1 July 1993 Electro-optic imaging chain for a biplanar fluorscope for neurosurgery: magnetic field sensitivity and contrast measurements
Author Affiliations +
Optical Engineering, 32(7), (1993). doi:10.1117/12.138719
Abstract
The effects of magnetic fields on the microchannel plate (MCP) image intensifiers to be used in a novel biplanar fluoroscope are studied along with the system's overall contrast as a function of beam energy. For a second-generation device with wrap-around power supply, B-3dB values for the gain roll-off were found to be approximately 0.08 T (axial field) and 0.06 T (transverse field). The maximum image shift resuiting from a 0.0035-T transverse field is found to be 0.065 mm, limited by centroid location error resulting from low-dose x-ray noise. The results of x-ray contrast studies suggests that the presently estimated 0.1-rad dose delivered to the patient (2-h magnetic stereotaxis procedure; 12% "fluoro-on" duty cycle) might be reduced by increasing the x-ray energy.
P. A. Ramos, Stephen W. Allison, J. A. Molloy, Michael A. Lawson, E. G. Quate, Rogers C. Ritter, George T. Gilles, Michael Sean Grady, Matthew A. Howard, "Electro-optic imaging chain for a biplanar fluorscope for neurosurgery: magnetic field sensitivity and contrast measurements," Optical Engineering 32(7), (1 July 1993). http://dx.doi.org/10.1117/12.138719
JOURNAL ARTICLE
13 PAGES


SHARE
KEYWORDS
Microchannel plates

Magnetism

Electrons

X-rays

Image intensifiers

X-ray imaging

Imaging systems

Back to Top