1 May 2000 A condition on the spatial resolution of IR collimators for testing of thermal imaging systems
Author Affiliations +
Abstract
A precise condition on the spatial resolution of the IR collimator for testing thermal imaging systems is presented. The condition can be used even if only the spatial resolution of the IR collimator and that of the system under test, measured using popular definitions, are known. It is shown that when the condition is fulfilled, the thermal image degradation caused by the IR collimator is negligible.
Krzysztof Chrzanowski, Krzysztof Chrzanowski, Hee Chul Lee, Hee Chul Lee, Wieslaw Wrona, Wieslaw Wrona, } "A condition on the spatial resolution of IR collimators for testing of thermal imaging systems," Optical Engineering 39(5), (1 May 2000). https://doi.org/10.1117/1.602507 . Submission: