Open Access
1 January 2008 Two-wavelength carbon dioxide laser application for in-vitro blood glucose measurements
Martina C. Meinke, Gerhard J. Müller, Hansjörg Albrecht, Christina Antoniou, H. Richter, Jürgen M. Lademann
Author Affiliations +
Abstract
To develop a fast and easy clinical method for glucose measurements on whole blood samples, changes in glucose spectra are investigated varying temperature, glucose concentration, and solvent using attenuated total reflection Fourier transform infrared (ATR- FTIR) measurements. The results show a stability of the spectra at different temperatures and wavelength shifts of the absorption bands when water is replaced by blood. Because the ATR measurements are influenced by sedimentation of the red blood cells, a two-wavelength CO2 laser is used to determine the glucose concentration in whole blood samples. For this purpose, the first laser wavelength λ1 is tuned to the maximum of the glucose absorption band in blood at 1080 cm-1, and the second laser wavelength λ2 is tuned to 950 cm-1 for background measurements. The transmitted laser power through the optical cell containing the whole blood sample at λ1 and λ2 is used to determine the ratio. This signal correlates well with the glucose concentration in the whole blood samples. The CO2 laser measurement is too fast to be influenced by the red blood cell sedimentation, and will be a suitable method for glucose determination in whole blood.

1.

Introduction

The concentration of glucose in the capillaries of the skin is a very important parameter in the field of medical diagnostics. Glucose plays a major role within the complex nutritive supply system of the tissue. In healthy persons, the concentration of glucose is stable within a narrow range (dependent on age and fasting: 70 to 110mgdl ). In patients with diabetes, the regulation system is disordered. The disease causes the metabolism to malfunction, and without therapeutic support, the probability of survival is low. Diabetic therapy requires that blood glucose concentration be measured several times a day. More than 10% of the costs in a clinical chemistry laboratory result from tests used to determine blood glucose concentration, thereby being the most frequently measured parameter. The standardized laboratory measurement methods include several work steps, and are therefore time- and cost-intensive methods.

A minimally invasive, fast, and reliable method for the determination of blood glucose is still of high economic relevance, because such a method has not yet been found. The development of a method for the quantitative determination of glucose in blood by means of infrared spectroscopy has been the subject of recurrent research and publication. Such research has involved investigations not only on dried blood,1, 2, 3 but also on serum and whole blood4 utilizing both transmission spectroscopy and ATR spectroscopy.5, 6 Cylindrical ZnSe crystals have been employed for these experiments. However, this approach has various practical shortcomings, such as the adsorption of proteins on the surface of the crystal.

In other methods, blood samples have been applied drop-wise on polyethylene cards and dried before measurement to overcome the interference of water absorption.7 One early report about the state of the art8 and a recently published review9 have discussed the use of analytical instrumentation for glycaemic control. For an overview of various other laboratory assays, see Heise and Abel.10 Multivariate methods11 and Raman spectroscopy have also been used to determine glucose concentration in vivo in blood.12 A recently published article used a system based on liquid-core optical fiber Raman and absorption spectroscopy, followed by partial least square regression to determine chemical concentrations in blood serum or urine.13 Glucose in water has proved to be a challenge to analysts in the past, because water has strong absorption bands in the infrared wavelength range of interest. The new FTIR spectrometers allow the detection of glucose quantitatively within a range of 1300 to 1000cm1 . Whole blood shows bands of other substances within the range of interest such as proteins, urea, cholesterol, and triglycerides, in addition to glucose absorption. Moreover, there is an additional loss of light caused by erythrocyte scattering.

Vonach examined the direct determination of glucose in whole blood, without any sample preparation, using FTIR spectroscopy.14 The measurements were carried out in a flow cell. The whole spectral region from 1182 to 964cm1 had to be analyzed for determination of glucose concentration. A clear correlation could be obtained between the measured data and clinically determined blood glucose concentration.

In the investigations described in this work, FTIR measurements were used to investigate the influence of temperature on glucose in water, the influence of concentration on glucose in water and in whole blood, and the influence of the surrounding media (water versus blood) on glucose spectra. The results were compared to glucose determination carried out by the application of a two-wavelength CO2 laser.

2.

Materials and Methods

2.1.

Material

Glucose samples were prepared by dissolving nine different amounts of glucose (Merck KGaA, Darmstadt, Germany) in 100-ml water. The sample concentration ranged from 40 to 400mgdl .

Blood samples of healthy volunteers were used in the experiments. Glucose was added prior to testing, and the concentration of these samples was determined in the laboratory of clinical chemistry using the glucose-UV-fluid test (GLUC, Hitachi 747, Biomed, Oberschleißheim, Germany). In healthy volunteers, the normal blood glucose concentration ranges between 60 and 117mgdl , and can be slightly increased by ingesting pure glucose.

2.2.

Fourier Transform Infrared Spectrometer

A FTIR spectrometer (Vector 22, Bruker Optik GmbH, Bremen, Germany) with a ZnSe ATR crystal was used for the analysis of aqueous glucose solutions and whole blood samples. The measurements were carried out at different temperatures between 15 and 28°C . The measurements were carried out in ATR mode, and therefore the ATR crystal and the samples were surrounded by a heating system. The temperature was measured before and after each measurement with a digital thermometer GTH 1200 A (Greisinger Electronics, Regenstauf, Germany).

2.3.

Two-Wavelength CO2 Laser

A two-wavelength CO2 laser was developed and manufactured at the Institute of Applied Physics in Nishny Novgorod, Russia. The two laser wavelengths were selected at λ1=1080cm1 and λ2=950cm1 , because λ1 is located within the glucose absorption band and λ2 is used for background corrections.

The two laser beams with λ1 and λ2 were superposed in the optical cell containing the blood sample (Fig. 1 ). The transmitted light was detected separately for each wavelength behind the sample. Part of both laser beams was detected on a reference detector. The ratio of the laser power detected at λ1 and λ2 representing the glucose absorption was determined and transferred to data analysis.

Fig. 1

Scheme of the two-wavelength CO2 laser setup.

014021_1_033801jbo1.jpg

The optical cell consisted of two CaF2 plates with a spacer of 50μm . The cell was filled with the blood sample using capillary action. The blood volume required for the measurements was about 1μl .

3.

Results

3.1.

Fourier Transform Infrared Attenuated Total Reflection Measurements

The absorbance values of nine different concentrations of glucose in water in the region from 40 up to 400mgdl are presented in Fig. 2 . The typical glucose absorption bands are shown at 1080, 1036, 1059, and 980cm1 . A continuous increase in the absorbance could be detected with increasing glucose concentration. Sufficient signal could be detected even for a 40mgdl glucose solution. No dependence on glucose concentration could be observed at 950cm1 .

Fig. 2

Mid-infrared spectra for nine aqueous glucose concentrations between 40 and 400mgdl obtained by ATR-FTIR spectroscopy.

014021_1_033801jbo2.jpg

The relationship between the absorbance and the glucose concentration of the solvents is demonstrated for the three main bands at 1036, 1059, and 1080cm1 in Fig. 3 . All three absorption bands show a linear dependence and could be used for the determination of glucose concentration in water.

Fig. 3

Linear correlation of three glucose absorption bands obtained by ATR-FTIR spectroscopy with the glucose concentration.

014021_1_033801jbo3.jpg

The influence of temperature on glucose absorption was determined in the range from 15 to 28°C for the glucose band. The results have shown that the glucose bands from 1160 to 980cm1 are not significantly influenced by temperature changes.

When analyzing a blood sample containing 92mgdl glucose, it was found that the glucose spectrum changes from water to blood. If glucose was added to a whole blood sample, its spectrum changed with time and resulted in a constant spectrum after 2min . The spectra of glucose in pure water, and of glucose added to a whole blood sample, measured at different times are presented in Fig. 4 . The glucose band at 1034cm1 has its maximum intensity in water (curve a). In blood, this absorption decreases (curve b), and after 100s , the highest intensity was obtained at 1080cm1 (curve c). Furthermore, the ATR signals changed when the red blood cells started to sediment. To investigate this in more detail, a blood sample was left to sediment and different parts of the sample were measured. Figure 5 shows a spectrum of the pure supernatant, the plasma, which gave the lowest absorbance (curve a). The sediment, consisting of the red blood cells (Rbc), resulted in a spectrum with the highest absorbance (curve d). A mixture of plasma with 5% red blood cells was measured directly after application on the crystal (curve b), and 2min later (curve c). The measurement taken immediately after application showed a spectrum slightly higher than the plasma spectrum, whereas the one taken after 2min came close to that of the high-concentrated red blood cells.

Fig. 4

ATR-FTIR absorbance spectra for 92mgdl glucose in water (curve a), 92mgdl glucose added to whole blood and measured immediately after application (curve b), and glucose added to whole blood and measured after 100sec (curve c).

014021_1_033801jbo4.jpg

Fig. 5

ATR-FTIR absorbance spectra from one blood sample (glucose concentration 92mgdl ); plasma and red blood cells were separated. Spectra are shown from (a) pure plasma, (b) the same plasma with 5% red blood cells (Rbc) added and measured immediately after application onto the ATR crystal, (c) the same sample as (b) but measured 2min after application onto the ATR crystal, and (d) red blood cell fraction.

014021_1_033801jbo5.jpg

3.2.

Two-Wavelength CO2 Laser Measurements

The applicability of a two-wavelength laser for the quick determination of glucose was tested in blood. First of all, measurements were performed on aqueous glucose solutions and analogical to the ATR measurements, the glucose concentration exhibited a linear dependency (not shown here). Then whole blood samples were investigated at different glucose concentrations. As a result of the FTIR measurements, one laser wavelength was tuned to 1080cm1 where the maximum absorption for glucose was found in whole blood. The other wavelength was tuned to the isosbestic point at 950cm1 where no dependence on glucose concentration could be observed. Glucose concentrations of whole blood samples were determined in the Dept. of Clinical Chemistry. The samples were then analyzed using the two-wavelength CO2 laser system. The correlation between the quotient of the signal intensity of λ1 and λ2 , and the glucose concentration determined in the Dept. of Clinical Chemistry, is presented in Fig. 6 . A correlation factor of R2=0.997 was obtained.

Fig. 6

The quotient of the signal intensities at wavelengths 1080 and 950cm1 , obtained using the CO2 laser correlated with the glucose concentration in the range of 40 to 400mgdl .

014021_1_033801jbo6.jpg

4.

Discussion

In contrast to NIR measurements, the investigations in the MIR spectral range have shown that the intensity of glucose absorption bands around 1035cm1 does not change with temperature between 15 and 28°C . This indicates that the absorption can be correlated with the glucose concentration and could be used for glucose determination. The linear dependency of the ATR signal could be shown for selected wavelengths. The intensities of the absorption bands changed when water was substituted by blood plasma. The changes were due to changes in the glucose structure induced by the different surrounding media.15

The maximum glucose absorption was found in whole blood samples at 1080cm1 . The background was determined close to the glucose absorption band at the isosbestic point at 950cm1 . These two wavelengths could be used for the spectroscopic determination of glucose concentration in whole blood samples. The influence of water absorption could be neglected, if this background is subtracted from the glucose absorption.

In contrast to the CO2 laser measurements, the ATR measurements show a strong dependency on red blood cell sedimentation. A part of the glucose is located in the red blood cells. This is known from the HbA1c value, which provides information about blood glucose level over the last 4 to 6 weeks. Hemoglobin is connected to glucose, a relationship that is dependent on the concentration of glucose in blood plasma. The glucose connected to hemoglobin could render the blood glucose determination faulty. To use this method for routine measurements, a very strict time schedule would be necessary, and a dependency on the hematocrit is anticipated. This strong time dependency was not observed when using the CO2 laser. Nevertheless, a time dependency due to red blood cell sedimentation could also be expected for this method, albeit to a lesser extent. The influence on this method due to red blood cells is mainly related to the scattering behavior of red blood cells. But the scattering of the cells is similar for both of the applied wavelengths and can be corrected using a ratio of the two. As a result, no marked dependency on the hematocrit is to be expected. The difference in these signals is only correlated to the glucose absorption and not influenced by changes in laser radiation intensity.

Measurement of low glucose concentrations using the CO2 laser is faster than using the ATR, as this requires several scans before a reliable signal-to-noise ratio can be attained. The measurements can be carried out on a ms time scale by using small blood samples of less than 10μl . In contrast to methods in the literature that require several preparation steps, such as separation of plasma and blood cells, diluting or drying the sample, this method provides a one-step procedure by filling whole blood into the cuvette or capillary. To develop a routine device based on the two-wavelength CO2 laser method, interindividual influences should be tested using blood from different persons.

5.

Conclusion

The investigations using ATR-FTIR spectrometry show that glucose in water is linearly dependent on absorbance. The absorbance maximum at 1036cm1 decreases when glucose is added to whole blood. All intensities change and the absorption band at 1080cm1 is the most prominent absorption band in whole blood. The signal intensity at 950cm1 is not dependent on glucose, and can be used for background correction such as water and red blood cells. Based on these results, a fast and easy method could be established without any prior preparation of whole blood by using a two-wavelength CO2 laser, and tuning the laser to the wavelengths 1080 and 950cm1 . The laser method is not affected by red blood cell sedimentation, and can be used to measure the glucose concentration in the range of interest 40 to 400mgdl on a ms timescale.

Acknowledgments

The work was supported by the German Ministry of Education and Research (BMBF, Kennziffer: 13N7531).

References

1. 

W. Petrich and G. Werner, “Erkennung von krankheitsmustern im infrarotspektrum von blutseren,” Physikalische Blätter, 55 49 –51 (1999). Google Scholar

2. 

G. Budínová, J. Salva, and K. Volka, “Application of molecular spectroscopy in the mid-infrared region to the determination of glucose and cholesterol in whole blood and in blood serum,” Appl. Spectrosc., 51 631 –635 (1997). https://doi.org/10.1366/0003702971941034 0003-7028 Google Scholar

3. 

D. Rohleder, G. Kocherscheidt, K. Gerber, W. Kiefer, W. Kohler, J. Mocks, and W. Petrich, “Comparison of mid-infrared and Raman spectroscopy in the quantitative analysis of serum,” J. Biomed. Opt., 10 031108 (2005). https://doi.org/10.1117/1.1911847 1083-3668 Google Scholar

4. 

H. Zeller, P. Novak, and R. Landgraf, “Blood glucose measurement by infrared spectroscopy,” Int. J. Artif. Organs, 12 129 –135 (1989). 0391-3988 Google Scholar

5. 

R. J. Robinson and S. D. McDonald, “Glucose-sensitive membrane and infrared absorption spectroscopy for potential use as an implantable glucose sensor,” ASAIO J., 38 M458 –M462 (1992). https://doi.org/10.1097/00002480-199207000-00076 1058-2916 Google Scholar

6. 

Y. Mendelson, A. C. Clermont, R. A. Peura, and B. C. Lin, “Blood glucose measurement by multiple attenuated total reflection and infrared absorption spectroscopy,” IEEE Trans. Biomed. Eng., 37 458 –465 (1990). https://doi.org/10.1109/10.55636 0018-9294 Google Scholar

7. 

H. M. Heise, R. Marbach, T. H. Koschinsky, and F. A. Gries, “Multicomponent assay for blood substrates in human plasma by mid-infrared spectroscopy and evaluation for clinical analysis,” Appl. Spectrosc., 48 85 –95 (1994). https://doi.org/10.1366/0003702944027598 0003-7028 Google Scholar

8. 

H. Brauner and G. Müller, “Determination of glucose – a state of the art report,” Biomed. Tech., 25 26 –32 (1980). 0013-5585 Google Scholar

9. 

V. R. Kondepati and H. M. Heise, “Recent progress in analytical instrumentation for glycemic control in diabetic and critically ill patients,” Anal. Bioanal. Chem., 390 125 –139 (2008). https://doi.org/10.1007/s00216-007-1651-y 1618-2642 Google Scholar

10. 

H. M. Heise and P. U. Abel, “Clinical analysis: glucose,” Encyclopedia of Analytical Science, Elsevier Science, Amsterdam (2005). Google Scholar

11. 

B. Eppich and G. Müller, “Mehrbandenfilteranalysator zur bestimmung kleinster konzentrationsänderungen in mehrkomponenten-substanzgemischen,” Google Scholar

12. 

G. Müller, “Nicht invasive glucosemessung,” Google Scholar

13. 

D. Qi and A. J. Berger, “Chemical concentration measurement in blood serum and urine samples using liquid-core optical fiber Raman spectroscopy,” Appl. Opt., 46 (10), 1726 –1734 (2007). https://doi.org/10.1364/AO.46.001726 0003-6935 Google Scholar

14. 

R. Vonach, J. Buschmann, R. Falkowski, R. Schindler, B. Lendl, and R. Kellner, “Application of mid-infrared transmission spectroscopy to the direct determination of glucose in whole blood,” Appl. Spectrosc., 52 820 –822 (1998). https://doi.org/10.1366/0003702981944553 0003-7028 Google Scholar

15. 

A. Pintar, J. Batista, and J. Levec, “In situ Fourier transform infrared spectroscopy as an efficient tool for determination of reaction kinetics,” Analyst (Cambridge, U.K.), 127 (11), 1535 –1540 (2002). https://doi.org/10.1039/b207204a 0003-2654 Google Scholar
©(2008) Society of Photo-Optical Instrumentation Engineers (SPIE)
Martina C. Meinke, Gerhard J. Müller, Hansjörg Albrecht, Christina Antoniou, H. Richter, and Jürgen M. Lademann "Two-wavelength carbon dioxide laser application for in-vitro blood glucose measurements," Journal of Biomedical Optics 13(1), 014021 (1 January 2008). https://doi.org/10.1117/1.2870093
Published: 1 January 2008
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Glucose

Blood

Gas lasers

Absorption

Carbon dioxide lasers

Carbon monoxide

FT-IR spectroscopy

Back to Top