1 August 2010 Monte Carlo simulation for microchannel plate framing camera
Houzhi Cai, Jinyuan Liu, Lihong Niu, Wenda Peng, Junlan Zhou, Li Gu, Hanben Niu, Jinghua Long
Author Affiliations +
Abstract
A microchannel plate (MCP) framing camera is simulated by using the Monte Carlo method. Considering that the laser pulse width has effect on the experimental exposure time, we simulate the experimental exposure time varying with the laser pulse width. The gain characteristics of the MCP framing camera are simulated while the MCP is applied with very short electrical gating pulses. A new method is presented using a nongain MCP instead of a gain MCP.

1.

Introduction

A microchannel plate (MCP) framing camera is a powerful diagnostic tool for laser-driven inertial confinement fusion and fast Z-pinch experiments.1 Typically, exposure time of a framing camera gated by a MCP 0.5mm thick is about 100ps . The shortest exposure time of 35ps was achieved by using a gated MCP 0.2mm thick.2

Theoretical simulation of the MCP framing camera has been widely studied. Eberhardt presented a discrete stage electron multiplier model for the MCP. He assumed that the stages of the electron multipliers were constant and independent of the applied voltage on the MCP.3 In this work, we assume that the stages are related to the applied voltage on the MCP.

Recently, characteristics of the electron cascade in a MCP are simulated based on the Monte Carlo method.4, 5 Considering that the laser pulse width has effect on the experimental exposure time, we simulate the experimental exposure time varying with laser pulse width. Furthermore, the gain characteristics of the MCP are simulated, while the gating pulse width is not more than the electron transit time. A new method is presented using a nongain MCP instead of a gain MCP. The temporal characteristics of the nongain MCP framing camera are simulated.

2.

Microchannel Plate Model

The MCP is an array of parallel continuous electron multipliers, assuming there is little influence between each channel and all the channels have the same gain characteristics. Then, the electron cascade in the MCP can be approximated by the behavior of a single microchannel.4 The secondary electron yields, energies, and emission angles are the most important parameters in the simulation. The secondary electron yield is determined by the energy and incident angle of the primary electron. While the primary electron incident angle is θ , assume that the primary electron with incident energy Vm(θ) will produce the maximum yield δm(θ) . The Vm(θ) and δm(θ) are given as5

Eq. 1

Vm(θ)=Vm(0)cosθ,

Eq. 2

δm(θ)=δm(0)exp[α(1cosθ)],
where Vm(0) , δm(0) , and α are determined by the MCP lead glass.6 In our simulation, we take the data Vm(0)=280eV , δm(0)=4 , and α=0.5 .

The mean secondary emission yield for the primary electron with energy V and incident angle θ is given as5

Eq. 3

δ(V,θ)=δm(θ)4VVm(θ)[1+VVm(θ)]2.

The actual secondary electron yield is acquired by random sampling from a Poisson distribution with the mean value given by Eq. 3.

The secondary electron energy is sampled randomly from a Maxwell probability distribution with the most probable energy of 2.5eV . Energy conservation is considered in this work to ensure that the sum of the secondary electron emission energy from any given impact is less than the energy of the primary electron.4 The emission angle of each secondary electron is assigned by sampling from a cosine distribution, and the most probable angle of emission is along the surface normal to the channel wall.

The trajectory and velocity of each secondary electron can be calculated using the nonrelativistic equations of motion. Therefore, we can obtain the energy, incident angle, and position when the electrons strike the MCP channel wall. This impact process is repeated until the electrons emerge from the output surface of the MCP.

An ultraviolet (UV) laser is used to excite photoelectrons from the photocathode coated on the input surface of the MCP in the traditional exposure time measurement.2 While the MCP framing camera is operated in the gating state, a short electrical pulse is applied to the MCP. The exposure time is defined as the full width at half maximum (FWHM) of the MCP gain G(t) . From an experiment point of view, the G(t) is hard to obtain due to the laser pulse width influence. We assume that the UV laser pulse is a temporal Gaussian distribution and converted to photoelectron pulse with the same distribution, given as

Eq. 4

n(t)=npexp[4ln2(tΔt)2],
where Δt is the laser pulse width, and np is the amplitude of the Gaussian distribution set to be 100,000.

The Gaussian photoelectron pulse excited by the laser pulse at time T is amplified by the MCP, and the number of electrons N emerging from the channel exit can be deduced by

Eq. 5

N(T)=n(T)G(T)=+n(Tt)G(t)dt.

It can be seen from the prior discussion that the measured result is the convolution between n(T) and G(T) in the traditional exposure time measurement.

3.

Simulation Results and Analysis

MCP with channel diameter d of 12μm , thickness L of 0.5mm , and bias angle θ of 6deg is simulated. The gating electrical pulse is a temporal Gaussian distribution the width of W and amplitude of VP , given as

Eq. 6

V(t)=VPexp[4ln2(tWW)2].

The gating state simulation is initiated by introducing three electrons near the input end of the channel, typical of what might be expected from a 1-keV x-ray.4 The introducing time for each of the three electrons range from 0ps to 2W ps, increased by a step of 0.1ps . While the gating pulse with a width of 300ps and an amplitude of 1kV is applied to the MCP microstrip line, the G(t) can be obtained with Monte Carlo simulation. The gain curve G(t) is plotted in Fig. 1. The FWHM of the G(t) is 90ps . It is the theoretical exposure time of the framing camera.

Fig. 1

Simulation results for a gating pulse with width of 300ps and amplitude of 1kV applied to the MCP. (a) The gain curve G(t) while the laser pulse width is zero. (b) Considering the laser pulse width, the FWHM of N(T) versus the laser pulse width.

080502_1_1.jpg

Since laser pulse width Δt is given, N(T) can be calculated using Eq. 5. Each curve N(T) is calculated with laser pulse width Δt ranging from 0to50ps . The FWHM of curves N(T) is presented in Fig. 1. It is clear from Fig. 1 that the laser pulse width strongly effects the exposure time measured result. If the laser pulse width is limited to zero, the measured result is close to the theoretical exposure time. Therefore, a laser pulse with larger width cannot be used in the exposure time measurement.

Figure 2 shows the collision numbers, the mean transit time, and the transit time spread (TTS) as a function of the direct current (dc) voltage applied to the MCP. The simulation is initiated by introducing single electron near the input end of the MCP. It can be seen from Fig. 2 that all the electron-channel wall collision numbers, the mean transit time of electrons traveling in the MCP, and the TTS are decreased while dc voltage is increased. While the acceleration is increased with increasing MCP voltage, the electrons travel a longer distance down the channel between collisions, and that leads to a decrease of collision numbers. These factors result in decreasing the mean transit time and the TTS. As the dc voltage is 1kV , the electron transit time is 185ps and the TTS is 42ps . These simulation results are consistent with existing measurements.5

Fig. 2

Simulation results for different voltage applied to the MCP. (a) The collision numbers. (b) The transit time and TTS.

080502_1_2.jpg

Figure 3 shows the G(t) with different gating pulse widths. The amplitude of the gating pulse applied to the MCP is 1kV , and the widths are 185 and 135ps , respectively. While the gating pulse width is 185ps , FWHM of the gain curve is 45ps and maximum gain is 60. While the gating pulse width is 135ps , the maximum gain is less than 10 and the FWHM of gain curve is 40ps . One can make a conclusion from Fig. 2 and Fig. 3 that if the gating pulse width is less than the electron transit time, the exposure time of the framing camera is approximately equal to the TTS, and the gain is low. The minimum exposure time is considered to be equal to the TTS. Therefore, although the gating pulse width is much shorter than the transit time, the exposure time cannot be shortened.

Fig. 3

The dynamic gain curves with different gating pulse widths: the broken line is the gain curve for the gating pulse with an amplitude of 1000V and a width of 185ps , and the solid line is that for the gating pulse with a width of 135ps .

080502_1_3.jpg

To shorten the exposure time, we present a new method using a nongain MCP instead of a gain MCP. The nongain MCP is an array of parallel channels. There is no material coated on the channel wall for secondary electron emission, so that the electrons impacting the wall will be absorbed. Only those photoelectrons with appropriate energy and emission angles can emerge from the output end of the channel without collision. The structure of the nongain MCP framing camera is the same as the previous MCP gated framing camera.1 The simulations of the nongain MCP are started by assuming some number of initial photoelectrons near the input end of the MCP. The photoelectron energy is sampled randomly from a Maxwell probability distribution, and the emission angle of each photoelectron is assigned by sampling from a cosine distribution. The trajectory of each photoelectron can be calculated using the nonrelativistic equations of motion. If the photoelectron strikes the MCP channel wall, it will be absorbed. Otherwise, it can emerge from the channel exit. The transit time and the TTS are plotted versus the applied voltage on the MCP in Fig. 4. The mean transit time and the TTS of the nongain MCP are decreased with the dc voltage increase. The TTS in the nongain MCP is one order of magnitude lower than that in the gain MCP. The exposure time in the nongain MCP framing camera is greatly shortened.

Fig. 4

The transit time and the TTS versus dc voltage in the nongain MCP.

080502_1_4.jpg

4.

Conclusions

Simulation for a MCP framing camera based on the Monte Carlo method is described. The simulated results show that the exposure time measurement error is increased with the laser pulse width increasing in the exposure time measurement. A laser pulse width far less than the FWHM of G(t) should be used in the experiment. The minimum exposure time is considered to be equal to the TTS. To shorten the exposure time, a new method is presented using a nongain MCP instead of the gain MCP. The TTS in the nongain MCP is much less than that in the gain MCP from the simulation results. Our future work is to measure the temporal resolution of the nongain MCP framing camera.

Acknowledgment

This work is supported by the Nature Science Foundation of China under grant number 10875080.

References

1. 

J. Liu, L. Niu, W. Peng, “Application of a fast electrical pulse in gated multichannel plate camera,” Rev. Sci. Instrum., 78 (5), 055104 (2007). https://doi.org/10.1063/1.2737750 0034-6748 Google Scholar

2. 

P. M. Bell, J. D. Kilkenny, R. Hanks, and O. Landen, “Measurements with a 35psec gate time microchannel plate camera,” Proc. SPIE, 1346 456 –464 (1990). https://doi.org/10.1117/12.23371 0277-786X Google Scholar

3. 

E. H. Eberhardt, “Gain model for microchannel plates,” Appl. Opt., 18 (9), 1418 –1423 (1979). https://doi.org/10.1364/AO.18.001418 0003-6935 Google Scholar

4. 

M. Wu, C. A. Kruschwitz, D. V. Morgan, and J. Morgan, “Monte Carlo simulations of microchannel plate detectors. I. Steady-state voltage bias results,” Rev. Sci. Instrum., 79 (7), 073104 (2008). https://doi.org/10.1063/1.2949119 0034-6748 Google Scholar

5. 

M. Ito, H. Kume, and K. Oba, “Computer analysis of the timing properties in micro channel plate photomultiplier tubes,” IEEE Trans. Nucl. Sci., NS-31 (1), 408 –412 (1984). https://doi.org/10.1109/TNS.1984.4333288 0018-9499 Google Scholar

6. 

A. Authinaraynan and R. W. Dudding, “Changes in secondary electron yield from reduced lead glasses,” Adv. Electron. Electron Phys., 40A 167 (1976). https://doi.org/10.1016/S0065-2539(08)61470-5 0065-2539 Google Scholar
©(2010) Society of Photo-Optical Instrumentation Engineers (SPIE)