1 November 1991 Enhanced temporal resolution with a scanning imaging radiometer
Author Affiliations +
Abstract
A system has been developed that reduces the effective integration time of an off-shelf RS-170 optomechanical scanning imaging radiometer by two orders of magnitude. The system can be used to image repetitive thermal events at frequencies up to 4 kHz, without aliasing. The effective integration time of the system is 125 μs/frame. Examples include in-cylinder engine combustion and heat transfer in a transistor operating at 2.5 kHz.
Steven M. Shepard, Steven M. Shepard, David T. Sass, David T. Sass, Thomas P. Imirowicz, Thomas P. Imirowicz, } "Enhanced temporal resolution with a scanning imaging radiometer," Optical Engineering 30(11), (1 November 1991). https://doi.org/10.1117/12.55992 . Submission:
JOURNAL ARTICLE
4 PAGES


SHARE
RELATED CONTENT


Back to Top