|
1.IntroductionDiffuse optical imaging is a noninvasive technique used to obtain the spatial distribution of optical properties and chromophore concentrations of tissue from measurements of near-infrared (NIR) light, which propagates through the tissue. The technique can be used to monitor cerebral oxygenation and hemodynamics, to provide information about cerebral pathological conditions, to measure functional brain activation during a physical or mental task, and for breast cancer detection.1, 2, 3, 4, 5, 6 Two different approaches are considered in NIR imaging: optical topography and optical tomography. Optical topography is used to describe systems that obtain diffuse reflectance measurements from an array of source-detector pairs placed on the surface of the tissue, whereas optical tomography involves reconstructing a 3-D volume using multiple measurement of transmitted light across the whole tissue volume.7 A few NIR spectroscopy studies have been published that explore the selection of an optimal wavelength pair that minimizes cross-talk, where a change in oxy-hemoglobin concentration may replicate a change in deoxy-hemoglobin concentration [HHb] and vice versa, and minimizes the influence of noise on calculated chromophore concentration changes.8, 9, 10, 11, 12, 13 These studies revealed that, for measurements performed at two different NIR wavelengths, one of the wavelengths should be in the range and the other . Image reconstruction using continuous wave (CW) measurements at a single wavelength is an inverse problem with nonunique solutions, where multiple sets of optical properties can yield similar data.14 As a consequence, cross-talk between the images of absorption and scattering occurs. Increasing the number of measurement wavelengths only increases the number of unknowns because absorption and scattering need to be reconstructed for each wavelength. However, the multispectral method proposed by Corlu 15, 16 directly reconstructs the chromophore concentrations and scattering factors that are wavelength independent, meaning that the number of unknowns remains constant when the number of measurement wavelengths increases, representing a clear advantage over the conventional method. Corlu 15, 16 introduced an approach for selecting the optimal wavelengths for optical tomography based on their multispectral method. This approach sought to maximize separation between absorption and scattering separation while also optimizing separation between the concentrations of main breast tissue chromophores ( , HHb, water, and lipid). However, this overlooks the possibility that the volumes over which measurements are sensitive to changes in absorption and scattering may be different at the selected wavelengths. This issue is of importance in order to accurately locate optical property changes occurring within the object of study, because different wavelengths may have different sensitivities to the same changes in the medium, and the same measured change could be associated with changes in different regions within the reconstructed volume. In this study, the optimal wavelength set is estimated by incorporating a third condition, which takes into account the interrogated volume. The spatial similarities between sensitivity profiles for different wavelengths are analyzed and quantified. 2.Theory2.1.Reconstruction MethodThe conventional image reconstruction method involves finding the optical properties of the medium from a set of measurements made at the surface. The optical properties are then combined to calculate the chromophore concentrations at wavelength λ using the following equation: where is the extinction coefficient and is the concentration of the ’th chromophore. The reduced scattering coefficient is approximately given by Mie scattering theory aswhere is the scatter amplitude and the scatter power. In this work, the latter is considered to be constant for simplicity.The multispectral method involves reconstructing chromophore concentrations and scatter parameters using all measurements at all wavelengths simultaneously.16 The problem is nonlinear but can be linearized and reduced to the so-called difference method, based on the assumption that the true chromophore concentrations are close to an initial guess. For CW measurements, only intensity is measured, hence changes in the measured data at measurement wavelengths are related to changes in the chromophore concentrations and scatter amplitude through matrix , where the elements of the matrix are given byandwhere is the diffusion coefficient. The Jacobian matrix represents the sensitivity of a measurement to changes in the optical properties within the medium. The Jacobian can be calculated by solving the diffusion equation using the finite element method, for both the forward and adjoint solutions.17 In the Rytov approximation, changes in the log intensity are used instead.17, 18Chromophore concentrations and scatter parameters can be obtained by solving an ill-posed inverse problem at all wavelengths. In the work presented here, this is aided using the zero-order Tikhonov regularization, where is the NOSER-type regularization parameter,19, 20 with , where is a scaling factor and is the maximum singular value of . The constant ensures that the regularization parameter is never too small and is approximately equal to the noise present in the data.2.2.Identification of the Optimal Wavelengths in Optical Topography2.2.1.Unique and simultaneous reconstructionThe nonuniqueness problem associated with using CW measurements results in cross-talk between absorption and scattering. Minimizing the nonuniqueness of the multispectral method by selecting the optimal wavelengths represents one of the conditions introduced by Corlu 16 Rewriting the nonuniqueness conditions introduced by Arridge and Lionheart14 in terms of the wavelength-independent variables in the multispectral method, it follows that the inverse problem has a nonunique solution when the residual norm16 : is zero. Thus, a maximization of is required. is a wavelength-dependent matrix of the formThe ability to separate chromophore concentrations can be evaluated from the matrix form of Eq. 1,16 using the condition number of the extinction coefficient matrix , which is the ratio of the largest to the smallest singular value of the matrix. A small indicates a smooth singular value decay resulting in a reduced sensitivity of the system to errors in the measurements, and therefore, chromophores contribute evenly to absorption. In order to distinguish between chromophore concentrations, the wavelength sets chosen should minimize the condition number , which defines the second condition introduced by Corlu 16 This condition number also describes the linear independence of the system of equations. When is small, the matrix has full rank and, therefore, the wavelength set provides distinct and independent information.2.2.2.Interrogated volumeIn optical topography brain activation studies using multiple wavelengths, it is invariably assumed that measurements at each wavelength contain information from the same cerebral region. If not, data at some wavelength could be measuring a region where the activation occurs, whereas the data at others could be more sensitive to surrounding areas and the study may return erroneous results. Each row of the Jacobian matrix represents the measurement sensitivity of a source-detector pair to a perturbation in the optical properties of the object of study, which is known as the photon measurement density function (PMDF).21 In terms of image reconstruction, if the forward model is calculated for the same optical properties but the measurement wavelengths have different sensitivities to changes in absorption and scattering, then for a perturbation in the optical properties of the tissue, the reconstructed perturbation will occur in different spatial locations. In practice, the optical properties of tissues beneath the probe are unknown, and in order to reconstruct changes in tissue properties as accurately as possible, it is necessary to acquire data at wavelengths where the sensitivities at each wavelength coincide. In this work, a method is proposed to find which combinations of wavelengths have overlapping sensitivities, which is an indicator of the similarity between interrogated volumes, by comparing PMDFs for a certain source-detector pair using different measurement wavelengths. The influence of using different NIR measurements is analyzed in terms of image reconstruction and volume sensitivities. This is less relevant when considering transmission measurements (e.g., across a head of breast) because there is an intrinsically greater overlap of the PMDFs. 3.MethodsThe software package TOAST (time-resolved optical absorption and scattering tomography)22 was used to model the propagation of light in tissue at wavelengths from spaced in intervals. To generate PMDFs for each wavelength, one needs to insert the corresponding optical properties of the medium in the forward problem. Initially, the medium is considered to be homogeneous with optical properties of gray matter. Subsequently, a three-layer model is used to generate PMDFs, which consists of two extracerebral layers with skin and skull optical properties and an intracerebral layer with gray-matter optical properties. A single source-detector pair apart was used in the forward model because simulations have indicated that this is sufficient to measure hemodynamic changes in adult gray matter.23 3.1.Adult Head Model3.1.1.Homogeneous modelThe first model consisted of a uniform slab with dimensions with intracerebral tissue optical properties derived from a combination of chromophores present in brain tissue: hemoglobin, water, lipid, and cytochromes. Brain absorption spectrumTotal absorption coefficient for brain tissue is calculated from Eq. 1 as where [HbT] is the total hemoglobin concentration [in micromolar (μM)], is the oxygen saturation, is the fraction of water content, is the absorption coefficient of 100% water ( ), is the fraction of lipid content, is the absorption coefficient of 100% fat ( ), is the specific absorption coefficient of HHb , is the specific absorption coefficient of , is the specific absorption coefficient of reduced cytochrome , is the specific absorption coefficient of oxidised cytochrome , and Ox is the fraction of oxidized Cyt . The parameter is the background absorption, which was set to so that the total absorption matches the value used by Okada and Delpy.31 The extinction coefficients used in this work were obtained from the literature.26, 32, 33, 34 The estimated concentrations in the gray matter for each chromophore are summarized in Table 1 .Table 1Chromophore concentrations used to calculate the absorption coefficients for the skin and gray matter. Brain scattering spectrumMatcher 35 measured of the forehead of different human adults over the wavelength range of . They fitted the experimental spectra to Eq. 2 and obtained and . Meanwhile, van der Zee27 measured for gray matter in vitro and found , which can be used as a reference number and included in the exponential relation found by Matcher, 35 which then becomes 3.1.2.Three-layer modelFigure 1 shows the second adult head model, which is a slab with the same dimensions as the homogeneous model, consisting of three layers that imitate the scalp, skull, and intracerebral tissue. Scalp is represented by a thickness layer, and the skull is thick. Absorption and scattering spectra of skull are described in Firbank 36 Skin absorption spectrumThree main chromophores were considered for scalp: hemoglobin, water, and melanin. Their estimated concentrations in skin are summarized in Table 1. The melanin absorption coefficient can be approximated by the expression30 in , where is the wavelength represented in units of nanometers. Considering the contributions from all skin chromophores, the total absorption coefficient of the skin layer can be represented aswhere is the fraction of melanin content. It was not necessary to consider a background absorption for the skin layer, because the absorption coefficient values already show a good agreement with the values presented by Simpson 37Skin scattering spectrumThe wavelength dependence of the reduced scattering coefficient of skin has been modeled by Bashkatov 38 as a combination of Mie and Rayleigh scattering given by in , where is the wavelength in nm. The first term corresponds to Mie scattering due to large collagen fibers and the second to Rayleigh scattering mainly due to small collagen fibrils.3.2.Similarity MeasuresThe sum of squared differences (SSD) can be used to measure the similarity between two sensitivity profiles. For PMDFs and with voxels , the SSD is defined as where is the number of voxels. The SSD will be zero when the distributions are identical.3.3.Optimal WavelengthsSensitivity profile maps (PMDFs) were obtained for the wavelength range of in steps of , making a total of 35 wavelengths. Initially, the PMDFs were combined in sets of three wavelengths , making a total of 6545 wavelength sets. Thereafter, wavelength sets of four wavelengths were investigated, which results in 52,360 sets. For a medium with absorption chromophores, at least wavelengths are required when the scattering parameter is allowed to vary.16 Therefore, in order to find the solution for both absorption and scattering coefficients for a medium with two chromophores, it is necessary to employ at least three wavelengths. It is common for optical topography systems to acquire data at just two wavelengths. Therefore, all possible wavelength pairs are compared (total of 595), to provide an insight into what can be achieved using only two measurement wavelengths. The aim of this study is to determine which wavelength set gives the best combination of: (i) good separation between absorption and scattering, (ii) good separation between and [HHb], and (iii) excellent overlap between sensitivity profiles. Quantitatively, we aim to find the wavelength sets that maximize the residual norm [Eq. 7], minimize the condition number of the matrix [Eq. 9], and minimize the SSD [Eq. 15]. The matrix in Eq. 8 for three wavelengths is reduced to where the scatter power is considered to be 0.53.353.4.PMDF ComparisonThe PMDFs for wavelengths from are compared by quantifying the differences in their sensitivity in terms of penetration depth, volume, and full width at half maximum (FWHM), in order to determine if these differences are significant. Absolute sensitivity values are used (because the absorption PMDF has negative sensitivity), and the threshold is set to 0.01. An analysis is performed of how these differences affect the measurements and image reconstructions. 3.4.1.VolumeThe volume is the number of voxels with sensitivity above the threshold times the voxel volume. The relative volume compares the PMDF volume at with all the other PMDF volumes , where is the number of voxels above the threshold in the PMDF at and is the number of voxels above the threshold for the remaining wavelengths.3.4.2.FWHMThe FWHM was calculated for the PMDF cross section at half-way between the source and detector for the depth direction. The PMDF profiles include the point of maximum sensitivity. The FWHM provides a measure of the spread of the PMDFs. 3.4.3.Penetration depthThe penetration depth (PD) is defined as the maximum distance into the medium at which perturbations in the optical properties can still be measured. The central region of the PMDF penetrates deeper into the tissue, which corresponds to the PMDF profile ( direction) that passes through the midpoint between the source and detector. Hence, the distance at which the sensitivity drops below the threshold for these profiles is considered to be the PD. 3.5.Image ReconstructionSimulations were performed to compare the cross-talk effects in images reconstructed from measurements at wavelengths selected using our method against those from measurements at wavelengths selected using Corlu ’s method.16 The simulated medium consisted of a slab with 59,202 nodes, resulting in 38,572 elements, and dimensions . An array with eight detectors and eight sources (described previously by Correia 39) is centrally placed at . However, here only source-detector separations from were considered. TOAST was used to generate CW model data at each wavelength, to which 5% random Gaussian noise was added. The background had brain optical properties, which were described in Section 3.1.1. Hence, the deoxy-hemoglobin concentration [HHb] is , the oxy-hemoglobin concentration is , and the scattering amplitude is . Three perturbations with dimensions were placed within the medium, as shown in Fig. 2 . The center of each perturbation was at below the imaging surface. A single parameter was changed for each perturbation, which consisted of either an increase in [HHb] of , in of , or in of . These perturbations are sufficiently small to ensure that the concentrations and scattering amplitude can be retrieved by linear reconstruction, and are roughly consistent with the magnitude of cerebral hemodynamic changes. Images were reconstructed using the multispectral method on a regular pixel basis. 4.Results4.1.Optimal WavelengthsFor simplicity, the absorption and diffusion sensitivity profiles are dealt with separately. The first results are shown in Fig. 3 for the homogeneous model, considering absorption sensitivity profiles only and three wavelength sets. Note that the SSD values in Fig. 3 represent the average for each wavelength set. Because of the large number of points, further analysis of the wavelength sets was limited to those for which , , and (because the SSD for two similar PMDFs is ) (dark circles in Fig. 3). Each criterion has a different order of magnitude; hence, the results contained in this new range were normalized. The distance between the criteria values of a certain set of wavelengths in this range and the point was also calculated, which is the point with the smallest SSD, smallest condition number criteria, and highest residual. The wavelength distribution of the ten sets of three wavelengths with the shortest distance are displayed in the histogram in Fig. 4 . Histograms are obtained by counting the number of times a certain wavelength occurs, which is done separately for the three wavelengths, and normalized to the most frequently occurring wavelength. For comparison, Fig. 5 shows the histogram for the ten wavelength sets within the limit and , whose distance is closest to the point . The means and standard deviations of each wavelength distribution were calculated in order to find the optimal wavelengths and measure their spread. A uncertainty error is combined with , due to the bin size. Using the three-wavelength selection criteria, the wavelengths are , , and . When only the high residual and low condition number criteria are used the optimal wavelengths are , , and . The results suggest that when the new PMDF overlap condition is included the smallest wavelength becomes larger, and the largest wavelength becomes smaller. The optimal four wavelengths that satisfy the three criteria are , , , and . Adding a fourth wavelength broadened the central wavelength range and shifted the highest wavelength to a smaller value. The wavelength set obtained without including the third condition is , , , and . The differences between these results and those presented by Corlu 16 are inevitably due to differences in the wavelength range and spacing, chromophore extinction coefficients, and criteria limits imposed. The SSD for diffusion PMDFs is always slightly smaller than for absorption PMDFs. This is probably because scattering does not have a strong wavelength dependence. Thus, the optimal wavelength sets found using the absorption SSDs also return small diffusion SSDs. The histograms for the three-layer adult head model for three and four wavelengths are shown in Figs. 6 and 7 , respectively. For the three wavelength distributions, the optimal wavelength values are , , and , and for four wavelengths these values are , , , and . The main difference between the optimal wavelengths found for the homogeneous model and the three-layer model occurs for the largest wavelength, which becomes larger for the latter. The other wavelengths are approximately the same and, similar to the homogeneous case, adding a fourth wavelength resulted in the merger of the two central wavelengths (i.e., the mean value is approximately the same but the range is larger). Also for this model, the diffusion SSDs are smaller than the corresponding values for absorption; hence, the previous results also apply for diffusion-sensitivity profiles. The wavelength distributions obtained using our method were similar when the scattering power was varied from 0 to 2.8, whereas for the two-criteria method similar wavelength distributions were obtained for a range of from 0 to 1.2. 4.1.1.Two wavelengthsIt is common to assume that scattering is homogeneous and constant over time. This assumption is convenient when data are acquired at only two wavelengths, enabling solutions to be obtained if only two chromophores are considered. The following analysis is aimed at identifying the two optimal wavelengths that give good separation between chromophores and good spatial overlap between sensitivity profiles. The solution is nonunique when only two wavelengths are used. Although the SSD and the condition numbers can be very small, the residual norm is never larger than , which confirms that one cannot separate absorption from scattering using only two measurement wavelengths. Chromophore concentrations can be distinguished if the condition number is small. The wavelengths whose distance is shortest to the point with the smallest condition number cluster at and . On the other hand, the wavelengths with shortest distance to the point cluster at and . The inclusion of the PMDF overlap condition has a clear effect on the largest wavelength. For the three-layer model, the wavelength pair with the shortest distance to the point is and , which is very similar to the optimal wavelength set found for the homogeneous case. The influence of on the absorption and diffusion PMDFs is identical; high absorption coefficients lead to narrower PMDFs. A similar tendency is observed when has a high value for the absorption PMDFs, although the diffusion PMDFs become broader. Nevertheless, the SSD for both absorption and diffusion PMDFs is similar. The wavelength sets obtained using the diffusion PMDFs are identical to those found using the absorption PMDFs. The largest SSD occurs for wavelengths paired with , which corresponds to the absorption peak of water. As mentioned previously, the sum of squared differences SSD for diffusion is slightly smaller than for absorption. 4.2.PMDF Comparison4.2.1.VolumeThe relative volume was calculated for the absorption PMDFs and diffusion PMDFs , for both head models. Figure 8 shows the relative volumes for the different wavelengths and the value of for each case. The volume of the is larger than that of the . For the homogeneous model, the volume resembles an inverted spectrum of brain tissue, because for large the PMDF volume becomes smaller and for small the volume becomes larger. For the , the volume decreases as expected, given that the volume decreases with decreasing . However, the volume for does not vary as much as for . The relative volumes for the layered model are similar to those obtained for the homogeneous case. 4.2.2.FWHMFigure 9 shows the difference between the depth FWHMs and that corresponding to the wavelength , . In general, the difference is , which indicates that the spread of regions of higher sensitivity does not vary much with wavelength. The FWHM differences are smaller for , in particular for the homogeneous model, which shows a relatively flat spectrum for wavelengths of . 4.2.3.PDPD is larger for the , as shown in Fig. 10 . The graphs show the differences with respect to the PMDF PD at , . For the homogeneous model, the maximum depth is and occurs for a wavelength of . The lowest PD is and occurs for a wavelength of . This means that the maximum PD difference is . For the layered model, the maximum penetration depth is for a wavelength of . For a wavelength of , the PD attains a minimum value of . The largest difference between PD is . As a consequence of the voxel size, which only allows discrete spatial steps, the relative PD spectra exhibit a steplike behavior. The exhibits a small variation in the PD for the different wavelengths (Fig. 10). For the homogeneous model, the combination of low and high for wavelengths from results in a broader with higher PD. The maximum PD is , and the largest difference between PD is . The for skull decays faster than for brain tissue, and at wavelengths of their values diverge significantly. When the external layers have lower and than the inner layer, the sensitivity profiles shift toward the surface. The high absorption of the brain tissue at reduces the interrogation depth to , compared to the maximum at . 4.3.Image ReconstructionImages were reconstructed from simulated measurements at wavelengths selected using our new proposed method, and from measurements at wavelengths selected using the method of Corlu 16 Figure 11 shows the cross-sectional images at [Fig. 11] and [Fig. 11], using measurement wavelengths at 680, 720, and , simultaneously. Figure 12 shows the cross-sectional images at [Fig. 12] and [Fig. 12], using measurement wavelengths at 650, 720, and , simultaneously. Note that the color scales are not the same to enhance the differences between the images. For both sets of wavelengths, the images exhibited cross-talk between chromophores and also scattering. Cross-talk is particularly significant between oxy- and deoxy-hemoglobin concentrations in the deoxy-hemoglobin reconstruction and more evident in the images reconstructed using the wavelengths selected by our method. However, the deoxy-hemoglobin perturbation appears closer to the real position in the images reconstructed from wavelengths selected by our method than in the images obtained from wavelengths selected using the two-condition criteria. The is underestimated in all the cases. The [HHb] is overestimated at and underestimated at . For comparison, images were reconstructed from simulated measurements at two wavelengths that are commonly used, 780 and , and a third wavelength located between these two at . Figure 13 shows the cross-sectional images for this wavelengths set at and Fig. 13 the reconstructed images at . Cross-talk is clearly visible in the oxy-hemoglobin concentration and scattering amplitude images. The calculated concentration and scattering amplitude values are underestimated, except for [HHb] calculated at . However, the [HHb] does not appear in the images in the correct position. The same simulation study was performed using four wavelengths, and the corresponding cross-sectional images at and obtained using the wavelengths found using our method are shown in Figs. 14 and 14 , respectively. Because of the limited sensitivity of the measurement to scattering to regions deeper than below the surface, the reconstructions were found to have less cross-talk at than at the true target depth . Figure 15 shows the cross-sectional images at [Fig. 15] and [Fig. 15], using the wavelengths selected by the two-criteria method. Reconstructions at for both sets of wavelengths were similar and showed very small cross-talk. Nevertheless, the scattering amplitude and deoxy-hemoglobin concentration calculated at were overestimated. The slightly better reconstructions at obtained using our wavelength set could be due to the higher sensitivity of these wavelengths to the depth at which the targets are located. Furthermore, even though the sensitivity to scattering is quite limited in deeper regions, the PD of the has its maximum at these wavelengths. For comparison, images were reconstructed using the wavelengths 750, 780, 820, and . Figure 16 shows the cross-sectional images at , and Fig. 16 shows the images at , obtained using the latter measurement set. Cross-talk is visible in the images; the calculated chromophore concentrations and scattering amplitude are underestimated; and the deoxy-hemoglobin perturbation appears misplaced in the image. The simulation results indicate that four measurement wavelengths are required, in order to reconstruct images with negligible cross-talk, from reflection measurements and using linear multispectral reconstruction. For the three-layer adult head model presented here, the sensitivity to scattering perturbation occurring in the brain region is very low; hence, it is not possible to separate the two chromophores and scattering amplitude simultaneously. Finally, simulations were performed using the same medium as before, but without the scattering perturbation and for two wavelengths only. Figure 17 shows the reconstructions at and obtained from measurements at wavelengths found using our method (690 and ). The images show a good separation between chromophores, but at , the calculated chromophore concentrations are overestimated and, at , are slightly underestimated. The images reconstructed from measurements at wavelengths found using the two criteria method (710 and ) exhibit a small amount of cross-talk (Fig. 18 ). The presence of cross-talk could be due to sensitivity differences between the two wavelengths. At , only the top of the target is interrogated, whereas the light at is likely to travel through the whole target volume. All the simulations were repeated for up to three times larger concentration and scattering amplitude changes, and the results showed a similar tendency. In applications of optical topography, the true optical properties of the medium are usually unknown; hence, the sensitivity matrix is usually calculated using an estimate of the average optical properties, and subsequently used to reconstruct images from measurements at different wavelengths. Given that the optical properties of the medium are wavelength dependent, the reconstructed images are affected by using sensitivity matrices generated for optical properties different from the true properties. If the estimate of the optical properties is significantly different from the true values, then the absorption perturbation is likely to appear misplaced in the image, particularly in the direction. 5.DiscussionA method of determining the optimum wavelengths for optical topography is presented. This method is based on the multispectral method and is an extension of Corlu’s wavelength selection method,16 where a third criterion is introduced that minimizes the differences between interrogated volumes at different wavelengths. Including the sensitivity profiles similarity criterion in the selection of the optimal wavelengths has a clear effect on the resultant wavelength sets. The insertion of two superficial layers in the adult head model had an influence on the results, in particular, for the largest wavelength mean value, which shows that selection depends on the accuracy of the head model. It is possible to separate and [HHb] from measurements performed at two wavelengths. However, one cannot distinguish between absorption and scattering. When the sensitivity profiles similarity condition is included in the wavelength selection method, the largest wavelength mean value changes in comparison to that obtained using the condition alone. The main characteristics of the optically interrogated volumes (volume, FWHM, and PD) were quantified, in order to determine how measurements at different NIR wavelengths may be combined to provide information about the volume. Particular attention is given to the wavelengths selected using our wavelength selection method and Corlu’s method.16 If we consider the thickness of the head superficial layers to be and the gray matter to be , then in theory for the homogeneous model, the lowest wavelength selected using the two-criteria method penetrates into the gray matter, whereas the midrange wavelengths penetrate , which means that some light could reach the white matter. Therefore, the information contained in the measurements at these wavelengths could be significantly different. Including our PMDF overlap condition in the selection of the optimal wavelengths reduces the difference in the PD from . The first and last wavelengths, with mean values 680 and , respectively, penetrate into the brain tissue. This value is closer to that obtained for the central wavelengths, which is . The volumes interrogated are also larger and more similar for the wavelengths selected using our method. For the , the changes in the PD are smaller and the maximum PD is only , hence confined to the gray matter. However, the differences between PD are smaller when our method is used. For the three-layer adult head model, the discrepancies between the interrogated volumes at different wavelengths are smaller. However, the volumes are more similar for the wavelengths selected using our method than using the two-selection criteria. In terms of PD, for the , the variations are smaller than for the homogeneous model. For the central wavelengths of the selected wavelength sets light penetrates into the gray matter. For the lowest wavelengths obtained with our method, light penetrates into the brain. For the remaining wavelengths, a maximum depth of from the brain surface is probed. Although the differences between PD for the different wavelengths are identical for both selection methods, when using our method it optimizes for the wavelength set with higher PD, simultaneously with the best separation between chromophores, absorption, and scattering. The PMDF spread (i.e., the FWHM and volume) is also larger and more similar for our wavelength set. For the , the maximum PD is only ; hence, light hardly reaches the brain surface. The differences are much smaller than for the absorption PMDFs. From this analysis, it was shown that for the adult human head it can be assumed that the volumes of interrogated tissue at different NIR wavelengths are identical. Hence, measurements performed at different NIR wavelengths provide information from the same brain regions. This method could be applied to other types of tissue. For example, for measurements of muscle, breast, or neonate brain. For muscle, the highest absorption occurs at and the lowest at .40 For these wavelengths, the PD is 15 and , respectively. This results in a PD maximum difference of . For studies of muscle exercise, of the forearm for example, which has a diameter of , a difference of is quite significant. The influence of the skin and fat layer on the light propagation is likely to increase the PD, and it should depend on the thickness of these layers. For breast tissue, using the optical properties from Cubeddu, 41 the largest differences in the absorption coefficient occur at wavelengths 720 and , with a PD of 22 and , respectively. Therefore, the maximum PD difference is . The main objective of breast diffuse optical imaging is tumor detection. It is generally assumed that a tumor may be distinguished from surrounding healthy tissue on the basis of its apparent blood volume and oxygenation. Overlapping sensitivities are required to provide accurate measures of these parameters and reduce the uncertainty in the measured spatial location of the tumour. Light propagation in the neonatal brain is different from the adult brain tissue, because the surrounding layers are thinner and the optical properties are different. The is lower for the neonatal brain tissue.27 Because of the high scattering of the neonatal brain, the sensitive regions of the PMDF should be confined to the gray matter. However, the cortical thickness is only , meaning that light is likely to penetrate into deeper regions of the white matter.42 The PMDFs of other types of measurements, either frequency or time domain, can also be analyzed using our method. For example, the region of higher sensitivity of the mean time or phase PMDF is located deeper in the tissue than that of the CW PMDFs.43 It is clear that the adult head model had a large influence on the results. A more complete model of the adult head should include the cerebrospinal fluid (CSF) surrounding the brain tissue. However, the diffusion equation-based model is only valid for high scattering tissues and cannot be applied to the low scattering CSF. Monte Carlo simulations of light transport can be used, but they require long computational times. An alternative method is the hybrid radiosity-diffusion method, where the light propagation in high scattering media is calculated using the diffusion equation and the low scattering region by radiosity theory.44, 45 The inclusion of a CSF layer in our adult head model would not significantly affect our results because the high sensitivity regions of the PMDF would still be confined to superficial regions of the brain and because this layer has very little wavelength dependence. The PD of the light in models with and without CSF is almost the same, for source-detector separations of 30 and .31 A further consideration is the error associated with the extinction coefficient for the different chromophores. The spectra found in the literature, for certain chromophores, show some variation. Based on Corlu’s method16 for wavelength optimization, Brendel and Nielsen46 added a criterion that evaluates the reconstruction errors due to uncertainty in the extinction coefficients. From simulations and breastlike phantom measurements, they found a set of optimum wavelengths almost independent on the uncertainties in the extinction coefficients spectra. Another potential source of error is the assumed values of the chromophore concentrations for each head layer. This influences the optical properties of the layers, which could possibly have an effect on the spatial sensitivity profiles. Contrary to previous findings, our simulation results indicate that four measurement wavelengths are required in order to reconstruct images with low cross-talk from reflection measurements and using linear multispectral reconstruction. Images reconstructed from simulated measurements at wavelengths selected using our method exhibit a smaller cross-talk than those reconstructed from measurements selected using the two criteria. Therefore, the use of measurements with overlapping sensitivities slightly improves the image reconstruction. AcknowledgmentsThe work has been supported by a scholarship awarded to T.C. by Fundação para a Ciência e a Tecnologia, Portugal. This work has been supported in part by funding from the ECs seventh framework programme under Grant Agreement No. FP7-HEALTH-2007-201076. ReferencesA. P. Gibson,
T. Austin,
N. L. Everdell,
M. Schweiger,
S. R. Arridge,
J. H. Meek,
J. S. Wyatt,
D. T. Delpy, and
J. C. Hebden,
“Three-dimensional whole-head optical tomography of passive motor evoked responses in the neonate,”
Neuroimage, 30 521
–528
(2006). https://doi.org/10.1016/j.neuroimage.2005.08.059 1053-8119 Google Scholar
L. C. Enfield,
A. P. Gibson,
N. L. Everdell,
D. T. Delpy,
M. Schweiger,
S. R. Arridge,
C. Richardson,
M. Keshtgar,
M. Douek, and
J. C. Hebden,
“Three-dimensional time-resolved optical mammography of the uncompressed breast,”
Appl. Opt., 46 3628
–3638
(2007). https://doi.org/10.1364/AO.46.003628 0003-6935 Google Scholar
B. W. Zeff,
B. R. White,
H. Dehghani,
B. L. Schlaggar, and
J. P. Culver,
“Retinotopic mapping of adult human visual cortex with high-density diffuse optical tomography,”
Proc. Natl. Acad. Sci. U.S.A., 104 12169
–12174
(2007). https://doi.org/10.1073/pnas.0611266104 0027-8424 Google Scholar
S. R. Hintz,
D. A. Benaron,
A. M. Zourabian,
D. K. Stevenson, and
D. A. Boas,
“Bedside functional imaging of the premature infant brain during passive motor activation,”
J. Perinat. Med., 29 335
–343
(2001). https://doi.org/10.1515/JPM.2001.048 0300-5577 Google Scholar
A. Custo,
D. A. Boas,
D. Tsuzuki,
I. Dan,
R. Mesquita,
B. Fischl,
W. E. L. Grimson, and
W. Wells,
“Anatomical atlas-guided diffuse optical tomography of brain activation,”
Neuroimage, 49 561
–567
(2010). https://doi.org/10.1016/j.neuroimage.2009.07.033 1053-8119 Google Scholar
R. Choe,
S. D. Konecky,
A. Corlu,
K. Lee,
T. Durduran,
D. R. Busch,
S. Pathak,
B. J. Czerniecki,
J. Tchou,
D. L. Fraker,
A. DeMichele,
B. Chance,
S. R. Arridge,
M. Schweiger,
J. P. Culver,
M. D. Schnall,
M. E. Putt,
M. A. Rosen, and
A. G. Yodh,
“Differentiation of benign and malignant breast tumors by in vivo three-dimensional parallel-plate diffuse optical tomography,”
J. Biomed. Opt., 14 024020
(2009). https://doi.org/10.1117/1.3103325 1083-3668 Google Scholar
J. C. Hebden,
“Advances in optical imaging of the newborn infant brain,”
Psychophysiology, 40 501
–510
(2003). https://doi.org/10.1111/1469-8986.00052 0048-5772 Google Scholar
K. Uludağ,
J. Steinbrink,
A. Villringer, and
H. Obriga,
“Separability and cross talk: optimizing dual wavelength combinations for near-infrared spectroscopy of the adult head,”
Neuroimage, 22 583
–589
(2004). https://doi.org/10.1016/j.neuroimage.2004.02.023 1053-8119 Google Scholar
Y. Yamashita,
A. Maki, and
H. Koizumi,
“Wavelength dependence of the precision of noninvasive optical measurement of oxy-, deoxy-, and total-hemoglobin concentration,”
Med. Phys., 28 1108
–1114
(2001). https://doi.org/10.1118/1.1373401 0094-2405 Google Scholar
G. Strangman,
M. A. Franceschini, and
D. A. Boas,
“Factors affecting the accuracy of near-infrared spectroscopy concentration calculations for focal changes in oxygenation parameters,”
Neuroimage, 18 865
–879
(2003). https://doi.org/10.1016/S1053-8119(03)00021-1 1053-8119 Google Scholar
N. Okui and
E. Okada,
“Wavelength dependence of crosstalk in dual-wavelength measurement of oxy- and deoxy-hemoglobin,”
J. Biomed. Opt., 10 011015
(2005). https://doi.org/10.1117/1.1846076 1083-3668 Google Scholar
S. Umeyama and
T. Yamada,
“A new crosstalk measure of near-infrared spectroscopy and its application to wavelength combination optimization,”
J. Biomed. Opt., 14 034017
(2009). https://doi.org/10.1117/1.3147402 1083-3668 Google Scholar
H. Sato,
M. Kiguchi, and
A. Maki,
“Wavelength dependence of effective pathlength factor in noninvasive optical measurements of human brain functions,”
Jpn. J. Appl. Phys., 45 L361
–L363
(2006). https://doi.org/10.1143/JJAP.45.L361 0021-4922 Google Scholar
S. R. Arridge and
W. R. B. Lionheart,
“Nonuniqueness in diffusion-based optical tomography,”
Opt. Lett., 23 882
–884
(1998). https://doi.org/10.1364/OL.23.000882 0146-9592 Google Scholar
A. Corlu,
T. Durduran,
R. Choe,
M. Schweiger,
E. M. C. Hillman,
S. R. Arridge, and
A. G. Yodh,
“Uniqueness and wavelength optimization in continuous-wave multispectral diffuse optical tomography,”
Opt. Lett., 28 2339
–2341
(2003). https://doi.org/10.1364/OL.28.002339 0146-9592 Google Scholar
A. Corlu,
R. Choe,
T. Durduran,
K. Lee,
M. Schweiger,
S. R. Arridge,
E. M. C. Hillman, and
A. G. Yodh,
“Diffuse optical tomography with spectral constrains and wavelength optimization,”
Appl. Opt., 44 2082
–2093
(2005). https://doi.org/10.1364/AO.44.002082 0003-6935 Google Scholar
S. R. Arridge,
“Optical tomography in medical imaging,”
Inverse Probl., 15 R41
–R93
(1999). https://doi.org/10.1088/0266-5611/15/2/022 0266-5611 Google Scholar
M. A. OLeary,
D. A. Boas,
B. Chance, and
A. G. Yodh,
“Experimental images of heterogeneous turbid media by frequency domain diffusing photon tomography,”
Opt. Lett., 20 4268
(1995). 0146-9592 Google Scholar
M. Cheney,
D. Isaacson,
J. C. Newell,
S. Simske, and
J. C. Goble,
“NOSER: an algorithm for solving the inverse conductivity problem,”
Int. J. Imaging Syst. Technol., 2 66
–75
(1990). https://doi.org/10.1002/ima.1850020203 0899-9457 Google Scholar
A. Adler,
T. Dai, and
W. R. B. Lionheart,
“Temporal image reconstruction in electrical impedance tomography,”
Physiol. Meas, 28 S1
–S11
(2007). https://doi.org/10.1088/0967-3334/28/7/S01 0967-3334 Google Scholar
S. R. Arridge,
“Photon measurement density functions. Part1: analytical forms,”
Appl. Opt., 34 7395
–7409
(1995). https://doi.org/10.1364/AO.34.007395 0003-6935 Google Scholar
S. R. Arridge,
J. C. Hebden,
M. Schweiger,
F. E. W. Schmidt,
M. E. Fry,
E. M. C. Hillman,
H. Dehghani, and
D. T. Delpy,
“A method for three-dimensional time-resolved optical tomography,”
Int. J. Imaging Syst. Technol., 11 2
–11
(2000). https://doi.org/10.1002/(SICI)1098-1098(2000)11:1<2::AID-IMA2>3.0.CO;2-J 0899-9457 Google Scholar
E. Okada,
M. Firbank,
M. Schweiger,
S. R. Arridge,
M. Cope, and
D. T. Delpy,
“Theoretical and experimental investigation of near-infrared light propagation in a model of the adult head,”
Appl. Opt., 36 21
–31
(1997). https://doi.org/10.1364/AO.36.000021 0003-6935 Google Scholar
T. S. Leung,
C. E. Elwell, and
S. T. Delpy,
“Estimation of cerebral oxy- and deoxy-haemoglobin concentration changes in a layered adult head model using near-infrared spectroscopy and multivariate statistical analysis,”
Phys. Med. Biol., 50 5783
–5798
(2005). https://doi.org/10.1088/0031-9155/50/24/002 0031-9155 Google Scholar
D. R. White,
E. M. Widdowson,
H. Q. Woodard, and
J. W. Dickerson,
“The composition of body tissues (II). fetus to young adult,”
Br. J. Radiol., 64 149
–159
(1991). https://doi.org/10.1259/0007-1285-64-758-149 0007-1285 Google Scholar
M. Cope,
“The application of near infrared spectroscopy to non invasive monitoring of cerebral oxygenation in the newborn infant,”
University College London,
(1991). Google Scholar
P. van der Zee,
“Measurement and modelling of the optical properties of human tissue in the near-infrared,”
University College London,
(1992). Google Scholar
I. V. Meglinski and
S. J. Matcher,
“Quantitative assessment of skin layers absorption and skin reflectance spectra simulation in the visible and near-infrared spectral regions,”
Physiol. Meas, 23 741
–753
(2002). https://doi.org/10.1088/0967-3334/23/4/312 0967-3334 Google Scholar
George Zonios,
Julie Bykowski, and
Nikiforos Kollias,
“Skin melanin, hemoglobin, and light scattering properties can be quantitatively assessed in vivo using diffuse reflectance spectroscopy,”
J. Invest. Dermatol., 117 1452
–1457
(2001). https://doi.org/10.1046/j.0022-202x.2001.01577.x 0022-202X Google Scholar
S. L. Jacques,
“Skin optics,”
(1998) http://omlc.ogi.edu/news/jan98/skinoptics.html Google Scholar
E. Okada and
D. T. Delpy,
“Near-infrared light propagation in an adult head model. I. Modeling of low-level scattering in the cerebrospinal fluid layer,”
Appl. Opt., 42 2906
–2914
(2003). https://doi.org/10.1364/AO.42.002906 0003-6935 Google Scholar
S. J. Matcher,
M. Cope, and
D. T. Delpy,
“Use of the water absorption spectrum to quantify tissue chromophore concentration changes in near-infrared spectroscopy,”
Phys. Med. Biol., 39 177
–196
(1994). https://doi.org/10.1088/0031-9155/39/1/011 0031-9155 Google Scholar
R. L. P. van Veen,
H. J. C. M. Sterenborg,
A. Pifferi,
A. Torricelli, and
R. Cubeddu,
“Determination of VIS- NIR absorption coefficients of mammalian fat, with time- and spatially resolved diffuse reflectance and transmission spectroscopy,”
(2004) Google Scholar
S. J. Matcher,
M. Cope, and
D. T. Delpy,
“In vivo measurements of the wavelength dependence of tissue-scattering coefficients between 760 and measured with time-resolved spectroscopy,”
Appl. Opt., 36 386
–396
(1997). https://doi.org/10.1364/AO.36.000386 0003-6935 Google Scholar
M. Firbank,
M. Hiraoka,
M. Essenpreis, and
D. T. Delpy,
“Measurement of the optical properties of the skull in the wavelength range ,”
Phys. Med. Biol., 38 503
–510
(1993). https://doi.org/10.1088/0031-9155/38/4/002 0031-9155 Google Scholar
C. R. Simpson,
M. Kohl,
M. Essenpreis, and
M. Cope,
“Near-infrared optical properties of ex vivo human skin and subcutaneous tissues measured using the monte carlo inversion technique,”
Phys. Med. Biol., 43 2465
–2478
(1998). https://doi.org/10.1088/0031-9155/43/9/003 0031-9155 Google Scholar
A. N. Bashkatov,
E. A. Genina,
V. I. Kochubey, and
V. V. Tuchin,
“Optical properties of human skin, subcutaneous and mucous tissues in the wavelength range from ,”
J. Phys. D, 38 2543
–2555
(2005). https://doi.org/10.1088/0022-3727/38/15/004 0022-3727 Google Scholar
T. Correia,
A. Banga,
N. L. Everdell,
A. P. Gibson, and
J. C. Hebden,
“A quantitative assessment of the depth sensitivity of an optical topography system using a solid dynamic tissue-phantom,”
Phys. Med. Biol., 54 6277
–6286
(2009). https://doi.org/10.1088/0031-9155/54/20/016 0031-9155 Google Scholar
J. Xia,
A. Weaver,
D. E. Gerrard, and
G. Yao,
“Monitoring sarcomere structure changes in whole muscule using diffuse light reflectance,”
J. Biomed. Opt., 11 040504
(2006). https://doi.org/10.1117/1.2234278 1083-3668 Google Scholar
R. Cubeddu,
C. D’Andrea,
A. Pifferi,
P. Taroni,
A. Torricelli, and
G. Valentini,
“Effects of the menstrual cycle on the red and near-infrared optical properties of the human breast,”
Photochem. Photobiol., 72 383
–391
(2000). https://doi.org/10.1562/0031-8655(2000)072<0383:EOTMCO>2.0.CO;2 0031-8655 Google Scholar
Y. Fukui,
Y. Ajichi, and
E. Okada,
“Monte carlo prediction of near-infrared propagation in realistic adult and neonatal head models,”
Appl. Opt., 42 2881
–2887
(2003). https://doi.org/10.1364/AO.42.002881 0003-6935 Google Scholar
S. R. Arridge and
M. Schweiger,
“Photon measurement density functions. Part2: finite element method calculations,”
Appl. Opt., 34 8026
–8037
(1995). https://doi.org/10.1364/AO.34.008026 0003-6935 Google Scholar
S. R. Arridge,
H. Dehghani,
Martin Schweiger, and
E. Okada,
“The finite element model for the propagation of light in scattering media: a direct method for domains with nonscattering regions,”
Med. Phys., 27 252
–264
(2000). https://doi.org/10.1118/1.598868 0094-2405 Google Scholar
H. Dehghani,
S. R. Arridge,
Martin Schweiger, and
D. T. Delpy,
“Optical tomography in the presence of void regions,”
J. Opt. Soc. Am., 17 1659
–1670
(2000). https://doi.org/10.1364/JOSAA.17.001659 0030-3941 Google Scholar
B. Brendel and
T. Nielsen,
“Wavelengths optimization in multispectral diffuse optical tomography considering uncertainties in absorption spectra,”
Proc. SPIE, 6629 66290A
(2007). https://doi.org/10.1117/12.728421 0277-786X Google Scholar
|