1 June 1981 Imaging CO2 High Energy Laser ,Usec Pulses With A Pyroelectric Vidicon Camera
Steven H. Mersch, James S. Harris, Donald L. Mullen
Author Affiliations +
Abstract
This paper discusses the capabilities and limitations of a pyroelectric vidicon camera used to profile 10 ,usec CO2 laser pulses. Points examined include dynamic range, signal-to-noise ratio capabilities, spatial resolution capabilities, uniformity of sensitivity, maximum faceplate irradiance, and required electro-magnetic (EM) and radio frequency (RF) shielding. The relationship between camera performance and pulse length is also discussed. Samples of data acquired and requirements of an adequate data processing system are presented.
Steven H. Mersch, James S. Harris, and Donald L. Mullen "Imaging CO2 High Energy Laser ,Usec Pulses With A Pyroelectric Vidicon Camera," Optical Engineering 20(3), 203464 (1 June 1981). https://doi.org/10.1117/12.7972742
Published: 1 June 1981
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Carbon dioxide lasers

Cameras

Carbon dioxide

Laser energy

Data acquisition

Data processing

Signal to noise ratio

Back to Top