1 December 1978 An Optical System for an Automatic Eye Refractor
Author Affiliations +
Abstract
Various instruments to measure the refractive state of the eye have been built based on a servosystem concept. This paper describes the optical system of the Dioptron automatic objective refractor. This autorefractor measures the modulation of a grating image formed on the retina while being controlled by a digital computer. The optical subsystems include the image projecting system, the image detecting system, the patient fixation target, and the patient alignment system. Accuracy is obtained by controlling the patient's accommodation and eliminating "instrument" myopia. The overall system is designed to be compact while maintaining very good signal-to-noise ratio and minimum operator controls.
C. R. Munnerlyn, C. R. Munnerlyn, } "An Optical System for an Automatic Eye Refractor," Optical Engineering 17(6), 176627 (1 December 1978). https://doi.org/10.1117/12.7972293 . Submission:
JOURNAL ARTICLE
4 PAGES


SHARE
RELATED CONTENT

Evaluation Of Optical Instruments Using Target Detection
Proceedings of SPIE (December 06 1978)
Evaluation Of Optical Instruments Using Target Detection
Proceedings of SPIE (December 21 1978)
An Optical System For An Automatic Eye Refractor
Proceedings of SPIE (November 20 1977)
Formula for the contrast sensitivity of the human eye
Proceedings of SPIE (December 17 2003)
AOSLO: from benchtop to clinic
Proceedings of SPIE (August 28 2006)