1 November 2002 Influence of aberrations of a Fourier transform infrared spectrometer optical system on measurement errors
Author Affiliations +
Abstract
The monochromatic response as a Fourier transform infrared (FTIR) spectrometer apparatus functions is determined to estimate the influence of optical system aberrations on the measurement precision. Exemplary results for the third-order spherical aberration are presented. General conclusions concerning the influence of wave aberrations in the spectrometer channel with moving mirror are formulated.
Leszek Wawrzyniuk, Romulad Jozwicki, "Influence of aberrations of a Fourier transform infrared spectrometer optical system on measurement errors," Optical Engineering 41(11), (1 November 2002). https://doi.org/10.1117/1.1512303