28 July 2014 Viscous optical clearing agent for in vivo optical imaging
Zijian Deng, Lijia Jing, Ning Wu, Pengyu lv, Xiaoyun Jiang, Qiushi Ren, Changhui Li
Author Affiliations +
Abstract
By allowing more photons to reach deeper tissue, the optical clearing agent (OCA) has gained increasing attention in various optical imaging modalities. However, commonly used OCAs have high fluidity, limiting their applications in in vivo studies with oblique, uneven, or moving surfaces. In this work, we reported an OCA with high viscosity. We measured the properties of this viscous OCA, and tested its successful performances in the imaging of a living animal’s skin with two optical imaging modalities: photoacoustic microscopy and optical coherence tomography. Our results demonstrated that the viscous OCA has a great potential in the study of different turbid tissues using various optical imaging modalities.

1.

Introduction

By smoothing the refractive index, optical clearing agents (OCAs) can effectively reduce tissue light scattering.1,2 Since OCAs can make the skin more “transparent,” they are commonly employed in many optical imaging modalities, such as optical coherence tomography (OCT),3,4 two-photon microscopy,5 and second harmonic generation imaging.6 Recently, OCA was also implemented in another emerging optical imaging method, photoacoustic microscopy (PAM),7,8 and in vivo photoacoustic flow cytometry.9 In addition to developments in applications, more chemical components and new preparation methods make OCAs less toxic and more suitable for in vivo study.

However, commonly used OCAs primarily have high fluidity, which can easily flow away from the oblique, uneven, or moving surfaces of living bodies. Therefore, repeatedly reapplying agents during the experiments is generally required. Practically, this operation not only probably alters the morphology of the imaged target but also has the risk to change the alignment of the imaging system. Therefore, a viscous OCA for long lasting application on oblique, uneven, or moving surfaces, like the water-based ultrasonic coupling gel used in ultrasound imaging, is highly desired.

In this work, we developed an OCA with high viscosity. This agent can stay on a skin surface for a much longer time with little sacrifice of the optical clearing effect. In the following, we first described the preparation of this agent, and then measured its physical properties, followed by in vivo animal experiments with two optical imaging modalities to demonstrate its performance. Finally, we discussed the toxicity and other alternative methods to make viscous OCAs.

2.

Methods and Results

2.1.

Preparation of the Viscous OCA

The viscous OCA is composed of an OCA,10 polyethylene glycol (PEG)400 (Sinopharm Chemical Reagent Co., Ltd., Beijing, China) mixed with thiazone (Yuancheng Gongchuang Technology Co., Ltd., Wuhan, China), a penetration enhancer, with the volume ratio of 91. To increase the viscosity, 5% (mass percent) polyvinylpyrrolidone (PVP) K88-92 (Aladdin Industrial Co., Ltd., Shanghai, China) was added to the mixture as a thickener.

Due to the low dissolution rate of PVP K88-92 in PEG-400, we used water bath heating for PVP dissolution. To avoid possible chemical reactions related to thiazone at high temperature, thiazone was not put into the mixture while heating. After pouring PEG-400 into a beaker containing PVP and a magnetic stirrer, we rotated the magnetic stirrer immediately to disperse PVP. The mixture was kept at 95°C with stirring until all PVP was dissolved. Then, we cooled it slowly down to room temperature (25°C). Thiazone was then added into the mixture with stirring. Due to the high viscosity of the mixture at room temperature, glass rod stirring by hand was performed until all the thiazone was dissolved.

2.2.

Characteristics of the Viscous OCA

We used a rheometer (AR-G2, TA Instruments, Elstree, Herts, United Kingdom) to measure the viscosity of this viscous OCA, and compared it with that of PEG-400+thiazone (PEG-T, volume ratio of 91). Newton’s law of internal friction, which is obeyed by all Newtonian fluids, indicates that the viscosity of a lowly viscous fluid is the ratio of the shear stress to the shear rate, which is independent on the shear rate. Different from invariable viscosity (0.09567Pa·s) of PEG-T, which was proven to be a Newtonian fluid, the viscosity of PEG-400+thiazone+PVP (PEG-T-PVP) depended highly on the changes of the shear rate [shown in Fig. 1(b)], which means that the substance is too viscous to obey Newton’s law of internal friction. Considering the low shear rate of the viscous OCA in practical experiment conditions, the viscosities at 10.00  1/s shear rate were chose for comparison. The viscosity of the new agent was 5.268, which is about 55-fold times that of PEG-T (0.09599). The viscosity of the mixture can be adjusted by changing the percentage of PVP in the mixture.

Fig. 1

(a) The photographs of PEG-T-PVP in viscous state. (b) The viscosities of PEG-T and PEG-T-PVP at different shear rate. (c) The absorption spectrum of PEG-T-PVP at 400 to 1100 nm.

JBO_19_7_076019_f001.png

Optical properties of this OCA were measured. The optical refractive index of PEG-T-PVP was measured to be 1.477, which was similar to PEG-T (1.474), and was a bit larger than that of PEG-400 (1.465). The UV-vis absorption spectrum of the new agent was measured on a UV-vis spectrophotometer (Evolution 220, Thermo Scientific, Shanghai, China) in Fig. 1(c), which shows a small optical absorption from 400 to 1100 nm.

We also measured the relative acoustic attenuation of PEG-T-PVP at 20 MHz. In this measurement, PEG-T-PVP, water-based ultrasonic coupling gel, and deionized (DI) water, were successively filled into a container with 14-mm inner bottom diameter, 13-mm inner height. An unfocused ultrasound transducer (central frequency: 20 MHz; V-212-BB-RM, Olympus, Waltham, Massachusetts) contacted the top surface of the agents and detected the pulse-echo ultrasound signals passing through the agents. The amplitude ratio of the signal from the PEG-T-PVP to that from the DI water is 0.9132, which was similar to the ratio for gel (0.9609), so the PEG-T-PVP has a very low ultrasound attenuation as does the gel.

PEG-T has been proven to have a good optical clearing effect.10 We tested the potential influence of additional PVP to the skin optical clearing effect. In this study, PEG-T-PVP and PEG-T were applied to two different pieces of ex vivo ventral skin from the same rat with thicknesses of about 1.13 and 1.09 mm, respectively. Figure 2(a) shows the setup of this experiment. A He-Ne laser at 632.8 nm was used as the light source. The light was expanded through a concave lens, then shone on the rat skin. Two optical power meters (sensors: OP-2 VIS; LabMax_TOP, Coherent, Santa Clara, California) were used to detect the intensity of the transmitted light through the skin and also to monitor the variation of light intensity, respectively. The relative change of transmittance at i min ΔTi was calculated as

Eq. (1)

ΔTi=TiT0T0,
where Ti=Pti/Psi, Pti is the power of transmitted light at i min; and Psi is the laser source power at i min. Figure 2(b) shows the skin optical clearing results of the two agents. The slope of the solid line (with PVP) is 0.8241-fold as much as that of the dashed line (without PVP), which means that the presence of PVP slightly decreases the skin optical clearing effect.

Fig. 2

(a) The experiment setup of PEG-T-PVP and PEG-T applied to ex vivo rat skin. (b) The ΔT of ex vivo rat skin with two different agents and corresponding fitted curves y=Ax.

JBO_19_7_076019_f002.png

2.3.

In Vivo Animal Study

We then employed this viscous OCA in the in vivo imaging of animal skin by PAM and OCT. Due to the high viscosity, no agent reapplying operation was employed in all the following experiments, which guaranteed system alignment and imaging consistency during the experiment. Furthermore, we tested massage which is a commonly used method to improve skin permeability, to help the optical clearing of this viscous OCA. All research complied with protocols approved by the Institutional Animal Care and Use Committee of Peking University.

2.3.1.

PAM study

In this study, two PAM modalities, acoustic-resolution PAM (AR-PAM)11 and optical-resolution PAM (OR-PAM),12 were used to study the effect of OCA on both rat and mouse skins, which are common models for skin optical clearing study. The laser wavelengths in following experiments are 532 nm.

Details of the AR-PAM system and its operation can be found in previous literature.13 The imaged rat (120g, SD) was first anesthetized, and then the imaging region was cleanly depilated and gently scrubbed with DI water to remove impurities on the skin. Anesthesia was maintained by an isoflurane machine (1.0% to 1.2% vaporized isoflurane with an airflow rate of 1L/min) throughout the experiment. An excessive amount of OCA was applied on its ventral skin before imaging. After a 5-min system alignment, we repeatedly scanned along a fixed line for 20 times to acquire a series of B-scan images. The interval between two successive scannings was 30 min. The wave-like patterns in Figs. 3(a) and 3(b) were induced by the rat’s breathing motion during scanning. In this study, a 20-frame-mean amplitude (Ai) from a selected subcutaneous blood vessel [shown in Figs. 3(a) and 3(b)] at i min was calculated. ΔAi, the relative change of Ai, was processed similar to Eq. (1). Figure 3(e) demonstrated the obvious increase in PA amplitudes from the selected vessel. In addition, we measured the depth from the skin top surface to the center of the selected vessel in the selected region. Figure 3(f) shows the decrease of the depth. It is consistent with the skin thinning effect due to dehydration caused by the hypertonicity of PEG-400. Maximal amplitude projection (MAP) images of another rat skin with an application of viscous OCA were also obtained. As shown in Figs. 3(g) and 3(h), not only do the PA amplitudes of shallow subcutaneous vessels become stronger, but more branching vessels in deeper regions were also visualized. The ΔA of five randomly selected subcutaneous blood vessels from MAP images (from 5 to 365 min, 60-min interval) monotonically increased as in Fig. 3(e).

Fig. 3

AR-PAM B-scan images of rat skin after (a) 5 min and (b) 95-min PEG-T-PVP application, respectively. Panels (c) before and (d) after 95-min PEG-T-PVP application, respectively. (e) The ΔA of selected vessel from B-scan images. (f) The depth between the surface of the skin and the selected vessel from B-scan images. Maximal amplitude projection (MAP) images of another rat skin after (g) 5 min and (h) 365-min PEG-T-PVP application, respectively. (i) Normalized PA amplitude along the red dashed line in (g) and (h).

JBO_19_7_076019_f003.png

We also compared the performance between the viscous OCA and water-based ultrasonic coupling gel in an in vivo mouse (25g, BALB/c) skin study. The content of vaporized isoflurane for mice was decreased from 0.8% to 1.0%. Contrary to the increases when using the viscous OCA, the PA amplitude from subcutaneous vessels decreased using the traditional water-based gel [shown in Fig. 4(e)], which demonstrated that the skin whitening induced by gel could impact the penetration of light. In addition, using water-based gel made the distances between epidermis and the subcutaneous blood vessel larger, as shown in Fig. 4(f). The gel was hypotonic versus skin, which fosters the diffusing of water from the gel into the skin, allowing the skin to become whiter and thicker.

Fig. 4

AR-PAM B-scan images of mice skin after (a) and (c) 5 min, (b) and (d) 245-min PEG-T-PVP application and water-based ultrasonic coupling gel application, respectively. (e) The ΔA of selected vessels from B-scan images. (f) The depth between the surface of the skin and the selected vessel from B-scan images. Panels (g) and (i) before, (h) and (j) after 245-min PEG-T-PVP and gel application, respectively.

JBO_19_7_076019_f004.png

In the OR-PAM, the lateral resolution is determined by t diffraction-limited optical focusing, while the focusing capability degrades due to tissue optical scattering. We tested the performance of this viscous OCA in an OR-PAM study of a living mouse (25g, BALB/c). Details of the OR-PAM system and its operation can be found in other literature.14 The increase of amplitudes from subcutaneous blood vessels was seen in Fig. 5(c). The MAP images were acquired after 5 and 65 min applications of the new OCA, as shown in Figs. 5(d) and 5(e). The FWHMs of the Gaussian fit was calculated along the red dashed lines on the two vessels [seen in Figs. 5(f) and 5(g)]. Not only did the PA amplitudes of the two vessels increase, but the FWHMs of the two vessels noticeably decreased, which verified that viscous OCA can improve the lateral resolution of OR-PAM.

Fig. 5

OR-PAM B-scan images of mouse skin after (a) 5 min, and (b) 125-min PEG-T-PVP application, respectively. (c) The ΔA of selected vessel from B-scan images. MAP images of another mouse skin after (d) 5 min and (e) 65-min PEG-T-PVP application, respectively. The FWHMs of Gaussian fit (f) and (g) along red dash lines in (d) and (e), respectively.

JBO_19_7_076019_f005.png

2.3.2.

OCT study

OCA has already been successfully implemented in OCT.15 To validate the performance of this viscous OCA, we used it in in vivo rat skin imaging with a commercial OCT system (central wavelength: 840 nm, OCT1000, New Vision, Dongguan, China). The preparation of rat (120g, SD) skin and animal anesthesia was similar to the aforementioned procedure. After application of the viscous OCA on rat skin, we scanned the rat skin every 15 min, and at each scanning 32 B-scan images were acquired. Five of each of the 32 B-scans were randomly selected on which to perform statistical analysis [as shown in Figs. 6(a) and 6(b)]. Details of the processing method for selected regions can be found in the previous literature.15 The total attenuation coefficient μt was gained from the fitted curve of the averaged OCT signals shown in Figs. 6(c) and 6(d). Δμt, the relative change of μt, was processed similarly to Eq. (1) using μ¯ti, the μt mean of 5 frames at i min. Figure 6(e) demonstrated the decrease of μt of rat skin.

Fig. 6

OCT B-scan images of rat skin after (a) 0 min and (b) 120-min PEG-T-PVP application, respectively. Panels (c) and (d) the averaged OCT in-depth signal of respectively selected regions from (a) and (b) and corresponding fitted curves R(z)=A+BeμtZ, where A0=4.156, B0=149.1, μt0=70.76cm1, and A1=1.987, B1=161.8, μt1=49.03cm1. (e) The Δμt of selected region from five random B-scans.

JBO_19_7_076019_f006.png

2.3.3.

Performance of massage for viscous OCA

As is commonly employed in the implementation of OCAs,8,9 massage was used as an effective way to foster the optical clearing effect. Therefore, we also tested the effect of massage in helping the penetration of the viscous OCA when using the AR-PAM imaging system. In this study, we first acquired a control image of the rat (120g, SD) skin [seen in Fig. 7(a)] without OCA and massage. Then, the animal was removed from the AR-PAM system, and we applied the viscous OCA on the same site and performed massage for 30 min. After that, we repositioned the rat in the PAM system and did another imaging [as seen in Fig. 7(b)]. Compared with Fig. 7(a), Fig. 7(b) showed a dramatic enhancement in the photoacoustic signal. In addition, the time for the OCA to take effect was greatly shortened compared with the results in Figs. 3(g) and 3(h). Since massage slightly altered the morphology of the skin and the repositioning of the animal caused variations in the imaging area, the imaging region in Fig. 7(b) does not exactly match that in Fig. 7(a).

Fig. 7

MAP images of another rat skin (a) before and (b) after 30-min PEG-T-PVP application with massage, respectively. Normalized PA amplitude along the (c) red dashed line and (d) red dotted line in (g) and (h), respectively. Panels (e) before and (f) after 30-min PEG-T-PVP application with massage, respectively.

JBO_19_7_076019_f007.png

3.

Conclusions and Discussions

In summary, we developed a viscous OCA composed of PEG-400, thiazone, and PVP. We demonstrated that this OCA has the combined good properties of high viscosity and an optical clearing effect. The high viscosity allows the OCA to be more stable on oblique, uneven, or moving skin surfaces, avoiding multiple reapplications of OCAs during the experiment. In addition, the new OCA overcomes the skin whitening effect of water-based ultrasonic coupling gel. In addition, our results also demonstrated that the massage operation successfully helps the performance of this viscous OCA. The toxicity of PEG-T has been tested in a previous study,10 concluding that no apparent inflammatory or toxic reactions were induced. In addition, it has been confirmed that the PVP is a safe component often used in daily cosmetics.16 As a nontoxic excipient to thicken the mixture without a chemical reaction, PVP does not affect the safety of the mixture.

In addition to the recipe and preparation method of the viscous OCA in this work, other chemical components and procedures are potential for viscous OCAs. Several OCAs,2 which are widely used for optical clearing, can be components of viscous OCAs. Different penetration enhancers or thickeners also have the potential for the preparation of viscous OCAs. However, the solubility of penetration enhancers or thickeners in OCAs is the main challenge to preparation. For example, agarose, another polymer for thickening, is poorly soluble in PEG-400 and propylene glycol; thiazone is poorly soluble in glycerol at room temperature. We can choose an appropriate proportion among these three substances to gain a mixture sited to a specific application.

To the best of our knowledge, this is the first time an OCA with a high viscosity has been prepared. Our results showed that the imaging qualities in PAM and OCT can be improved using this viscous OCA. Other operations17 for improving the skin permeability to OCA can be used for penetration enhancement before imaging. This viscous OCA has a great potential in the study of different turbid tissues using other optical imaging modalities, such as confocal or multiphoton microscopy and fluorescence imaging.

Acknowledgments

This work was sponsored by National Natural Science Foundation of China (Grant No. 61078073), the National Basic Research Program of China (973 Program, 2011CB707502), and the Guangdong Innovative Research Team Program (No. 2011S090).

References

1. 

V. V. Tuchinet al., “Light propagation in tissues with controlled optical properties,” J. Biomed. Opt., 2 (4), 401 –417 (1997). http://dx.doi.org/10.1117/12.281502 JBOPFO 1083-3668 Google Scholar

2. 

D. Zhuet al., “Recent progress in tissue optical clearing,” Laser Photonics Rev., 7 (5), 732 –757 (2013). http://dx.doi.org/10.1002/lpor.2013.7.issue-5 1863-8880 Google Scholar

3. 

S. G. ProskurinI. V. Meglinski, “Optical coherence tomography imaging depth enhancement by superficial skin optical clearing,” Laser Phys. Lett., 4 (11), 824 –826 (2007). http://dx.doi.org/10.1002/lapl.200710056 1612-2011 Google Scholar

4. 

X. Wenet al., “Enhanced optical clearing of skin in vivo and optical coherence tomography in-depth imaging,” J. Biomed. Opt., 17 (6), 066022 (2012). http://dx.doi.org/10.1117/1.JBO.17.6.066022 JBOPFO 1083-3668 Google Scholar

5. 

R. Cicchiet al., “Contrast and depth enhancement in two-photon microscopy of human skin ex vivo by use of optical clearing agents,” Opt. Express, 13 (7), 2337 –2344 (2005). http://dx.doi.org/10.1364/OPEX.13.002337 OPEXFF 1094-4087 Google Scholar

6. 

S. Plotnikovet al., “Optical clearing for improved contrast in second harmonic generation imaging of skeletal muscle,” Biophys. J., 90 (1), 328 –339 (2006). http://dx.doi.org/10.1529/biophysj.105.066944 BIOJAU 0006-3495 Google Scholar

7. 

Y. ZhouJ. J. YaoL. H. V. Wang, “Optical clearing-aided photoacoustic microscopy with enhanced resolution and imaging depth,” Opt. Lett., 38 (14), 2592 –2595 (2013). http://dx.doi.org/10.1364/OL.38.002592 OPLEDP 0146-9592 Google Scholar

8. 

Y. Y. Liuet al., “Optical clearing agents improve photoacoustic imaging in the optical diffusive regime,” Opt. Lett., 38 (20), 4236 –4239 (2013). http://dx.doi.org/10.1364/OL.38.004236 OPLEDP 0146-9592 Google Scholar

9. 

Y. A. Menyaevet al., “Optical clearing in photoacoustic flow cytometry,” Biomed. Opt. Express, 4 (12), 3030 –3041 (2013). http://dx.doi.org/10.1364/BOE.4.003030 BOEICL 2156-7085 Google Scholar

10. 

J. WangR. ShiD. Zhu, “Switchable skin window induced by optical clearing method for dermal blood flow imaging,” J. Biomed. Opt., 18 (6), 061209 (2013). http://dx.doi.org/10.1117/1.JBO.18.6.061209 JBOPFO 1083-3668 Google Scholar

11. 

K. MaslovG. StoicaL. V. H. Wang, “In vivo dark-field reflection-mode photoacoustic microscopy,” Opt. Lett., 30 (6), 625 –627 (2005). http://dx.doi.org/10.1364/OL.30.000625 OPLEDP 0146-9592 Google Scholar

12. 

S. HuK. MaslovL. V. Wang, “Second-generation optical-resolution photoacoustic microscopy with improved sensitivity and speed,” Opt. Lett., 36 (7), 1134 –1136 (2011). http://dx.doi.org/10.1364/OL.36.001134 OPLEDP 0146-9592 Google Scholar

13. 

S. Yeet al., “Studying murine hindlimb ischemia by photoacoustic microscopy,” Chin. Opt. Lett., 10 (12), 121701 (2012). COLHBT 1671-7694 Google Scholar

14. 

N. Wuet al., “High-resolution dual-modality photoacoustic ocular imaging,” Opt. Lett., 39 (8), 2451 –2454 (2014). http://dx.doi.org/10.1364/OL.39.002451 OPLEDP 0146-9592 Google Scholar

15. 

E. A. Geninaet al., “Optical coherence tomography monitoring of enhanced skin optical clearing in rats in vivo,” J. Biomed. Opt., 19 (2), 021109 (2013). http://dx.doi.org/10.1117/1.JBO.19.2.021109 JBOPFO 1083-3668 Google Scholar

16. 

W. F. Bergfeldet al., “Final report on the safety assessment of polyvinylpyrrolidone (PVP),” Int. J. Toxicol., 17 (Suppl. 4), 95 –130 (1998). http://dx.doi.org/10.1177/109158189801700408 IJTOFN 1091-5818 Google Scholar

17. 

S. P. Chernovaet al., “On the dynamics of optical clearing of human skin in vivo,” Proc. SPIE, 4162 227 –235 (2000). http://dx.doi.org/10.1117/12.405947 PSISDG 0277-786X Google Scholar

Biography

Zijian Deng is a graduate student at Peking University.

Lijia Jing is a graduate student at Harbin Institute of Technology.

Ning Wu is a graduate student at Peking University.

Pengyu lv is a graduate student at Peking University.

Xiaoyun Jiang is a graduate student at Peking University.

Qiushi Ren is a professor and chair of the Department of Biomedical Engineering, Peking University.

Changhui Li is an associate professor of the Department of Biomedical Engineering, Peking University. He obtained his PhD in physics from Texas A&M University in 2006. His research interests include biomedical optics and molecular imaging.

© 2014 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2014/$25.00 © 2014 SPIE
Zijian Deng, Lijia Jing, Ning Wu, Pengyu lv, Xiaoyun Jiang, Qiushi Ren, and Changhui Li "Viscous optical clearing agent for in vivo optical imaging," Journal of Biomedical Optics 19(7), 076019 (28 July 2014). https://doi.org/10.1117/1.JBO.19.7.076019
Published: 28 July 2014
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Skin

Optical clearing

Optical imaging

In vivo imaging

Optical coherence tomography

Imaging systems

Tissue optics

RELATED CONTENT


Back to Top