4 June 2012 Laser exposure analysis for a near-infrared ocular interferometer
Author Affiliations +
Abstract
Ocular interferometry has potential value in a variety of ocular measurement applications, including measuring ocular thicknesses, topography of ocular surfaces or the wavefront of the eye. Of particular interest is using interferometry for characterizing corneal shape and irregular corneal features, making this technology attractive due to its inherent accuracy and spatial resolution. A particular challenge of designing an ocular interferometer is determining safe laser exposure levels to the eye, including both the retina and anterior segment. Described here are the laser exposure standards relevant in the interferometer design and the corresponding calculations and results. The results of this work can be used to aid in the design of similar laser-based systems for ocular evaluation.
© 2012 Society of Photo-Optical Instrumentation Engineers (SPIE)
Brian C. Primeau, Brian C. Primeau, Goldie L. Goldstein, Goldie L. Goldstein, John E. Greivenkamp, John E. Greivenkamp, } "Laser exposure analysis for a near-infrared ocular interferometer," Optical Engineering 51(6), 064301 (4 June 2012). https://doi.org/10.1117/1.OE.51.6.064301 . Submission:
JOURNAL ARTICLE
8 PAGES


SHARE
RELATED CONTENT

Corneal injury to ex vivo eyes exposed to a 3.8...
Proceedings of SPIE (April 24 2005)
Biophysical and medical safety basis of laser emission
Proceedings of SPIE (January 07 1996)
Safe use of ultrashort lasers
Proceedings of SPIE (June 03 1999)
Retinal Thermal Injury
Proceedings of SPIE (October 06 1980)
Laser Safety In Manufacturing
Proceedings of SPIE (February 27 1978)

Back to Top