31 May 2013 Measuring the effective pulse duration of nanosecond and femtosecond laser pulses for laser-induced damage experiments
Author Affiliations +
Abstract
We report on our approach to measure the quantity named effective pulse duration as defined in the ISO 21254-1:2011 standard, which deals with laser-induced damage (LID) threshold measurements. The approach is applied to measure pulses from two laser sources: an injection-seeded electro-optically Q-switched Nd:YAG nanosecond system with 10-Hz pulse repetition frequency, and a fully integrated Ti:sapphire laser with 150–400 fs and 2-kHz pulse repetition frequency. For comparison, the full-width-half-maximum (FWHM) of the same pulses is also measured. The analysis and description of the measurement process, the experimental results, and the corresponding uncertainties are presented. A smaller combined uncertainty is obtained for the effective pulse duration than for the FWHM-defined pulse duration for each time scale involved in experiments. This suggests that the effective pulse duration is the appropriate parameter to characterize the pulse duration in LID experiments.
© 2013 Society of Photo-Optical Instrumentation Engineers (SPIE)
Alexandru Zorila, Alexandru Zorila, Laurentiu Rusen, Laurentiu Rusen, Aurel Stratan, Aurel Stratan, George Nemes, George Nemes, } "Measuring the effective pulse duration of nanosecond and femtosecond laser pulses for laser-induced damage experiments," Optical Engineering 52(5), 054203 (31 May 2013). https://doi.org/10.1117/1.OE.52.5.054203 . Submission: