245 (2016). http://dx.doi.org/10.1016/j.bbr.2016.06.037 BBREDI 0166-4328 Google Scholar

11. 

M. A. Yucel et al., “Short separation regression improves statistical significance and better localizes the hemodynamic response obtained by near-infrared spectroscopy for tasks with differing autonomic responses,” Neurophotonics, 2 (3), 035005 (2015). http://dx.doi.org/10.1117/1.NPh.2.3.035005 Google Scholar

12. 

K. J. Stone et al., “The validity and reliability of continuous-wave near-infrared spectroscopy for the assessment of leg blood volume during an orthostatic challenge,” Atherosclerosis, 251 234 –239 2016). http://dx.doi.org/10.1016/j.atherosclerosis.2016.06.030 Google Scholar

13. 

J. Lankford et al., “Cerebral blood flow during HUTT in young patients with orthostatic intolerance,” Clin. Auton. Res., 25 (5), 277 –284 (2015). http://dx.doi.org/10.1007/s10286-015-0295-9 Google Scholar

14. 

A. Tsubaki et al., “Effect of Valsalva maneuver-induced hemodynamic changes on brain near-infrared spectroscopy measurements,” Oxygen Transport to Tissue XXXV, 789 97 –103 Springer, New York (2013). Google Scholar

15. 

R. Saager and A. Berger, “Measurement of layer-like hemodynamic trends in scalp and cortex: implications for physiological baseline suppression in functional near-infrared spectroscopy,” J. Biomed. Opt., 13 (3), 034017 (2008). http://dx.doi.org/10.1117/1.2940587 JBOPFO 1083-3668 Google Scholar

16. 

F. Pott et al., “Middle cerebral artery blood velocity during intense static exercise is dominated by a Valsalva maneuver,” J. Appl. Physiol., 94 (4), 1335 –1344 (2003). http://dx.doi.org/10.1152/japplphysiol.00457.2002 Google Scholar

17. 

M. Laures et al., “Assessing the role of cerebral autoregulation during intrathoracic pressure changes by near infrared spectroscopy (NIRS),” Eur. Respir. J., 46 (59), PA2355 (2015). http://dx.doi.org/10.1183/13993003.congress-2015.PA2355 ERJOEI 0903-1936 Google Scholar

18. 

A. H. van Beek et al., “Cerebral autoregulation: an overview of current concepts and methodology with special focus on the elderly,” J. Cereb. Blood Flow Metab., 28 (6), 1071 –1085 (2008). http://dx.doi.org/10.1038/jcbfm.2008.13 Google Scholar

19. 

P. A. Low et al., “Prospective evaluation of clinical characteristics of orthostatic hypotension,” Mayo Clin. Proc., 70 (7), 617 –622 (1995). http://dx.doi.org/10.4065/70.7.617 MACPAJ 0025-6196 Google Scholar

20. 

J. B. Kim et al., “Utility of corrected QT interval in orthostatic intolerance,” PLoS One, 9 (9), e106417 (2014). http://dx.doi.org/10.1371/journal.pone.0106417 POLNCL 1932-6203 Google Scholar

21. 

D. M. Sletten et al., “COMPASS 31: a refined and abbreviated composite autonomic symptom score,” Mayo Clin. Proc., 87 (12), 1196 –1201 (2012). http://dx.doi.org/10.1016/j.mayocp.2012.10.013 MACPAJ 0025-6196 Google Scholar

22. 

J. B. Kim et al., “Autonomic dysfunction according to disease progression in Parkinson’s disease,” Parkinsonism Relat. Disord., 20 (3), 303 –307 (2014). http://dx.doi.org/10.1016/j.parkreldis.2013.12.001 Google Scholar

23. 

E. R. Vogel, P. Sandroni and P. A. Low, “Blood pressure recovery from Valsalva maneuver in patients with autonomic failure,” Neurology, 65 (10), 1533 –1537 (2005). http://dx.doi.org/10.1212/01.wnl.0000184504.13173.ef NEURAI 0028-3878 Google Scholar

24. 

C. C. Huang et al., “Effect of age on adrenergic and vagal baroreflex sensitivity in normal subjects,” Muscle Nerve, 36 (5), 637 –642 (2007). http://dx.doi.org/10.1002/mus.20853 Google Scholar

25. 

I. Palamarchuk et al., “Baroreflex sensitivity: reliability of baroreflex components of the Valsalva maneuver,” Auton. Neurosci., 185 138 –140 (2014). http://dx.doi.org/10.1016/j.autneu.2014.05.002 Google Scholar

26. 

S. Roy et al., “Cardiovagal baroreflex sensitivity in Parkinson’s disease and multiple-system atrophy,” J. Clin. Neurol., 12 (2), 218 –223 (2016). http://dx.doi.org/10.3988/jcn.2016.12.2.218 Google Scholar

27. 

C. Schrezenmaier et al., “Adrenergic and vagal baroreflex sensitivity in autonomic failure,” Arch. Neurol., 64 (3), 381 –386 (2007). http://dx.doi.org/10.1001/archneur.64.3.381 ARNEAS 0003-9942 Google Scholar

28. 

T. Goto et al., “Lateralized frontal activity for Japanese phonological processing during child development,” Front. Hum. Neurosci., 9 417 (2015). http://dx.doi.org/10.3389/fnhum.2015.00417 Google Scholar

29. 

Y. Kita et al., “Self-face recognition in children with autism spectrum disorders: a near-infrared spectroscopy study,” Brain Dev., 33 (6), 494 –503 (2011). http://dx.doi.org/10.1016/j.braindev.2010.11.007 NTHAA7 0029-0831 Google Scholar

30. 

E. Okada et al., “Theoretical and experimental investigation of near-infrared light propagation in a model of the adult head,” Appl. Opt., 36 (1), 21 –31 (1997). http://dx.doi.org/10.1364/AO.36.000021 APOPAI 0003-6935 Google Scholar

31. 

G. Strangman, M. A. Franceschini and D. A. Boas, “Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters,” NeuroImage, 18 (4), 865 –879 (2003). http://dx.doi.org/10.1016/S1053-8119(03)00021-1 NEIMEF 1053-8119 Google Scholar

32. 

K. Sakatani et al., “Cerebral blood oxygenation changes induced by auditory stimulation in newborn infants measured by near infrared spectroscopy,” Early Hum. Dev., 55 (3), 229 –236 (1999). http://dx.doi.org/10.1016/S0378-3782(99)00019-5 EHDEDN 0378-3782 Google Scholar

33. 

H. Liu et al., “Noninvasive investigation of blood oxygenation dynamics of tumors by near-infrared spectroscopy,” Appl. Opt., 39 (28), 5231 –5243 (2000). http://dx.doi.org/10.1364/AO.39.005231 APOPAI 0003-6935 Google Scholar

34. 

S. Prahl, Optical Absorption of Hemoglobin, Oregon Medical Laser Center(1999). Google Scholar

35. 

A. Villringer and B. Chance, “Non-invasive optical spectroscopy and imaging of human brain function,” Trends Neurosci., 20 (10), 435 –442 (1997). http://dx.doi.org/10.1016/S0166-2236(97)01132-6 TNSCDR 0166-2236 Google Scholar

36. 

A. H. Rothschild et al., “Sensitivity of R-R variation and Valsalva ratio in assessment of cardiovascular diabetic autonomic neuropathy,” Diabetes Care, 10 (6), 735 –741 (1987). http://dx.doi.org/10.2337/diacare.10.6.735 DICAD2 0149-5992 Google Scholar

37. 

H. Goldberg, E. I. Elisberg and L. N. Katz, “The effects of the Valsalva-like maneuver upon the circulation in normal individuals and patients with mitral stenosis,” Circulation, 5 (1), 38 –47 (1952). http://dx.doi.org/10.1161/01.CIR.5.1.38 CIRCAZ 0009-7322 Google Scholar

38. 

F. P. Tiecks et al., “Effects of the Valsalva maneuver on cerebral circulation in healthy adults. A transcranial Doppler study,” Stroke, 26 (8), 1386 –1392 (1995). http://dx.doi.org/10.1161/01.STR.26.8.1386 SJCCA7 0039-2499 Google Scholar

39. 

F. Pott et al., “Middle cerebral artery blood velocity during a Valsalva maneuver in the standing position,” J. Appl. Physiol., 88 (5), 1545 –1550 (2000). http://dx.doi.org/10.1152/jappl.2000.88.5.1545 Google Scholar

40. 

R. C. Duvoisin, “The Valsalva maneuver in the study of syncope,” Electroencephalogr. Clin. Neurophysiol., 13 622 –626 (1961). http://dx.doi.org/10.1016/0013-4694(61)90178-X ECNEAZ 0013-4694 Google Scholar

41. 

W. K. Shen et al., “2017 ACC/AHA/HRS guideline for the evaluation and management of patients with syncope: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines and the Heart Rhythm Society,” Heart Rhythm, 14 (8), e155 –e217 (2017). http://dx.doi.org/10.1016/j.hrthm.2017.03.004 Google Scholar

42. 

G. Gulli et al., “Prolonged latency in the baroreflex mediated vascular resistance response in subjects with postural related syncope,” Clin. Auton. Res., 15 (3), 207 –212 (2005). http://dx.doi.org/10.1007/s10286-005-0273-8 Google Scholar

43. 

C. Hock et al., “Age dependency of changes in cerebral hemoglobin oxygenation during brain activation: a near-infrared spectroscopy study,” J. Cereb. Blood Flow Metab., 15 (6), 1103 –1108 (1995). http://dx.doi.org/10.1038/jcbfm.1995.137 Google Scholar

44. 

C. Hock et al., “Decrease in parietal cerebral hemoglobin oxygenation during performance of a verbal fluency task in patients with Alzheimer’s disease monitored by means of near-infrared spectroscopy (NIRS)–correlation with simultaneous rCBF-PET measurements,” Brain Res., 755 (2), 293 –303 (1997). http://dx.doi.org/10.1016/S0006-8993(97)00122-4 BRREAP 0006-8993 Google Scholar

45. 

K. Sakatani et al., “Language-activated cerebral blood oxygenation and hemodynamic changes of the left prefrontal cortex in poststroke aphasic patients: a near-infrared spectroscopy study,” Stroke, 29 (7), 1299 –304 (1998). http://dx.doi.org/10.1161/01.STR.29.7.1299 SJCCA7 0039-2499 Google Scholar

46. 

J. S. Wyatt et al., “Quantification of cerebral oxygenation and haemodynamics in sick newborn infants by near infrared spectrophotometry,” Lancet, 328 (8515), 1063 –1066 (1986). http://dx.doi.org/10.1016/S0140-6736(86)90467-8 LANCAO 0140-6736 Google Scholar

47. 

D. Canova et al., “Inconsistent detection of changes in cerebral blood volume by near infrared spectroscopy in standard clinical tests,” J. Appl. Physiol., 110 (6), 1646 –1655 (2011). http://dx.doi.org/10.1152/japplphysiol.00003.2011 Google Scholar

48. 

G. E. Strangman, Q. Zhang and Z. Li, “Scalp and skull influence on near infrared photon propagation in the Colin27 brain template,” NeuroImage, 85 (1), 136 –149 (2014). http://dx.doi.org/10.1016/j.neuroimage.2013.04.090 NEIMEF 1053-8119 Google Scholar

49. 

S. Brigadoi and R. J. Cooper, “How short is short? Optimum source–detector distance for short-separation channels in functional near-infrared spectroscopy,” Neurophotonics, 2 (2), 025005 (2015). http://dx.doi.org/10.1117/1.NPh.2.2.025005 Google Scholar

50. 

L. A. Henderson, “Brain responses associated with the Valsalva maneuver revealed by functional magnetic resonance imaging,” J. Neurophysiol., 88 3477 –3486 (2002). http://dx.doi.org/10.1152/jn.00107.2002 Google Scholar

Biographies for the authors are not available.

© 2018 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-423X/2018/$25.00 © 2018 SPIE
Yoo Hwan Kim, Zephaniah Phillips V, Seung-ho Paik, Nam-Joon Jeon, Beop-Min Kim, and Byung-Jo Kim "Prefrontal hemodynamic changes measured using near-infrared spectroscopy during the Valsalva maneuver in patients with orthostatic intolerance," Neurophotonics 5(1), 015002 (26 January 2018). https://doi.org/10.1117/1.NPh.5.1.015002
Received: 20 July 2017; Accepted: 4 January 2018; Published: 26 January 2018
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Near infrared spectroscopy

Hemodynamics

Blood

Neurophotonics

Sensors

Statistical analysis

Brain

RELATED CONTENT