1 July 1993 Electro-optic imaging chain for a biplanar fluorscope for neurosurgery: magnetic field sensitivity and contrast measurements
Author Affiliations +
Abstract
The effects of magnetic fields on the microchannel plate (MCP) image intensifiers to be used in a novel biplanar fluoroscope are studied along with the system's overall contrast as a function of beam energy. For a second-generation device with wrap-around power supply, B-3dB values for the gain roll-off were found to be approximately 0.08 T (axial field) and 0.06 T (transverse field). The maximum image shift resuiting from a 0.0035-T transverse field is found to be 0.065 mm, limited by centroid location error resulting from low-dose x-ray noise. The results of x-ray contrast studies suggests that the presently estimated 0.1-rad dose delivered to the patient (2-h magnetic stereotaxis procedure; 12% "fluoro-on" duty cycle) might be reduced by increasing the x-ray energy.
P. A. Ramos, P. A. Ramos, Stephen W. Allison, Stephen W. Allison, J. A. Molloy, J. A. Molloy, Michael A. Lawson, Michael A. Lawson, E. G. Quate, E. G. Quate, Rogers C. Ritter, Rogers C. Ritter, George T. Gilles, George T. Gilles, Michael Sean Grady, Michael Sean Grady, Matthew A. Howard, Matthew A. Howard, } "Electro-optic imaging chain for a biplanar fluorscope for neurosurgery: magnetic field sensitivity and contrast measurements," Optical Engineering 32(7), (1 July 1993). https://doi.org/10.1117/12.138719 . Submission:
JOURNAL ARTICLE
13 PAGES


SHARE
RELATED CONTENT

2D magnetic field warp reversal in images taken with DIXI...
Proceedings of SPIE (September 25 2013)
A Focussed Mcp Image Intensifier Tube For High Speed Gating
Proceedings of SPIE (February 03 1988)
A novel x-ray intensifier
Proceedings of SPIE (January 17 2005)
The Low Intensity X-Ray Imaging Scope (LIXISCOPE)
Proceedings of SPIE (September 14 1978)

Back to Top