|
Optical coherence tomography (OCT) is a noninvasive, noncontact imaging modality that uses coherent gating to obtain high-resolution cross-sectional images of tissue microstructure.1 The 3-D microstructure image volume reveals more morphological information than 2-D images, which is critical in clinical applications such as early-stage cancer detection. 2, 3, 4, 5 Currently, a swept laser source for OCT can attain a scan rate higher than , which is suitable to record and analyze a 3-D image volume.6, 7, 8 To acquire a 3-D image volume, a fast 2-D image frame-rate, which is determined by the swept laser scan rate, the imaging endoscope rotational speed, and the data processing speed, is needed. Our swept laser scan rate was , which was sufficient to realize a video-rate 2-D image. The OCT imaging endoscope rotational speed is determined by the high spinning speed of the microelectromechanical (MEMS) motor. The MEMS motor-based endoscope design has several advantages.5, 9 It decouples the rotational torque transfer from the optical fiber to the image probe, thus eliminating the related polarization effect of the spinning image probe. With the small size of the MEMS motor, the whole endoscope package is small and flexible enough to be suitable for clinical applications. Compared to the previous one, the motor diameter decreased 30%, from to ; and the length decreased 27%, from to .5 Currently, the bottleneck of the entire system speed is the data processing speed, which is limited by the CPU’s processing power. Postprocessing is commonly used. A real-time technique enables onsite diagnosis and is preferable for endoscopic clinical examinations. Furthermore, a real-time technique has been demonstrated in time-domain OCT. One method is to accelerate signal processing based on hardware, including the digital autocorrelator, the field-programmable gate array (FPGA), and the digital signal processor (DSP).10, 11 An alternative method is to fully utilize the CPU’s processing power based on a multithreading software technique.12 In our study, the DSP board and the coupled digitizer were used for real-time data acquisition and image processing for the high-speed swept-source OCT. The digitizer streamed the data directly to the DSP board by a custom-designed auxiliary bus. The computer bus, which was the potential speed bottleneck for data streaming, was bypassed. Parallel data processing was applied with the DSP board. Specifically, each DSP processed the image frame independently. Although we demonstrated real-time processing with a scan rate of only , an estimated 90-kHz scan rate real-time processing can be achieved by combining a digitizer and a DSP board. A real-time image display of the in vivo human airway was achieved with a DSP and a MEMS motor-based endoscope. Our swept-source OCT system and MEMS motor endoscope are shown in Fig. 1 . The swept light source (center wavelength , FWHM bandwidth , scan rate; Santec Corporation, Komaki, Aichi, Japan) output was . The coupler split the output power. We coupled 80% of the output power into the sample arm and the remaining 20% into the reference arm. The reference power was attenuated by an adjustable neutral density attenuator to obtain maximum sensitivity. Two circulators were used in both the reference and sample arms to redirect the back-reflected light to a fiber coupler (50/50 split ratio) for balanced detection. The measured axial and lateral resolutions of our OCT system in the air were and , respectively. The image range depth was . The measured sensitivity of the OCT system with a MEMS motor endoscope was . The 6-dB sensitivity rolloff occured at . A software dispersion compensation algorithm was applied to cancel the dispersion generated by the endoscope optics.13 In the sample arm, the MEMS rotational motor-based endoscope was connected. Figure 1 shows the endoscope. The 1310-nm single-mode optical fiber, the 1.3-mm grin lens, the 1.5-mm MEMS motor, and the reflected mirror were packaged inside the biocompatible fluorinated ethylene-propylene (FEP) tube of the endoscope. The size of the driving coils, a key component of the MEMS motor, was reduced due to the improvement of manufacturing technology. Thus, the whole motor became smaller. The diameter was and the length was . A total of four wires was used to connect the motor and outside circuit. Two wires were used to supply the 3 volts of DC voltage. One wire transferred the encoded revolutions per minute (RPM) signal, and one wire controlled the rotational direction. The outside circuit implemented the feedback control by adjusting the supply voltage according to the encoded RPM signal, which increased the stability of the rotational speed. A two-stage, 40:1 ratio micro gear head was placed in front of the rotational shaft to smooth the step motion, which was designed to achieve an optimal rotational speed of 30 frames/second. In the experiment, the motor speed was set to 19.5 frames/second to synchronize with the OCT system speed. The outside linear transversal stage pulled the entire endoscope back at /sec to realize a 3-D helix scan. A total of 400 image slices were processed and displayed in 20.5 seconds. In the detection arm, the signal collected by the photodetectors was digitized by a 14-bit data acquisition board (PDA14, Data Acquisition Signal Waveform Digitizer, Signatec Inc., Newport Beach, California) sampling at samples/second, and the number of data points for each axial line (A-line) data acquisition was 1024. The A-line data acquisition start trigger for the digitizer was generated by the swept source. The digitizer outputted the A-line data directly to the DSP board (PMP1000, Parallel Digital Signal Processing Board, Signatec Inc., Newport Beach, California) through a custom-designed auxiliary bus. The complex analytical depth encoded signal was converted from the detected fringe signal by calibration, resampling, dispersion compensation, and fast Fourier transform. The structure image was reconstructed from the amplitude term of the complex depth encoded signal. In the DSP board, eight DSPs (C6414T, Texas Instruments Inc.) were programmed to work in parallel to implement the above algorithm. Figure 2a shows the data flow in the DSP board. The digitized data were fed sequentially into the DSP board from the digitizer. Each DSP handled a 1024 A-line image frame independently with the help of a scheduler. The resulting structure image was sent to the monitor in the same order as the input. The processing power from the DSPs was able to handle a 20-kHz A-line rate. In a simulation test, a square waveform of was set as the A-line trigger. An arbitrary waveform was acquired by the data acquisition board at samples/second. The 2048 A-line image frames were processed. The 43.9 frames/second image output rate was monitored. The test result indicated that a 90-kHz scan rate real-time processing could be achieved. Figure 2b shows the comparison performance test between the DSP and the Intel Xeon core duo CPU. The estimated performance from the DSP board was 3 times faster than the Dell Precision Workstation 690 with Intel 3-GHz Quad Core CPU in terms of fast Fourier transformation. The MEMS motor-based endoscope was introduced into the human airway under visual guidance from a rigid bronchoscope. The outer diameter of the endoscope was . Figure 3 shows the segmental human lower left lobe bronchus. In Fig. 3a, the layer-like structure from the 11 o’clock to 5 o’clock positions shows the posterior membrane. The motor wire blocked part of the signal and caused the dent artifact. Figure 3b shows the 3-D reconstruction image volume, which was rebuilt from 400 slices. The stepping distance between 2-D frames was . Figure 3c shows the reconstructed longitudinal image slice between the two arrows in Fig. 3b. The epithelium, lamina propria, and elastic lamina can be seen. Figure 4 shows segmental human distal trachea. In Fig. 4a, the notch between the trachea and the posterior membrane can be seen. Some parts of the trachea were outside the OCT image range. A mask was applied to the image to suppress the FEP tube mirror artifact, which generated a constant radial line at a depth of around beneath the surface. From the 3-D reconstruction in Fig. 4b, the notch is clearer. The longitudinal length of the image volume was . The stepping distance between the 2-D frames was . Figure 4c shows the reconstructed longitudinal image slice between the two arrows in Fig. 4b; the cartilage arc can be seen in the middle. In conclusion, a real-time swept source OCT system with a 20-kHz A-line speed was demonstrated by using multiple DSPs. An in vivo human airway 3-D image volume was acquired by a MEMS motor-based endoscope. The estimated performance from the pair of digitizers and the DSP board could support the swept laser with a 90-kHz scan rate. AcknowledgmentsThis work was supported by research grants from the National Institutes of Health (EB-00293, CA-91717, and RR-01192), the Air Force Office of Scientific Research (FA9550-04-1-0101), and the Beckman Laser Institute Endowment. ReferencesD. Huang,
E. A. Swanson,
C. P. Lin,
J. S. Schuman,
W. G. Stinson,
W. Chang,
M. R. Hee,
T. Flotte,
K. Gregory,
C. A. Puliafito, and
J. G. Fujimoto,
“Optical coherence tomography,”
Science, 254 1178
–1181
(1991). https://doi.org/10.1126/science.1957169 0036-8075 Google Scholar
M. Wojtkowski,
V. Srinivasan,
J. G. Fujimoto,
T. Ko,
J. S. Schuman,
A. Kowalczyk, and
J. S. Duker,
“Three-dimensional retinal imaging with high-speed ultrahigh-resolution optical coherence tomography,”
Ophthalmology, 112 1734
–1746
(2005). https://doi.org/10.1016/j.ophtha.2005.05.023 0161-6420 Google Scholar
S. H. Yun,
G. J. Tearney,
B. J. Vakoc,
M. Shishkov,
W. Y. Oh,
A. E. Desjardins,
M. J. Suter,
R. C. Chan,
J. A. Evans,
I.-K. Jang,
N. S. Nishioka,
J. F. de Boer, and
B. E. Bouma,
“Comprehensive volumetric optical microscopy in vivo,”
Nat. Med., 12 1429
–1433
(2006). 1078-8956 Google Scholar
D. C. Adler,
Y. Chen,
R. Huber,
J. Schmitt,
J. Connolly, and
J. G. Fujimoto,
“Three-dimensional endomicroscopy using optical coherence tomography,”
Nat. Photonics, 1 709
–716
(2007). 1749-4885 Google Scholar
J. Su,
J. Zhang,
L. Yu, and
Z. Chen,
“In vivo three-dimensional microelectromechanical endoscopic swept source optical coherence tomography,”
Opt. Express, 15 10390
–10396
(2007). https://doi.org/10.1364/OE.15.010390 1094-4087 Google Scholar
W. Y. Oh,
S. H. Yun,
G. J. Tearney, and
B. E. Bouma,
“ tuning repetition rate ultrahigh-speed wavelength-swept semiconductor laser,”
Opt. Lett., 30 3159
–3161
(2005). https://doi.org/10.1364/OL.30.003159 0146-9592 Google Scholar
R. Huber,
M. Wojtkowski, and
J. G. Fujimoto,
“Fourier domain mode locking (FDML): A new laser operating regime and applications for optical coherence tomography,”
Opt. Express, 14 3225
–3237
(2006). https://doi.org/10.1364/OE.14.003225 1094-4087 Google Scholar
R. Huber,
D. C. Adler, and
J. G. Fujimoto,
“Buffered Fourier domain mode locking: unidirectional swept laser sources for optical coherence tomography imaging at ,”
Opt. Lett., 31 2975
–2977
(2006). https://doi.org/10.1364/OL.31.002975 0146-9592 Google Scholar
P. H. Tran,
D. S. Mukai,
M. Brenner, and
Z. Chen,
“In vivo endoscopic optical coherence tomography by use of a rotational microelectromechanical system probe,”
Opt. Lett., 29 1236
–1238
(2004). https://doi.org/10.1364/OL.29.001236 0146-9592 Google Scholar
V. Westphal,
S. Yazdanfar,
A. M. Rollins, and
J. A. Izatt,
“Real-time, high velocity-resolution color Doppler optical coherence tomography,”
Opt. Lett., 27 34
–36
(2002). https://doi.org/10.1364/OL.27.000034 0146-9592 Google Scholar
S. Yan,
D. Piao,
Y. Chen, and
Q. Zhu,
“Digital signal processor-based real-time optical Doppler tomography system,”
J. Biomed. Opt., 9 454
–463
(2004). https://doi.org/10.1117/1.1695409 1083-3668 Google Scholar
B. Hyle Park,
M. C. Pierce,
B. Cense, and
J. F. de Boer,
“Real-time multi-functional optical coherence tomography,”
Opt. Express, 11 782
–793
(2003). 1094-4087 Google Scholar
B. Cense,
N. Nassif,
T. Chen,
M. Pierce,
S.-H. Yun,
B. Park, B. Bouma,
G. Tearney, and
J. de Boer,
“Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography,”
Opt. Express, 12 2435
–2447
(2004). https://doi.org/10.1364/OPEX.12.002435 1094-4087 Google Scholar
|