1 February 2013 Fiber-optic manipulation of urinary stone phantoms using holmium:YAG and thulium fiber lasers
Author Affiliations +
Abstract
Fiber-optic attraction of urinary stones during laser lithotripsy may be exploited to manipulate stone fragments inside the urinary tract without mechanical grasping tools, saving the urologist time and space in the ureteroscope working channel. We compare thulium fiber laser (TFL) high pulse rate/low pulse energy operation to conventional holmium:YAG low pulse rate/high pulse energy operation for fiber-optic suctioning of plaster-of-paris (PoP) stone phantoms. A TFL (wavelength of 1908 nm, pulse energy of 35 mJ, pulse duration of 500 μs, and pulse rate of 10 to 350 Hz) and a holmium laser (wavelength of 2120 nm, pulse energy of 35 to 360 mJ, pulse duration of 300 μs, and pulse rate of 20 Hz) were tested using 270-μm-core optical fibers. A peak drag speed of ∼2.5  mm/s was measured for both TFL (35 mJ and 150 to 250 Hz) and holmium laser (210 mJ and 20 Hz). Particle image velocimetry and thermal imaging were used to track water flow for all parameters. Fiber-optic suctioning of urinary stone phantoms is feasible. TFL operation at high pulse rates/low pulse energies is preferable to holmium operation at low pulse rates/high pulse energies for rapid and smooth stone pulling. With further development, this novel technique may be useful for manipulating stone fragments in the urinary tract.

1.

Introduction

Our research group has previously studied the experimental thulium fiber laser (TFL) as a potential alternative to the clinical holmium:YAG laser for lithotripsy.15 The superior TFL Gaussian beam profile has been focused to a diameter as small as 70 μm, providing efficient coupling of higher laser energy into smaller fibers, thus leaving more irrigation space in the working channel, without hindering the flexibility of the ureteroscope.1,3 Furthermore, the TFL wavelength (1908 nm) more closely matches a major water absorption peak in tissue than does the holmium laser wavelength (2120 nm), leading to lower ablation thresholds and higher ablation rates.6 Finally, the TFL parameters (e.g., pulse length, pulse rate, and duty cycle) are more adjustable, thus providing more efficient stone ablation with reduced retropulsion and fiber tip degradation.24

During our previous TFL lithotripsy studies, we noticed that urinary stones were periodically attracted toward the fiber tip rather than pushed away from the fiber tip as would be expected from the cavitation-induced pressure wave. This phenomenon appeared more frequently as we conducted studies at higher TFL pulse rates. Previous investigators have reported movement of stones using the holmium laser that could be a result of this “suction effect,”79 but until now, this phenomenon has not been studied in detail. The TFL is capable of operating at low pulse energies and variable pulse rates and is therefore ideally suited to reproduce this effect without risk of stone retropulsion, which has previously been demonstrated to be minimal using the TFL at pulse rates up to 150 Hz.4

Currently, stone stabilization devices utilizing numerous designs, including baskets, grasping tools, and “backstop” polymer material approaches, are being used to reduce stone retropulsion and increase ablation efficiency during holmium laser lithotripsy.10 However, these devices occupy valuable space inside the single working channel of the ureteroscope and add considerable expense to the surgical procedure. The objective of this study is to analyze the suction effect and determine its dependence on holmium laser pulse energy and TFL pulse rate. This study also explores the mechanism, presumably the pressure wave resulting from cavitation bubble collapse, that is responsible for the suction effect.

2.

Methods

2.1.

Laser Parameters

An experimental thulium fiber laser (TLR 110-1908, IPG Photonics, Oxford, Massachusetts) was operated at a wavelength of 1908 nm and externally modulated with a function generator (DS345, Stanford Research Systems, Sunnyvale, California) to produce a pulse duration of 500 μs, similar to previous TFL lithotripsy studies.4,5 The TFL was operated at a constant pulse energy of 35 mJ while varying the pulse rate from 10 to 350 Hz. A clinical holmium:YAG laser (TwoPointOne XE, Coherent, Santa Clara, California) was operated at a wavelength of 2120 nm and fixed pulse duration of 300 μs. The holmium laser was limited to operation at a relatively low pulse rate of 20 Hz while varying the pulse energy from 35 to 360 mJ.

2.2.

Stone Suction Experiments

Spherical, 4-mm-diameter, plaster-of-paris stone phantoms with an average mass of 40.4±2.0mg were formed using a mold and sandpaper to smooth rough spots. These stones were used as an idealized stone model to eliminate potential variability due to stone shape and density. Previous investigators have also used PoP stone phantoms to model urinary stones because of their comparable tensile strength.11,12 Each stone was dried for at least 24 h and weighed using an analytical balance (AB54-S, Mettler-Toledo, Switzerland). The stones were monitored throughout the suction experiments, and any damaged stone was removed and replaced. A minimum of five PoP stone phantoms was used for each set of laser parameters in all of the studies.

Stones were placed in a saline bath on a level, flat surface with ruler markings. Laser energy was delivered through a 270-μm-core optical fiber (Holmium Lightguide 270D, Olympus Gyrus ACMI, Southborough, Massachusetts) to the stone. The experiment was recorded with a camera at a frame rate of 30 Hz (73K3HN-YC, Mintron, Fremont, California). The fiber was positioned parallel to the surface of the saline bath, with energy being delivered slightly off-center, above the stone. The fiber was then pulled away from the stone at the maximum speed allowable to maintain stone movement without detachment. Figure 1 shows the experimental setup used to measure the stone velocity. Figure 2(a) and 2(b) shows representative images of the initial and final stone locations after pulling for 16 s using the holmium laser at a pulse energy of 70 mJ. The velocity for each set of laser parameters was plotted using the recorded distance and time traveled by the stone. For the TFL, velocity versus pulse rate was plotted. For the holmium laser, velocity versus pulse energy was recorded. The fiber was placed slightly off-center and above the stone phantom for optimal manipulation [Fig. 2(c)].

Fig. 1

Experimental setup used to record stone movement during laser fiber-optic manipulation.

JBO_18_2_028001_f001.png

Fig. 2

Before (a) and after (b) snapshots of the 4-mm-diameter PoP stone being pulled by the 270-μm-diameter fiber across a ruled surface in a saline bath using the holmium laser, operating at a pulse rate of 20 Hz and a pulse energy of 70 mJ. (c) Side view of fiber-to-stone orientation. Positioning of the fiber tip slightly off-center, above the stone, provided optimal stone manipulation.

JBO_18_2_028001_f002.png

2.3.

Particle Image Velocimetry

Polymer microspheres (Duke Standards 4000 Series, Thermo Fisher Scientific, Waltham, Massachusetts) with diameters ranging from 30 to 50 μm, refractive index of 1.59, and density of 1.05g/cm3 were suspended in water. The microsphere density closely matched that of the water to ensure suspension in the water bath. The microspheres were illuminated from the side by a fiber-optic lamp. A 270-μm-core fiber was inserted into the bath with the length of the fiber level with the surface. Laser energy was delivered in the water bath while videos of particle flow were recorded under magnification. The recorded particle flow was used to map the flow of water as a function of the laser parameters. This method examined the macroscopic effects of a train of laser pulses on the water flow.

2.4.

Thermal Imaging

A thermal camera (SC655, FLIR, Billerica, Massachusetts) capable of a 50-Hz frame rate and 640×480 resolution was used to track the flow of heated water as laser energy was delivered. The water bath was maintained near body temperature (37°C). Similar to the particle image velocimetry (PIV) experiments described above, a 270-μm-core fiber was inserted into the water bath with the length of the fiber level with the surface. Videos of the thermal signatures from heated water were recorded for TFL pulse rates of 10, 50, 100, 200, and 350 Hz. The thermal signature videos were compared to those using the microspheres as an alternative method to visualize the water flow caused by laser energy delivered in the water bath.

3.

Results

3.1.

Stone Suction Experiments

For the TFL, the effect of a net force pulling the stone toward the trunk end of the fiber was noticeable even at a low pulse energy and pulse rate of 35 mJ and 10 Hz, respectively [Fig. 3(a)]. The effect was weak, and no stone retropulsion was observed. As the pulse rate was increased, the effect of the force pulling the stone toward the fiber tip became stronger. The stone pull velocity increased up to 250 Hz (although the increase between 75 and 250 Hz was not statistically significant). Each pulse caused the water to flow axially in two different directions. The water flow in both directions increased as the laser pulses were delivered at higher rates. The compounding effects of forces from each pulse acted on the stone by either pushing or pulling it away from the fiber tip, depending on where the fiber tip was placed with respect to the stone.

Fig. 3

(a) Stone pull velocity as a function of pulse rate for TFL pulse energy of 35 mJ. (b) Stone pull velocity as a function of pulse energy for a holmium laser pulse rate of 20 Hz.

JBO_18_2_028001_f003.png

Furthermore, as the pulse rate was increased up to 250 Hz, strong retropulsion forces were observed. This was a problem when trying to keep the stone attracted to the fiber tip. Because the effect of pulling or pushing the stone is dependent on where the fiber tip is placed, when the net retropulsive force begins to dominate, the location of the fiber tip is critical. This may be the cause for the high error bars in the midrange pulse rate data points shown in Fig. 3(a). Both retropulsive and attractive forces were acting on the stone, and it was difficult to control which force dominated in the 200-Hz range.

As the pulse rate was increased to 350 Hz, retropulsion forces dominated. The stone was not pulled as far without eventually being rapidly pushed away. One of the reasons for this effect is that the retropulsion forces were so strong that a tightly wound vortex (to be seen in the PIV experiments) developed in close proximity to the fiber tip. Because the increase in water flow away from the fiber was stronger than the increase toward the fiber, the microspheres caught in the vortex were swept in the jet flowing away from the fiber tip.

The trend was similar for the holmium laser [Fig. 3(b)]. As the pulse energy was increased, the stone velocity also increased up to 210mJ (although the increase was not statistically significant for pulse energies between 70 and 210 mJ). The stone velocity then decreased as the pulse energy increased beyond 210 mJ. For operation at 35 mJ, the suction effect was observed without any retropulsion effects. As the pulse energy was increased, the stone was pulled more rapidly, but in a discrete, choppy pattern of motion. The process of pulling the stone was not smooth, as experienced with the TFL. Use of higher pulse energies >70mJ resulted in greater retropulsion, which required the fiber to be held further away from the stone to maintain the suction effect. At pulse energies >210mJ, the stone bounced randomly in all directions and retropulsion was increasingly difficult to avoid. Pulse energies at 300 mJ and higher yielded a much different result than lower energies. The stone was either pushed far ahead of the fiber tip or pushed far behind the fiber tip in a quick, discrete motion. The attractive force at higher pulse energies required the fiber to be placed completely past the stone, with the potential clinical risk of the laser energy delivered by the fiber being absorbed by tissue structures directly in front of the fiber.

Figure 4 shows the stone pull velocity as a function of average power. This plot provides a direct comparison between holmium and TFL. The TFL was able to exploit the suction effect at more than double the power produced by the holmium laser. More power delivery could make it possible for the TFL to potentially exploit the suction effect to maximize stone ablation rates while minimizing stone retropulsion.

Fig. 4

Stone pull velocity for both lasers as a function of average power.

JBO_18_2_028001_f004.png

3.2.

Particle Image Velocimetry

Microspheres ranging from 30 to 50 μm in diameter were used to track the water flow during laser irradiation with both the TFL and holmium. The results of our study show two dominating but opposing forces during lithotripsy. One force, a retropulsive force, is away from the fiber tip. The other force, an attractive force, is toward the fiber tip. The attractive force is suspected to be the cause of the suction effect observed during laser irradiation with both the thulium and holmium lasers.

Figure 5 illustrates the change in flow of the microspheres as the TFL pulse rate was increased. At low pulse rates, the microspheres flowed in two directions, toward and away from the fiber tip [Fig. 5(a)]. As pulse rates were increased, a vortex began to form around the sides of the fiber, causing microspheres pulled toward the fiber to return and become trapped in the jet flowing away from the fiber [Fig. 5(b)]. At pulse rates >250Hz, the vortex surrounding the end of the fiber became tightly wound around the fiber tip [Fig. 5(c)]. This caused microspheres flowing toward the fiber tip to get trapped and then pushed back into the flow away from the fiber tip. This chaotic, tightly wound vortex is suspected to be the primary cause of decreasing stone pull velocity at higher pulse rates in the suction effect experiments.

Fig. 5

Flow of microspheres for TFL pulse rates of 20 (a) 200 (b) and 350 (c) Hz at 35 mJ per pulse.

JBO_18_2_028001_f005.png

The effects of the forces on the microspheres were more clear and consistent using the holmium laser than those of the TFL operating at 20 Hz with regard to the flow of microspheres. The vortices seen at higher pulse rates using the TFL were not seen using the holmium as pulse energies were increased. This could be due to the attractive force becoming stronger at roughly the same rate as the retropulsive force. The experiments using the holmium showed that the speed at which the spheres moved away from the fiber tip, in both directions, increased as the pulse energy was increased.

Although the frame rate of the camera used to capture the flow of microspheres was limited to 30 Hz, it was able to capture the shockwave-induced luminescence resulting from the collapse of the cavitation bubbles created by both TFL and holmium (Fig. 6). Such a distinct signature of cavitation bubbles has been previously reported as well.13,14 Figure 6 also shows distinct differences in the shape and formation of cavitation bubbles created using the TFL [Fig. 6(a)] and the holmium laser [Fig. 6(b)].

Fig. 6

Shockwave-induced luminescence resulting from the collapse of cavitation bubbles for TFL pulse rate of 350 Hz at 35 mJ pulse energy (a) and holmium pulse rate of 20 Hz at 350 mJ pulse energy (b).

JBO_18_2_028001_f006.png

The modulated TFL operates at a longer pulse length (500 μs) than the holmium (300 μs). Furthermore, the absorption coefficient of water at the TFL wavelength (μa=160cm1) is five times greater than that of the holmium (μa=28cm1) (Ref. 1). Previous investigators have shown that for both increased absorption and longer pulse durations, the cavitation bubble becomes elongated and multiple cavitation bubbles may form.1517 Indirect evidence of this is seen in both the results for the TFL microsphere experiments and the shockwave-induced luminescence observed emanating from two distinct points in front of the fiber tip [Fig. 6(a)].

The holmium-induced cavitation bubbles have been shown to be less elongated and pear shaped. This is also supported by the results of the microsphere study performed with the holmium laser [Fig. 6(b)]. The area of luminescence appeared oval and resembled the shape of holmium-induced cavitation bubbles reported by other investigators.12,15,16,18,19

3.3.

Thermal Imaging

A thermal camera was used to track the flow of heated water as laser energy was delivered just beneath the surface of the water bath. For the TFL, the camera recorded a flow of heated water away from the fiber tip and toward the trunk end of the fiber. The thermal camera also recorded small thermal eddies near the fiber tip. Figure 7 shows a recorded video frame. As the TFL pulse rate was increased, the area of thermal flow also increased. Thermal images (not shown here) were also acquired while heating water using the holmium laser, but no discernible eddies or flow pattern were observed. This may be due to the longer optical penetration depth of the holmium laser wavelength in water, resulting in a wider spread of water heating. The videos captured with the thermal camera reinforced the validity of trends observed in the suction effect and particle velocimetry studies.

Fig. 7

Frame of a thermal video captured during fiber-optic delivery of the TFL energy at 35 mJ and 200 Hz in a water bath.

JBO_18_2_028001_f007.png

4.

Discussion

Our laboratory has studied the thulium fiber laser as a potential alternative to the holmium laser in lithotripsy. The TFL Gaussian spatial beam profile is better suited for coupling into smaller optical fibers. Use of smaller fibers in turn allows more flexible ureteroscope deflection and increased irrigation rates during lithotripsy. Furthermore, the TFL wavelength closely matches a major water absorption peak. Because a major absorber of laser energy in kidney stones is water, the higher absorption coefficient translates to more efficient urinary stone ablation. Finally, the TFL is capable of operating at variable pulse durations and rates. We have previously demonstrated that increased TFL pulse rates and complex modulation of pulse trains can lead to more efficient stone ablation while limiting negative effects such as retropulsion or fiber tip degradation. This flexible TFL operation with variable parameters makes it a suitable candidate for exploitation of the suction effect as well.

Previous investigators have reported the role of cavitation bubble dynamics during holmium laser lithotripsy.8,12,16,2022 Some of these reports have shown images of small residual effects of the bubble collapse behind the fiber tip.12,19 This is presumably due to a collapsing bubble–induced pressure wave propagating in the axial direction of the fiber, both toward and away from the fiber tip. Although this pressure wave has been observed to be too weak to induce photomechanical ablation of urinary stones, it has been shown to be sufficiently strong to push stones away from the fiber tip. This study investigated in further detail the pressure wave reported in previous studies that may be responsible for the suction effect.19,20,2325 A recent study has also reported that during the collapse of micrometer-sized bubbles the maximum pressure of the shock wave emitted during bubble rebound can be an order of magnitude smaller than that induced by millimeter-sized bubbles.26 Future studies utilizing a high-speed camera for direct imaging of cavitation bubble dimensions and dynamics during both Ho:YAG and TFL lithotripsy may address how large the bubble must be for the suction effect to be observed.

There are many potential clinical applications for the suction effect. The most direct use would be to manipulate urinary stones in the kidney or bladder. Because the TFL is able to produce the suction effect at pulse rates that do not result in retropulsion, a stone could be trapped at the fiber tip and then transported to a more desirable location for stone ablation. If the stone is sufficiently small, the suction effect could also potentially be used to navigate the stone out of the urinary tract. Furthermore, when using the TFL, the suction effect is dominant over retropulsion at pulse rates ideal for stone ablation (up to 150 Hz).4 This may provide the possibility of trapping the stone at the fiber tip during stone ablation. Ideally, the stone would bounce around the fiber tip while it is broken down. Overall, exploitation of the suction effect would theoretically provide an urologist with greater control during laser lithotripsy without the need for stone stabilization devices.

Although the reproduction and quantification of the suction effect is possible, it requires further study. Our results show high error bars, mainly due to human error in positioning the fiber tip in the most efficient location relative to the stone to utilize the suction effect. A urologist may have even less control positioning the fiber around the stone. This limitation must be overcome before the suction effect can be viable for surgical application. Also, the suction effect is dependent on stone size. The effect is not as strong as stone size increases. The shape of the stone may also play a role in the strength of the suction effect. Finally, our studies were performed in a stable environment. The drag force of motionless saline was the only major force acting against the stone. Other forces, such as saline flow through the working channel of the ureteroscope, may also play a significant role in whether the stone can be trapped by the attractive forces responsible for this suction effect.

Our knowledge of the suction effect is not yet sufficient for use in practical applications. Further studies need to be conducted to overcome these current limitations. Pulse duration, pulse energy, pulse distribution, stone size, stone shape, and environment all need to be further explored in more detail. Computer simulations may also be necessary to assist in the optimization of this large matrix of parameters.

5.

Conclusions

This study has demonstrated the ability of both the holmium:YAG and thulium fiber lasers to rapidly and reproducibly pull stone phantoms with proper placement of the optical fiber and optimal choice of laser parameters. Future studies may focus on the role of this “suction effect” as a tool to manipulate urinary stones during laser lithotripsy. This phenomenon may also ideally be used to limit the movement of stone fragments during ablation, thus potentially eliminating the need for a stone stabilization device.

Acknowledgments

This research was supported, in part, by a Collaborative Research Grant between UNC-Charlotte and the Carolinas Medical Center and a Faculty Research Grant from UNC-Charlotte. Richard Blackmon is supported by a National Science Foundation Graduate Fellowship.

References

1. 

N. J. ScottC. M. CilipN. M. Fried, “Thulium fiber laser ablation of urinary stones through small-core optical fibers,” IEEE J. Sel. Top. Quantum Electron., 15 (2), 435 –440 (2009). http://dx.doi.org/10.1109/JSTQE.2008.2012133 IJSQEN 1077-260X Google Scholar

2. 

R. L. BlackmonP. B. IrbyN. M. Fried, “Holmium:YAG (λ=2, 120 nm) versus thulium fiber (λ=1, 908 nm) laser lithotripsy,” Lasers Surg. Med., 42 (3), 232 –236 (2010). http://dx.doi.org/10.1002/lsm.v42:3 LSMEDI 0196-8092 Google Scholar

3. 

R. L. BlackmonP. B. IrbyN. M. Fried, “Thulium fiber laser lithotripsy using tapered fibers,” Lasers Surg. Med., 42 (1), 45 –50 (2010). http://dx.doi.org/10.1002/lsm.v42:1 LSMEDI 0196-8092 Google Scholar

4. 

R. L. BlackmonP. B. IrbyN. M. Fried, “Comparison of holmium:YAG and thulium fiber laser lithotripsy: ablation thresholds, ablation rates, and retropulsion effects,” J. Biomed. Opt., 16 (7), 071403 (2011). http://dx.doi.org/10.1117/1.3564884 JBOPFO 1083-3668 Google Scholar

5. 

R. L. BlackmonP. B. IrbyN. M. Fried, “Enhanced thulium fiber laser lithotripsy using micro-pulse train modulation,” J. Biomed. Opt., 17 (2), 028002 (2012). http://dx.doi.org/10.1117/1.JBO.17.2.028002 JBOPFO 1083-3668 Google Scholar

6. 

E. D. Jansenet al., “Temperature-dependence of the absorption-coefficient of water for midinfrared laser-radiation,” Lasers Surg. Med., 14 (3), 258 –268 (1994). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

7. 

P. KalraN. B. LeD. Bagley, “Effect of pulse width on object movement in vitro using holmium:YAG laser,” J. Endourol., 21 (2), 228 –231 (2007). http://dx.doi.org/10.1089/end.2005.1130 JENDE3 0892-7790 Google Scholar

8. 

H. Leeet al., “Stone retropulsion during holmium:YAG lithotripsy,” J. Urol., 169 (3), 881 –885 (2003). http://dx.doi.org/10.1097/01.ju.0000046367.49923.c6 JOURAA 0022-5347 Google Scholar

9. 

T. D. Whiteet al., “Evaluation of retropulsion caused by holmium:YAG laser with various power settings and fibers,” J. Endourol., 12 (2), 183 –186 (1998). http://dx.doi.org/10.1089/end.1998.12.183 JENDE3 0892-7790 Google Scholar

10. 

C. J. PagnaniM. El AkkadD. H. Bagley, “Prevention of stone migration with the accordion during endoscopic ureteral lithotripsy,” J. Endourol., 26 (5), 484 –488 (2012). http://dx.doi.org/10.1089/end.2011.0386 JENDE3 0892-7790 Google Scholar

11. 

C. G. Marguetet al., “In vitro comparison of stone retropulsion and fragmentation of the frequency doubled, double pulse Nd:YAG laser and the holmium:YAG laser,” J. Urol., 173 (5), 1797 –1800 (2005). http://dx.doi.org/10.1097/01.ju.0000154341.08206.69 JOURAA 0022-5347 Google Scholar

12. 

H. W. Kanget al., “Dependence of calculus retropulsion on pulse duration during holmium:YAG laser lithotripsy,” Lasers Surg. Med., 38 (8), 762 –772 (2006). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

13. 

C. D. Ohlet al., “Bubble dynamics, shock waves and sonoluminescence,” Phil. Trans. R. Soc. London A, 357 (1751), 269 –294 (1999). http://dx.doi.org/10.1098/rsta.1999.0327 0264-3820 Google Scholar

14. 

A. A. BuzukovV. S. Teslenko, “Sonoluminescense following focusing of laser radiation into liquid,” JETP Lett., 14 (5), 189 –191 (1971). JTPHES 1063-7761 Google Scholar

15. 

M. Frenzet al., “Comparison of the effects of absorption coefficient and pulse duration of 2.12-μm and 2.79-μm radiation on laser ablation of tissue,” IEEE J. Quantum Electron., 32 (12), 2025 –2036 (1996). http://dx.doi.org/10.1109/3.544746 IEJQA7 0018-9197 Google Scholar

16. 

P. Zhonget al., “Transient cavitation and acoustic emission produced by different laser lithotripters,” J. Endourol., 12 (4), 371 –378 (1998). http://dx.doi.org/10.1089/end.1998.12.371 JENDE3 0892-7790 Google Scholar

17. 

K. F. Chanet al., “Erbium:YAG laser lithotripsy mechanism,” J. Urol., 168 (2), 436 –441 (2002). http://dx.doi.org/10.1016/S0022-5347(05)64653-8 JOURAA 0022-5347 Google Scholar

18. 

T. Luet al., “Cavitation effect of holmium laser pulse applied to ablation of hard tissue underwater,” J. Biomed. Opt., 15 (4), 048002 (2010). http://dx.doi.org/10.1117/1.3470092 JBOPFO 1083-3668 Google Scholar

19. 

T. AsshauerK. RinkG. Delacretaz, “Acoustic transient generation by holmium-laser-induced cavitation bubbles,” J. Appl. Phys., 76 (9), 5007 –5013 (1994). http://dx.doi.org/10.1063/1.357212 JAPIAU 0021-8979 Google Scholar

20. 

K. F. Chanet al., “Holmium:YAG laser lithotripsy: a dominant photothermal ablative mechanism with chemical decomposition of urinary calculi,” Lasers Surg. Med., 25 (1), 22 –37 (1999). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

21. 

S. S. Sporeet al., “Holmium:YAG lithotripsy: optimal power settings,” J. Endourol., 13 (8), 559 –566 (1999). http://dx.doi.org/10.1089/end.1999.13.559 JENDE3 0892-7790 Google Scholar

22. 

A. Vogel, “Nonlinear absorption: intraocular microsurgery and laser lithotripsy,” Phys. Med. Biol., 42 (5), 895 –912 (1997). http://dx.doi.org/10.1088/0031-9155/42/5/011 PHMBA7 0031-9155 Google Scholar

23. 

B. Hanet al., “Mechanical effects of laser-induced cavitation bubble on different geometrical confinements for laser propulsion in water,” Opt. Lasers Eng., 49 (3), 428 –433 (2011). http://dx.doi.org/10.1016/j.optlaseng.2010.11.012 OLENDN 0143-8166 Google Scholar

24. 

E. A. Brujanet al., “The final stage of the collapse of a cavitation bubble close to a rigid boundary,” Phys. Fluids, 14 (1), 85 –92 (2002). http://dx.doi.org/10.1063/1.1421102 PHFLE6 1070-6631 Google Scholar

25. 

A. VogelW. Lauterborn, “Acoustic transient generation by laser-produced cavitation bubbles near solid boundaries,” J. Acoust. Soc. Am., 84 (2), 719 –731 (1988). http://dx.doi.org/10.1121/1.396852 JASMAN 0001-4966 Google Scholar

26. 

A. BrujanY. Matsumoto, “Collapse of micrometer-sized cavitation bubbles near a rigid boundary,” Microfluid. Nanofluid., 13 (6), 957 –966 (2012). http://dx.doi.org/10.1007/s10404-012-1015-6 1613-4982 Google Scholar
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Richard L. Blackmon, Jason R. Case, Susan R. Trammell, Pierce B. Irby M.D., and Nathaniel M. Fried "Fiber-optic manipulation of urinary stone phantoms using holmium:YAG and thulium fiber lasers," Journal of Biomedical Optics 18(2), 028001 (1 February 2013). https://doi.org/10.1117/1.JBO.18.2.028001
Published: 1 February 2013