22 May 2014 Modified forward model for eliminating the time-varying impact in fluorescence molecular tomography
Wei He, Guanglei Zhang, Huangsheng Pu, Fei Liu, Xu Cao, Jianwen Luo, Jing Bai
Author Affiliations +
Abstract
In conventional fluorescence molecular tomography, the distribution of fluorescent contrast agents is reconstructed with the assumption of constant concentration during data acquisition for each image frame. However, the concentration of fluorescent contrast target is usually time-varying in experiments or in-vivo studies. In this case, the reconstruction methods cannot be directly applied to the fluorescence measurements without considering the time-varying effects of concentration. We propose a modified forward model by dividing the fluorescence yield distribution into two parts: one is a constant representing the spatial distribution of the fluorescent target and the other is an impact factor representing the effects of the concentration change and other possible factors. By extracting spatial distribution information from the reconstruction result, the location and volume of the fluorescent target can be obtained accurately. Both simulation and phantom experiments are carried out and the results indicate that, by using the modified forward model, the quality of reconstruction could be significantly improved in terms of accurate localization and strong anti-noise ability.

1.

Introduction

Fluorescence molecular tomography (FMT) is an important molecular imaging modality developed to reveal three-dimensional (3-D) distribution of the fluorescent targets using fluorescent measurements and appropriate image reconstruction methods. It has the advantages of low cost, high sensitivity, noninvasiveness, and no ionization radiation.13 During the past years, FMT has made significant progress with developments the in imaging systems,46 mathematical models,710 and fluorescent agents.11 FMT has been extensively applied to visualization of biological and medical processes in vivo at cellular and molecular levels and provides a powerful technical tool for disease detection, drug discovery, etc.1214

Conventionally, the images of FMT are reconstructed based on the (implicit) assumption that the fluorophore concentration is constant during the acquisition process of measurement data for each image frame.13 The temporal resolution of FMT is about 1 to 2 min depending on the sum of rotation time and total exposure time of projection images.1517 The assumption holds even in advanced reconstruction methods that attempt to obtain dynamic FMT images frame by frame.1620

However, many reports have demonstrated that the concentration of fluorescent target is time-varying in experiments or in-vivo studies due to physical degradation of fluorophores21 (photodegradation, thermal degradation, etc.) or the process of metabolism,22 which means that the concentration may exhibit significant changes during the time of measurement data acquisition for one image frame.16,17,22 In this case, if the reconstruction methods are directly applied to the fluorescence measurements without considering the time-varying effects of concentration, inaccurate localization of fluorescent target and a high level of noise will be obtained.23,24 To acquire high quality reconstruction results, the time-varying information of the concentration should be taken into account.

In this study, we propose a modified forward model based on the compartment model theory,18 a well-known approach for pharmacokinetic analysis, to eliminate the impacts due to the concentration change of fluorescent targets. In this model, the unknown fluorescence yield distribution is divided into two parts. One part is a constant representing the spatial distribution (location and volume) of the fluorescent target and the other is an impact factor representing the effects of the concentration change and other possible factors. This is because the concentration of fluorescent targets in the region of interest (ROI) may vary significantly with time, but the location and volume of the ROI filled with fluorescent targets remains invariant. Based on the modified model, a new linear relationship between the unknown and the boundary measurement can be constructed, from which an accurate and constant spatial distribution of the fluorescent target can be obtained with Tikhonov regularization. Simulation studies and phantom experiments are conducted to qualitatively and quantitatively illustrate the improvement in the localization of fluorophore and anti-noise abilities with the proposed model.

The outline of this paper is as follows. In Sec. 2, the details of the proposed model are described. The materials used in the simulation studies and phantom experiments are described in Sec. 3. In Sec. 4, the reconstruction results are presented to show the performance of the proposed model. Finally, discussions and conclusion are provided in Sec. 5.

2.

Method

2.1.

Conventional Forward Model

A suitable model for photon propagation plays a crucial role in fluorescence tomographic reconstruction. Diffusion equation (DE), as an approximation to the radiative transfer equation, has been successfully used to describe the light transport in the near-infrared spectral window.25 For FMT, coupled DEs, which describe the photon migration in the excitation and emission spectrum, can be written as follows:26

Eq. (1)

{·[Dx(r)Φx(r)]+μax(r)Φx(r)=s(r)·[Dm(r)Φm(r)]+μam(r)Φm(r)=Φx(r)x(r),
where the subscripts x and m denote the excitation and emission wavelengths, respectively; s(r) is the excitation source; Dx,m is the diffusion coefficient given by Dx,m=1/3[μsx,m(r)+μax,m(r)], where μsx,m(r) is the reduced scattering coefficient; μax,m(r) stands for the absorption coefficient of the medium; Φx,m(r) represents the photon flux density; x(r) denotes the fluorescence yield, which is directly proportional to the fluorophore concentration.27

The corresponding Robin-type boundary condition25 is given by

Eq. (2)

Φx,m(r)+2ρDx,m(r)Φx,m(r)n=0,
where n denotes the outward normal of the boundary and ρ is the boundary mismatch parameter and accounts for the light reflection on the boundary surface.

In this paper, the absorption coefficient and reduced scattering coefficient are assumed to be the same at the excitation and emission wavelengths, which is reasonable because the excitation and emission wavelengths are close to each other.28

In the finite element framework,29 a linear relationship between the unknown fluorescence distribution and the boundary measurements of the s’th projection can be constructed as

Eq. (3)

Wsxs=Φs(s=1,2,,L),
where L is the number of projections for one image frame. Ws=(Ws,d) is the subsystem matrix, which is composed of the mapping vector of all source-detect pairs in the s’th projection. Φs=[Φ(rs)] is the vector composed of the fluorescence data of the s’th projection. xs is the distribution of the fluorescence in the s’th projection.

In the conventional FMT, the concentration of the fluorescent target is assumed to be constant during data acquisition for one image frame. It means that xs is constant at all the projections (s=1,2,,L). Thus, the final system equation of FMT can be constructed with the subsystem equations of all the projections

Eq. (4)

Wx=Φ,
where x is an N×1 vector representing the fluorescent probe distribution to be reconstructed. Φ is an M×1 vector which contains the emission measurements at the boundary based on charge-coupled device (CCD) camera photon detection, and W is an M×N weight matrix.

2.2.

Modified Forward Model

As described in Sec. 1, xs is usually time-varying during measurement data acquisition at different projections. It means the final system equation cannot be obtained by directly combining the subsystem equations together.

In the modified forward model, xs consists of two components as follows:

Eq. (5)

xs=ks*x0,
where x0 is an N×1 vector representing the spatial distribution of the fluorescent targets. Here, the spatial distribution specifically refers to the location and volume of the fluorescent target, which is changeless with all the projections. ks(s=1,2,,L) is defined as an impact factor in the s’th projection (ks>0), representing the effects of the concentration change and other possible factors.

Then, the new subsystem equation can be obtained by substituting Eq. (5) into Eq. (3):

Eq. (6)

Ws*x0=Φsks.

In order to make the mathematical solution possible, we define ks as

Eq. (7)

ks=11ps(ps<1),
where ps is meaningless in physics. For simplicity, ps is used to indicate the impact factor in the following part of this paper.

By substituting Eq. (7) into Eq. (6), we obtain

Eq. (8)

Ws*x0+Φs*ps=Φs.

Thus, we can obtain the modified system matrix of FMT from Eq. (8)

Eq. (9)

A=(W1Φ10WsΦsWL0ΦL).

Then, the new unknown composed of the spatial distribution of the fluorescent target and the impact factor can be written as

Eq. (10)

y=(x0p),
where p=(p1,p2,,pL) is an L×1 vector and y is an (N+L)×1 vector.

The final system equation of the FMT can be written as

Eq. (11)

A*y=Φ,
where Φ is still an M×1 vector and A is an M×(N+L) weight matrix. After extracting spatial distribution information x0 from y, the location and volume of the fluorescent target can be obtained.

2.3.

Tikhonov Regularization

For Eqs. (4) or (11), the estimation of x or y from the measurement vector Φ is an ill-posed inverse problem due to the highly diffusive nature of the photon propagation in tissues.29 Here, the Tikhonov regularization method30 is implemented to address this problem, as it could often provide a reasonable solution and could be efficiently solved by now-standard minimization tools. In the time-varying case as described by Eq. (11), the objective function can be stated as

Eq. (12)

E(y)=ΦAy2+λLy2,
where λ is the regularization parameter (λ>0) and is empirically set, and L is an (N+L)×(N+L) identity regularization matrix.

By minimizing the objective function E(y) given in Eq. (12) with respect to y, the fluorescence distribution [i.e., x0 in Eq. (10)] can be obtained. The minimization of E(y) is performed using the least-squares algorithm.31 The process for solving Eq. (4) is the same as that for Eq. (11).

2.4.

Evaluation Metrics

In this paper, the contrast-to-noise ratio (CNR)32 and relative error (RE) are selected as the evaluation indices to quantitatively assess the quality of the reconstructed images.

The contrast of the regions with the fluorescent target to the background can be quantified by the CNR, and a higher CNR implies a better image quality. Given an ROI where the fluorescent target is confined, the CNR is defined by

Eq. (13)

CNR=μROIμBCK(ωROIσROI2+ωBCKσBCK2)1/2,
where μROI and μBCK are the mean fluorophore concentration values in the ROI and background, respectively; σROI2 and σBCK2 are the variances in the ROI and background, respectively; ωROI and ωBCK are the weighting factors determined by the relative volumes of the ROI and background, respectively.

The RE can comprehensively quantify the reconstruction accuracy and a smaller RE implies a better image quality. The RE is defined as

Eq. (14)

RE=XXactualXactual,
where Xactual is the actual distribution of the fluorescent target and X is the reconstruction result.

3.

Materials

3.1.

Simulation Studies

In the simulation studies, a virtual mouse atlas was employed to provide the 3-D anatomical information.33,34 We focused on imaging the kinetic behaviors of indocyanine green (ICG) in the lungs and liver, and the mouse torso from the neck to the base of the liver was selected as the investigated region, with a length of 2.9 cm. The rotational axis of the mouse was defined as the Z-axis with the bottom plane set as Z=0cm.

To simulate the photons’ propagation in biological tissues, a heterogeneous mouse model was set up. It was composed of the heart, lungs, liver, and kidneys, as shown in Fig. 1(a). The optical parameters from Ref. 35 were assigned to the corresponding organs. To simplify the problem, optical properties outside the organs were regarded as homogeneous.

Fig. 1

Schematic diagram of the simulation model. (a) The three-dimensional (3-D) geometric model of the digital mouse used in the simulations with a length of 2.9 cm from the neck to the base of the live (surface: gray; heart: yellow; lungs: cyan; liver: pink; and kidneys: purple). The red points in (a) indicate the position (Z=1.5cm) of 24 isotropic point sources. For each excitation source, the fluorescence is measured from the opposite side within a 160 field-of-view. (b) Time course of ICG in the lungs and liver. The cyan and pink curves are the actual time course of ICG in the lungs and liver, respectively, which are obtained from Ref. 22. Case 1 (lungs: red point and liver: blue point) represents the concentration of ICG in a nearly steady state (from 1000 to 1072 s). Case 2 (lungs: black point and liver: green point) represents the concentration of ICG in a steep rise period (from 60 to 132 s).

JBO_19_5_056012_f001.png

With the image model, the time course of ICG as it circulated in the lungs and liver following the tail vein injection was simulated. The true time course of ICG in the lungs and liver was obtained from Ref. 22 and is shown in Fig. 1(b). Twenty-four projections evenly distributed over 360 deg were used to acquire the measurement data of one image frame. The speed of the rotation stage was set to 6deg/s and the CCD exposure time in each projection was 0.5 s. Thus, the interval of each projection was 3 s and the time resolution of the FMT system was about 72 s. Here, two periods in the time course of ICG were selected: one from 1000 to 1072 s, when the concentrations of ICG in the lungs and liver are in a nearly steady state within 24 projections [case 1 in Fig. 1(b)]; the other from 60 to 132 s, when the concentrations of ICG in the lungs and liver are in a steep rise period [case 2 in Fig. 1(b)].

3.2.

Phantom Experiments

The performance of the proposed model was further evaluated in phantom experiments based on a noncontact full-angle FMT system previously developed by our laboratory.36 The sketch of the system is shown in Fig. 2. Briefly, the imaged object was placed on a 360 deg rotation stage working under a step-by-step mode. Around the rotational stage, a noncontact full-angle FMT system was positioned to acquire the fluorescence measurement data.

Fig. 2

The schematic of the noncontact full-angle fluorescence molecular tomography system.

JBO_19_5_056012_f002.png

Figure 3 shows the schematic diagram of the phantom experiments. A glass cylinder (3.0 cm in diameter and 4.5 cm in height) filled with a mixture of water and intralipid was employed as the homogeneous phantom (μa=0.02cm1 and μs=10.0cm1). Two small transparent glass tubes (0.5 cm in diameter) filled with different concentrations of ICG were immerged in the cylinder phantom and used as the fluorescent targets. A point excitation source was located at approximately half the height of the phantom. For the acquisition of the fluorescence images, a 775±6nm excitation filter and an 840±6nm emission filter were separately placed in the front of the point excitation source and the CCD camera.

Fig. 3

The schematic diagram of the phantom experiments. (a) The side view of the phantom. (b) The top view of the phantom. A glass cylinder (3.0 cm in diameter and 4.5 cm in height) containing a mixture of water and intralipid was employed as the phantom. Two transparent glass tubes (0.5 cm in outer diameter) filled with different concentrations of ICG in each projection were immerged into the phantom to serve as the fluorescent targets. (c) ICG concentration curves in the two tubes. Each point in (c) denotes the ICG concentration in each projection.

JBO_19_5_056012_f003.png

During the data acquisition process, the phantom was placed on the rotation stage and rotated for 360 deg. For the whole circle, 18 fluorescence images were acquired with an angular increment of 20 deg. In each projection angle, the two tubes shown in Fig. 3(b) were both filled in ICG with the predefined concentrations shown in Fig. 3(c), and one fluorescence image was acquired. After finishing the data acquisition of a projection angle, the cylinder phantom was rotated to the next projection angle, the two tubes were filled in ICG with the next predefined concentrations shown in Fig. 3(c), and another fluorescence image was acquired for the new projection angle. The curves shown in Fig. 3(c) were the predefined concentrations of the two fluorescent targets in the phantom experiments, and were used to imitate the time-varying ICG concentrations of the mouse lungs and liver in the steep rise period (case 2 of the simulation studies) described in Sec. 3.1. For simplicity, the concentrations of the two fluorescent targets were set to increase linearly in the 18 projection angles because the concentrations of the mouse lungs and liver in case 2 of the simulation studies also increased approximate linear.

4.

Results

All computations in this paper were carried out on a personal computer with Intel(R) Core(TM) i7 CPU (2.80 GHz) and 8 GB RAM. In the simulation studies and phantom experiments, the reconstruction results of the spatial distribution x0 were shown while the impact fact p was not provided because this work focused on obtaining accurate locations and volumes of the fluorescent targets. Since the locations and volumes of the fluorescent targets should be the same under different time-varying concentration situations, the spatial distribution x0, which represented the locations and volumes of the fluorescent targets, was normalized by its maximum in both the simulation studies and phantom experiments.

4.1.

Reconstruction Results of Simulations

The reconstruction results based on the conventional and modified forward models in the simulation studies are illustrated in Figs. 4 and 5. The results are normalized by the maximum and presented in the form of the sections. The two red solid lines on the 3-D image indicate the positions of the two investigated section images which are the representative slices of the lungs and the liver, respectively. In both Figs. 4 and 5, (b) and (e) are the actual distributions of the fluorescent targets, (c) and (f) are the reconstruction results based on the conventional forward model, and (d) and (g) are the reconstruction results based on the modified forward model.

Fig. 4

The reconstruction result of case 1 in the simulation studies. (a) Schematic diagram of the simulation model. (b and e) The actual spatial distribution of the fluorescent targets. (c and f) The reconstruction results based on the conventional forward model. (d and g) The reconstruction results based on the modified forward model. Panels (b)–(d) are the representative slice of the lungs and panels (e)–(g) are the representative slice of the liver. All the reconstructed images are normalized by the maximum of the results.

JBO_19_5_056012_f004.png

Fig. 5

The reconstruction result of case 2 in the simulation studies. (a) Schematic diagram of the simulation model. (b and e) The actual spatial distribution of the fluorescent targets. (c and f) The reconstruction results based on the conventional forward model. (d and g) The reconstruction results based on the modified forward model. Panels (b)–(d) are the representative slice of the lungs and panels (e)–(g) are the representative slice of the liver. All the reconstructed images are normalized by the maximum of the results.

JBO_19_5_056012_f005.png

Figure 4 shows the results of case 1, where the concentrations of ICG in the lungs and liver are in a nearly steady state within 24 projections. For this case, both models could obtain high quality results with accurate locations of the fluorescent targets and small noise level. Table 1 shows the quantitative results of case 1. The CNR and RE of the reconstruction results obtained by the two models are similar to each other. In addition, the computational times of the two models are also similar to each other.

Table 1

Quantitative analysis of the reconstruction results in the simulation studies.

ModelContrast-to-noise ratio (CNR)Relative error (RE)Time (s)
Case 1Conventional3.770.3140.9
Modified4.520.2741.7
Case 2Conventional1.470.8741.1
Modified3.330.4142.3

Figure 5 shows the results of case 2, where the concentrations of ICG in the lungs and liver are in a steep rise period. Due to the significant change of concentrations, the reconstruction results based on the conventional forward model show inaccurate locations and volumes of the fluorescent targets, and suffer from a high level of noise [see Figs. 5(c) and 5(f)]. In contrast, a significant improvement in locations and volumes of the fluorescent targets and anti-noise ability is obtained with the modified forward model [see Figs. 5(d) and 5(g)]. Table 1 also shows the quantitative results of case 2. It can be seen that the CNR obtained with the modified model is higher than that with the conventional model, while the RE obtained with the modified model is smaller than that with the conventional model. The computational times of the two models are similar to each other.

4.2.

Reconstruction Results of Phantom Experiments

The reconstruction results based on the conventional and modified forward models in the phantom experiments are illustrated in Fig. 6. The reconstruction results are normalized by the maximum and presented in the form of the sections and 3-D views. The red solid line on the 3-D image indicates the position of the investigated section image. Figures 6(a) and 6(d) are the actual fluorescent targets. Figures 6(b) and 6(e) are the reconstruction results based on the conventional forward model. Figures 6(c) and 6(f) are the reconstruction results based on the modified forward model.

Fig. 6

Sections and 3-D views of the reconstruction result in the phantom experiments. (a and d) The actual fluorescent targets. (b and e) The reconstruction result based on the conventional forward model. (c and f) The reconstruction result based on the modified forward model. The red solid line on the 3-D image (d–f) indicates the position of the investigated two-dimensional image (a–c). All the images are normalized by the maximum of the results.

JBO_19_5_056012_f006.png

Due to the significant change of ICG concentration in the tube, the reconstruction results based on the conventional forward model show wrong locations and volumes of the fluorescent targets, and suffer from a high level of noise [Figs. 6(b) and 6(e)]. On the contrary, accurate locations and volumes of the fluorescent targets and less noisy results are obtained with the proposed model [Figs. 6(c) and 6(f)]. Table 2 shows the quantitative results of the phantom experiments. The CNR of the reconstruction results obtained with the modified model is significantly higher than that with the conventional model. The RE obtained with the modified model is much smaller than that with the conventional model. The computational times of the two models are similar to each other.

Table 2

Quantitative analysis of the reconstruction results in the phantom experiments.

ModelCNRRETime (s)
Conventional4.911.1149.8
Modified8.630.3450.9

5.

Discussion

FMT plays an important role in revealing biological processes in vivo by 3-D visualization of fluorescent targets. However, when the concentrations of the fluorescent targets are time-varying during data acquisition for one image frame, the reconstruction of FMT based on the assumption of constant fluorophore concentration could not obtain accurate locations and volumes of the fluorescents targets, but instead bring a lot of noise into the results. The main aim of this paper is to improve the performance of FMT when the concentrations of the fluorescent targets exhibit significant changes during data acquisition. This is achieved with a modified forward model, which divides the fluorescence yield distribution into two parts. x0 is a constant representing the spatial distribution of the fluorescent target and ks is an impact factor representing the effects of the concentration change and other possible factors. Based on the modified forward model, a new linear relationship between the unknown and the boundary measurement (A*y=Φ) can be constructed for reconstruction. Afterward, the spatial distribution x0 can be exacted from the reconstruction result y.

In the simulations, when the concentrations of ICG in the lungs and liver are nearly constant within different projections during data acquisition [case 1 in Fig. 1(b)], high quality results [Figs. 4(c) and 4(f)] could be obtained with both the conventional and modified forward models. In particular, the difference between the reconstruction results obtained with both models could be neglected [see Fig. 4 and Table 1 (case 1)]. This suggests that the reconstruction based on both models is appropriate for the case with time-invariant concentrations of fluorescent targets.

Nevertheless, when the fluorophore concentrations severely change at different projections, the reconstruction based on the conventional forward model would be seriously affected. Figures 5(c), 5(f), 6(b), and 6(e) showed that the reconstruction results suffered from a high level of noise, and a very serious deviation both in the locations and volumes of the fluorescent targets, in both simulations and phantom experiments. In contrast, the results in Figs. 5(d), 5(g), 6(c), and 6(f) demonstrated that the modified forward model could significantly improve the localization and anti-noise capability of the reconstruction. In addition, the quantitative results of CNRs and REs in Tables 1 (case 2) and 2 also illustrated a higher image quality by using the proposed model.

It should be noted that this work focuses on obtaining an accurate spatial distribution x0 which represents the locations and volumes of the fluorescent targets, rather than the time-varying concentrations of the fluorescent targets, because the scale factor ks is a single value which represents the overall effects of the concentration change. In order to obtain the time-varying concentration curves, ks should be made spatially varying and it will become a vector with the same length as x0. This will significantly aggravate the ill-posedness of the FMT inverse problem, since the number of unknowns is doubled. Nevertheless, with the time-varying concentration curves obtained, the proposed modified model will be further perfected. Furthermore, the kinetic parameters which have significant physiological and pathological meanings can also be estimated by analyzing the time-varying concentration curves.37,38

In conclusion, we proposed a modified forward model for the case with fluorophore concentration changes during the acquisition process of measurement data, which could significantly improve the quality of the reconstruction in terms of accurate localization and strong anti-noise ability. Reconstruction results of both simulation and phantom studies have verified the effectiveness of the proposed model. Future work will focus on validation of this model in in-vivo experiments.

Acknowledgments

This work is supported by the National Basic Research Program of China (973) under Grant No. 2011CB707701; the National Natural Science Foundation of China under Grant Nos. 81227901, 81271617, 61322101, and 61361160418; the National Major Scientific Instrument and Equipment Development Project under Grant No. 2011YQ030114.

References

1. 

V. Ntziachristoset al., “Looking and listening to light: the evolution of whole-body photonic imaging,” Nat. Biotechnol., 23 313 –320 (2005). http://dx.doi.org/10.1038/nbt1074 NABIF9 1087-0156 Google Scholar

2. 

V. Ntziachristos, “Going deeper than microscopy: the optical imaging frontier in biology,” Nat. Methods, 7 603 –614 (2010). http://dx.doi.org/10.1038/nmeth.1483 1548-7091 Google Scholar

3. 

G. LukerK. Luker, “Optical imaging: current applications and future directions,” J. Nucl. Med., 49 (1), 1 –4 (2007). http://dx.doi.org/10.2967/jnumed.107.045799 JNMEAQ 0161-5505 Google Scholar

4. 

V. Ntziachristoset al., “In vivo tomographic imaging of near-infrared fluorescent probes,” Mol. Imaging, 1 (2), 82 –88 (2002). http://dx.doi.org/10.1162/153535002320162732 MIOMBP 1535-3508 Google Scholar

5. 

G. HuJ. YaoJ. Bai, “Full-angle optical imaging of near-infraredfluorescent probes implanted in small animals,” Prog. Nat. Sci., 18 (6), 707 –711 (2008). http://dx.doi.org/10.1016/j.pnsc.2007.11.021 PNASEA 1002-0071 Google Scholar

6. 

T. LasserV. Ntziachristos, “Optimization of 360° projection fluorescence molecular tomography,” Med. Image Anal., 11 (4), 389 –399 (2007). http://dx.doi.org/10.1016/j.media.2007.04.003 MIAECY 1361-8415 Google Scholar

7. 

K. RenG. BalA. H. Hielscher, “Transport- and diffusion-based optical tomography in small domains: a comparative study,” Appl. Opt., 46 (27), 6669 –6679 (2007). http://dx.doi.org/10.1364/AO.46.006669 APOPAI 0003-6935 Google Scholar

8. 

A. SoubretJ. RipollV. Ntziachristos, “Accuracy of fluorescent tomography in the presence of heterogeneities: study of the normalized born ratio,” IEEE Trans. Med. Imaging, 24 (10), 1377 –1386 (2005). http://dx.doi.org/10.1109/TMI.2005.857213 ITMID4 0278-0062 Google Scholar

9. 

D. Wanget al., “A novel finite-element-based algorithm for fluorescence molecular tomography of heterogeneous media,” IEEE Trans. Inf. Technol. Biomed., 13 (5), 766 –773 (2009). http://dx.doi.org/10.1109/TITB.2009.2015144 ITIBFX 1089-7771 Google Scholar

10. 

D. Hydeet al., “Data specific spatially varying regularization for multi-modal fluorescence molecular tomography,” IEEE Trans. Med. Imaging, 29 (2), 365 –374 (2010). http://dx.doi.org/10.1109/TMI.2009.2031112 ITMID4 0278-0062 Google Scholar

11. 

M. C. Pierceet al., “Optical contrastagents and imaging systems for detection and diagnosis of cancer,” Int. J. Cancer, 123 (9), 1979 –1990 (2008). http://dx.doi.org/10.1002/ijc.v123:9 IJCNAW 1097-0215 Google Scholar

12. 

E. GravesR. WeisslederV. Ntziachristos, “Fluorescence molecular imaging of small animal tumor models,” Curr. Mol. Med., 4 (4), 419 –430 (2004). http://dx.doi.org/10.2174/1566524043360555 CMMUBP 1566-5240 Google Scholar

13. 

N. C. Deliolaniset al., “In-vivo imaging of murine tumors using complete-angle projection fluorescence molecular tomography,” J. Biomed. Opt., 14 (3), 030509 (2009). http://dx.doi.org/10.1117/1.3149854 JBOPFO 1083-3668 Google Scholar

14. 

M. RudinR. Weissleder, “Molecular imaging in drug discovery and development,” Nat. Rev. Drug Discov., 2 123 –131 (2003). http://dx.doi.org/10.1038/nrd1007 NRDDAG 1474-1776 Google Scholar

15. 

X. Caoet al., “Accelerated image reconstruction in fluorescence molecular tomography using dimension reduction,” Biomed. Opt. Exp., 4 (1), 1 –14 (2013). http://dx.doi.org/10.1364/BOE.4.000001 BOEICL 2156-7085 Google Scholar

16. 

X. Liuet al., “Unmixing dynamic fluorescence diffuse optical tomography images with independent component analysis,” IEEE Trans. Med. Imaging, 30 (9), 1591 –1604 (2011). http://dx.doi.org/10.1109/TMI.2011.2134865 ITMID4 0278-0062 Google Scholar

17. 

X. Liuet al., “Principal component analysis of dynamic fluorescence diffuse optical tomography images,” Opt. Exp., 18 (6), 6300 –6314 (2010). http://dx.doi.org/10.1364/OE.18.006300 OPEXFF 1094-4087 Google Scholar

18. 

B. AlacamB. Yazici, “Direct reconstruction of pharmacokineticrate images of optical fluorophores from NIR measurements,” IEEE Trans. Med. Imaging, 28 (9), 1337 –1353 (2009). http://dx.doi.org/10.1109/TMI.2009.2015294 ITMID4 0278-0062 Google Scholar

19. 

B. Alacamet al., “Pharmacokinetic-rate images of indocyanine green for breast tumors using near-infrared optical methods,” Phys. Med. Biol., 53 (4), 837 –859 (2008). http://dx.doi.org/10.1088/0031-9155/53/4/002 PHMBA7 0031-9155 Google Scholar

20. 

S. V. Patwardhanet al., “Time-dependent whole-body fluorescence tomography of probe bio-distributions in mice,” Opt. Exp., 13 (7), 2564 –2577 (2005). http://dx.doi.org/10.1364/OPEX.13.002564 OPEXFF 1094-4087 Google Scholar

21. 

V. SaxenaM. SadoqiJ. Shao, “Indocyanine green–loaded biodegradable nanoparticles: preparation, physiochemical characterization, and in vitro release,” Int. J. Pharm., 278 (2), 293 –301 (2004). http://dx.doi.org/10.1016/j.ijpharm.2004.03.032 IJPHDE 0378-5173 Google Scholar

22. 

E. M. C. HillmanA. Moore, “All-optical anatomical co-registration for molecular imaging of small animals using dynamic contrast,” Nat. Photonics, 1 (9), 526 –530 (2007). http://dx.doi.org/10.1038/nphoton.2007.146 1749-4885 Google Scholar

23. 

V. Kolehmainenet al., “State-estimation approach to the nonstationary optical tomography problem,” J. Opt. Soc. Am. A, 20 (5), 876 –889 (2003). http://dx.doi.org/10.1364/JOSAA.20.000876 JOAOD6 0740-3232 Google Scholar

24. 

S. Princeet al., “Time-series estimation of biological factors in optical diffusion tomography,” Phys. Med. Biol., 48 (11), 1491 –1504 (2003). http://dx.doi.org/10.1088/0031-9155/48/11/301 PHMBA7 0031-9155 Google Scholar

25. 

M. Schweigeret al., “The finite element method for the propagation of light in scattering media: boundary and source conditions,” Med. Phys., 22 (11), 1779 –1792 (1995). http://dx.doi.org/10.1118/1.597634 MPHYA6 0094-2405 Google Scholar

26. 

M. A. O’Leary, “Imaging with diffuse photon density waves,” Univ. of Pennsylvania, (1996). Google Scholar

27. 

M. L. Landsmanet al., “Light absorbing properties, stability, and spectral stabilization of indocyanine green,” J. Appl. Physiol., 40 (4), 575 –583 (1976). JAPYAA 0021-8987 Google Scholar

28. 

X. Gaoet al., “In vivo cancer targeting and imaging with semiconductor quantum dots,” Nat. Biotechnol., 22 (8), 969 –976 (2004). http://dx.doi.org/10.1038/nbt994 NABIF9 1087-0156 Google Scholar

29. 

S. R. Arridge, “Optical tomography in medical imaging,” Inverse Probl., 15 R41 –R93 (1999). http://dx.doi.org/10.1088/0266-5611/15/2/022 INPEEY 0266-5611 Google Scholar

30. 

P. K. Yalavarthyet al., “Structural information within regularization matrices improves near infrared diffuse optical tomography,” Opt. Exp., 15 (13), 8043 –8058 (2007). http://dx.doi.org/10.1364/OE.15.008043 OPEXFF 1094-4087 Google Scholar

31. 

C. C. PaigeM. A. Saunders, “LSQR: an algorithm for sparse linear equations and sparse least squares,” ACM Trans. Math. Software, 8 (1), 43 –71 (1982). http://dx.doi.org/10.1145/355984.355989 ACMSCU 0098-3500 Google Scholar

32. 

J.-C. Baritauxet al., “Sparsity-driven reconstruction for FDOT with anatomical priors,” IEEE Trans. Med. Imaging, 30 (5), 1143 –1153 (2011). http://dx.doi.org/10.1109/TMI.2011.2136438 ITMID4 0278-0062 Google Scholar

33. 

B. Dogdaset al., “Digimouse: a 3D whole body mouse atlas from CT and cryosection data,” Phys. Med. Biol., 52 (3), 577 –587 (2007). http://dx.doi.org/10.1088/0031-9155/52/3/003 PHMBA7 0031-9155 Google Scholar

34. 

D. Stoutet al., “Creating a whole body digital mouse atlas with PET, CT and cryosection images,” Mol. Imaging Biol., 4 (4), S27 (2002). Google Scholar

35. 

G. AlexandrakisF. R. RannouA. F. Chatziioannou, “Tomographic bioluminescence imaging by use of a combined optical-PET (OPET) system: a computer simulation feasibility study,” Phys. Med. Biol., 50 (17), 4225 –4241 (2005). http://dx.doi.org/10.1088/0031-9155/50/17/021 PHMBA7 0031-9155 Google Scholar

36. 

F. Liuet al., “A parallel excitation based fluorescence molecular tomography system for whole-body simultaneous imaging of small animals,” Ann. Biomed. Eng., 38 (11), 3440 –3448 (2010). http://dx.doi.org/10.1007/s10439-010-0093-4 ABMECF 0090-6964 Google Scholar

37. 

J. T. Elliottet al., “Model-independent dynamic constraint to improve the optical reconstruction of regional kinetic parameters,” Opt. Lett., 37 (13), 2571 –2573 (2012). http://dx.doi.org/10.1364/OL.37.002571 OPLEDP 0146-9592 Google Scholar

38. 

S. C. Daviset al., “Dynamic dual-tracer MRI-guided fluorescence tomography to quantify receptor density in vivo,” Proc. Natl. Acad. Sci. U. S. A., 110 9025 –9030 (2013). http://dx.doi.org/10.1073/pnas.1213490110 PNASA6 0027-8424 Google Scholar

Biography

Wei He received a BS degree in biomedical engineering from Hebei University of Technology, Tianjin, China, in 2008 and received an MS degree in mechanical engineering from China Agricultural University, Beijing, China, in 2010. Currently, he is a PhD candidate in the Department of Biomedical Engineering, Tsinghua University, Beijing, China. His research interests focus on fluorescence molecular tomography for small animal imaging.

Guanglei Zhang received an MS degree in biomedical engineering from Northwestern Polytechnical University, Xi’an, China, in 2007. Currently, he is a PhD candidate in the Biomedical Engineering Department, Tsinghua University, Beijing, China. His research interests focus on fluorescence molecular tomography for small animal imaging.

Huangsheng Pu received an MS degree in biomedical engineering from Fourth Military Medical University, Xi’an, China, in 2010. He is currently a PhD candidate in the Biomedical Engineering Department, Tsinghua University, Beijing, China. His research interest is inverse problems of fluorescence molecular tomography.

Fei Liu received a BS degree in biomedical engineering from Zhejiang University, Zhejiang, China, in 2008 and a PhD degree in biomedical engineering from Tsinghua University, Beijing, China, in 2013. She is currently a postdoctoral fellow in the Biomedical Engineering Department, Tsinghua University, Beijing, China. Her research interest is fluorescence molecular tomography for small animal imaging.

Xu Cao received a BS degree in biomedical engineering from Northwest Polytechnical University, Xi’an, China, in 2007 and received a PhD degree in the Biomedical Engineering Department, Tsinghua University, Beijing, China, in 2013. Now, he is a lecturer in the School of Life Sciences and Technology, Xidian University, Xi’an, China. His research interest is the optical molecular imaging.

Jianwen Luo received his BS and PhD degrees in biomedical engineering from Tsinghua University in 2000 and 2005, respectively. He was a postdoctoral research scientist from 2005 to 2009, and an associate research scientist from 2009 to 2011, in the Department of Biomedical Engineering at Columbia University. He became a professor at Tsinghua University in 2011 and was enrolled in the Thousand Young Talents Program in 2012. His research interest is biomedical imaging.

Jing Bai received an MS and a PhD degree from Drexel University, in 1983 and 1985, respectively. From 1985 to 1987, she was a research associate and assistant professor at the Biomedical Engineering and Science Institute, Drexel University. In 1991, she became a professor in the Biomedical Engineering Department of Tsinghua University. Her current research interests include medical ultrasound and infrared imaging. She has authored or coauthored ten books and more than 300 journal papers.

© 2014 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2014/$25.00 © 2014 SPIE
Wei He, Guanglei Zhang, Huangsheng Pu, Fei Liu, Xu Cao, Jianwen Luo, and Jing Bai "Modified forward model for eliminating the time-varying impact in fluorescence molecular tomography," Journal of Biomedical Optics 19(5), 056012 (22 May 2014). https://doi.org/10.1117/1.JBO.19.5.056012
Published: 22 May 2014
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Luminescence

Liver

Lung

3D modeling

Data acquisition

Image quality

Tomography

Back to Top