1.IntroductionLaser Doppler perfusion imaging (LDPI) is a bio-optical technique that can be used to generate two-dimensional maps of superficial tissue perfusion.1, 2 LDPI has successfully been used to investigate microvascular blood perfusion in many applications, e.g., assessment of irritant skin reactions,3 for evaluating treatment of basal cell carcinoma,4 and for investigating perfusion in burn wounds.5 Although the technique has found widespread use as a research tool for physiological studies, it has not been frequently used in the clinical setting on a daily basis. Partly due to technical reasons such as lack of possibility to estimate absolute perfusion values, influence of tissue optical properties on the perfusion estimate,6 and the sensitivity to tissue motion not related to blood flow,7 the added clinical value has not yet proven substantial. In order to achieve an accurate LDPI estimation of the microvascular perfusion it is not only assumed but also a prerequisite that the Doppler broadening of the backscattered laser light is caused by scattering in moving blood cells only. If other tissue motion is present, caused by, e.g., patient movements during skin perfusion assessment or by the movement of the heart during studies of myocardial perfusion, an overestimation of the true blood perfusion is inevitable. When the tissue being assessed is moving relative to the LDPI system, both specularly reflected and scattered photons will become Doppler shifted and erroneously contribute to the perfusion estimate. A possible hypothesis is that the perfusion estimate, derived from the LDPI system during tissue motion, is generated by Doppler broadening of: specularly reflected light , light scattered in the tissue , and light scattered by moving blood cells . During well-controlled measurement conditions, i.e., without any tissue motion, and are zero. By eliminating either of or , or both, movement artifacts may be reduced. The use of polarized light in bio-optical research has attracted significant interest in recent years. Jacques et al. have developed a system for imaging superficial tissue layers based on linearly polarized light and proposed that the system may be used to, e.g., identify skin cancer margins.8, 9 Yaroslavsky and colleagues used polarized light combined with either a multispectral dye-enhanced technique10 or fluorescence11 to study the possibility of delineating margins of different types of nonmelanoma skin cancers. The above examples use polarization techniques to enhance backscattered light that has retained the polarization of the incident light, in order to visualize superficial tissue layers only. However, it is also possible to use polarization techniques to block backscattered and reflected light, which has retained its original polarization, from being detected. Polarized light gradually loses its initial polarization state when being multiply scattered in tissue.12, 13, 14, 15 However, light being specularly reflected from the surface retains its original polarization state.16, 17 Inserting an analyzing polarization filter, with the polarization axis perpendicular to the polarization state of the incident laser light, in front of the detectors makes it possible to exclude the specularly reflected laser light from being detected, and still be able to detect photons scattered by moving blood cells. The aim of this study was to investigate if the perpendicular polarization technique could be used to reduce the influence of movement-induced artifacts on the LDPI perfusion signal. 2.Material and Methods2.1.Polarized Laser Doppler Perfusion ImagerThe scanner head of an LDPI system2, 18 (PIM I, Lisca Development AB, Sweden) was modified in order to investigate possible differences in tissue motion influence, for polarized and nonpolarized detection. A linearly polarized laser (C2-220T, , , Saven AB, Sweden) was used as a light source. An analyzing linear polarization filter (HN38, Neutral linear polarizer, Polaroid Corporation, Japan), with the polarization direction perpendicular to the polarization state of the laser light, was inserted in front of the detectors. The polarization filter could easily be removed from the scanner head, allowing measurements to be performed both with and without polarized detection. 2.1.1.LDPI signal generation and processingBackscattered light impinging on the two photodetectors in the scanner head generated two signals: the photocurrent Doppler signal and the total backscattered light intensity signal (dc signal), formed by subtraction and addition of the two photocurrents, respectively.19 The photocurrent Doppler signal was bandpass filtered ( – , ), generating an ac signal. The ac and dc signals were sampled ( samples ) and stored in a computer using a data acquisition board (DAQ-Card 6024E, National Instruments Corp., USA). Software developed in ( 6.1, The MathWorks, Inc., USA) was used for off-line processing of the signals. The discrete digital LDPI perfusion signal was calculated as the first-order moment of the ac signal power spectral density estimate, followed by noise compensation:2, 20 where is an estimate of the power spectral density, calculated from points of the sampled ac signal, and are the lower and upper frequency limits, respectively, and is a noise compensation function, describing the linear relationship between the detector noise and the dc signal level, established by calibration measurements on a stationary (polyacetal plastic) object. The normalization and the noise compensation are performed using an average of the sampled dc signal from the sample points. was set to 512 samples during all analysis (giving one perfusion signal value approximately every millisecond) and and were set to and , respectively.2.2.Movement, Surface, and Flow ModelsTo generate a well-controlled movement, a setup consisting of a minishaker connected to an electrical power amplifier (Type 4810 and Type 2706, respectively, Brüel & Kjær, Copenhagen, Denmark) was used. The output from the electrical power amplifier (movement signal) controlled the position of the shaker. The movement signal was sampled ( samples ) and stored simultaneously with the ac and dc signals, making it possible to correlate the LDPI signals to the movement of the shaker. By feeding the amplifier with a signal from a signal generator, the type of movement, the amplitude, and the frequency could easily be changed. 2.2.1.Flow model with varying surface structuresA flow model with varying surface structures and reflecting properties was created using a piece of . One half of it was prepared with sandpaper (granularity 80) and the other half with plexiglas polish (Fakopol, Röhm & Hass GmBH, Germany) in order to get a rough surface structure (denoted ) and a smooth surface structure (denoted S), respectively. The smooth surface gave rise to a large amount of specular reflection, compared to the rough one. Adding one layer of semitransparent tape ( tape 810, 3M, USA) to the smooth surface created a smooth surface structure with a lower amount of specular reflection (denoted T). Flow channels, parallel to the surface, were created by drilling a hole, 1.3 mm below each of the two surfaces (R and S/T), and inserting a tube (Polyethylene tube, ClayAdams, Division of Becton, Dickingson & Company, USA, and , outer and inner diameter, respectively) through the holes [Fig. 1(a) ]. Flow measurements were enabled by connecting one end of the tube to a syringe pump (M362, , Orion Research, Inc., USA), with a well-controlled flow. The piece could be attached to the minishaker in a way that enabled movement in two directions; horizontal and vertical (Fig. 1). 2.3.Flow Model and Skin MeasurementsThe scanner head was placed above the measurement object and each measurement procedure, performed in darkness, lasted for . A sinusoidal shaker movement (amplitude ) was used and measurements were performed at four different shaker frequencies; 0, 1, 2, and , corresponding to an approximate maximum velocity of 0, 1, 2, and , respectively. All measurements were performed both with and without the polarization filter in front of the detectors and both in the horizontal and the vertical direction (Fig. 1). 2.3.1.Flow model measurementsMeasurements on the flow model were performed for all three surface structures. Milk with 3% fat concentration was diluted (1:500) in water at room temperature, and perfused through the tube at nine different flow velocities ranging from approximately 0.6 to . The total number of measurements was 432 (three surfaces, with and without polarization filter, horizontal and vertical movement, four shaker frequencies, and nine flow velocities). 2.3.2.Skin measurementsTwelve healthy Caucasian subjects (six male and six female, mean age 30, range 24–48), denoted A–L, were recruited for this study, which was approved by the regional Human Ethics Committee (Number 15-04). All subjects were nonsmokers and nondiabetic and were not allowed to drink beverages containing caffeine during the two hours immediately prior to the measurements. Before measurements were performed at the dorsal sides of the index fingers, proximal to the fingernail [Fig. 1(b)], these sites were inspected in order to ensure that no wounds were present. All subjects gave informed, written consent. Before the initial measurements, the subject was acclimatized to the room temperature (24– ) during . To achieve maximal support and stability, a vacuum pillow was positioned around the lower arm of the subject. Measurements were performed both during a normal (unprovoked) blood flow state and, thereafter, at an occluded, blood-emptied state. The latter was accomplished by rolling several elastic rubber rings in proximal direction, from the distal end of the fingertip. Measurements were also performed after applying nondrying immersion oil (Type A, Nikon, Japan) to the skin, in order to mimic tissue surfaces with high specular reflection, e.g., ulcers and skin burns. Initial shaker movement direction and finger (right/left) for oil measurements were randomly chosen for each subject and kept constant throughout the experiment. For normal blood flow, three measurements were carried out at each shaker frequency , both with and without the polarization filter. Initially a baseline recording without tissue motion was performed (I), thereafter a measurement during shaker movement (II), and finally a baseline measurement without shaker movement (III). Before changing shaker frequency and repeating measurements (I–III), there was a resting period of . At the blood-emptied state, measurements were performed sequentially, without interruption according to: a baseline measurement without tissue motion, measurements during shaker movement , and finally a baseline measurement without tissue motion. The total number of measurements for each subject was 112 (normal blood flow—blood-emptying, with and without polarization filter and immersion oil, horizontal and vertical movement direction, three shaker frequencies, movement and baseline measurements). 2.4.Data AnalysisAn averaged perfusion signal value [mean and standard deviation ] was calculated using the entire measurement sequence of . Further, an averaged perfusion signal extending over a complete sinusoidal movement period was formed. This was achieved by using the movement signal as a trigger and averaging the perfusion signal for all movement periods recorded during the 15-s measurement sequence. Thereby, it was possible to relate the perfusion signal to the velocity of the measurement object (Fig. 2 ). The velocity of the measurement object during a movement period was calculated by differentiation of the movement signal. Optical effects, caused by the insertion of the polarization filter in front of the detectors, changed the perfusion signal amplification. Therefore, a normalization factor was calculated for each combination of the measurement conditions (oil, flow, surface) for each subject and in the flow model. This was performed by using all baseline perfusion signals available, without shaker movement. All perfusion signals recorded using the polarization filter were normalized with the corresponding normalization factor. The averaged perfusion signal extending over a complete movement period was investigated at the highest shaker frequency , where the influence from tissue motion is at a maximum. The two maximum perfusion signal values during the reciprocating motion (one in each direction), without the polarization filter, were identified and averaged, resulting in one value for each measurement sequence of (Fig. 2). This value, together with the corresponding value recorded using the polarization filter calculated at identical tissue velocity, was used as an input to the statistical analysis. 2.4.1.Statistical analysisThe statistical analysis was performed according to full factorial ANOVA models, using v. 13.32 (Minitab, Inc., USA) and Statistica v. 6.0 (StatSoft, Inc., USA). Flow model data was analyzed using the factors polarization (with and without the polarization filter), movement direction (horizontal or vertical) and surface structure (R, S, and T), with tube flow (v1–v9) as a block. For skin measurements, the factors polarization, movement direction, and flow (normal or blood-emptied), with an additive random effect for subject (A–L), were used. Homogeneity of variances and normality was tested using the Brown-Forsythe modification of Levene’s test,21 considering the distances of the observations from their sample median rather than their sample mean, and the Ryan-Joiner test,22 respectively. Residuals were plotted against all factors, variables, and fitted values, and were visually studied to reveal any patterns and relations.23 Transformation of data was applied for the analysis of the flow model measurements, in order to fulfill ANOVA assumptions. A p-level was regarded as significant. Pairwise comparison of group means was performed using 95% confidence intervals (double-sided, distribution) with Bonferroni correction for multiple comparisons. Oil measurements performed on the skin were excluded from the statistical analysis because this factor was difficult to control; a large amount of specularly reflected light hit the detector in some measurements, but not in others, leading to large variances in the measurements. 3.Results3.1.Flow Model MeasurementsFlow model recordings showed significant main effects for all three factors and a significant three-factor interaction effect between polarization, movement direction, and surface structure. Measurements with the polarization filter resulted in a lower perfusion signal. This was independent of surface structure and movement direction. Examples of perfusion signals, with and without polarization filter, for the three surface structures during horizontal movement, are presented in Fig. 3 . The three surface structures gave rise to different perfusion signal levels without the polarization filter, in the horizontal movement direction. With the polarization filter, however, no significant difference was found (Fig. 3). In the vertical movement direction, no significant difference between any of the three surfaces was found. This was independent of the use of the polarization filter or not. Vertical movement resulted in lower perfusion signal levels, compared to horizontal. This was independent of polarization and surface structure. Perfusion signals for surface structure S, calculated as an average of the entire measurement period , are presented in Figs. 4(a) –4(d). It includes measurements with and without the polarization filter at both horizontal and vertical movement (0 to ) for different tube flow velocities. The figure illustrates the lower movement influence in the vertical direction, particularly evident without the polarization filter [Figs. 4(a) and 4(c)]. It also shows the effects of the polarization filter, reducing the movement influence, noticeable in the horizontal movement direction [Figs. 4(a)–4(b)]. 3.2.Skin MeasurementsSkin perfusion measurements showed significant main effects for all factors (polarization, movement direction, and flow) and a significant two-factor interaction effect between flow and movement direction. Measurements with the polarization filter resulted in a lower perfusion signal level compared to without , 95% confidence interval of the difference (a.u.): , because movement artifacts were reduced. This is exemplified in Fig. 5 , showing a skin measurement performed with and without the polarization filter, during horizontal movement and normal blood flow, at three different shaker frequencies. Vertical tissue movement resulted in a lower perfusion signal compared to horizontal, 95% confidence interval of the difference (a.u.): Blood-emptying resulted in a significantly lower signal compared to normal during vertical movement, 95% confidence interval of the difference (a.u.):
Application of the immersion oil caused a large amount of light being specularly reflected backward to the scanner head. During movement, the perfusion signal level was substantially reduced by the polarization filter in cases when the specular reflection hit the detector [Fig. 6(a) ], otherwise no effects were seen [Fig. 6(b)]. When specular reflection hit the detector, oil measurements gave rise to large movement artifacts compared to normal skin measurements. 4.DiscussionIn this study, a possibility to reduce movement-induced artifacts in the LDPI perfusion signal by blocking specularly reflected light, utilizing a polarization technique, has been investigated. Measurements on the skin and in a flow model were performed during well-controlled movement of the measurement objects. The polarized LDPI technique reduced the perfusion signal levels significantly. At a tissue velocity of , the perfusion signal was strongly influenced by motion not related to blood flow [Fig. 5(a) and Fig. 6(a)]. Therefore, the lower signal levels found when using the polarization filter could be explained by reduced movement artifacts. This was supported by the similar results from the flow model measurements. A possible explanation for the reduction of the movement artifacts is that photons reflected at the skin surface retain their original polarization,16, 17 thus being effectively blocked by the polarization filter. This prevents the formation of unwanted beat notes from appearing on the detector surface, because the Doppler-shifted photons carrying no blood flow-related information cannot mix with randomly polarized backscattered photons. Hence, no influence on the detected photocurrent Doppler signal is obtained. These results support the hypothesis that movement artifacts in the perfusion signal can be reduced by blocking specularly reflected light from being detected. Flow model measurements, without the polarization filter, confirmed previous findings of different magnitudes of movement-induced artifacts from varying surface structures.7, 24 Since the polarization filter blocks the specularly reflected light, the perfusion signal is expected to become unaffected by the surface structure and its reflecting properties. This was also confirmed in the flow model measurements. Immersion oil was applied to the skin in order to mimic a shiny, smooth surface structure, found at, e.g., skin burns, ulcers, and internal organs. During tissue motion, the glare generated by the specularly reflected light gave rise to large movement artifacts when the reflection hit the detectors [Fig. 6(a)]. In these cases, the light intensity at the tissue surface was large at the site of the incident laser beam, as seen from the detectors, caused by the specular reflection. Thus, the light spot generated by the reflected and backscattered light was considerably different compared to normal skin. The number of speckle areas on the detector is dependent on the spot size;2, 25 smaller spot size means fewer speckle areas and larger amplification of the perfusion signal, possibly explaining the large signals registered. Similar results were found in the flow model measurements during horizontal movement, where the movement influence on the perfusion signals was largest for the smooth surface structure, which gave rise to a large amount of specular reflection impinging on the detectors. Applying the polarization filter in front of the detectors, however, resulted in markedly reduced movement artifacts [Fig. 3(b) and Fig. 6(a)]. Even without tissue motion, the surface structure can influence the LDPI perfusion estimate due to the amplification effect of specular reflection. Hence, different readings for the same tissue blood perfusion could be recorded. Further, specular reflection has been reported to give rise to overestimated perfusion values during measurements on burn scars, caused by saturated total backscattered light intensity (dc signal) levels.26 These false perfusion readings can be avoided by using polarized LDPI, an additional advantage besides reducing movement artifacts, thus increasing system usability and making LDPI more useful in the clinical setting. However, by using the polarization filter not only specularly reflected light is blocked from being detected, but also approximately 50% of the randomly polarized, multiply scattered light. This results in a decreased signal-to-noise ratio if laser light intensity is not increased. The use of polarized light and a perpendicular analyzing polarization filter also affects the amplitude and interpretation of the perfusion signal, due to optical effects. Besides a changed tissue spot size, as seen from the detectors, the relation between homodyne and heterodyne contribution19 is affected, since static specularly reflected and weakly scattered light are blocked from being detected.16, 17 Additionally, polarization may also affect the sampling volume.13, 14, 15, 27 Hence, perfusion readings from an ordinary LDPI system cannot be directly compared to readings from a polarized LDPI system. In this study, a normalization factor, dependent on the measurement object and measurement conditions, was used to compensate for these effects. Horizontal and vertical movement is expected to influence the LDPI signal differently.7 The size of the Doppler shift generated by a moving object equals , where is the scattering object velocity and is the scattering vector, the latter defined as the difference between the incident and the scattered light wave direction.28 Assuming tissue movement at a single velocity in the vertical direction, when mixing of the light occurs at the surface of the square-law detector, no reference beam (unshifted light) is present. Rather, all light mixed at the detector has been Doppler shifted by the moving tissue, which means shifting all light slightly in frequency. This effect is cancelled out by the square-law detector, which explains why the small tube flow velocity variations can be detected despite the large shaker movement, resulting in a linear response for lower tube flow velocities (Fig. 4). If a static beam reference is introduced at the detector in this case, large movement artifacts would be registered. A possible explanation to the different influence of horizontal and vertical movement is illustrated in Fig. 7 . Differences in the direction of incident light and backscattered light , impinging on the detector, define a -distribution, which gives rise to a detected Doppler broadening. The -distribution depends on, e.g., tissue spot size, detector size, and the tissue-detector distance. Therefore, in a typical LDPI setup most scattering vectors are expected to have approximately the same direction as the incident light. During vertical movement , the detected Doppler broadening becomes smaller compared to during horizontal movement , due to smaller variations of the projection of the different on the corresponding [Fig. 7(b)]. Additionally, variations in surface structure at the site of the impinging laser light will give rise to a larger Doppler broadening during horizontal movement, making the difference even larger. The perfusion signal in the flow model measurements was significantly lower during vertical, compared to horizontal, movement in all cases. For skin measurements, however, this result holds true only in the blood-emptied case. One explanation to this discrepancy may be that the tissue of the finger is compressible, unlike the piece. Therefore, in our movement model, the compression of the tissue may cause redistribution of blood cells influencing the blood flow at the measurement site. 5.ConclusionsThis study has shown the possibility to reduce movement-induced artifacts in the perfusion signal by blocking specularly reflected light, utilizing a polarization technique. Polarized LDPI is particularly appropriate when applied to shiny tissues, such as internal organs, ulcers, and skin burns. AcknowledgmentsThis study was supported by the Competence Centre for Non-invasive Medical Measurements (NIMED). The authors would like to thank Eva Enqvist, Senior Lecturer, Department of Mathematics, Linköping University, for invaluable help and fruitful discussions regarding the statistical analysis and Håkan Rohman, Department of Biomedical Engineering, Linköping University, for skilful help when building the flow model. ReferencesT. J. Essex and
P. O. Byrne,
“A laser Doppler scanner for imaging blood flow in skin,”
J. Biomed. Eng., 13
(3), 189
–194
(1991). 0141-5425 Google Scholar
K. Wårdell,
A. Jakobsson, and
G. E. Nilsson,
“Laser Doppler perfusion imaging by dynamic light scattering,”
IEEE Trans. Biomed. Eng., 40
(4), 309
–316
(1993). https://doi.org/10.1109/10.222322 0018-9294 Google Scholar
A. Fullerton,
B. Rode, and
J. Serup,
“Skin irritation typing and grading based on laser Doppler perfusion imaging,”
Skin Res. Technol., 8
(1), 23
–31
(2002). 0909-752X Google Scholar
A. M. Enejder,
C. af Klinteberg,
I. Wang,
S. Andersson-Engels, N. Bendsoe,
S. Svanberg, and
K. Svanberg,
“Blood perfusion studies on basal cell carcinomas in conjunction with photodynamic therapy and cryotherapy employing laser-Doppler perfusion imaging,”
Acta Derm Venereol, 80
(1), 19
–23
(2000). 0001-5555 Google Scholar
F. W. Kloppenberg,
G. I. Beerthuizen, and
H. J. ten Duis,
“Perfusion of burn wounds assessed by laser Doppler imaging is related to burn depth and healing time,”
Burns, 27
(4), 359
–363
(2001). 0305-4179 Google Scholar
M. Larsson,
W. Steenbergen, and
T. Strömberg,
“Influence of optical properties and fiber separation on laser Doppler flowmetry,”
J. Biomed. Opt., 7
(2), 236
–243
(2002). https://doi.org/10.1117/1.1463049 1083-3668 Google Scholar
M. G.D. Karlsson,
M. Larsson,
T. Strömberg, and
K. Wårdell,
“Influence of tissue movements on laser Doppler perfusion imaging,”
Proc. SPIE, 4624 106
–114
(2002). 0277-786X Google Scholar
S. L. Jacques,
J. R. Roman, and
K. Lee,
“Imaging superficial tissues with polarized light,”
Lasers Surg. Med., 26
(2), 119
–129
(2000). https://doi.org/10.1002/(SICI)1096-9101(2000)26:2<119::AID-LSM3>3.0.CO;2-Y 0196-8092 Google Scholar
S. L. Jacques,
J. C. Ramella-Roman, and
K. Lee,
“Imaging skin pathology with polarized light,”
J. Biomed. Opt., 7
(3), 329
–340
(2002). https://doi.org/10.1117/1.1484498 1083-3668 Google Scholar
A. N. Yaroslavsky,
V. Neel, and
R. R. Anderson,
“Demarcation of nonmelanoma skin cancer margins in thick excisions using multispectral polarized light imaging,”
J. Invest. Dermatol., 121
(2), 259
–266
(2003). https://doi.org/10.1046/j.1523-1747.2003.12372.x 0022-202X Google Scholar
A. N. Yaroslavsky,
V. Neel, and
R. R. Anderson,
“Fluorescence polarization imaging for delineating nonmelanoma skin cancers,”
Opt. Lett., 29
(17), 2010
–2012
(2004). https://doi.org/10.1364/OL.29.002010 0146-9592 Google Scholar
F. C. MacKintosh,
J. X. Zhu,
D. J. Pine, and
D. A. Weitz,
“Polarization memory of multiply scattered light,”
Phys. Rev. B, 40
(13), 9342
–9345
(1989). https://doi.org/10.1103/PhysRevB.40.9342 0163-1829 Google Scholar
J. M. Schmitt,
A. H. Gandjbakhche, and
R. F. Bonner,
“Use of polarized light to discriminate short-path photons in a multiply scattering medium,”
Appl. Opt., 31
(30), 6535
–6546
(1992). 0003-6935 Google Scholar
V. Sankaran,
K. Schönenberger,
J. T. Walsh Jr., and
D. J. Maitland,
“Polarization discrimination of coherently propagating light in turbid media,”
Appl. Opt., 38
(19), 4252
–4261
(1999). 0003-6935 Google Scholar
V. Sankaran,
J. T. Walsh Jr., and
D. J. Maitland,
“Comparative study of polarized light propagation in biologic tissues,”
J. Biomed. Opt., 7
(3), 300
–306
(2002). https://doi.org/10.1117/1.1483318 1083-3668 Google Scholar
R. R. Anderson,
“Polarized light examination and photography of the skin,”
Arch. Dermatol., 127
(7), 1000
–1005
(1991). https://doi.org/10.1001/archderm.127.7.1000 0003-987X Google Scholar
S. P. Morgan and
I. M. Stockford,
“Surface-reflection elimination in polarization imaging of superficial tissue,”
Opt. Lett., 28
(2), 114
–116
(2003). 0146-9592 Google Scholar
G. E. Nilsson,
A. Jakobsson, and
K. Wårdell,
“Tissue perfusion monitoring and imaging by coherent light scattering,”
Proc. SPIE, 1524 90
–109
(1991). 0277-786X Google Scholar
G. E. Nilsson,
T. Tenland, and
P. Å. Öberg,
“A new instrument for continuous measurement of tissue blood flow by light beating spectroscopy,”
IEEE Trans. Biomed. Eng., 27
(1), 12
–19
(1980). 0018-9294 Google Scholar
G. E. Nilsson,
T. Tenland, and
P. Å. Öberg,
“Evaluation of a laser Doppler flowmeter for measurement of tissue blood flow,”
IEEE Trans. Biomed. Eng., 27
(10), 597
–604
(1980). 0018-9294 Google Scholar
M. B. Brown and
A. B. Forsythe,
“Robust tests for the equality of variances,”
J. Am. Stat. Assoc., 69
(346), 364
–367
(1974). 0162-1459 Google Scholar
J. T.A. Ryan and
B. L. Joiner,
“Normal probability plots and tests for normality,”
(1976) Google Scholar
D. C. Montgomery, Design and Analysis of Experiments, 4th ed.John Wiley & Sons, New York (1997). Google Scholar
M. Arildsson,
G. E. Nilsson, and
K. Wårdell,
“Critical design parameters in laser Doppler perfusion imaging,”
Proc. SPIE, 2678 401
–408
(1996). 0277-786X Google Scholar
A. Serov,
W. Steenbergen, and
F. de Mul,
“Dynamic speckle patterns formed by a mixture of moving and stationary scatterers: theory and experiment,”
Proc. SPIE, 3915 158
–169
(2000). 0277-786X Google Scholar
R. Bray,
K. Forrester,
C. Leonard,
R. McArthur,
J. Tulip, R. Lindsay,
“Laser Doppler imaging of burn scars: a comparison of wavelength and scanning methods,”
Burns, 29
(3), 199
–206
(2003). 0305-4179 Google Scholar
I. M. Stockford,
S. P. Morgan,
P. C. Chang, and
J. G. Walker,
“Analysis of the spatial distribution of polarized light backscattered from layered scattering media,”
J. Biomed. Opt., 7
(3), 313
–320
(2002). https://doi.org/10.1117/1.1483316 1083-3668 Google Scholar
G. E. Nilsson,
E. G. Salerud,
N. O.T. Strömberg, and
K. Wårdell,
“Laser Doppler perfusion monitoring and imaging,”
Biomedical Photonics Handbook, 15:1
–24 CRC Press, Boca Raton, FL (2003). Google Scholar
|