Open Access
1 March 2006 Fluorophore-mediated, fiber-optic, multi-analyte, immunosensing system for rapid diagnosis and prognosis of cardiovascular diseases
Liang Tang, Yongjie Ren, Bin Hong, Kyung A. Kang
Author Affiliations +
Abstract
A prototype of a fiber-optic, multi-analyte, immunobiosensing system was developed to simultaneously quantify disease-representing biomarkers in blood plasma. This system was for simultaneous quantification of two different groups of multi-biomarkers related to cardiovascular diseases (CVD): anticoagulants (protein C, protein S, antithrombin III, and plasminogen) for deficiency diagnosis; and cardiac markers (B-type natriuretic peptide, cardiac troponin I, myoglobin, and C-reactive protein) for coronary heart disease diagnosis. As an initial effort towards the development of a disposable and easy-to-use sensing cartridge as a rapid diagnostic tool for CVD related diseases, a prototype of a flow control system was also developed to automatically perform simultaneous four-analyte quantification. Currently, the system is capable of quantifying the multiple anticoagulants in their clinically significant sensing ranges within 5 minutes, at an average signal-to-noise (S/N) ratio of 25. A simultaneous assay of the four cardiac markers can be performed within 10 min, at an average S/N ratio of 20. When this highly portable multi-analyte sensing system is completed and successfully tested for CVD patient's plasma, it can provide rapid (<10 min) and reliable diagnostic and prognostic information at a patient's bedside.

1.

Introduction

Cardiovascular disease (CVD) is the leading cause of death in the US with an average of one death every 34 sec and approximately $400 billion of direct and indirect cost in the US alone.1 Among the CVD related diseases, coronary heart disease (CHD) is the single largest killer of Americans. Early and accurate diagnosis of CHD, especially in the emergency room, is crucial to design an appropriate patient care strategy. By measuring the levels of particular biomolecules, i.e., cardiac markers, in the blood, emergency room physicians can quickly determine whether patients have actually suffered a coronary event. Several important cardiac marker proteins have been identified and routinely used in the current clinical practice. Cardiac troponin I (cTnI) is widely used as a standard biomarker based on its absolute cardiac specificity and its long serum half-life ( 710days ; ESC∕ACC, 2002). A fast increase of myoglobin (MG) level in bloodstream following heart attack allows for a rapid patient evaluation. B-type natriuretic peptide (BNP) is useful for the emergency diagnosis of heart failure and for the prognosis in patients with acute coronary syndromes (ACS).2, 3 C-reactive protein (CRP) is an important prognostic indicator of CHD and ACS.3 A simultaneous quantification of these four cardiac markers allows clinicians to diagnose CHD quickly and to accurately design a patient care strategy, especially in an emergency room.3 A fast and reliable detection of these cardiac markers will also help medical professionals differentiate diseases among those showing similar symptoms. Maisel have evaluated the value of rapid bedside measurement of plasma BNP to distinguish coronary heart failure from a pulmonary cause of dyspnea for patients presenting a major complaint of acute dyspnea in the emergency room.2

The third leading cause of CVD in the US is venous thrombo-embolism (VTE) including deep vein thrombosis, lung embolism, cerebral venous thrombosis, and purpura fulminans. The incidence of VTE is approximately 1 per 1000 persons (1% for the elderly), and it accounts for 100,000 to 150,000 deaths annually.4, 5 When the haemostatic system in human body is unregulated, due to either coagulation problem or more commonly, impaired capacity of natural anticoagulant mechanism caused by anticoagulant deficiencies, the body has a predisposition to fatal venous thrombo-embolism.6, 7, 8 Protein C (PC), protein S (PS), antithrombin III (ATIII), and plasminogen (PLG) are four major anticoagulants in blood.9 Deficiency in PC, PS, or ATIII was reported to significantly increase the risks of VTE.10 Altogether, they account for about 15–20% of VTE cases.11 Therefore, accurate diagnostic tools for early detection of these deficiencies are invaluable to prevent the fatal VTE complications. A simultaneous quantification of these four anticoagulants in blood is beneficial for the accurate diagnosis of the actual cause of an abnormal clotting.

Due to the extremely low levels of these biomarkers ( pMnM ; Table 1 ) and the presence of other structurally similar biomolecules in blood, a frequently used assay method is enzyme linked immunosorbent assay (ELISA). Although very accurate, it is time-consuming (hours to days), expensive, and technically complicated. Commercially available test kits for anticoagulants (Sigma; St. Louis, MO), BNP (Biosite; San Diego, CA), cTnI (Roche; Basel, Switzerland), and CRP (Dade Behring; Deerfield, IL) can provide fast, easy, and point-of-care assays. However, they usually provide only qualitative single biomarker information and most of the assay kits are rather expensive. The need for an accurate, rapid, and cost-effective disease diagnosis and prognosis has been the motivation for developing a fluorophore-mediated, fiber-optic immuno-biosensing system. This technique performs a fluorophore mediated sandwich immunoassay within the evanescent field on the surface of an optical fiber,12 and has been successful for rapid clinical diagnostics.13, 14, 15 In our research group, a fiber-optic immuno-biosensing system was developed for rapid (5min) PC deficiency diagnosis. 16, 17, 18, 19, 20, 21, 22, 23 Compared to the conventional analytical methods such as DNA analysis, polymerase chain reaction, and ELISA, the immuno-optical biosensor has the advantages of rapid response time, user-friendliness, and cost-effectiveness.24

Table 1

Clinically significant sensing ranges of the anticoagulants and cardiac markers.

BiomarkersNormal Concentration[ μ g/ml (nM)]Target Sensing Range[ μ g/ml (nM)]
AnticoagulantPC4 (64)0.5–2.5 (8–40)
PS10 (143)1.5–5 (21–70)
ATIII150 (2300)45–105 (700–1600)
PLG200 (2000)60–120 (600–1200)
Cardiac markerBNP <0.0001 (0.026)0.0001–0.001 (0.026–0.26)
cTnI <0.001 (0.04)0.0007–0.007 (0.03–0.3)
MG <0.03 (1.7)0.07–0.7 (4–40)
CRP <0.8 (6.4)0.7–7 (5.6–56)

For simultaneous quantification of multi-cardiac-markers or multi-anticoagulants, a single fiber-optic biosensing system needs to be expanded to multi-analyte detection. A common method for multi-analyte biosensing is using multiple labels, such as multiple radioactive markers,25 fluorophores,26 or enzymes.27 However, a loss of sensitivity in the multi-analyte format is usually observed, compared to the individual assay format, due to the poor discrimination of signals generated by distinct labels. Another approach is by spatially separating the assay zone with same or different labels. The simultaneous quantification of multiple analytes in one sample is usually carried out on an antibody array format where different antibodies specific for each of the targeted analytes are immobilized in discrete areas.28 A simple approach for multi-analyte detection with spatially resolved methods may be connecting the single analyte assay system in series. Bakaltcheva have reported multi-explosive detection using fiber optic biosensing.29 RAPTOR™, manufactured by the Research International, Inc. (Monroe, WA), uses four polystyrene fiber sensors and is capable of simultaneous monitoring of four analytes.30 For our application, four individual sensors bearing antibodies specific for each biomarker are placed, in series, in an analyte sensing unit, similar to RAPTOR™. We, however, use quartz fibers and immobilize the first antibodies via biotin-streptavidin reaction to achieve low background signals and high sensor reusability.16, 17, 18 The convective flow application during the incubation of the sample and other reagents was also employed in our system to enhance the mass transport of these chemicals from the bulk media to the sensor surface. Hence, a special microfluidic unit was added for flow velocity control. In addition, a novel fluorescence enhancer developed in our research group is currently used in our sensing protocol that is not compatible with plastic fibers. Our multi-analyte sensing system consists of a sensing unit with four sensors, a fluorometer, and an automatic flow control unit. The results on using this system for the simultaneous quantification of multi-anticoagulants and multi-cardiac-markers with the application of the fluorescence enhancers are presented in this paper. The optimization of the sensing protocol was also performed to realize a rapid and accurate sensing system. To make the sensing procedures more user-friendly and more consistent with minimal human errors, automation of this multi-sensing system operation was also attempted.

2.

Materials and Methods

2.1.

Materials

Human B-type natriuretic peptide (BNP; a peptide of 32 amino acids) was purchased from Bachem (Torrance, CA) and two different murine, monoclonal IgG against human BNP, from Strategic Biosolutions (Newark, DE). Purified cardiac troponin I (cTnI; MW=29,000 ), myoglobin (MG; MW=16,000 ), and C-reactive protein (CRP; MW=114,000 ) from human heart and the respective, two different murine, monoclonal antibodies were obtained from Fitzgerald Industries (Concord, MA).

Human protein C (PC; MW=62,000 ) and two types of murine monoclonal anti-human PC (1°Mab-PC and 2°Mab-PC; MW=150,000 ) were provided by the American Red Cross (Rockville, MD). PC-free human plasma containing less than 1% PC was obtained from American Diagnostica, Inc. (Greenwich, CT). Human protein S (PS; MW=70,000 ), antithrombin III (ATIII; MW=65,000 ), plasminogen (PLG; MW=92,000 ) and two different types of murine monoclonal antibodies against respective anticoagulants (1°Mabs and 2°Mabs; MW=150,000 ) were purchased from Haematologic Technology (Essex Junction, VT).

Human serum albumin (HSA; MW=68,000 ), ImmunoProbe™ Biotinylation kit, and Sigmacote® were from Sigma (St. Louis, MO). Fluorolink™ Cyanine 5 (Cy5) reactive dye (the maximum excitation and emission at 649 and 670nm , respectively) and Alexa Fluor 647 reactive dye (AF647; the maximum excitation and emission at 650 and 668nm , respectively) were from Amersham Pharmacia Biotech (Uppsala, Sweden) and Molecular Probes (Eugene, OR), respectively. Plastic clad silica (PCS) optical fibers ( 600μm core diameter) were purchased from Research International, Inc. (Monroe, WA). LabView™ software package (version 7.1) and data acquisition card (6024E PCMCIA) were obtained from National Instruments (Austin, TX). Electronically controllable micro peristaltic pumps ( 35rpm , 12 VDC, 140mA ) and 2-way (normally open) and 3-way micro solenoid valves (12 volts DC) were from APT Instruments (Litchfield, IL) and ASCO Scientific (Florham Park, NJ), respectively.

2.2.

Methods

2.2.1.

Samples and reagents preparation

The previous study demonstrated that a 103mgHSAmlPBS solution can be used to emulate analyte-free human plasma with the same viscosity (1.9cP) .18 Therefore, the sample was prepared by adding a known amount of the analyte (cardiac markers or anticoagulants) to the HSA solutions at 103mgml . Conjugation of biotin to 1°Mabs was achieved using the ImmunoProbe™ Biotinylation kit and the conjugates were separated from free biotin by the size exclusion chromatography (Amersham-Pharmacia Biotech; Piscataway, NJ) with P-10 Gel (cutoff MW=20,000 ), according to manufacturer’s instructions. 2°Mab was reacted with bisfunctional NHS-ester Cy5 dye for 30min to obtain the optimum ratio of 1:5 and the Cy5-2°Mab conjugate was separated from unreacted dye by the size-exclusion chromatography.17 For the Alexa Fluor 647 (AF647), the reaction time was 1hour , following the manufacturer’s instructions. Sensor washing buffer (PBST; pH 7.4) was PBS buffer with 0.01% (v∕v) Tween 20. Sensor regeneration buffer was 0.1M triethylamine solution at pH 11.21, 31

2.2.2.

Fiber-optic biosensor preparation

An optical fiber (Research International; Monroe, WA) was tapered to maximize the internal reflection of photons within the fiber.12 The cladding of an optical fiber was removed at a predetermined length from one distal end. The exposed fiber optic core was tapered in hydrofluoric acid using the automatic tapering machine for 1h .31 Next, the tapered fiber surface was chemically treated to be streptavidin-coated, as described by Bhatia 32 in case a capillary was used as a sensing chamber. Before inserting the fiber into the chamber, Sigmacote® was applied to its inner surface to form a tight film of silicone on glass to minimize analyte adsorption. The chamber is connected to two nylon T connectors at both ends to form a functional sensor as described by Spiker 21 Biotin-conjugated 1°Mab (65μgml) was then injected into the sensing chamber and incubated at 4°C for 24h . The 1°Mab was immobilized via the avidin-biotin bridge on the fiber surface.22 1°Mab for respective analytes was immobilized on the fiber surface to form functional sensors. Before performing assays, 0.1M ethanolamine was applied as a blocking buffer to minimize possible nonspecific bindings of biomolecules. If not used immediately, the sensor was stored at 4°C in phosphate buffered saline (PBS; pH 7.4) with 0.02% (w∕v) sodium azide.

2.2.3.

Assay procedures

During an assay, the antibody-coated sensors were connected to the fluorometer, Analyte 2000™ (Research International; Monroe, WA). It has four channels capable of simultaneously quantifying four factors. The assay procedures with static incubation (i.e., no flow during sample∕reagent incubation) were performed as described by Spiker and Kang, unless otherwise specified.20, 22 For the assay with convection, during reaction (incubation), sample and reagent solutions were injected into the sensor unit via a peristaltic pump (Ismatec, Inc.; Glattbrugg-Zürich, Switzerland) and circulated within the sample circulation unit at a predetermined flow velocity.33 Currently, the total sample volume for an assay for four analytes is approximately 1ml . The assay procedures are briefly described as follows:

  • step 1: sample injection and incubation;

  • step 2: washing the sensor to remove non adsorbed molecules;

  • step 3: AF647-2°Mab injection and incubation;

  • step 4: washing the sensor to remove non reacted AF647-2°Mab; and

  • step 5: regenerating the sensor using the regeneration buffer, for a next assay cycle.

At the end of steps 2 and 4, readings were taken using the Analyte 2000™, where optical signal is converted to photocurrent (pA). The difference in the signal intensity between these two steps (ΔpA) is a direct measure of the fluorescence produced by the 1°Mab∕analyte∕AF647-2°Mab complex on the sensor surface, which is correlated with the target analyte concentration in the sample.

For multi-analyte quantification, four sensors were placed in the four-analyte sensing unit (Fig. 1 ). Each sensor was inserted into one of the four microchannels which were connected in series. The sensing procedures were the same as the individual analyte quantification.

Fig. 1

(a) Schematic diagram and (b) photograph of a prototype four-channel sensing unit with four 3cm sensors inserted into the channels, each (designed by Kang,; the microchannels are micromilled by Keynton, University of Louisville).

021011_1_030602jbo1.jpg

3.

Results and Discussion

3.1.

Multianticoagulant Sensing System

As stated in Table 1, the target sensing ranges for the anticoagulants are: PC ( 0.52.5μgml ; 840nM ), PS ( 1.55μgml ; 2170nM ), ATIII ( 45105μgml ; 7001600nM ), and PLG ( 60120μgml ; 6001200nM ). Individual fiber-optic biosensors were developed separately and optimized. For a multi-analyte sensing, four sensors were connected in the four-analyte sensing unit and simultaneous quantification of four analytes was performed.

3.1.1.

Single sensor study with protocol optimization

Blood plasma is frequently the sample for an assay in the clinical practice. When the incubation time for the sample and reagent was 5 and 3min , respectively, the signal intensity generated by PC molecules in plasma sample was decreased by approximately 70%, compared to that in the buffer based sample.18 Theoretical and experimental analyses demonstrated that the main cause for the signal reduction is the high viscosity of plasma.17, 31 The reaction kinetics of analyte (here, PC) and 1°Mab in a fiber-optic biosensing system involves two steps: the analyte transport from the liquid medium to the sensor surface; and the reaction between the transported analyte and the 1°Mab molecule on the surface. During the static incubation period, analyte molecules are transported to the sensor surface by diffusion. The analyte diffusion coefficient (D;cm2s) is inversely related to the sample medium viscosity:34

Eq. 1

D=kT6πμRA,
where k is the Boltzmann constant (1.38×1023JK) ; T is the sample medium temperature (room temperature, 298K ); μ is the sample medium viscosity ( 1.0cP for PBS buffer and 1.9cP for plasma); and RA is the radius of the analyte. Since the viscosity of plasma (1.9cP) is almost twice that of the PBS, the PC diffusion coefficient in plasma is approximately 50% less. Kwon has used the Damkohler number (NDa) for analyzing reaction kinetics: a system with an NDa value of greater than 100 is considered to be diffusion-limited; less than 0.01, reaction-limited.31, 35, 36 The values for the PC systems in both PBS and plasma samples are on the order of 100, indicating that the systems are diffusion-limited and the system with plasma samples is more diffusion-limited than the one with PBS buffer.

To enhance the sensing performance by improving the analyte mass transport to the sensor surface, another mechanism, in addition to diffusion, was sought. By applying convective flow during the sample incubation, the PC sensor performance for viscous samples (e.g., plasma) was significantly enhanced.17 Kwon has analyzed the effect of the convective flow on the fiber-optic biosensors using the thin film theory.31, 36 On the surface of a sensor, there is a film resistant to the analyte mass transport from the bulk solution to the sensor surface to react with 1°Mab immobilized on the surface. The thickness of this film ( δ ; cm) is closely related to the analyte mass transport rate:

Eq. 2

δ=Dkm,
where km is the effective mass transfer coefficient (cm∕s). When a convective flow is applied during the incubation (reaction) of sample, the film thickness is significantly reduced. For the PC sensing system, compared to the system with the convection at 0.1cms , the km value at 0.7 and 1.2cms was increased by 157 and 232%, respectively. As a result, the film thickness was reduced with the increase in the flow velocity. At 0.1cms , the thickness was 12×103cm . When the flow velocity was increased to 0.7cms , the film thickness was reduced by 60% (4.5×103cm) . At 1.2cms , the thickness was only 30% (3.5×103cm) of that of 0.1cms . As shown in the experimental analysis, the signal intensity at 0.7cms increased by 110% for 1μgPCml -plasma, compared to that with the static incubation.23, 31 At very higher flow velocities, although the PC mass transport is faster, the reaction rate is not fast enough to consume all of the PC molecules transported to the sensor surface. Therefore, the signal intensity remained steady at the velocities higher than 0.7cms . The optimal flow velocity for the PC sensing was then determined to be 0.7cms , where the reaction kinetics changes from the mass-transport-limited to the reaction-limited. At this velocity, the PC assay time was reduced to 5 min including 0.5 and 2 min for the sample and second antibody incubations, respectively.23

During the initial development of the anticoagulant sensor, fluorophore Cy5 was used as the signal mediator. However, the photobleaching and low quantum yield of this fluorophore have been issues for improvement. Alexa Fluor 647 (AF647) has a quantum yield more than twice that of Cy5 (0.28) and is more photostable. The excitation and emission wavelengths of AF647 are also compatible with our fluorometer, Analyte 2000™ and it has been demonstrated that AF647 can improve the sensing performance of the fluoro-immunoassays significantly.37 Therefore, a study was performed to investigate the efficacy of the AF647-conjugated second antibody for the PC sensing. AF647 was conjugated with the 2°Mab-PC (AF647-2°Mab-PC) at various dye-to-protein (D∕P) ratios from 1 to 8. 1μgml PC was reacted with the 1°Mab-PC immobilized on the sensor surface and then AF647-2°Mab at various D∕P ratios was applied to the sensor (Fig. 2 ). The signal intensity increased steadily up to the D∕P ratio value of 3 and then, at above 3, decreased with the increase in the ratio. Since the fluorescence for free forms of AF647-2°Mab at higher D∕P ratios increased (data not shown), one possible reason for the signal intensity reduction in the PC sensing system is the reduced relative affinity of AF647-2°Mab to the PC at higher D∕P ratios. The result from ELISA showed that, at the D∕P ratio of 2 and 3, the relative affinity of AF647-2°Mab-PC was similar to the 2°Mab-PC without conjugates (i.e., DP=0 ), and at D∕P ratio of 4, the relative affinity of the 2°Mab-PC to PC molecules decreased by 40%. The optimal AF647 D∕P ratio for the PC sensing was, therefore, determined to be 3. The signal intensity of PC measurement using AF647-2°Mab at D∕P ratio of 3 increased by approximately 115%, compared to the Cy5-2°Mab-PC at the optimal D∕P ratio of 5.

Fig. 2

Fluorescence of AF647-2°Mab-PC in the PC sensing system with change in the D∕P ratio and comparison to that of Cy5-2°Mab-PC at D∕P ratio of 5 (experimental conditions: 6cm PC sensors; 1μgml PC in plasma; 5μgml 2°Mab-PC; 0.52min incubation times for the sample and AF647-2°Mab-PC; 0.7cms convective flow velocity).

021011_1_030602jbo2.jpg

In the clinical practice, a minimal bio-sample volume for an assay is desirable and, therefore, a study was performed to minimize the sensor size. The performance of a 3cm PC sensor was tested in the target sensing range and compared with that of a 6cm sensor (data not shown). The signal intensity of the 3cm PC sensor also showed a linear relationship with the PC concentration, at an average signal-to-noise (S∕N) ratio of approximately 25. This result demonstrated that a 3cm PC sensor is capable of accurately quantifying PC in plasma (sample volume: 500μl ) in the clinically important sensing range for the PC deficiency diagnosis within 5min .

The sensing protocol of the PC sensor was further extended for the development of PS, ATIII, and PLG sensing system. The optimal flow velocity for the PS sensing was determined to be 0.5cms (data not shown), which is less than the optimal velocity of the PC sensing ( 0.7cms for 1μgml PC). It is probably because the PS sensing range is higher than the PC’s and therefore, less mass-transfer-limited during the reaction between PS and 1°Mab-PS, assuming that antibodies for both systems have similar affinity to respective antigens. The effects of the convective flow on the ATIII and PLG sensing performance were also studied (data not shown). Unlike the PC sensing, with a static incubation (i.e., no flow), the signal intensity was high enough to clearly differentiate ATIII or PLG concentration in the sensing range. It indicates that the diffusional mass transport is sufficient for the ATIII or PLG molecular supply to the sensor surface, since the sensing ranges are hundreds of times higher than that of PC. Similar to the PC sensor, the standard curves of the PS, ATIII, and PLG sensing were in linear relationship (r2=0.99) with the respective analyte concentration in the sensing ranges, at a S∕N ratio of 30, capable of quantifying each concentration in the blood plasma in the deficiency range within 5min (data not shown).

3.1.2.

Simultaneous quantification of the four anticoagulants

For the simultaneous four-anticoagulant quantification, each sensor, with the specific antibodies against respective anticoagulants immobilized on its surface, was placed in one of the four channels in the four-analyte sensing unit (Fig. 1). The unit connects the four sensors in series and when a sample flows over the four channels, each anticoagulant is captured by one of the four sensors for detection.

Cross-reactivity of the four sensors

The cross-reactivity of each anticoagulant sensor to the other three analytes was first investigated. To determine the maximum extent of the possible cross-reactivity, each sensor was tested with the other three analytes at their upper sensing limits. For example, for a PC sensor, PS (5μgml) , ATIII (105μgml) , and PLG (120μgml) were tested and the fluorescent signals were compared to those of the PC measurement at 0.5μgml (lower sensing limit) without other factors (Fig. 3 ). The responses from the PC sensor for the PS, ATIII, and PLG in the sample were less than 5% (almost not shown) of the positive response for PC at its lower sensing limit. The PS, ATIII, and PLG sensors showed similar results, indicating that all four sensors were highly specific with little cross-reactivity.

Fig. 3

Minimal cross-reactivity of PC, PS, ATIII, and PLG sensors with three other analytes (experimental conditions: 3cm sensors; respective analyte at its lower sensing limit; the other three analytes at their upper sensing limits; incubation times for the sample and reagent were 0.5 and 2min , respectively; flow velocity at 0.7cms ).

021011_1_030602jbo3.jpg

Effect of second antibody mixture on the sensing performance

For the multi-anticoagulant sensing, after the sample incubation, four different AF647-2°Mabs need to be introduced into the respective sensor chambers for quantification by fluoresence. This can be done by either applying each antibody to the respective chamber or applying a mixture of the four different AF647-2°Mabs through the four-analyte sensing unit if the mixture does not affect the sensing performance. The second case is more desirable for a multi-sensing unit because it requires an easier design for the fluid manipulation and automation of the sensing procedures. The effect of the AF647-2°Mabs mixture was, therefore, investigated for the sensing performance of each sensor.

Figure 4 shows the signal intensities generated by a PC sensor using only AF647-2°Mab-PC (squares) and the AF647-2°Mabs mixture (diamonds). Both measurements were performed by a single PC sensor to eliminate possible fiber-to-fiber sensing variation. The concentration of the total AF647-2°Mabs in the mixture was four times that of individual sensing. The signal intensity generated by the mixture slightly decreased, but the reduction was less than 6%, with a similar standard deviation (510%) for both measurements. The cross-reactivity among the four analytes was already proven to be minimal. The cause for the signal reduction could be that the presence of other molecules (here, AF647-2°Mab-PS, AF647-2°Mab-ATIII, and AF647-2°Mab-PLG) causes a slower mass transport rate for the AF647-2°Mab-PC in the sensing system within a limited reaction time. A more systematic investigation for the actual cause is currently ongoing. The background signal generated by the AF647-2°Mab mixture slightly increased (1015pA) for all the tested concentrations, compared to that by AF647-2°Mab-PC only, with a possible increase in the non-specific binding by the increase in the 2°Mab concentration.

Fig. 4

Effect of the AF647-2°Mab mixture on PC sensing (experimental conditions: 3cm PC sensor; 0.52min incubation for the samples and reagents, respectively; 0.7cms convective flow velocity).

021011_1_030602jbo4.jpg

The effect of the antibody mixture on the PS, ATIII, and PLG sensor performance was also studied (data not shown). Similar to the PC sensor, the signal intensities of the measurements with the AF647-2°Mabs mixture were decreased by less than 4.5%. For all the four sensors, the standard curves with the AF647-2°Mabs mixture were linear with the respective anticoagulant concentration (r2=0.99) , at an average S∕N ratio of 25. These results demonstrated that the simultaneous quantification of the four anticoagulants is possible with the mixed AF647-2°Mabs with little changes in the sensitivity, especially if the standard curve is also obtained with the AF647-2°Mabs mixture. Therefore, it was decided to apply the mixture to the multi-sensing unit for simultaneous PC, PS, ATIII, and PLG detection.

Simultaneous four-factor quantification

The sensing performance of the four-analyte sensing unit for samples with PC, PS, ATIII, and PLG at their lower (0.5, 1.5, 45, and 60μgml , respectively) and higher (2.5, 5, 105, and 120μgml , respectively) sensing limits was first investigated (data not shown). The signal intensities of the simultaneous multi-anticoagulant sensing were decreased by 5–10% compared to those of single anticoagulant sensing at their lower sensing limits, at their upper sensing limits, approximately 3–7%. Since the specificity of the sensor to the respective analyte has been already well proven (Fig. 3), this signal reduction was probably due to the reduced diffusion of the analyte with the presence of the other analytes in the sample. This signal reduction causes little problem if the patient has only one anticoagulant deficiency because the extent of the signal reduction will be consistent. If the patient has more than one anticoagulant deficiency then it may slightly overestimate the analyte concentration in the sample. However, even with this concern, the error is less than 10%. A more systematic study is in progress to identify the actual cause for the signal reduction. Convective flow at a higher flow velocity may improve the analyte mass transport to the sensor surface to reduce the interference of the other analyte in the sample. Increased sample incubation time may also be helpful for the mass transfer in mixed sample.

Despite the slight signal reduction, the sensitivities of the four anticoagulant sensors in multi-sensing are similar to the individual anticoagulant sensing (Fig. 5 ). The standard curves are linear with the respective analyte concentration in the sensing ranges (r2=0.970.99) . The study results demonstrated that the four-analyte sensing unit is capable of performing a rapid (5min) , accurate, and simultaneous quantification of PC, PS, ATIII, and PLG in blood plasma, at an average S∕N ratio of 25.

Fig. 5

Sensitivity of (a) PC (squares) and PS (diamonds), (b) ATIII (circles) and PLG (triangles) sensors in the four-analyte sensing unit for simultaneous four factor quantification. Solid lines: single anticoagulant quantification; dashed lines: anticoagulant in mixture (experimental conditions: 3cm sensors; PC, PS, ATIII and PLG in plasma; 0.52min incubation times for the sample and reagent, respectively; 0.7cms convective flow velocity; mixed AF647-2°Mabs).

021011_1_030602jbo5.jpg

3.2.

Multi-Cardiac-Marker Sensing

The fiber-optic, multi-analyte immunobiosensing system can also be applied for simultaneous quantification of multiple cardiac markers in blood plasma as a diagnostic and prognostic tool for coronary heart disease. As stated before (Table 1), the clinically significant sensing ranges for the BNP [0.11ngml(26260pM)] and cTnI [0.77ngml(30300pM)] are hundreds of times lower than the anticoagulants. The MG [70700ngml(440nM)] and CRP [7007000ngml(5.656nM)] are comparable to the anticoagulants, therefore, little difficulty was expected for the MG and CRP sensing.

Due to the extremely low target BNP and cTnI sensing range, the initial sensing study was performed with a 12cm BNP sensor, and 10 min, each, for the sample and AF647-2°Mab incubation. With a static incubation (i.e., no flow), the signal intensity of the BNP sensing in blood plasma was only 510pA . Evaluation of the Damkohler number (NDa) shows that the reaction kinetics of the BNP and cTnI sensing system is mass-transfer-limited with diffusional mass transport only. To improve the sensing performance by accelerating the analyte mass transport, a convective flow was applied and the optimal flow velocity was determined to be 1.2cms (data not shown). This flow velocity is much higher than the anticoagulant sensing system, because of the extremely low sensing range. The film thickness for the BNP sensing at 0.1cms was calculated to be 25.9×103cm , using the theoretical analysis by Kwon,31 twice of that of the PC sensing system (12×103cm) .

The signal intensity for many disease representing biomarkers, such as BNP and cTnI, is extremely low due to the low target sensing range. Since the fiber-optic biosensing system is fluorophore mediated, fluorescence enhancers were applied to improve the sensor performance by enhancing the fluorescent signal. Our research group has developed novel nanometal particle reagents (NMPRs) that can improve the sensitivity of the biosensing system as high as 10 times. These reagents are specially prepared with nanometal particles in a selective, biocompatible organic solvent. The fluorescence enhancement was by (1) the transfer of the electrons, normally involved in intramolecular fluorescence quenching, from a fluorophore to a nanometal particle, via the plasmon rich, electromagnetic field surrounding the nanometal particle; and (2) the increase of the energy gap of a fluorophore between the ground and the excitation states in a solvent.38, 39, 40 Figure 6a shows the standard curve of the BNP sensing with and without nanogold particle reagent (NGPR). The signal intensity increased by approximately 100–300%, with the sensor size reduced by 75% and a less than 50% of the assay time. Similar results were obtained for the cTnI, MG, and CRP sensing [Fig. 6b, 6c, 6d], indicating that the 3cm cardiac marker sensors were capable of accurately quantifying the cardiac marker concentrations in blood plasma within 10 min, at an average S∕N ratio of 20.

Fig. 6

Standard curves of (a) BNP, (b) cTnI, (c) MG, and (d) CRP sensing with (squares) and without (triangles) NGPR (experimental conditions: 3cm sensors; incubation time for the sample and AF647-2°Mabs were 3 and 4min , respectively; 1.2cms convective flow velocity).

021011_1_030602jbo6.jpg

Similar to the multi-anticoagulant sensing, the four cardiac marker sensors were placed in the four-analyte sensing unit for simultaneous multi-cardiac-marker quantification. The four cardiac markers have no known structural homology. However, to determine the possible interference to the sensing performance in the multi-sensing format, each sensor was tested with samples containing the other three analytes at their upper limit sensing ranges (data not shown), as in multi-anticoagulant sensing. The signal intensities generated by the BNP sensor for cTnI (7ngml) , MG (700ngml) , and CRP (7000ngml) in the sample were less than 5% of the response for the BNP at its lower sensing limit (0.1ngml) . Similar results were shown for the cTnI, MG, and CRP sensors when they were probed with the other three analytes present in the sample. AF647-2°Mabs mixture was also tested and it was found that each sensor can be as accurate as that with only the respective AF647-2°Mab, with a slight signal reduction. The sensing performance of the multi-cardiac-marker quantification was comparable to the single cardiac marker sensing, at a similar S∕N ratio (data not shown). Similar to the multi-anticoagulant sensing, the signal intensity for mixed sample was slightly decreased (510%) due to the presence of other three analytes in the sample. This preliminary study result showed that the multi-sensing unit was capable of simultaneously quantifying the four cardiac markers within 10 min, at an average S∕N ratio of 20.

3.3.

Automation of the Sensing Procedures with Automatic Flow Control System

The ultimate goal of this research is to develop a rapid, automatic, and easy-to-use sensing cartridge (or chip) for CVD diagnosis and∕or providing prognostic information to the surgeons at patient’s bedside. The automatic operation is beneficial for a more reliable patient evaluation since the assay procedures can be well controlled with minimal human errors. As an initial effort, an automatic flow control system consisting of electrically controllable pump and valves was developed. A computer program (LabView™) electronically operates the micro peristaltic pump and six micro solenoid valves in the system. To test the sensing performance of the multi-sensing unit integrated with the automatic flow control system, a simultaneous quantification of the multi-anticoagulants was performed at their lower and upper sensing limits with manual and automatic operations. For both limits, the signal intensities of the measurements with the automatic flow control system were comparable to the manual operation.

Current effort is focused on converting this prototype system to a compact microfluidic system where all the electronic control components, micro-pump, and micro-valves are integrated on a small board inside an instrument with a space where a sensing cartridge can be inserted. The final version of the sensing system as a rapid diagnostic tool for CVD diseases is envisioned as follows: when the sample and reagent solutions are filled in appropriate reservoirs and the sensing cartridge is inserted on the instrument, an assay will be performed automatically and the quantitative results are shown in the LCD screen within 10 min with a printout.

4.

Conclusions

A prototype of a semi-automatic, fluorophore mediated, fiber-optic, multi-analyte immunobiosensing system has been developed for simultaneous quantification of multiple CVD related biomarkers in blood plasma. In particular, a multi-anticoagulant and a multi-cardiac-marker sensing systems were developed for anticoagulant deficiency diagnosis and coronary heart disease diagnosis and prognosis, respectively.

First, individual sensors ( 3cm long) were developed and the sensing protocols were optimized. For simultaneous quantification of four factors in blood, the four sensors bearing antibodies specific for respective biomarkers were connected in series in a four-analyte sensing unit. The multi-anticoagulant sensing system was capable of performing a rapid (5min) , accurate, and simultaneous quantification of PC, PS, ATIII, and PLG in blood plasma, at an average S∕N ratio of 25. A simultaneous assay of the four cardiac markers (BNP, cTnI, MG, and CRP) in their clinically significant sensing ranges can be performed within 10 min, at an average S∕N ratio of 20. Efforts were also made to develop an automatic flow control system to precisely and automatically perform the assay procedures for simultaneous four analytes quantification. This semi-automatic sensing system is an initial effort towards the ultimate goal of a disposable and automatic sensing cartridge as a portable diagnostic tool for CVD related diseases. This fast and portable assay system can be used not only in the emergency room for a rapid CHD diagnosis, but also for monitoring of patients at the bedside to provide prognostic information of coronary events. The principle of this multi-analyte sensing system can also be applied for multi-sensing of other disease-representing biomarkers, such as cancer markers.

Acknowledgments

We would like to thank the National Science Foundation (CAREER Award BES-9733207; BES-0330075) for the financial support. American Heart Association and SigmaXi Honor Society are acknowledged for Liang Tang’s Predoctoral Fellowship and Grants-in-Aid of Research award, respectively. Special thanks are given to the American Red Cross for providing the protein C and related monoclonal antibodies.

References

1. 

, “Heart disease and stroke statistics—2005 update,” 3 –13 (2005) Google Scholar

2. 

A. S. Maisel, P. Krishnaswamy, R. M. Nowak, J. McCord, “Rapid measurement of B-type natriuretic peptide in the emergency diagnosis of heart failure,” N. Engl. J. Med., 347 161 –167 (2002). 0028-4793 Google Scholar

3. 

M. S. Sabatine, D. A. Morrow, J. A. de Lemos, C. M. Gibson, S. A. Murphy, N. Rifai, C. McCabe, E. M. Antman, C. P. Cannon, and E. Braunwald, “Multimarker approach to risk stratification in non-ST elevation acute coronary syndromes: simultaneous assessment of troponin I, c-reactive protein, and b-type natriuretic peptide,” Circulation, 105 1760 –1763 (2002). 0009-7322 Google Scholar

4. 

F. A. Anderson, H. B. Wheeler, R. J. Goldberg, D. W. Hosmer, N. A. Patwardhan, B. Jovanovic, A. Forcier, and J. E. Dalen, “A population-based perspective of the hospital incidence and case-fatality rates of deep vein thrombosis and pulmonary embolism, the Worcester DVT study,” Arch. Intern Med., 151 933 –938 (1991). 0003-9926 Google Scholar

5. 

F. R. Rosendaal, “Venous thrombosis: A multicausal disease,” Lancet, 353 1167 –1173 (1999). 0140-6736 Google Scholar

6. 

J. Amiral and J. Fareed, “Thromboembolic diseases: biochemical mechanisms and new possibilities of biological diagnosis,” Semin Thromb Hemost, 22 41 –48 (1996). 0094-6176 Google Scholar

7. 

R. D. Appleby and R. J. Olds, “The inherited basis of venous thrombosis,” Pathology, 29 341 –347 (1997). 0031-3025 Google Scholar

8. 

M. Makris, F. R. Rosendaal, and F. E. Preston, “Familial thrombophilia: genetic risk factors and management,” J. Intern Med., 740 9 –15 (1997). 0954-6820 Google Scholar

9. 

D. F. Bruley and W. N. Drohan, Advances in Applied Biotechnology Series (Vol. 11): Protein C and Related Anticoagulants, Gulf Publication Company, London (1991). Google Scholar

10. 

H. R. Martinez, R. A. Rangel-Guerra, and L. J. Marfil, “Ischemic stroke due to deficiency of coagulation inhibitors: Report of 10 young adults,” Stroke, 24 19 –25 (1993). 0039-2499 Google Scholar

11. 

C. S. Kitchens, “Thrombophilia and thrombosis in unusual sites,” Hemostasis and Thrombosis: Basic Principles and Clinical Practice, 1255 –1263 Lippincott Company, Philadelphia (1994). Google Scholar

12. 

F. S. Ligler, J. P. Golden, L. C. Shriver-Lake, R. A. Ogert, D. Wijesuria, and G. P. Anderson, “Fiber-optic biosensor for the detection of hazardous materials,” Immunomethods, 3 122 –127 (1993). 1058-6687 Google Scholar

13. 

J. R. Astles and W. G. Miller, “Measurement of free phenytoin in blood with a self-contained fiber optic immunosensor,” Anal. Chem., 66 1675 –1682 (1994). 0003-2700 Google Scholar

14. 

P. B. Daniels, J. E. Fletcher, P. M. O’Neill, C. G. Stafford, T. Bacarese-Hamilton, and G. A. Robinson, “A comparison of three fluorophores for use in an optical biosensor for the measurement of prostate-specific antigen in whole blood,” Sens. Actuators B, 27 (1–3), 447 –451 (1995). 0925-4005 Google Scholar

15. 

K. Schult, A. Katerkamp, D. Trau, F. Grawe, K. Cammann, and M. Meusel, “Disposable optical sensor chip for medical diagnostics: new ways in bioanalysis,” Anal. Chem., 71 5430 –5435 (1999). https://doi.org/10.1021/ac9907686 0003-2700 Google Scholar

16. 

H. I. Balcer, J. O. Spiker, and K. A. Kang, “Sensitivity of protein C immunosensor with and without human serum albumin,” Adv. Exp. Med. Biol., 471 605 –612 (1999). 0065-2598 Google Scholar

17. 

H. I. Balcer, H. J. Kwon, and K. A. Kang, “Assay optimization of a rapid, reusable protein C immunosensor for physiological samples,” Ann. Biomed. Eng., 30 141 –147 (2002). https://doi.org/10.1114/1.1433487 0090-6964 Google Scholar

18. 

H. J. Kwon, H. I. Balcer, and K. A. Kang, “Sensing performance of protein C immuno-biosensor for biological samples and sensor minimization,” Comp. Biochem. Physiol. A, 132 231 –238 (2002). 0300-9629 Google Scholar

19. 

J. O. Spiker, K. A. Kang, W. N. Drohan, and D. F. Bruley, “Protein C detection via fluorophore mediated immuno-optical biosensor,” Adv. Exp. Med. Biol., 428 621 –627 (1997). 0065-2598 Google Scholar

20. 

J. O. Spiker, K. A. Kang, W. N. Drohan, and D. F. Bruley, “Preliminary study of biosensor optimization for the detection of protein C,” Adv. Exp. Med. Biol., 454 681 –688 (1998). 0065-2598 Google Scholar

21. 

J. O. Spiker, W. N. Drohan, and K. A. Kang, “Reusability study of fiber optic based protein C biosensor,” Adv. Exp. Med. Biol., 471 731 –739 (1999). 0065-2598 Google Scholar

22. 

J. O. Spiker and K. A. Kang, “Preliminary study of real-time fiber optic based protein C biosensor,” Biotechnol. Bioeng., 66 158 –163 (1999). https://doi.org/10.1002/(SICI)1097-0290(1999)66:3<158::AID-BIT3>3.0.CO;2-G 0006-3592 Google Scholar

23. 

L. Tang, H. J. Kwon, and K. A. Kang, “Theoretical and experimental analysis of analyte transport in a fiber optic, protein C immuno-biosensor,” Biotechnol. Bioeng., 88 869 –879 (2004). 0006-3592 Google Scholar

24. 

P. B. Luppa, L. J. Sokoll, and D. W. Chan, “Immunosensors-principles and applications to clinical chemistry,” Clin. Chim. Acta, 314 1 –26 (2001). 0009-8981 Google Scholar

25. 

J. P. Yvert and J. F. Delagneau, “Simultaneous multiple immunoassays: current data,” Ann. Biol. Clin., 51 43 –48 (1993). 0003-3898 Google Scholar

26. 

M. Sauer, K. T. Han, R. Müller, A. Schulz, R. Tadday, S. Seeger, J. Wolfrum, J. Arden-Jacob, G. Deltau, N. I. Marx, and K. H. Drexhage, “New fluorescent labels for time-resolved detection of biomolecules,” J. Fluoresc., 3 131 –139 (1993). https://doi.org/10.1007/BF00862730 1053-0509 Google Scholar

27. 

M. N. Nanjee and N. E. Miller, “Sequential microenzymatic assay of cholesterol, triglycerides, and phospholipids in a single aliquot,” Clin. Chem., 42 915 –926 (1996). 0009-9147 Google Scholar

28. 

D. J. Pritchard, H. Morgan, and J. M. Cooper, “Micron-scale patterning of biological molecules,” Angew. Chem., Int. Ed. Engl., 34 91 –93 (1995). 0570-0833 Google Scholar

29. 

I. B. Bakaltcheva, F. S. Ligler, C. H. Patterson, and L. C. Shriver-Lake, “Multi-analyte explosive detection using a fiber optic biosensor,” Anal. Chim. Acta, 399 13 –20 (1999). 0003-2670 Google Scholar

30. 

G. P. Anderson, K. D. King, K. L. Gaffney, and L. H. Johnson, “Multianalyte detection using the fiber optic biosensor,” Biosens. Bioelectron., 14 (10–11), 771 –777 (2000). https://doi.org/10.1016/S0956-5663(99)00053-6 0956-5663 Google Scholar

31. 

H. J. Kwon, “Theoretical and experimental investigation on sensing performance of protein C immuno-optical sensor for physiological samples,” (2002) Google Scholar

32. 

S. K. Bhatia, L. C. Shriver-Lake, K. J. Prior, J. H. Georger, J. M. Calvert, R. Bredehorst, and F. S. Ligler, “Use of thiol-terminal silanes and heterobifunctional crosslinkers for immobilization of antibodies on silica surfaces,” Anal. Biochem., 178 408 –413 (1989). https://doi.org/10.1016/0003-2697(89)90662-3 0003-2697 Google Scholar

33. 

L. Tang, H. J. Kwon, and K. A. Kang, “Studies on effect of sample circulation on sensing performance of protein C immunosensor,” 1805 –1806 (2002). Google Scholar

34. 

R. B. Bird, W. E. Stewart, and E. N. Lightfoot, Transport Phenomena, John Wiley & Sons, New York (1960). Google Scholar

35. 

M. Stenberg, L. Stilbert, and H. Hygren, “External diffusion in solid-phase immunoassay,” J. Theor. Biol., 120 129 –140 (1986). 0022-5193 Google Scholar

36. 

J. Welty, C. E. Wicks, and R. E. Wilson, Fundamentals of Momentum, Heat and Mass Transfer, John Wiley & Sons, Canada (1984). Google Scholar

37. 

G. P. Anderson and N. L. Nerurkar, “Improved fluoroimmunoassays using the dye Alexa Fluor 647 with the RAPTOR, a fiber optic biosensor,” J. Immunol. Methods, 271 17 –24 (2002). 0022-1759 Google Scholar

38. 

B. Hong and K. A. Kang, “Biocompatible, nano-gold-particle fluorescence enhancer for fluorophore mediated, optical immunosensor,” Biosens. Bioelectron., 21 (7), 1333 –1338 (2006). 0956-5663 Google Scholar

39. 

B. Hong and K. A. Kang, “Fluorescence enhancers for fluorophore mediated biosensors for cardiovascular disease diagnosis,” Adv. Exp. Med. Biol., 578 Google Scholar

40. 

K. A. Kang and B. Hong, “Biocompatible nano-metal particle fluorescence enhancers,” Crit. Rev. Eukaryot. Gene Expr., 16 (1), 45 –60 (2006) Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Liang Tang, Yongjie Ren, Bin Hong, and Kyung A. Kang "Fluorophore-mediated, fiber-optic, multi-analyte, immunosensing system for rapid diagnosis and prognosis of cardiovascular diseases," Journal of Biomedical Optics 11(2), 021011 (1 March 2006). https://doi.org/10.1117/1.2192529
Published: 1 March 2006
Lens.org Logo
CITATIONS
Cited by 28 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Sensing systems

Statistical analysis

Fiber optics

Plasma

Picosecond phenomena

Blood

Back to Top