1 November 2002 Influence of aberrations of a Fourier transform infrared spectrometer optical system on measurement errors
Leszek Wawrzyniuk, Romulad Jozwicki
Author Affiliations +
The monochromatic response as a Fourier transform infrared (FTIR) spectrometer apparatus functions is determined to estimate the influence of optical system aberrations on the measurement precision. Exemplary results for the third-order spherical aberration are presented. General conclusions concerning the influence of wave aberrations in the spectrometer channel with moving mirror are formulated.
©(2002) Society of Photo-Optical Instrumentation Engineers (SPIE)
Leszek Wawrzyniuk and Romulad Jozwicki "Influence of aberrations of a Fourier transform infrared spectrometer optical system on measurement errors," Optical Engineering 41(11), (1 November 2002). https://doi.org/10.1117/1.1512303