1 July 2009 Broadband diffuse optical spectroscopy assessment of hemorrhage- and hemoglobin-based blood substitute resuscitation
Jangwoen Lee, Jae Gwan Kim, Sari B. Mahon, Bruce Jason Tromberg, David Sho Mukai, Kelly A. Kreuter, Darin J. Saltzman, Renee Patino M.D., Robert Goldberg M.D., Matthew Brenner M.D.
Author Affiliations +
Abstract
Hemoglobin-based oxygen carriers (HBOCs) are solutions of cell-free hemoglobin (Hb) that have been developed for replacement or augmentation of blood transfusion. It is important to monitor in vivo tissue hemoglobin content, total tissue hemoglobin [THb], oxy- and deoxy-hemoglobin concentrations ([OHb], [RHb]), and tissue oxygen saturation (StO2=[OHb]/[THb]×100%) to evaluate effectiveness of HBOC transfusion. We designed and constructed a broadband diffuse optical spectroscopy (DOS) prototype system to measure bulk tissue absorption and scattering spectra between 650 and 1000 nm capable of accurately determining these tissue hemoglobin component concentrations in vivo. Our purpose was to assess the feasibility of using DOS to optically monitor tissue [OHb], [RHb], StO2, and total tissue hemoglobin concentration ([THb]=[OHb]+[RHb]) during HBOC infusion using a rabbit hypovolemic shock model. The DOS prototype probe was placed on the shaved inner thigh muscle of the hind leg to assess concentrations of [OHb], [RHb], [THb], as well as StO2. Hemorrhagic shock was induced in intubated New Zealand white rabbits (N=6) by withdrawing blood via a femoral arterial line to 20% blood loss (10-15 cc/kg). Hemoglobin glutamer-200 (Hb-200) 1:1 volume resuscitation was administered following the hemorrhage. These values were compared against traditional invasive measurements, serum hemoglobin concentration (sHGB), systemic blood pressure, heart rate, and blood gases. DOS revealed increases of [THb], [OHb], and tissue hemoglobin oxygen saturation after Hb-200 infusion, while blood total hemoglobin values continued did not increase; we speculate, due to hyperosmolality induced hemodilution. DOS enables noninvasive in vivo monitoring of tissue hemoglobin and oxygenation parameters during shock and volume expansion with HBOC and potentially enables the assessment of efficacy of resuscitation efforts using artificial blood substitutes.

1.

INTRODUCTION

Hemorrhage remains a leading cause of death in civilian and military trauma patients, and a major cause of morbidity and mortality in patients with medical illnesses, vascular events, and patients undergoing surgery. Limitations, availability, costs, infection risks, logistical issues, and complications of blood transfusions for resuscitation of hemorrhage have led to extensive efforts to develop safe and effective blood substitutes.1 A number of blood substitute approaches have been pursued, including development of an approved commercially available polymerized hemoglobin-based veterinary blood oxygen carrier (HBOC) product, hemoglobin glutamer-200 (Hb-200, Oxyglobin®, Biopure Corporation, Cambridge, Massachusetts). 1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14

The goals of resuscitation from severe hemorrhage include restoration of hemodynamic stability and tissue oxygen delivery.15 Standard methods for assessment of adequacy of resuscitation during hemorrhage, including hematocrit, serum hemoglobin concentration, cardiac output, blood pressure, and pulse rate, are insensitive and nonspecific due to dynamic physiologic and reflex responses of the hemorrhaging patient.16, 17, 18, 19 These difficulties in assessing adequacy of hemorrhage resuscitation are further compounded during blood substitute resuscitation due to interaction of complex factors, including differences between blood and cell-free blood substitutes.1, 10, 13 Some of the factors that may differentially affect tissue perfusion and oxygenation during hemoglobin-based blood substitute treatment include polymerized heme structures, heterogeneous polymer size distribution, NO scavenging, gas transfer from cell stromal free composition, and osmolality effects.1, 6, 12, 13, 20, 21

Recently developed noninvasive diffuse optical spectroscopy (DOS) technology has been demonstrated to have the potential to noninvasively assess the physiologic changes associated with hemorrhage-induced reductions in tissue perfusion and the ability to indicate the degree of peripheral tissue hemoglobin oxygenation.15, 22, 23, 24, 25, 26, 27 DOS is a near-infrared (NIR) optical spectroscopic approach that can measure complex tissue-scattering induced light signal attenuation and use this information in the determination of broadband NIR absorption spectra in tissues, resulting in improved quantification due to its ability to measure absolute tissue chromophore concentrations compared to standard NIR methods, particularly in conditions where tissue scattering properties are changing dynamically, such as hemorrhage. This information can then be translated into quantitative measurements of the NIR absorbing components of the tissue including [OHb], [RHb], [THb], and tissue oxygen saturation (StO2=[OHb][THb]×100%) . Because absorption spectra of the oxygenated and deoxygenated Hb-200 are nearly identical to those of OHb and RHb,2, 3, 4, 5 we proposed that noninvasive DOS should be capable of assessing the tissue response to hemorrhage and hemoglobin-based blood substitute resuscitation.

In this study, we investigate the ability of DOS to quantitatively detect changes in peripheral tissue perfusion occurring in vivo during hemorrhage and hemoglobin-based blood substitute resuscitation. We compare noninvasive DOS measures of [THb], StO2 , [OHb], and [RHb] to standard hemorrhage assessment measures of heart rate, arterial pressures, and stroke volume during graded hemorrhage and blood substitute resuscitation to determine the ability of DOS to assess resuscitation response at the target tissue level. The major purpose of this study is to exemplarily demonstrate how DOS technology can assess tissue oxygenation restoration during HBOC infusion.

2.

Materials and Methods

The study was approved by the Institutional Laboratory Animal Care and Use Committee, University of California, Irvine, and by the U.S. Air Force Animal Research programs, and performed in accordance with Federal, Department of Defense, and State Regulations.

The methods for animal preparation, anesthesia, monitoring, and DOS measurements have been previously described15, 26, 28 and are summarized below.

2.1.

Anesthesia and Intubation

Male New Zealand white rabbits (N=6) (Western Oregon Rabbit Supply, Philomath, OR) weighing 35kg (median 3.9kg , mean 3.9kg , standard error 0.166 and range 0.97kg ) were anesthetized with 2:1 ratio of ketamine HC1 (100mgmL) (Ketaject, Phoenix Pharmaceutical Inc., St. Joseph, Michigan):Xylazine (20mgmL) (Anased, Lloyed Laboratories, Shenandoa, Iowa) at a dose of 0.75mLkg IM. After the IM injection a 22–24 gauge 1-in. catheter was placed in the animal’s marginal ear vein to administer intraverious (IV) anesthesia and secured with 1-in. standard porous adhesive tape. Maintenance anesthetic was dosed at 0.3mL of 1:1 mixture of ketamine:xylazine IV (ketamine 100mgmL :xylazine 20mgmL ) or continuous infusion of the 20cc of the 1:1 solution diluted to 50cc total in NS infused at 0.17ccmin as needed. Depth of anesthesia was monitored according to established guidelines. Animals were intubated with a 3.0 endotracheal tube and mechanically ventilated (Harvard Apparatus dual phase control respirator, Holliston, Massachusetts) at respiration rate of 32min and a tidal volume of 50cc and FiO2 of 100%. Pulse oximetry was accomplished with a probe placed on the tongue to measure SaO2 (Biox 3700 Pulse Oximeter, Ohmeda, Boulder, Colorado) and compared to arterial blood gas measurements.

2.2.

Vascular Access, Blood Pressure, Blood Gasses

Left femoral arterial and femoral venous central lines (Cook 4.0 Fr, 18 Ga femoral artery catheter set, Cook, Inc., Bloomington, Indiana) were placed for arterial blood draws and systemic pressures. Pressures were obtained using a calibrated pressure transducer (TSD104A transducer and MP100 WSW System, Biopac Systems, Inc, Santa Barbara, California) and stored digitally. Mean, systolic, and diastolic pressures were determined from 510s tracings. After each blood draw, lines were flushed with <0.5cc of heparin (Elkins-Sinn, Inc., Cherry Hill, New Jersey) to prevent line thrombus occlusion. Arterial blood samples were measured with a blood gas analyzer calibrated to rabbit blood (IRMA Series 2000 Blood Analysis System, Diametrics Medical Inc., St Paul, Minnesota). Central venous blood samples were drawn from the femoral vein. Complete blood counts (CBC) were obtained from collected samples of venous blood and sent to an outside facility (Antech Diagnostics, Irvine, California) for analysis.

2.3.

Noninvasive Measurements (Broadband Diffuse Optical Spectroscopy)

Detailed analysis of broadband DOS and the prototype system we constructed in our laboratory has been described previously.23, 29, 30, 31 A multiwavelength, frequency domain (FD) photon migration instrument32, 33 was combined with a steady-state (SS) NIR spectrometer for the noninvasive in vivo measurement of tissue chromophore concentration. A plastic probe containing the source and detector fibers was placed on the anterior medial surface of the right hind thigh for broadband DOS measurement. The source and detector separation of 10mm is used for both FD and SS acquisitions. The broadband DOS prototype we used employs five laser diodes (661, 681, 783, 805, and 850nm ) and a fiber-coupled avalanche photodiode (APD) detector (Hamamatsu high-speed APD module C5658, Bridgewater, New Jersey). The APD detects the intensity-modulated diffuse reflectance signal at modulation frequencies between 50 and 300MHz after propagating through the tissue. The absorption and reduced scattering coefficients are measured directly at each of the six laser diode wavelengths using the frequency-dependent phase and amplitude data. The reduced scattering coefficient is calculated throughout the NIR by fitting a power law to these five reduced scattering coefficients ( μs=aλSP , where λ is wavelength, a is prefactor, and SP is scattering power).34, 35, 36 The acquired reduced scattering spectrum characterizes the scattering properties of tissue throughout the NIR. The SS acquisition is a broadband NIR reflectance measurement from 650to1000nm that follows the FD measurements using a tungsten-halogen light source (Ocean Optics HL-2000, Dunedin, Florida) and a miniature spectrometer (BWTEK BTC611E, Newark, Delaware). It takes 30s to make a broadband DOS measurement. The intensity of the SS reflectance measurements are calibrated to the FD values of absorption and scattering to establish the absolute reflectance intensity. The absolute SS reflectance spectra are then analyzed to calculate tissue absorption coefficient (μa) spectra. Finally, the tissue concentrations of OHb, RHb, and H2O are calculated by a linear least-squares fit of the wavelength-dependent extinction coefficient spectra of each chromophore. The concentration calculation was conducted after each experiment using real-time data acquired during the study.

2.4.

Experimental Procedures

A three-step, 20% hemorrhage model was used for these studies.15, 26

After animal preparation, anesthesia, and intubation, baseline physiologic variable measurements were obtained, along with noninvasive DOS assessment of THb, OHb, and RHb. A 20% total hemorrhage goal was calculated (15cckg) . One-third of the calculated total hemorrhage volume was removed by withdrawal from the femoral arterial line over 3060s . Repeat physiologic and DOS measurements were obtained following the blood removal steps. The blood drawing and measurement process was repeated at 5 and 10min until the 20% blood-loss goal was achieved after the third hemorrhage. Measurements took approximately 1030s to obtain.

After completion of the hemorrhage phase measurements, the animals were begun on a slow continuous infusion of Hb-200, Oxyglobin at a rate of 0.5ccmin . Oxyglobin is a solution of 13gdL of purified, polymerized bovine hemoglobin diluted in a modified Lactated Ringers solution. The Hb-200 infusion was continued until an infused volume equal to the total amount of blood that had been previously withdrawn was attained. Physiologic and DOS measurements were performed at one-third, two-thirds, and completion of the Oxyglobin infusion.

At completion of the experiment, each animal was euthanized with an intravenous injection of Eutha-6 (1.02.0mL) through the marginal ear-vein catheter according to animal laboratory guidelines and the approved IACUC protocol.

2.4.1.

Statistical analysis

One-way repeated measures analysis of variance (ANOVA) was used for comparison of outcome variables using SYSTAT® version 10 (Systat Software, Inc., Point Richmond, CA). All data are presented as mean±standard error (SE), and statistical significance were inferred when p<0.05 .

3.

Results

Table 1 lists the concentrations of broadband DOS OHb, RHb, THb, and StO2 as well as sHgb during each hemorrhage step and Oxyglobin resuscitation step.

Table 1

Broadband DOS OHb, RHb, THb concentrations, and tissue oxygen saturation (StO2) and sHgb during hemorrhage and Oxyglobin resuscitation. Data are from six animals and are presented as mean±standard error.

OHb (mM)RHb (mM)THb (mM) StO2 (%)sHgb (g/dL)
Baseline 35.3±6.5 18.9±2.6 54.2±8.5 63.4±4.2 14.8±0.4
Hem 1 31.4±5.5 19.2±2.7 50.6±7.7 60.8±3.6 13.9±0.4
Hem 2 28.4±5.4 20.0±2.7 48.4±7.4 56.9±4.4 13.0±0.4
Hem 3 25.0±4.7 20.1±2.8 45.1±6.7 53.7±4.8 12.1±0.4
Res 1 26.6±4.7 20.6±3.2 47.0±7.3 55.3±4.1 12.1±0.5
Res 2 28.3±4.8 20.5±3.7 48.8±8.0 57.3±4.0 11.9±0.4
Res 3 ±29.0±4.8 21.4±3.9 50.4±8.1 56.9±4.0 11.9±0.5

Figure 1 illustrates the hemoglobin concentrations ([OHb], [RHb], and [THb]=([OHb]+[RHb]) and StO2 during hemorrhage steps and continuous infusion of Oxyglobin as resuscitation fluid in a single animal. In this example, broadband DOS acquired the measurement every 10s during the entire procedure.

Fig. 1

Continuous monitoring of tissue hemoglobin concentrations and StO2 using noninvasive DOS. DOS measurements were taken every 10s . Data shown above are from a single rabbit.

044027_1_029904jbo1.jpg

Figure 2 depicts the absolute concentrations of RHb, OHb, and tissue O2 saturation (StO2) during hemorrhage and resuscitation using Oxyglobin. [RHb] stayed relatively constant throughout hemorrhage and resuscitation phases ( p=0.207 , repeated measures ANOVA with all seven steps). [OHb] decreased significantly within subjects after each hemorrhage step (p<0.001) and increased during Oxyglobin infusion (p<0.001) . StO2 decreased during the hemorrhage phase (p<0.001) , but changes in StO2 were not statistically significant during the resuscitation (p=0.147) , although the trend was similar to that of the increase in [OHb]. Venous oxygen saturation (SvO2) values from blood gas at baseline, the completion of hemorrhage and after Oxyglobin resuscitation are also plotted in Fig. 2, and they showed a similar trend. SvO2 significantly dropped during hemorrhage phase from the baseline value (21%) and recovered back to the baseline after Oxyglobin infusion (p<0.05) .

Fig. 2

DOS RHb and OHb concentrations and StO2 during hemorrhage and resuscitation using Oxyglobin. Venous oxygen saturation values at baseline, at the end of hemorrhage, and at the end of Oxyglobin resuscitation are also noted. Error bars represent standard error. Data are from six animals.

044027_1_029904jbo2.jpg

Figure 3 shows the normalized changes in DOS THb and sHgb during and hemorrhage steps and resuscitation with continuous infusion of Hb-200, Oxyglobin. Both DOS THb and sHgb were normalized by their respective baseline concentrations and fractional changes are plotted. Both DOS THb and sHgb decreased during hemorrhage phase (p<0.001) . In order to determine the effect of Oxyglobin infusion on DOS THb and sHgb, repeated measures ANOVA were performed using the Hem3 time points (completion of hemorrhage) as the posthemorrhage baseline values. Changes in sHgb were insignificant (p=0.92) , while DOS THb increased significantly after Oxyglobin infusion (p=0.002) .

Fig. 3

Fractional changes of DOS THb and sHgb from baseline during hemorrhage and resuscitation using Oxyglobin. [RHb] stayed relatively constant throughout hemorrhage and resuscitation phases ( p=0.207 , repeated measures ANOVA with all seven steps). [OHb] decreased significantly within subjects after each hemorrhage step (p<0.001) and increased during Oxyglobin infusion (p<0.001) . StO2 decreased during the hemorrhage phase (p<0.001) , but changes in StO2 were not statistically significant during the resuscitation (p=0.147) , although the trend was similar to that of the increase in [OHb]. Error bars represent standard error. Data are from six animals.

044027_1_029904jbo3.jpg

In Fig. 4 scattering changes were assessed during hemorrhage and Oxyglobin resuscitation. SP is determined by obtaining a power law fit of five individual μs at five laser diode wavelengths ( μs=aλSP , where λ is wavelength, a is prefactor). Scattering properties at the five discreet wavelengths sampled, as well as the power-fit SP showed demonstrated substantial changes during hemorrhage and blood substitute resuscitation.

Fig. 4

Example of changes in reduced scattering coefficient, μs (mm1) during hemorrhage and Oxyglobin resuscitation. SP stands for scattering power of a power law fit of five individual μs at five laser diode wavelengths ( μs=aλSP , where λ is wavelength and a is a prefactor) in a representative rabbit. These data show the importance of accounting for the substantial scattering changes occurring with hemorrhage and blood substitute resuscitation when utilizing near-infrared spectroscopic methods.

044027_1_029904jbo4.jpg

4.

Discussion

Improvements in diagnosis, monitoring, and optimization of treatment for patients with hemorrhage are essential for advances in medical, surgical, and trauma care. Advances in resuscitation products, including development of artificial blood substitutes and methods for assessing the adequacy of resuscitation at the target tissue level, are needed. Early changes that have been shown to occur with reduction in central blood volume during hemorrhage include reduction in cardiac output and decreased delivery of oxygen in the blood to peripheral tissues.18, 19, 37, 38

In this study, as in previous studies, we have demonstrated the ability of noninvasive DOS to detect changes in tissue perfusion resulting from reduction in blood volume induced by progressive hemorrhage.15, 26, 27 The ability of DOS to quantitatively determine the extent of scattering-induced light attenuation provides for the ability to accurately obtain measurements of [THb], [OHb], [RHb], as well as tissue hemoglobin oxygen saturation under conditions of changing scattering properties as experienced during hemorrhage and resuscitation.15, 23, 24, 26, 27 Because DOS measures the average concentration of these components over the region of interrogation, the contributions of the arterioles, capillaries, and venules are included in the composite average tissue hemoglobin component concentration measurements.

As previously shown,15, 26, 27 during blood loss from hemorrhage, the [THb] measured by DOS decreases, likely due to a combination of vasoconstriction and decrease in intravascular hemoglobin (as a fraction of intravascular volume).15, 27 During hemorrhage, the tissue [OHb] decreases, probably from a combination of physiologic factors including to vasoconstriction, decreased intravascular hemoglobin fraction, and increased conversion of OHb to RHb by the underperfused tissues.15 In contrast to decreasing [THb] and [OHb] during hemorrhage, tissue [RHb] levels do not change substantially during hemorrhage. This difference in response15 likely results in the decreased [RHb] from vasoconstriction, and decreased hematocrit are offset to varying degrees by increased [RHb] resulting from increased conversion of [OHb] to [RHb] at the tissue level.15 During resuscitation with whole blood, these hemorrhage-induced changes are reversed, as demonstrated in prior studies by our group15 using DOS and other groups using invasive methods.39, 40

In the present study, a number of differences in DOS-measured responses are seen during blood substitute resuscitation compared to the previously demonstrated whole blood resuscitation effects. During the resuscitation phase, the total tissue hemoglobin measured by DOS increased toward baseline values with administration of hemoglobin-based blood substitute. This finding was expected due to the increased intravascular volume associated with blood substitute resuscitation, as well as the added hemoglobin content within the blood substitute.10, 39 One might have predicted some blunting of this increase in total tissue hemoglobin concentration due to a possible component of NO scavenging-induced vasoconstriction reported with poly-heme-based blood substitutes, such as Oxyglobin, and other potential mechanisms leading to increased vascular resistance.1, 20, 21, 39, 41, 42, 43, 44, 45 However, such effects, if present, did not appear to substantially alter the increase in DOS-measured total tissue hemoglobin, suggesting that increased vascular volume predominates during blood substitute resuscitation as administered. These findings are consistent with findings from invasive hemodynamic monitoring studies in animals during HBOC resuscitation.40

In marked contrast to the DOS measured [THb], concurrently measured serum hemoglobin concentrations obtained from invasive blood samples did not rise during blood substitute resuscitation. This is not surprising given the fact that the HBOC, Oxyglobin, has a hemoglobin concentration of 13gdL , an osmolality of 300, and a colloid osmotic pressure of 43mmHg .1 Administration of replacement fluid with this combination of hyper-osmolality and a 13gdL hemoglobin concentration, which is only slightly above that of the animals following acute hemorrhage (intravascular post hemorrhage hemoglobin concentration, averaged 12.1±0.4gdL compared to prehemorrhage baseline values of 14.8±0.4gdL ), would be expected to draw fluid into the vascular space resulting in substantial increases in intravascular volume, but should cause only slight (if any) increases in intravascular hemoglobin concentration.6, 7, 8, 20

These findings are consistent with prior studies where Oxyglobin infusion has been shown to restore hemodynamic parameters, including cardiac output and mesenteric arterial blood flow following hemorrhage, with increases in mean arterial blood pressure above baseline, but not to restore total intravascular hemoglobin and arterial blood oxygen content to baseline.20 The hyperoncotic properties of Oxyglobin result in rapid volume expansion with homogeneous microvascular perfusion and the ability to facilitate diffusive oxygen transfer, accelerating metabolic recovery.20 This would explain the discrepancy between DOS [THb] measurements (which include both intravascular hemoglobin concentration and tissue vascular volume) versus invasively obtained blood sample hemoglobin concentrations (which include only the intravascular hemoglobin concentration component) during blood substitute resuscitation. We propose that the DOS measurements therefore may reflect a more accurate measure of adequacy of tissue perfusion than serum hemoglobin concentrations alone10 under these conditions.

DOS measurements of tissue [OHb] demonstrate a reversal of the hemorrhage-induced tissue OHb reductions during resuscitation. These findings are paralleled by an improvement in measured venous blood oxygenation obtained by invasive blood gas analysis and suggest the ability to detect reversal of the increased oxygen extraction accompanied by increased conversion of OHb to RHb that had been induced during hypoperfusion. Using invasive muscle blood flow and muscle oxygen partial pressure measurements, Dreissen 39 and Cheung 40 had shown similar improvements in muscle tissue perfusion and oxygenation with Oxyglobin HBOC resuscitation following hemorrhage. The degree of reversal appears to be dependent on HBOC concentrations achieved during the resuscitation process.39, 40

Substantial scattering changes occur in vivo during hemorrhage and resuscitation26 as a result of vascular changes, including vasoconstriction, vasodilation, and concentration of scatter-inducing red blood cells within the vasculature. During blood substitute resuscitation, the vasoactive effects of HBOCs, volume loading effects on the vasculature, as well as the absence of scattering cellular stroma all may possibly contribute to complex scattering alterations in bulk tissue during the blood substitute resuscitation phase. These scattering changes result in inaccuracies in standard near-infrared spectroscopy hemoglobin measurements that are overcome using DOS technology.

Stroma-free hemoglobin solutions have also been susceptible to methemoglobin (Met-Hgb) formation.20 Because of the marked spectral shifts in Met-Hgb, DOS is very sensitive in quantitatively detecting Met-Hgb levels in vivo,25 providing another potential advantage of DOS in monitoring blood substitute resuscitation. In the current studies, no significant increases in DOS-measured Met-Hgb levels were seen during short term follow-up of acute resuscitation with Oxyglobin HBOC.

There are a number of limitations in this study. DOS measures average tissue constituents to a depth of 5mm (at the source detector separation of 10mm used in the studies). We have chosen to monitor peripheral muscle region as an indicator of central organ blood flow as suggested in previous studies.15, 17, 18, 26 However, the DOS system that we employed uses a single-source detector separation that has the disadvantage of including superficial skin and fat regions in the interrogation field, which may behave differently during hemorrhage and resuscitation compared to the deeper muscle tissues.17, 18, 46

In addition, the current system measures at only one peripheral site. Regional variability, local traumatic injury, ischemia or hemorrhage at the site of measurement, or specific organ dysfunction would not be detected with the current design. Future planned designs of DOS device prototypes with collection in parallel from a number of sites and source detector separations could improve the quality and practicality of multisite data collection.

Within the DOS hemoglobin signals, there may be some component of interference by other NIR absorbers, such as tissue myoglobin. Although this may be of theoretical interest, from a practical standpoint it may not be necessary to fully separate these component contributions when examining tissue hemoglobin and hemoglobin oxygenation clinical effects of hemorrhage because they may be all be collective indicators of adequacy of perfusion.

We did not perform a specific set of control animal experiments for this particular study because previous studies from our laboratory have shown that Hb-O2, Hb-R, [THb] concentrations, and StO2 remain within 5% of baseline values at comparable time points. Likewise, sHgb, SaO2 , SvO2 , CO, and mean arterial pressure remained all within 6% of baseline values at the end of the previously reported study period, with the minimal changes seen being attributed to exposure to anesthesia effects and blood withdrawals for blood gas and sHgb analysis.26

5.

Conclusions

Although there are a number of limitations of the current study, these promising initial findings suggest that noninvasive DOS measurements may be very useful for detection and optimization of the effects of blood substitute resuscitation on peripheral perfusion. Assessment of effectiveness of blood substitute resuscitation by conventional methods is currently limited because of the complexities of reflex responses to acute hemorrhage, discrepancies between volume of blood loss and changes in serum hemoglobin, as well as differences between stroma-free hemoglobin-based blood substitutes compared to red cell–based blood transfusions. This study demonstrates that DOS measurements may overcome some of these issues by providing information on the adequacy of restoration of tissue hemoglobin concentration and oxygenation. Future studies will be needed to determine the role of DOS in assessing the effects of hemorrhage and blood substitute resuscitation in humans.

Acknowledgments

This research was supported by funding from Grants No. AF-9550-04-1-0101, No. FOS-2004-0011A, and No. FOS-2004-0012A, and LAMMP No. 445474-30136. Blood substitute Hemoglobin glutamer-200 (Oxyglobin) was kindly supplied by Biopure, Inc, Cambridge, MA.

References

1. 

B. Driessen, J. S. Jahr, F. Lurie, and R. A. Gunther, “Inadequacy of low-volume resuscitation with hemoglobin-based oxygen carrier hemoglobin glutamer-200 (bovine) in canine hypovolemia,” J. Vet. Pharmacol. Ther., 24 (1), 61 –71 (2001). https://doi.org/10.1046/j.1365-2885.2001.00307.x 0140-7783 Google Scholar

2. 

J. S. Jahr, F. Lurie, B. Driessen, J. A. Davis, R. Gosselin, and R. A. Gunther, “The HemoCue, a point of care B-hemoglobin photometer, measures hemoglobin concentrations accurately when mixed in vitro with canine plasma and three hemoglobin-based oxygen carriers (HBOC),” Can. J. Anaesth., 49 (3), 243 –248 (2002). https://doi.org/10.1007/BF03020522 0832-610X Google Scholar

3. 

J. S. Jahr, B. Driessen, F. Lurie, Z. Tang, R. F. Louie, and G. Kost, “Oxygen saturation measurements in canine blood containing hemoglobin glutamer-200 (Bovine): in vitro validation of the NOVA CO-Oximeter,” Vet. Clin. Pathol., 30 (1), 39 –45 (2001). https://doi.org/10.1111/j.1939-165X.2001.tb00255.x Google Scholar

4. 

F. Lurie, J. S. Jahr, and B. Driessen, “The novel HemoCu plasma/low hemoglobin system accurately measures small concentrations of three different hemoglobin-based oxygen carriers in plasma: hemoglobin glutamer-200 (bovine) (Oxyglobin), hemoglobin glutamer-250 (bovine) (Hemopure), and hemoglobin-Raffimer (Hemolink),” Anesth. Analg. (Baltimore), 95 (4), 870 –873 (2002). https://doi.org/10.1097/00000539-200210000-00014 0003-2999 Google Scholar

5. 

J. S. Jahr, F. Lurie, B. Driessen, Z. Tang, R. F. Louie, and G. Kost, “Validation of oxygen saturation measurements in a canine model of hemoglobin-based oxygen carrier infusion,” Am. J. Ther., 10 (1), 21 –28 (2003). https://doi.org/10.1097/00045391-200301000-00006 Google Scholar

6. 

B. Driessen, J. S. Jahr, F. Lurie, M. S. Golkaryeh, and R. A. Gunther, “Arterial oxygenation and oxygen delivery after hemoglobin-based oxygen carrier infusion in canine hypovolemic shock: a dose-response study,” Crit. Care Med., 31 (6), 1771 –1779 (2003). https://doi.org/10.1097/01.CCM.0000063476.79749.C1 0090-3493 Google Scholar

7. 

L. P. Posner, P. F. Moon, S. P. Bliss, R. D. Gleed, and H. N. Erb, “Colloid osmotic pressure after hemorrhage and replenishment with Oxyglobin Solution, hetastarch, or whole blood in pregnant sheep,” Vet. Anaesth. Analg., 30 (1), 30 –36 (2003). https://doi.org/10.1046/j.1467-2995.2003.00088.x Google Scholar

8. 

M. B. Callan and V. T. Rentko, “Clinical application of a hemoglobin-based oxygen-carrying solution,” Vet. Clin. North Am. Small Anim. Pract., 33 (6), 1277 –1293 (2003). https://doi.org/10.1016/S0195-5616(03)00119-0 0195-5616 Google Scholar

9. 

M. Lichtenberger, “Transfusion medicine in exotic pets,” Clin. Tech. Small Anim. Pract., 19 (2), 88 –95 (2004). https://doi.org/10.1053/j.ctsap.2004.01.006 1096-2867 Google Scholar

10. 

A. T. Cheung, B. Driessen, J. S. Jahr, P. L. Duong, S. Ramanujam, P. C. Chen, and R. A. Gunther, “Blood substitute resuscitation as a treatment modality for moderate hypovolemia,” Artif. Cells Blood Substit Immobil Biotechnol., 32 (2), 189 –207 (2004). https://doi.org/10.1081/BIO-120037827 1073-1199 Google Scholar

11. 

H. March, A. Barger, S. McCullough, D. Schaeffer, and P. Macwilliams, “Use of the ADVIA 120 for differentiating extracellular and intracellular hemoglobin,” Vet. Clin. Pathol., 34 (2), 106 –109 (2005). https://doi.org/10.1111/j.1939-165X.2005.tb00021.x Google Scholar

12. 

P. W. Buehler, R. A. Boykins, Y. Jia, S. Norris, D. I. Freedberg, and A. I. Alayash, “Structural and functional characterization of glutaraldehyde-polymerized bovine hemoglobin and its isolated fractions,” Anal. Chem., 77 (11), 3466 –3478 (2005). https://doi.org/10.1021/ac050064y 0003-2700 Google Scholar

13. 

D. R. King, S. M. Cohn, and K. G. Proctor, “Resuscitation with a hemoglobin-based oxygen carrier after traumatic brain injury,” Sitzungsber. K. Preuss. Akad. Wiss., 59 (3), 553 –560 (2005). 0371-196X Google Scholar D. R. King, S. M. Cohn, and K. G. Proctor, Sitzungsber. K. Preuss. Akad. Wiss., 59 (3), 560 –552 (2005). 0371-196X Google Scholar

14. 

A. T. Cheung, P. L. Duong, B. Driessen, P. C. Chen, J. S. Jahr, and R. A. Gunther, “Systemic function, oxygenation and microvascular correlation during treatment of hemorrhagic shock with blood substitutes,” Clin. Hemorheol Microcirc, 34 (1–2), 325 –334 (2006). 1386-0291 Google Scholar

15. 

J. Lee, A. E. Cerussi, D. Saltzman, T. Waddington, B. J. Tromberg, and M. Brenner, “Hemoglobin measurement patterns during noninvasive diffuse optical spectroscopy monitoring of hypovolemic shock and fluid replacement,” J. Biomed. Opt., 12 (2), 024001 (2007). https://doi.org/10.1117/1.2715189 1083-3668 Google Scholar

16. 

C. C. Wo, W. C. Shoemaker, P. L. Appel, M. H. Bishop, H. B. Kram, and E. Hardin, “Unreliability of blood pressure and heart rate to evaluate cardiac output in emergency resuscitation and critical illness,” Crit. Care Med., 21 (2), 218 –223 (1993). https://doi.org/10.1097/00003246-199302000-00012 0090-3493 Google Scholar

17. 

B. R. Soller, Y. Yang, O. O. Soyemi, K. L. Ryan, C. A. Rickards, J. M. Walz, S. O. Heard, and V. A. Convertino, “Noninvasively determined muscle oxygen saturation is an early indicator of central hypovolemia in humans,” J. Appl. Physiol., 104 (2), 475 –481 (2008). https://doi.org/10.1152/japplphysiol.00600.2007 8750-7587 Google Scholar

18. 

B. R. Soller, K. L. Ryan, C. A. Rickards, W. H. Cooke, Y. Yang, O. O. Soyemi, B. A. Crookes, S. O. Heard, and V. A. Convertino, “Oxygen saturation determined from deep muscle, not thenar tissue, is an early indicator of central hypovolemia in humans,” Crit. Care Med., 36 (1), 176 –182 (2008). https://doi.org/10.1097/01.CCM.0000295586.83787.7E 0090-3493 Google Scholar

19. 

A. Gosain, J. Rabkin, J. P. Reymond, J. A. Jensen, T. K. Hunt, and R. A. Upton, “Tissue oxygen tension and other indicators of blood loss or organ perfusion during graded hemorrhage,” Surg. Today, 109 (4), 523 –532 (1991). 0941-1291 Google Scholar

20. 

B. Driessen, J. S. Jahr, F. Lurie, and R. A. Gunther, “Effects of isovolemic resuscitation with hemoglobin-based oxygen carrier Hemoglobin glutamer-200 (bovine) on systemic and mesenteric perfusion and oxygenation in a canine model of hemorrhagic shock: a comparison with 6% hetastarch solution and shed blood,” Vet. Anaesth. Analg., 33 (6), 368 –380 (2006). https://doi.org/10.1111/j.1467-2995.2005.00280.x Google Scholar

21. 

D. Kubulus, H. Rensing, M. Paxian, J. T. Thierbach, T. Meisel, H. Redl, M. Bauer, and I. Bauer, “Influence of heme-based solutions on stress protein expression and organ failure after hemorrhagic shock,” Crit. Care Med., 33 (3), 629 –637 (2005). https://doi.org/10.1097/01.CCM.0000156295.48075.49 0090-3493 Google Scholar

22. 

B. J. Tromberg, O. Coquoz, J. B. Fishkin, E. R. Anderson, D. Pham, M. Brenner, and L. O. Svaasand, “Frequency-domain photon migration (FDPM) measurements of normal and malignant cell and tissue optical properties,” Biomedical Optical Spectroscopy and Diagnostics, 111 –116 , (1996). Google Scholar

23. 

T. H. Pham, O. Coquoz, J. B. Fishkin, E. Anderson, and B. J. Tromberg, “Broad bandwidth frequency domain instrument for quantitative tissue optical spectroscopy,” Rev. Sci. Instrum., 71 (6), 2500 –2513 (2000). https://doi.org/10.1063/1.1150665 0034-6748 Google Scholar

24. 

C. Albert, W. Richard Van, W. Feizal, and T. Bruce, “Noninvasive monitoring of red blood cell transfusion in very low birthweight infants using diffuse optical spectroscopy,” J. Biomed. Opt., 10 (5), 051401 (2005). https://doi.org/10.1117/1.2080102 1083-3668 Google Scholar

25. 

J. Lee, N. El-Abaddi, A. Duke, A. E. Cerussi, M. Brenner, and B. J. Tromberg, “Noninvasive in vivo monitoring of methemoglobin formation and reduction with broadband diffuse optical spectroscopy,” J. Appl. Physiol., 100 (2), 615 –622 (2006). https://doi.org/10.1152/japplphysiol.00424.2004 8750-7587 Google Scholar

26. 

J. Lee, D. J. Saltzman, A. E. Cerussi, D. V. Gelfand, J. Milliken, T. Waddington, B. J. Tromberg, and M. Brenner, “Broadband diffuse optical spectroscopy measurement of hemoglobin concentration during hypovolemia in rabbits,” Physiol. Meas, 27 (8), 757 –767 (2006). https://doi.org/10.1088/0967-3334/27/8/009 0967-3334 Google Scholar

27. 

T. H. Pham, R. Hornung, H. P. Ha, T. Burney, D. Serna, L. Powell, M. Brenner, and B. J. Tromberg, “Noninvasive monitoring of hemodynamic stress using quantitative near-infrared frequency-domain photon migration spectroscopy,” J. Biomed. Opt., 7 (1), 34 –44 (2002). https://doi.org/10.1117/1.1427046 1083-3668 Google Scholar

28. 

J. Lee, J. Armstrong, K. Kreuter, B. J. Tromberg, and M. Brenner, “Non-invasive in vivo diffuse optical spectroscopy monitoring of cyanide poisoning in a rabbit model,” Physiol. Meas, 28 (9), 1057 –1066 (2007). https://doi.org/10.1088/0967-3334/28/9/007 0967-3334 Google Scholar

29. 

F. Bevilacqua, A. J. Berger, A. E. Cerussi, D. Jakubowski, and B. J. Tromberg, “Broadband absorption spectroscopy in turbid media by combined frequency-domain and steady-state methods,” Appl. Opt., 39 (34), 6498 –6507 (2000). https://doi.org/10.1364/AO.39.006498 0003-6935 Google Scholar

30. 

D. J. Jakubowski, “Developement of broadband quantitative tissue optical spectroscopy for the non-invasive characterization of breast disease,” (2002). Google Scholar

31. 

B. J. Tromberg, N. Shah, R. Lanning, A. Cerussi, J. Espinoza, T. Pham, L. Svaasnad, and J. Butler, “Non-invasive in vivo characterization of breast tumors using photon migration spectroscopy,” Opt. Eng., 2 (1–2), 26 –40 (2000). 0091-3286 Google Scholar

32. 

S. Fantini, M. Franceschini, J. S. Maier, S. A. Walker, B. Barbieri, and E. Gratton, “Frequency-domain multichannel optical detector for noninvasive tissue spectroscopy and oximetry,” Opt. Eng., 34 32 –42 (1995). https://doi.org/10.1117/12.183988 0091-3286 Google Scholar

33. 

B. W. Pogue and M. S. Patterson, “Frequency-domain optical absorption spectroscopy of finite tissue volumes using diffusion theory,” Phys. Med. Biol., 39 1157 –1180 (1994). https://doi.org/10.1088/0031-9155/39/7/008 0031-9155 Google Scholar

34. 

R. Graaff, J. G. Aarnoose, J. R. Zijp, P. M. A. Sloot, F. F. M. de Mul, J. Greve, and M. H. Koelink, “Reduced light-scattering properties for mixtures of spherical particles: a simple approximation derived from Mie calculation,” Appl. Opt., 31 1370 –1376 (1992). https://doi.org/10.1364/AO.31.001370 0003-6935 Google Scholar

35. 

J. R. Mourant, T. Fuselier, J. Boyer, T. Johnson, and I. J. Bigio, “Predictions and measurements of scattering and absorption over broad wavelength ranges in tissue phantoms,” Appl. Opt., 36 949 –957 (1997). https://doi.org/10.1364/AO.36.000949 0003-6935 Google Scholar

36. 

J. M. Schmitt and G. Kumar, “Optical scattering properties of soft tissue: a discrete particle model,” Appl. Opt., 37 2788 –2797 (1998). https://doi.org/10.1364/AO.37.002788 0003-6935 Google Scholar

37. 

W. H. Cooke, K. L. Ryan, and V. A. Convertino, “Lower body negative pressure as a model to study progression to acute hemorrhagic shock in humans,” J. Appl. Physiol., 96 (4), 1249 –1261 (2004). https://doi.org/10.1152/japplphysiol.01155.2003 8750-7587 Google Scholar

38. 

V. A. Convertino, “Gender differences in autonomic functions associated with blood pressure regulation,” Am. J. Physiol., 275 (6), R1909 –1920 (1998). 0002-9513 Google Scholar

39. 

B. Driessen, L. Zarucco, R. A. Gunther, P. M. Burns, S. V. Lamb, S. E. Vincent, R. A. Boston, J. S. Jahr, and A. T. Cheung, “Effects of low-volume hemoglobin glutamer-200 versus normal saline and arginine vasopressin resuscitation on systemic and skeletal muscle blood flow and oxygenation in a canine hemorrhagic shock model,” Crit. Care Med., 35 (9), 2101 –2109 (2007). https://doi.org/10.1097/01.CCM.0000277040.31978.3D 0090-3493 Google Scholar

40. 

A. T. Cheung, P. L. To, D. M. Chan, S. Ramanujam, M. A. Barbosa, P. C. Chen, B. Driessen, J. S. Jahr, and R. A. Gunther, “Comparison of treatment modalities for hemorrhagic shock,” Artif. Cells Blood Substit Immobil Biotechnol., 35 (2), 173 –190 (2007). https://doi.org/10.1080/10731190601188257 1073-1199 Google Scholar

41. 

T. Standl, “Haemoglobin-based erythrocyte transfusion substitutes,” Expert Opin. Biol. Ther., 1 (5), 831 –843 (2001). https://doi.org/10.1517/14712598.1.5.831 1471-2598 Google Scholar

42. 

M. Intaglietta, P. C. Johnson, and R. M. Winslow, “Microvascular and tissue oxygen distribution,” Cardiovasc. Res., 32 (4), 632 –643 (1996). 0008-6363 Google Scholar

43. 

D. R. Spahn, B. J. Leone, J. G. Reves, and T. Pasch, “Cardiovascular and coronary physiology of acute isovolemic hemodilution: a review of nonoxygen-carrying and oxygen-carrying solutions,” Anesth. Analg. (Baltimore), 78 (5), 1000 –1021 (1994). 0003-2999 Google Scholar

44. 

R. M. Winslow, “Current status of blood substitute research: towards a new paradigm,” J. Intern Med., 253 (5), 508 –517 (2003). https://doi.org/10.1046/j.1365-2796.2003.01150.x 0954-6820 Google Scholar

45. 

R. M. Winslow, “Current status of oxygen carriers (‘blood substitutes’): 2006,” Vox Sang., 91 (2), 102 –110 (2006). https://doi.org/10.1111/j.1423-0410.2006.00789.x 0042-9007 Google Scholar

46. 

G. Gutierrez, H. D. Reines, and M. E. Wulf-Gutierrez, “Clinical review: hemorrhagic shock,” Crit. Care, 8 (5), 373 –381 (2004). https://doi.org/10.1186/cc2851 1364-8535 Google Scholar
©(2009) Society of Photo-Optical Instrumentation Engineers (SPIE)
Jangwoen Lee, Jae Gwan Kim, Sari B. Mahon, Bruce Jason Tromberg, David Sho Mukai, Kelly A. Kreuter, Darin J. Saltzman, Renee Patino M.D., Robert Goldberg M.D., and Matthew Brenner M.D. "Broadband diffuse optical spectroscopy assessment of hemorrhage- and hemoglobin-based blood substitute resuscitation," Journal of Biomedical Optics 14(4), 044027 (1 July 2009). https://doi.org/10.1117/1.3200932
Published: 1 July 2009
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Blood

Tissues

Rhodamine B

Tissue optics

Scattering

Oxygen

Diffuse optical spectroscopy

RELATED CONTENT


Back to Top