1 January 2006 Measurement of dye diffusion in scattering tissue phantoms using dual-wavelength low-coherence interferometry
Trude Støren, Arne Røyset, Lars Othar Svaasand, Tore Lindmo
Author Affiliations +
Abstract
We demonstrate low-coherence interferometry (LCI) for dye diffusion measurements in scattering tissue phantoms. The diffusion coefficient of a phthalocyanine dye in 1.5% agar gel containing scattering Intralipid was measured using a dual-wavelength interfero-meter. One wavelength was matched to the absorption peak of the dye at 675 nm. The other, 805 nm, was not affected by the dye, and was used to correct for varying sample scattering as a function of depth, assuming a constant ratio between scattering at the two wavelengths. The same wavelength dependence of scattering is assumed for the entire sample, but no a priori knowledge about the amount of scattering is needed. The dye diffusion coefficient was estimated by fitting a mathematical model of the interferometer signal to the measured LCI envelope. We compare results obtained using both a constant-scattering and a depth-resolved-scattering approach to determine the sample scattering. The presented method provides robust estimation of the diffusion coefficient when spatial resolution in determining the depth-resolved scattering is varied. Results indicate that the method is valid for samples having continuous spatial variations in the scattering coefficient over lengths as short as the coherence length of the probing light. The method allows in situ characterization of diffusion in scattering media.

1.

Introduction

Since its introduction1 optical coherence tomography (OCT) has become a well-established technique for obtaining high-resolution structural images of biological and other semitransparent tissue. The introduction of ultrabroadband spatially coherent sources2, 3 and full-field OCT instruments using affordable spatially incoherent white-light sources4 has brought the resolution toward that obtained on histological sections. In recent years several groups have worked on expanding OCT to functional imaging. Promising results have been reported for Doppler flow imaging,5 polarization sensitive OCT,6 and spectroscopic OCT.7, 8, 9 Several authors have addressed the challenge of extracting quantitative information on optical properties from scattering samples using low-coherence interferometry (LCI) and OCT. To obtain high precision in the estimated optical properties, knowledge is required about the relationship between optical properties of a scattering sample and the signal measured by LCI,10, 11, 12, 13, 14, 15 and about how speckle noise affects imaging of optical properties.9, 16, 17

The motivation for our work is to be able to monitor the concentration of photosensitizers used in photodynamic therapy (PDT). Apart from in vivo fluorescence monitoring, the photosensitizer part of in vivo PDT dosimetry is still in its infancy.18 The requirement for in situ, noninvasive monitoring of drug concentration rules out methods that measure diffusion as a steady-state molecular mobility, and methods based on special measurement or sample geometries.19 OCT represents a promising method to meet this challenge. Hand-held or endoscope-based OCT systems can access a variety of sites within the body and provide high-resolution cross-sectional images from which it is possible to determine the optical properties and thus the photosensitizer concentration and the effect of photosensitizer diffusion.

We have previously presented results from a first step toward the goal of noninvasive concentration monitoring by OCT, by measuring diffusion of a PDT-related dye in 1.5% agar gel.20 A model of the OCT signal as a function of sample depth and time in the presence of a diffusing dye was then fitted to experimental data in order to obtain an estimate of the diffusion coefficient. The model assumed homogeneous samples having a constant scattering coefficient. The scattering coefficient of 1.5% agar gel was measured using OCT prior to the diffusion measurements, and found to be 0.06mm1 and thus one to two orders of magnitude below realistic values for the scattering coefficient of tissue. In the present work we further develop the method presented in Ref. 20 toward application on realistic tissue samples by performing measurements on agar samples having different amounts of Intralipid (IL) added to increase scattering. Measurements are done simultaneously at two wavelengths. One wavelength matches the absorption peak of the dye, and the other, which is unaffected by the dye, is used to determine the scattering coefficient of the samples, using a simple but realistic model for the relationship between scattering at the two wavelengths. An estimate of the diffusion coefficient is obtained using a more general version of the parameter-fitting approach presented in Ref. 20. The model of the LCI signal used in the present paper is extended to handle slow variations in the scattering coefficient with depth whereas the model used in Ref. 20 required a constant scattering coefficient known prior to the diffusion measurements. While presently only tested on samples having homogeneous scattering properties, this method should be suitable for diffusion measurements on real tissue where the scattering coefficient is generally a function of position.

Reported values for the scattering coefficient of biological tissue are in the range10 110mm1 . We present results on samples having scattering coefficients up to 3mm1 thus lying within the range of realistic values for tissue scattering coefficients.

2.

Model

The previous letter20 gives a description of the model used to determine the diffusion coefficient of a dye which diffuses into a gel. We showed that the logarithm of the interference signal envelope, Aλ(z,t) recorded at wavelength λ , can be expressed as a function of the depth- (z) and time-resolved (t) attenuation coefficient of the gel-dye sample, μtλ(z,t) :

Eq. 1

ln[Aλ(z,t)]=ln(rλ(z)Gλ)20z12μtλ(z,t)dz=ln(rλ(z)Gλ)0z[μs,gelλ(z)+εaλC(z,t)]dz.
The envelope, Aλ(z,t) , is proportional to the field of the light back-scattered from geometrical depth z in the sample.21 Since the field is the square root of the intensity, the field attenuation coefficient is 12 the intensity-based attenuation coefficient μt while the factor 2 arises from double-pass of the light which probes the sample. Gλ is a constant which depends on the intensity of the incident light in the interfero-meter and gain factors in the detection system, rλ(z) is the average field reflectivity of the sample, μt,gel,λ(z) is the total attenuation coefficient of the Intralipid-containing agar gel, εaλ is the extinction coefficient of the absorbing dye, and C(z,t) is the depth- and time-dependent dye concentration. In Eq. 1 we assume that the Intralipid-containing agar gel has negligible absorption compared to scattering, and that the dye has negligible scattering compared to absorption. The scattering coefficient of a sample is related to its field reflectivity through the relation22

Eq. 2

μs,λ(z)rλ(z)2.
When μs,λ(z) and thus rλ(z) is constant or slowly varying as a function of depth, the first term in Eq. 1 can be approximated by a constant. This approximation is used throughout the present work, and we define Kλln(rλ(z)Gλ) .

For the case where dye is distributed evenly on the surface of the gel at time t=0 , over an area much larger than the diameter of the probing light beam, a good approximation for the dye concentration as a function of depth and time, is given by the one-dimensional delta-source solution of the diffusion equation for a semi-infinite medium:23

Eq. 3

C(z,t)=MSπDtexp[z24Dt]
where M is the total mass of dye deposited on the top surface area S of the gel.

Equations 1, 3 constitute the model which we use for the interferometer signal. Measurements are carried out at two wavelengths where one (λ1) is absorbed by the dye, and the other (λ2) is not affected by the dye. An estimate of the diffusion coefficient, D , is determined by fitting of the model of Eqs. 1, 3 to experimental data obtained at λ1 using D , M , and Kλ as fitting parameters. In order to carry out the parameter fitting an estimate of the scattering coefficient at λ1 is required. In the present work we use two approaches to estimating the scattering coefficient. The first approach assumes a constant scattering coefficient, while the second approach can be used for samples having a scattering coefficient which varies slowly as a function of sample position. Both methods are based on a single-scattering model of the LCI signal first used for estimating optical properties by Schmitt and Knüttel.10 For samples where the LCI signal also is influenced by multiple scattering, the simple single-scattering model must be replaced by more advanced models taking the effect of multiple scattering into account.21, 24, 25

When λ2 is unaffected by the dye, Eq. 1 is reduced to

Eq. 4

ln[Aλ2(z,t)]=Kλ20zμs,gel,λ2(z)dz,
which in the case of constant scattering is further simplified to

Eq. 5

ln[Aλ2(z)]=Kλ2μs,gel,λ2z.
In this case linear regression on the logarithm of the recorded envelope can be used to find an estimate of the scattering coefficient of the gel, μs,gel,λ2 .

In tissue, the scattering coefficient is generally a function of position, and more complicated procedures must be used to estimate the scattering properties. In the present work we limit the study to samples having a slowly varying depth-dependent scattering. In this case, one approach to obtaining a depth-resolved estimate of the scattering coefficient is by spatially filtering and differentiating the logarithm of the recorded envelope:9

Eq. 6

μ̂s,gel,λ2(z)=ddzln[Aλ2(z)hz(z)].
Spatial filtering is necessary in order to reduce the large uncertainty in a depth-resolved estimate of the attenuation coefficient due to inherent speckle noise in an LCI signal. Following the approach of Ref. 9, we carry out the filtering by convolving the envelope with a filter function hz(z) . The full width at half maximum of the filter function is denoted Lz and determines the spatial resolution of the depth-resolved attenuation-coefficient estimator. The symbol ⊗ denotes the convolution operation and the hat over the scattering coefficient indicates that it is an estimated value. Throughout this paper a Gaussian function truncated at ±2σ , where σ is the standard deviation of the Gaussian function, is used for the filter function. The Gaussian is a common filter function due to its lack of side lobes both in the position and spatial-frequency planes. Truncating it introduces some side lobes in the frequency plane but provides strict control over which parts of the filtered data that are influenced by edge effects. The properties and limitations of the attenuation coefficient estimator in Eq. 6, including its variance, is thoroughly discussed in Ref. 9.

Once the depth-resolved scattering coefficient at λ2 is determined, the scattering coefficient at λ1 can be estimated. In this paper we use a simple but realistic model of the scattering properties, assuming that there is a constant relationship between the scattering coefficient at the two wavelengths independent of Intralipid concentration and the presence of dye in the gel. Following a standard approach used by several authors,26, 27 we assume that there is a constant ratio between the scattering coefficient at the two wavelengths. We denote the ratio between the two scattering coefficients F :

Eq. 7

F=μs,λ1μs,λ2.
For samples having a more complicated relation between the scattering coefficient at the two wavelengths, more sophisticated approaches to separating the effects of scattering and absorption is required. Recently Yang 28 proposed a novel method based on spectral triangulation for determining the absorption properties in a scattering sample without any a priori knowledge of wavelength dependence of the scattering properties. This is achieved by using a third wavelength, and assuming a linear, but unknown, relationship between scattering at the three wavelengths. Another novel method for separating absorption from scattering was presented by Xu 15 They analyzed the spectrum of an OCT signal recorded using a broad-bandwidth source as a function of depth and separated the spectrum change due to absorption and scattering using a least squares fitting algorithm.

The model for the LCI signal from a sample containing a diffusing dye described above assumes that the diffusion coefficient does not vary as a function of position in the sample. It should be possible to extend the model in order to include variations in the diffusion coefficient as a function of the sample scattering, but this is beyond the scope of the present study and requires more knowledge about dye diffusion in agar gel and the effect of Intralipid on the diffusion properties.

3.

Experimental Setup

For the measurements, we use a bulk Michelson interfero-meter, as shown in Fig. 1 . Light from two pigtailed superluminescent diodes from Superlum Diodes is multiplexed in a fiber coupler before it is launched into the free-space interferometer through a collimating lens. According to data from the manufacturer, the diodes have center wavelengths 675 and 805nm , and spectral FWHMs 10 and 18nm , respectively. We have measured the coherence lengths in air to be lc=18.1μm for the 675-nm source and lc=14.2μm for the 805-nm source. The coherence lengths in a sample are found by dividing the coherence lengths in air by the group refractive index of the sample. These coherence lengths correspond to spectral FWHM 11 and 20nm for the 675- and 805-nm source, respectively, assuming Gaussian spectra. The power at the interferometer input was measured to 0.4 and 0.8mW for the 675- and 805-nm source, respectively. The light is focused into the sample by a focusing lens having focal length f=14.5mm (Melles Griot 06GLC003). An identical lens is used in front of the reference mirror in order to match dispersion in the two interferometer arms. We use focus tracking29 to ensure that the focus of the probing beam overlaps with the coherence volume of the interferometer. The focusing lens is scanned along with the reference mirror using lens velocity vl=vrng2 , where vr is the velocity of the reference mirror and ng is the group refractive index of the sample. Linear translation stages are used for scanning of the focusing lens and reference mirror. Between each A-scan the sample was displaced transversally using a third linear translation stage. For all the experiments presented, the reference-mirror scanning velocity was vr=1mms and the transversal distance between each A-scan was 10μm . Demultiplexing is carried out using a ruled gold grating, and the envelope of the interferometer signal is recorded simultaneously at both wavelengths. The system has a dynamic range of 90dB .

Fig. 1

Schematic illustration of the wavelength-multiplexed interfero-meter used for diffusion measurements. Light from two pigtailed SLDs having center wavelengths of λ1=675nm and λ2=805nm is multiplexed in a fiber coupler. Light from one of the fibers from the fiber coupler is launched into the free-space Michelson interferometer through a collimating lens. Depth scanning is obtained by adjusting the position (zr) of the reference mirror using a linear translation stage. Dynamic focus tracking is obtained by scanning the position (zl) of the sample-arm focusing lens using another linear translation stage (see text). A ruled gold grating is used for wavelength demultiplexing, and light from the two sources is recorded by separate detectors.

014017_1_002601jbo1.jpg

Fig. 6

Same as Fig. 4 but using filter length Lz=2μm when estimating the scattering coefficient. The black curves are experimental data recorded at 675nm while the dark-gray curves represent the fitted model. The statistical variations in the fitted model arise from fluctuations in the depth-resolved estimated scattering coefficient obtained from the data set recorded at 805nm . For clarity the experimental data recorded at 805nm have, however, been omitted from the figure.

014017_1_002601jbo6.jpg

Table 1

Overview of the measurement results.

CIL [%] n μ̂s : methodof estimation Lz [μm] ⟨D⟩ [10−10m2∕s] σD [10−10m2∕s] 95% CI [10−10m2∕s] MSE[dB]
0.0010Støren 202.50.30 (2.5±0.20)
0.1519CS12.10.30 (2.1±0.15) 1.4
0.308CS3.10.98 (3.1±0.80) 1.8
0.1519FOD21002.80.57 (2.8±0.27) 1.4
0.1519FOD202.70.51 (2.7±0.24) 1.9
0.1519FOD23.00.62 (3.0±0.30) 3.2
0.308FOD1003.81.0 (3.8±0.84) 1.7
0.308FOD203.71.3 (3.7±1.04) 2.5
0.308FOD23.81.5 (3.8±1.20) 3.5

1 μ̂s determined using the constant-scattering approach.

2 μ̂s determined using the function-of-depth approach.

4.

Tissue Phantoms

We have studied diffusion of aluminum phthalocyanine tetrasulfonate chloride dye (AlPcS834, Porphyrin Products, Inc.) in 1.5% agar gel containing various amounts of scattering Intralipid. The dye has molecular weight MW=895.19Daltons . It has high solubility in water and is thus not expected to be present in the IL phase of the samples. Using a commercial absorption spectrometer (Agilent 8452 UV-Visible Spectrophotometer), we measured the dye extinction coefficient at 675nm to be εa,675=44mlmgmm , about three orders of magnitude larger than at 805nm . Thus with λ1=675nm and λ2=805nm , the equations in Sec. 2 are valid.

Prior to the diffusion measurements, initial LCI measurements were carried out at both wavelengths on agar gels having different concentrations of Intralipid and no absorbing dye. The scattering coefficient of the agar-IL samples was determined using linear regression on the logarithm of the recorded interferometer envelope. Based on these initial measurements the value of F , defined in Eq. 7, was found to be 1.5±0.05 over a wide range of Intralipid concentrations.

5.

Experimental Method

Agar-gel samples having Intralipid concentrations CIL=0.15% and CIL=0.30% were prepared in 10-mm -deep cylindrical cuvettes having inner diameter 18mm . In this first demonstration of diffusion measurements on scattering samples, we wanted to limit the variation in the scattering coefficient and chose to prepare samples having only two different predefined Intralipid concentrations. A concentration of CIL=0.30% in agar gel resulted in a scattering coefficient of 3mm1 , and we found this to be the highest practical scattering coefficient for the given dynamic range of our system.

For convenience we prepared a stock solution of dissolved dye having a concentration of 4mgml . Using a micro pipette 20μl of the stock solution was deposited onto the gel, distributing the solution as evenly as possible. Details on the choice of amount of deposited dye is given below. For each agar-IL sample, the dye was deposited 30min after sample preparation. After depositing the dye, the cuvette was covered with a glass slide, and placed in the sample arm of the interferometer. The LCI envelope was recorded as a function of depth and time at both wavelengths. For each sample, measurement started immediately after deposition of the dye, acquiring a series of 40 A-scans over a time period of about 2.5min . The sample was displaced 10μm transversally between each A-scan in order to obtain A-scans having decorrelated speckles, and thus enable speckle averaging. To study the time development due to diffusion, data series at the same transversal positions on the sample were also acquired at three later time points. Thus, a total of four data series were recorded during 30min after dye-deposition for each of the agar samples. The depth information was converted from optical depth to geometrical depth, z , using n=1.34 for the refractive index.

At 675nm back-scattering just underneath the surface of the Intralipid-containing agar gel gives an LCI signal about 30dB above the instrument noise level. The signal decreases due to scattering and absorption as we probe into the sample. For the model in Eqs. 1, 3 to be valid, care must be taken that the signal is well above the noise level of the instrument throughout the probing depth. The amount of dye deposited on the sample surface and the amount of scattering thus limits the maximum probing depth. When designing the experiments, we determined the approximate maximum probing depth for each level of sample scattering by finding the depth where signal reduction due to scattering alone was approximately 1015dB compared to the signal level at the sample surface. By trial we then determined the amount of deposited dye that would reduce the signal further due to absorption to about 5dB above the noise level at the maximum probing depth. Following this procedure signal reductions due to scattering and absorption were about equal.

6.

Experimental Results and Discussion

Depth- and time-resolved data sets were recorded from diffusion measurements using 19 samples having an Intralipid concentration of CIL=0.15% , and 8 samples having CIL=0.30% following the procedure outlined above. The diffusion measurements are time consuming, and the series of measurements on 0.30%-IL samples was terminated after 8 parallel samples were measured. The difference in number of samples is taken into account in the statistical analysis of the experimental results.

Figure 2 shows gray-scale images of the logarithm of the LCI envelope from one of the CIL=0.15% -samples recorded at both wavelengths and plotted as a function of geometrical depth and time. In the plot t and z correspond to t and z in Eq. 3. The black areas in the images correspond to time periods in which no data was acquired. In the image recorded at 805nm the average envelope is constant as a function of time whereas the image recorded at 675nm shows a time development: Shortly after dye deposition the backscattered intensity falls rapidly as a function of imaging depth due to a high dye concentration near the glass-gel interface. Later, diffusion of the dye into the gel results in lower concentration levels and a slower decrease in backscattered intensity with depth.

Fig. 2

Gray-scale images showing the logarithm of the interferometer envelope of four series of A-scans recorded from the same 0.4×1.0mm vertical section of an Agar sample having Intralipid concentration CIL=0.15% . The abscissa represents the time t after dye deposition, and the ordinate axis is the geometrical depth in the sample, where z=0 is the glass-gel interface. Panels (a) and (b) show data recorded at wavelengths 675 and 805nm , respectively. The black areas in the plots represent time periods where no data is acquired.

014017_1_002601jbo2.jpg

Data in each of the four series were averaged transversally, resulting in depth-resolved speckle-averaged envelopes recorded at four different times for each of the two wavelengths. In order to obtain an estimate of the diffusion coefficient, the model in Eqs. 1, 3 was fitted to the logarithm of the speckle-averaged LCI envelope at 675nm , with D , M , and K as fitting parameters, using a least-squares algorithm. The data recorded at 805nm was used to find an estimate of the sample scattering coefficient prior to fitting.

As described in Sec. 2, we used two approaches for estimating the scattering coefficient, and parameter fitting was carried out using both approaches for all recorded data sets. For the first approach we assumed constant scattering as a function of depth, and used linear regression on the logarithm of the averaged envelope recorded at 805nm , to obtain an estimate of the gel-IL scattering coefficient, μs,805,gel , as described in relation to Eq. 5. Using Eq. 7 we thus found an estimate of μs,675,gel , the scattering coefficient at 675nm , and used this value as a constant parameter in Eq. 1. We denote this the constant-scattering approach. In the second approach, which we denote the function-of-depth approach, we allow depth-dependent scattering and use Eqs. 6, 7 to find a depth-resolved estimate of the scattering coefficient at 675nm which then is used in Eq. 1. Since the samples in the present work are relatively homogeneous, we expect similar values for the diffusion coefficient for the two approaches.

6.1.

Assuming Constant Scattering

Figure 3 shows speckle-averaged LCI envelopes from two typical data sets recorded at both wavelengths on samples prepared using the two different predefined Intralipid concentrations. The constant-scattering approach to estimating the scattering coefficient was used, and a value for the scattering coefficient at 805nm was obtained from the slope of the regression line of the logarithm of the envelope recorded at 805nm . Both the regression lines and the logarithm of the 805-nm envelope are plotted in the figure. For each sample we used the average of the scattering coefficient obtained from all four data series as an estimate for μs,gel,805 . Model fitting was carried out on the data set recorded at 675nm using the constant value for μs,gel,675 , obtained from μs,gel,805 for each sample. Figure 3 shows the fitted model for the LCI signal at 675nm as well as experimental data. For both IL concentrations there is good agreement between experimental data and the fitted model.

Fig. 3

Typical experimental data plotted together with fitted model based on estimating the scattering coefficient as a constant, using the constant-scattering approach. Speckle-averaged LCI envelopes are plotted as a function of depth for each of the four time series (t1t4) obtained from two samples having Intralipid concentration CIL=0.15% [panel (a)] and CIL=0.30% [panel (b)]. The jagged curves are experimental data recorded at 675 and 805nm . To avoid influence from the noise level (at 90dB ) and the reflection from the glass-gel interface (at z=0 ), data in a limited depth interval is used for parameter fitting, as described in Sec. 5. The experimental data within this interval is plotted in black and dark gray for 675nm and 805nm , respectively, while the experimental data outside this interval is plotted using light gray. The mean time after dye deposition for each of the curves is t1 (bottom 675-nm curve) to t4 (top 675-nm curve). The smooth black lines show the fitted model of Eqs. 1, 3 for 675nm . For the model plotted in this figure, the scattering coefficient at 675nm was assumed constant, and was determined from the envelope recorded at 805nm using the constant-scattering approach. For 805nm the linear regression lines are plotted in dark-grey together with the experimental data.

014017_1_002601jbo3.jpg

Results from n=19 samples having CIL=0.15% and n=8 samples having CIL=0.30% are summarized in Table 1 . Also presented in the table are 95% confidence intervals for the diffusion coefficient based on results from the two series. For the constant-scattering approach we obtain 95% confidence intervals of D=(2.1±0.15)×1010m2s and D=(3.1±0.80)×1010m2s for samples having CIL=0.15% and CIL=0.30% , respectively. These two mean values are statistically different down to a 1.8% significance level. Combining the results from the two series of sample IL concentrations we obtain D=(2.6±0.51)×1010m2s as the mean value and standard deviation of the diffusion coefficient. At a 5% significance level this result is not statistically different from the result presented in our previous letter20 obtained from measurements on agar without Intralipid, and thus having very low scattering.

6.2.

Allowing Depth-dependent Scattering

As a first step toward using the method presented in this paper for estimating the diffusion coefficient of a dye diffusing in a sample having unknown, and possibly spatially varying, scattering, we used the function-of-depth approach described in Sec. 5 on the relatively homogeneous agar-IL samples. Prior to the diffusion measurements we suspected that there could be local variations in the scattering coefficient of the samples due to a nonperfect distribution of Intralipid in the agar gel. The objective of this analysis was to test the feasibility of the function-of-depth approach to estimating the scattering properties, thus allowing local variations in scattering properties.

Figures 4 and 5 show the same experimental data as Fig. 3. In Fig. 6 the experimental data at 805nm is omitted from the figure for clarity. All three figures show the fitted model for the LCI signal at 675nm where μs,gel,675(z) is determined from the depth-resolved estimate of μs,gel,805(z) obtained using Eq. 6. In Figs. 4, 5, 6 filter lengths of Lz=100 , 20, and 2μm , respectively, are used in estimating the depth-resolved scattering coefficients. For each of the speckle-averaged envelopes at 675nm the depth-resolved scattering coefficient obtained from the speckle-averaged depth scan at the corresponding time point recorded at 805nm was used in Eq. 1.

Fig. 4

Same experimental data as Fig. 3, but using the function-of-depth approach to estimate the scattering coefficient, which is allowed to depend on the position in the sample. Using Eqs. 6, 7, a depth-resolved estimate for the scattering coefficient at 675nm is obtained from the envelope recorded at 805nm . A filter length of Lz=100μm is used in the attenuation-coefficient estimator. For the transversally averaged A-scans at each time point, μs,gel,675(z) obtained from the corresponding averaged A-scans at 805nm is then used in the model of Eqs. 1, 3 for the LCI envelope at 675nm . The smooth black lines show the fitted model plotted together with experimental data used in the parameter fitting procedure. Dashed vertical lines indicate the depth interval used for estimating the scattering coefficient. The part of the experimental data plotted using black and dark-gray lines indicate the part of the experimental data used for parameter fitting. The parameter-fitting interval is shorter than the estimation interval to avoid edge effects arising from the filtering described in Eq. 6. For clarity the experimental data and fitted model for the time t3 are omitted from the figure, but data series recorded at all four time points were used in parameter fitting throughout the paper.

014017_1_002601jbo4.jpg

Fig. 5

Same as Fig. 4 but using filter length Lz=20μm when estimating the scattering coefficient.

014017_1_002601jbo5.jpg

For Lz=100μm (Fig. 4) there is a good fit between the model and experimental data at 675nm . The model follows some of the slow variations in the experimental data as a function of depth. This indicates that the agar-IL samples may have local variations in the scattering coefficient and that the function-of-depth approach is able to recognize these. It also indicates that the model of Eqs. 1, 3 tolerates slow variations in the scattering coefficient of the samples under study. Average numerical results obtained from parameter fitting for all samples are summarized in Table 1. The average mean square errors (MSE) per sample point for all the samples, analyzed with Lz=100μm , are MSE015=1.4dB and MSE030=1.7dB for samples having CIL=0.15 and CIL=0.30% , respectively. These are about the same values as those obtained using the constant-scattering approach.

Reducing the filter width to 20μm (Fig. 5) and 2μm (Fig. 6) results in a fitted model that more and more resembles experimental data. Some of the local variations in the experimental data may seem to be reproduced by the model for both of these short filter lengths, and thus might be due to local variations in the scattering coefficient, but this may be coincidental. The MSE increases as the filter length decreases and is approximately doubled at Lz=2μm compared to Lz=100μm .

Table 1 shows that the average values for the diffusion coefficient using the function-of-depth approach are 20–30% higher than the values obtained using the constant-scattering approach. However, the differences between average diffusion coefficient values within the function-of-depth approach are less than 10% for the three filter widths. These results show that the function-of-depth approach is robust to changes in filter width when it comes to modeling the interference signal. Since the depth resolution in LCI equals the coherence length, it is reasonable to choose the filter length equal to the coherence length, which for λ1=675nm is lc=13.5μm in the sample.

When increasing the filter length beyond Lz=100μm we have found that the standard deviation, σD , in the estimated diffusion coefficient increases with the filter length. We believe this to be a result of the decrease in the depth interval used for model fitting due to the filter-width increase rather than the increase in Lz itself.

In order to obtain an accurate determination of D , the effect of diffusion of the dye, i.e., the time development of the speckle-averaged LCI envelope, must be visible in the measurements. When the depth dependency of the averaged envelopes is dominated by scattering in the depth interval used for parameter fitting, we expect a lower accuracy in determining D than when dye absorption dominates the signal. This assumption is supported by the above-mentioned increase in σD and also by the difference in σD between the two concentration series. We see that the standard deviation is 2–3 times higher within the results from the CIL=0.30% series compared to the CIL=0.15% series. When we designed the experiments, our emphasis was on ensuring that the absorption and scattering contributed equally to the decrease in signal level at the maximum probing depth compared to the signal just below the sample surface. The probing depth was determined as the depth where scattering reduced the signal, in decibels, to about half of the dynamic range. This resulted in the same amount of dye being deposited for both IL-concentration series and thus the same absorption coefficient. The ratio of the scattering coefficient to the absorption coefficient is thus doubled from the CIL=0.15% series to the CIL=0.30% series. This may explain the larger σD for the latter series compared to the former.

It is worth noting that the increase in diffusion coefficient with Intralipid concentration which was commented in Sec. 6.1 is also seen when the data are analyzed using the function-of-depth approach. At the present time we do not understand the cause of this increase. Further studies of this effect should be carried out, but it is beyond the scope of the present paper.

7.

Conclusion

The present work has demonstrated how the diffusion coefficient of a dye diffusing into a gel having unknown scattering properties may be determined using dual-wavelength LCI measurements. The method is based on simultaneous measurements at the two probing wavelengths where one is unaffected by the presence of the dye, and the other strongly absorbed by the dye.

The diffusion coefficient, D , is determined by fitting a mathematical model of the LCI envelope to the envelope recorded at the wavelength absorbed by the diffusing dye, using D as one of the fitting parameters. Measurement at the other wavelength, which is unaffected by the dye, is used for estimating the scattering coefficient at the absorbed wavelength. Two approaches to estimating the scattering properties, the constant-scattering approach and the function-of-depth approach, were tested. The first approach assumes homogeneous scattering properties while the second also can handle slow variations in scattering properties as a function of depth using a depth-resolved attenuation coefficient estimator previously presented by Støren 9 As expected, the two approaches gave similar estimates of the diffusion coefficient for the samples used in the present work, namely agar gel having various concentrations of Intralipid added to increase scattering. The numerical value of the estimated diffusion coefficient was also found to be in reasonable agreement with a previously published value obtained from measurements on agar without Intralipid. The numerical values for the average diffusion coefficient determined from measurements on samples having two different Intralipid concentrations and using the two approaches to determine the scattering coefficient were in the interval D=[2.1,3.8]×1010m2s .

The estimate of D obtained using the function-of-depth approach to determine sample scattering was, for these relatively homogeneous samples, found to be robust to changes in filter length (Lz) in the attenuation coefficient estimator. Spatial filtering is required when estimating the scattering coefficient based on LCI measurements due to inherent speckle noise in the measurements, and Lz limits spatial resolution in the depth-resolved estimate of the scattering coefficient. Filter lengths as short as 15% of the coherence length (lc) in the samples were tested, and no significant deviations from the value of D obtained using longer filter lengths were found. A reasonable choice of filter length is thus Lz=lc , since lc limits spatial resolution in an LCI measurement. We expect the method to give reliable results also for samples having depth-dependent scattering, as long as the diffusion coefficient is relatively independent of scattering. Nevertheless further studies involving measurements using animal models or preferably measurements of dye diffusion in real PDT scenarios are necessary in order to verify that the presented method can be used for in situ monitoring of photosensitizer concentration in PDT.

The model of the LCI envelope used in the present work is only valid for slow variations in the scattering coefficient and must be modified if the method is to be used on samples or tissue having large, or discontinuous, variations in scattering as a function of depth.

Acknowledgments

This work is supported by project 111737/432 and 129104/420 financed by the Research Council of Norway. The authors wish to thank Prof. Ole J. Løkberg for helpful discussions on instrumentation.

References

1. 

D. Huang, E. A. Swanson, C. P. Lin, J. S. Schuman, W. G. Stinson, W. Chang, M. R. Hee, T. Flotte, K. Gregory, C. A. Pulifito, and J. G. Fujimoto, “Optical coherence tomography,” Science, 254 1178 –1181 (1991). 0036-8075 Google Scholar

2. 

J. G. Fujimoto, C. Pitris, S. Boppart, and M. Brezinski, “Optical coherence tomography, an emerging technology for biomedical imaging and optical biopsy,” Neoplasia, 2 9 –25 (2000). https://doi.org/10.1038/sj.neo.7900071 1522-8002 Google Scholar

3. 

W. Drexler, “Ultrahigh-resolution optical coherence tomography,” J. Biomed. Opt., 9 47 –74 (2004). https://doi.org/10.1117/1.1629679 1083-3668 Google Scholar

4. 

A. Dubois, K. Grieve, G. Moneron, R. Lecaque, L. Vabre, and C. Boccara, “Ultrahigh-resolution full-field optical coherence tomography,” Appl. Opt., 43 2874 –2883 (2004). https://doi.org/10.1364/AO.43.002874 0003-6935 Google Scholar

5. 

Z. Chen, T. Milner, S. Srinivas, X. Wang, A. Malekafzali, M. van Gemert, and J. Nelson, “Non-invasive imaging of in vivo blood flow velocity using optical Doppler tomography,” Opt. Lett., 22 1 –3 (1997). 0146-9592 Google Scholar

6. 

J. de Boer, T. Milner, M. van Gemert, and J. Nelson, “Two dimensional birefringence imaging in biological tissue by polarisation sensitive optical coherence tomography,” Opt. Lett., 22 934 –936 (1997). 0146-9592 Google Scholar

7. 

J. Schmitt, S. Xiang, and K. Yung, “Differential absorption imaging with optical coherence tomography,” J. Opt. Soc. Am. A, 15 2288 –2296 (1998). 0740-3232 Google Scholar

8. 

U. Morgner, W. Drexler, F. Kärtner, X. Li, C. Pitris, E. Ippen, and J. Fujimoto, “Spectroscopic optical coherence tomography,” Opt. Lett., 25 111 –113 (2000). 0146-9592 Google Scholar

9. 

T. Støren, A. Røyset, L. O. Svaasand, and T. Lindmo, “Functional imaging of dye concentration in tissue phantoms by spectroscopic optical coherence tomography,” J. Biomed. Opt., 10 024037 (2005). https://doi.org/10.1117/1.1898242 1083-3668 Google Scholar

10. 

J. M. Schmitt, A. Knüttel, and R. F. Bonner, “Measurement of optical properties of biological tissues by low-coherence interferometry,” Appl. Opt., 32 6032 –6042 (1993). 0003-6935 Google Scholar

11. 

Y. Pan, R. Birngruber, J. Rosperich, and R. Engelhardt, “Low-coherence optical tomography in turbid tissue: theoretical analysis,” Appl. Opt., 34 6564 –6574 (1995). 0003-6935 Google Scholar

12. 

S. R. Thurber, A. M. Brodsky, and L. W. Burgess, “Characterization of random media by low-coherence interferometry,” Appl. Spectrosc., 54 1506 –1514 (2000). https://doi.org/10.1366/0003702001948420 0003-7028 Google Scholar

13. 

L. Thrane, M. H. Frosz, T. M. Jørgensen, A. Tycho, H. T. Yura, and P. E. Andersen, “Extraction of optical scattering parameters and attenuation compensation in optical coherence tomography images of multilayered tissue structures,” Opt. Lett., 29 1641 –1643 (2004). https://doi.org/10.1364/OL.29.001641 0146-9592 Google Scholar

14. 

D. J. Faber, F. J. van der Meer, M. C. G. Aalders, and T. G. van Leeuwen, “Quantitative measurement of attenuation coefficients of weakly scattering media using optical coherence tomography,” Opt. Express, 12 4353 –4365 (2004). https://doi.org/10.1364/OPEX.12.004353 1094-4087 Google Scholar

15. 

C. Xu, D. L. Marks, M. N. Do, and S. A. Boppart, “Separation of absorption and scattering profiles in spectroscopic optical coherence tomography using a least-squares algorithm,” Opt. Express, 12 4790 –4803 (2004). https://doi.org/10.1364/OPEX.12.004790 1094-4087 Google Scholar

16. 

A. I. Kholodnykh, I. Y. Petrova, K. V. Larin, M. Motamedi, and R. O. Esenaliev, “Precision of measurement of tissue optical properties with optical coherence tomography,” Appl. Opt., 42 3027 –3037 (2003). 0003-6935 Google Scholar

17. 

M. Pircher, E. Götzinger, R. Leitgeb, A. F. Fercher, and C. K. Hitzenberger, “Speckle reduction in optical coherence tomography by frequency compounding,” J. Biomed. Opt., 8 565 –569 (2003). https://doi.org/10.1117/1.1578087 1083-3668 Google Scholar

18. 

W. Star, “Light dosimetry in vivo,” Phys. Med. Biol., 42 763 –787 (1997). https://doi.org/10.1088/0031-9155/42/5/003 0031-9155 Google Scholar

19. 

B. Westrin, A. Axelsson, and G. Zacchi, “Diffusion measurements in gels,” J. Controlled Release, 30 189 –199 (1994). 0168-3659 Google Scholar

20. 

T. Støren, A. Simonsen, O. Løkberg, T. Lindmo, L. Svaasand, and A. Røyset, “Measurement of dye diffusion in agar gel by use of low-coherence interferometry,” Opt. Lett., 28 1215 –1217 (2003). 0146-9592 Google Scholar

21. 

J. M. Schmitt and A. Knüttel, “Model of optical coherence tomography of heterogeneous tissue,” J. Opt. Soc. Am. A, 14 1231 –1242 (1997). 0740-3232 Google Scholar

22. 

A. Ishimaru, Wave Propagation and Scattering in Random Media, 2 Academic Press, New York (1978). Google Scholar

23. 

J. Crank, The Mathematics of Diffusion, 11 –16 2nd ed.Oxford University Press, New York (1976). Google Scholar

24. 

Y. Pan, R. Birngruber, and R. Engelhardt, “Contrast limits of coherence-gated imaging in scattering media,” Appl. Opt., 36 2979 –2983 (1997). 0003-6935 Google Scholar

25. 

P. E. Andersen, L. Thrane, H. T. Yura, A. Tycho, T. M. Jørgensen, and M. H. Frosz, “Advanced modelling of optical coherence tomography systems,” Phys. Med. Biol., 49 1307 –1327 (2004). https://doi.org/10.1088/0031-9155/49/7/017 0031-9155 Google Scholar

26. 

Y. Pan and D. L. Farkas, “Noninvasive imaging of living human skin with dual-wavelength optical coherence tomography in two and three dimensions,” J. Biomed. Opt., 3 446 –454 (1998). https://doi.org/10.1117/1.429897 1083-3668 Google Scholar

27. 

U. Sathyam, B. Colston, L. DaSilva, and M. Everett, “Evaluation of optical coherence quantitation of analytes in turbid media by use of two wavelengths,” Appl. Opt., 38 2097 –2104 (1999). 0003-6935 Google Scholar

28. 

C. Yang, L. E. L. McGuckin, J. D. Simon, M. A. Choma, B. E. Applegate, and J. A. Izatt, “Spectral triangulation molecular contrast optical coherence tomography with indocyanine green as the contrast agent,” Opt. Lett., 29 2016 –2018 (2004). https://doi.org/10.1364/OL.29.002016 0146-9592 Google Scholar

29. 

Z. Chen, T. Milner, and J. Nelson, “Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media,” Opt. Lett., 22 64 –66 (1997). 0146-9592 Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Trude Støren, Arne Røyset, Lars Othar Svaasand, and Tore Lindmo "Measurement of dye diffusion in scattering tissue phantoms using dual-wavelength low-coherence interferometry," Journal of Biomedical Optics 11(1), 014017 (1 January 2006). https://doi.org/10.1117/1.2159000
Published: 1 January 2006
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scattering

Diffusion

Data modeling

Absorption

Statistical analysis

Statistical modeling

Interferometers

Back to Top