Open Access
18 December 2014 Differences in fluorescence profiles from breast cancer tissues due to changes in relative tryptophan content via energy transfer: tryptophan content correlates with histologic grade and tumor size but not with lymph node metastases
Author Affiliations +
Abstract
The correlation between histologic grade, an increasingly important measure of prognosis for patients with breast cancer, and tryptophan levels from tissues of 15 breast carcinoma patients was investigated. Changes in the relative content of key native organic biomolecule tryptophan were seen from the fluorescence spectra of cancerous and paired normal tissues with excitation wavelengths of 280 and 300 nm. Due to a large spectral overlap and matching excitation–emission spectra, fluorescence resonance energy transfer from tryptophan-donor to reduced nicotinamide adenine dinucleotides-acceptor was noted. We used the ratios of fluorescence intensities at their spectral emission peaks, or spectral fingerprint peaks, at 340, 440, and 460 nm. Higher ratios correlated strongly with high histologic grade, while lower-grade tumors had low ratios. Large tumor size also correlated with high ratios, while the number of lymph node metastases, a major factor in staging, was not correlated with tryptophan levels. High histologic grade correlates strongly with increased content of tryptophan in breast cancer tissues and suggests that measurement of tryptophan content may be useful as a part of the evaluation of these patients.

1.

Introduction

Tumor histologic grade is a recognized, but underappreciated, predictor of prognosis in breast cancer patients.13 Despite the importance of grade, it has historically not been included as one of the criteria for staging of this cancer. The current TNM staging system for breast carcinoma utilizes only tumor size and extent of local invasion (T), number of axillary lymph nodes with metastases (N), and presence of distant metastases (M) as measures of the stage of the cancer.4 As a result, patient prognosis (and thus the need for adjuvant chemotherapy) is often assessed using only these measures. Rakha et al.1 noted, however, that tumor histologic grade is at least as important an indicator of patient prognosis as the number of lymph node metastases and is more important than tumor size. In a review of 22,616 patients with breast cancer, Henson et al.5 found that patients with low-grade cancers and sizes less than 2 cm had a 99% 5-year survival despite the presence of lymph node metastases. In another study of patients with estrogen-receptor positive breast cancer and one to three positive axillary lymph nodes, the 10-year relapse rate was 43% for high-grade tumors, but only 24% and 5% for intermediate- and low-grade cancers.6

In 1984, Alfano et al.7,8 showed that optical spectroscopy could be used as an alternative technique to detect cancer by determining the different emission properties of key organic biomolecules in the tissue. Fluorescence spectroscopy using excitation wavelengths from the ultraviolet light region, for example, can highlight key biomolecules such as tryptophan, reduced nicotinamide adenine dinucleotides (NADH), and collagen in the tissue. It has been reported that cancerous tissue samples have an increased amount of the amino acid tryptophan (emission intensity peak maximum at 340nm) compared with collagen or NADH (emission intensity peaks maxima at 380 and 440nm, respectively).917 Most recently, Zhang et al.17 showed, using fluorescence spectroscopy, that increased tryptophan is linked to an aggressive cancer cell line. Tryptophan is required for protein synthesis and thus increased fluorescent levels can be expected in cancerous breast tissue due to higher cell density and uncontrolled cell division. Recently, the spectral fingerprints of tissues from patients with different breast cancer histologies were examined using a novel fluorescence device known as the S3-LED ratiometer unit and an LED excitation wavelength of 280 nm.15,16 In those reports, we showed consistently higher ratios of the 340 over 440 nm and 340 over 460 nm emission intensity peaks in malignant samples of breast cancer patients compared with their paired normal samples, regardless of the characteristics of the patient’s cancer.15

At an excitation of 280 nm, aromatic amino acids (phenylalanine, tyrosine, and tryptophan residues) can be excited; however, the emission is most likely from the indole ring of tryptophan. Tyrosine and phenylalanine will contribute significantly less to the overall emission at this excitation due to fluorescent quenching attributed to the energy transfer of phenylalanine to tyrosine and tyrosine to tryptophan. Apart from an intraprotein energy transfer from phenylalanine to tyrosine, phenylalanine is unlikely to produce fluorescence energy transfer due to a low-quantum yield. Similarly, tyrosine can be quenched due to an energy transfer from tyrosine to tryptophan residues. Therefore, tryptophan residues dominate emission and are ideal for fluorescence resonance energy transfer. At a selective longer-wavelength excitation of 300 nm, only tryptophan residues in proteins will be excited.18 From these excitation wavelengths, the longer-wavelength (red-shifted) L1a emitting state of tryptophan will occur at a maximum emission intensity peak of 340nm. Additional contributions to the emission spectra at these excitations can occur from proximal fluorophores in the tissue such as NADH (a highly fluorescent coenzyme with absorption and emission peaks at 340 and 440 to 460 nm, respectively). Table 2 highlights the absorption and emission peaks of key organic biomolecules in tissue. Due to a large spectral overlap and matching excitation–emission spectra, fluorescent resonance energy transfer from excited tryptophan (donor) to NADH (acceptor) (a dipole–dipole interaction shown by Förster to be an interaction of coupled electronic dipoles known as donor and acceptor molecules) can occur and affect the overall fluorescent spectrum of tryptophan. Studies by Torikata et al.19,20 have shown that energy transfer from tryptophan to neighboring NADH in pig lactate dehydrogenase and malate dehydrogenase causes a decrease in its fluorescence lifetimes and leads to quenching of tryptophan fluorescence.

In this study, the spectral emission profiles from 15 patients with breast carcinoma were obtained using the conventional LS 50 Perkin-Elmer luminescence spectrometer with excitations at 280 and 300 nm. Dual excitation wavelengths were used to show energy transfers among key fluorophores. These spectral profiles were analyzed using a combination of emission intensity peak ratios. We used the ratios of fluorescence intensities at their spectral emission peaks, or spectral fingerprint peaks, at 340, 440, and 460 nm. The relative intensities or ratios (R) of emission intensity peaks, from these excitations, at 340, 440, and 460 nm from malignant (M) tissue samples (RM1=I340/I440 and RM2=I340/I460) and from paired normal (N) tissue samples (RN1=I340/I440 and RN2=I340/I460) were calculated. The extent of the increase in tryptophan was measured from the ratios of cancerous ratios (RM1 and RM2) over normal ratios (RN1 and RN2) to give R1=RM1/RN1 and R2=RM2/RN2. We report that increased tryptophan levels demonstrated by increases in R1 and R2 correlate strongly with high histologic grade and with large tumor size, but not with lymph node metastases, or with estrogen receptor, progesterone receptor, or HER-2-Neu receptor statuses. This suggests that measurement of tryptophan content may be useful as a part of the evaluation of patients with breast cancer.

2.

Experiment

2.1.

Tissue Specimen

Spectral profiles were acquired from 30 breast tissue samples from 15 breast cancer patients (15 cancerous and 15 paired normal samples). Tissue samples were supplied by the National Disease Research Interchange and the Cooperative Human Tissue Network under an Institutional Review Board protocol. None of the samples were chemically treated (no phosphate buffer saline immersion). The samples were kept on ice except during the experiments. The tissues from paired normal and malignant samples were done under the same exact conditions. In general, the tissue samples were analyzed within 24 h of resection. The malignant sample and its paired normal sample were always done at the same time. The patient ages ranged from 48 to 76 years old (Table 1). Three were premenopausal, 11 were post-menopausal, and 1 was peri-menopausal. Fourteen were Caucasian and one was Black. One had received prior neoadjuvant chemotherapy without shrinkage of the tumor, while 14 were treatment naïve. One patient (n=10) had had a prior surgery and the sample analyzed was a recurrent cancer. Characteristics of the 15 cancers patient are given in Table 1. Measurements were acquired from multiple random locations along the surfaces of the samples. Cuvettes were used to hold tissue samples.

Table 1

Characteristics of 15 breast cancer patients.

Patient number (n)Age (years)GradeTumor size (cm)Number of nodesHistology
16334.81Invasive ductal carcinoma
26632.54Invasive ductal carcinoma
35932.84Invasive ductal carcinoma
45412.2>10Invasive ductal carcinoma
55935.4>10Invasive ductal carcinoma
67438.516Micropapillary breast carcinoma
76134.011Multifocal invasive ductal carcinoma
87634.5Not sampledInvasive ductal carcinoma
94823.52Invasive lobular carcinoma, pleomorphic variant
10513**Recurrent invasive ductal carcinoma
116221.8>10Invasive ductal carcinoma, lobular features
125333.01Invasive ductal carcinoma
136010.91Invasive ductal carcinoma
145732.70Invasive ductal carcinoma
156424.11Invasive carcinoma with ductal and lobular features
* Not assessed, recurrent cancer.

2.2.

Optical Setups

2.2.1.

LS 50 fluorescence spectrometer

The conventional LS 50 Perkin Elmer luminescence spectrometer is a computer controlled device which uses a Xenon (Xe) discharge lamp light source. It is equipped with excitation, emission, and synchronous scan functions in the spectral range of 200 to 800 nm and has ±1-nm wavelength accuracy. The slit widths can be varied to give resolutions between 2 and 15 nm. The emission spectra from the tissue samples were acquired using excitation wavelengths of 280 and 300 nm, with a 280- and 300-nm full-width half-maximum 25±5nm bandpass filters. The focal spot size of the excitation beam is approximately 7 mm in length and 2 mm in width.

3.

Results and Discussion

Spectral profiles from patients with breast carcinoma were acquired using the LS 50 fluorescence spectrometer at 280 and 300 nm excitations. Table 2 highlights the optimal absorption and emission peak maxima of key organic biomolecules tyrosine, tryptophan, collagen, elastin, NADH, and flavins. With light at excitation peak wavelength maxima of 280 to 300 nm, biomolecules in tissue such as tryptophan, collagen, and NADH will absorb this Gaussian-shaped excitation spectrum of light and emit a spectrum of light due to a combination of biomolecules with peak maxima at 340, 380, and 440 to 460 nm (largely due to tryptophan, collagen, and NADH, respectively). The most intense of these peaks (due to tryptophan) is located at 340nm. A weak secondary peak at 440 to 460 nm is due to NADH and is said to occur due to resonance energy transfer of protein–NADH complexes.1822 A large spectral overlap and matching excitation–emission peaks at 340 nm from tryptophan and NADH, respectively, make tryptophan and NADH a good donor–acceptor pair. The Förster distance R¯0 for this pair in solution has been reported as 25 Å [where R¯0 is obtained using the average value of the orientation factor (angle between the donor and acceptor dipoles) κ2, where κ2=κ2¯=2/3 to account for the fast-rotatory Brownian motion which occurs during the lifetime of the donor].23

Table 2

Absorption and emission peak maxima of key organic biomolecules.

MoleculesTyrosineTryptophanCollagenElastinNADHFlavins
Absorption peak maximum (nm)275280 to 300325325340450
Emission peak maximum (nm)280 to 300340400410440 to 460525

Figure 1(a) is a microscopic image of the pathology slide (20× magnified) from malignant breast tissue from patient n=10. The breast cancer sample from this patient with invasive ductal carcinoma was grade 3 (high grade) and this correlates with higher ratios of emission peaks. Figures 1(b) and 1(c) show the corresponding average fluorescence spectra from normal and malignant breast tissue samples with excitations of 280 and 300 nm. A peak emission maximum can be seen at 340nm in both the malignant and normal samples. This result is consistent with previous studies.15,16 The emission spectra show similar peak maxima at 340nm using an excitation of 280 and 300 nm and suggest that the fluorescence is dominated by tryptophan residues.

Fig. 1

(a) Microscopic image of the pathology slide (20× magnified) from malignant breast tissue from patient n=10; and (b) and (c) corresponding fluorescent spectra from malignant (MAL) and normal (NORM) breast tissues after 280 and 300 nm excitations.

JBO_19_12_125002_f001.png

Ratios (R) of emission intensity peak maxima from biomolecules in malignant (RM1=I340/I440and RM2=I340/I460) and normal (RN1=I340/I440 and RN2=I340/I460) samples were calculated. Figures 2(a) and 2(b) show that RM1 and RM2 from patients with high-grade breast cancer have higher bar graphs than RN1 and RN2 from excitations 280 and 300 nm. These figures demonstrate that malignant tissue samples have higher 340 over 440 nm and 340 over 460 nm ratios compared with corresponding ratios from their paired normal tissue samples.

Fig. 2

(a) and (b) Normalized ratios RM1, RM2, RN1, and RN2 from high-grade tumors with intensity peaks at excitations 280 and 300 nm, respectively, using the LS 50 spectrometer.

JBO_19_12_125002_f002.png

Normalized, averaged, and cancerous over normal ratios (R1=RM1/RN1 and R2=RM2/RN2) are given in Tables 3 and 4, respectively. Tables 3 and 4 compare the histologic grade with ratio results R1 and R2. Patients with high histologic grade breast cancer had higher R1, whereas patients with intermediate-/low-grade breast cancer had lower R1 results. R1 for high-grade tumors, from 280 and 300 nm excitations, had average values of 2.77608 (with standard deviation of 1.36389 and p-value of 0.00138<0.05) and 2.910345 (with standard deviation of 1.49592 and p-value of 0.00260<0.05).

Table 3

Normalized ratios (R1) of emission intensity peaks from RM1=I340/I440 over RN1=I340/I440.

Patient number (n)GradeExcitation 280 nmExcitation 300 nm
1High3.555594.24868
2High2.979661.81906
3High1.942093.68094
5High1.778441.45143
6High2.972602.62358
7High1.067600.98885
8High1.144531.50294
10High3.045032.97847
12High5.580904.30446
14High3.694365.50504
4Low or intermediate0.961241.03272
9Low or intermediate0.516361.10069
11Low or intermediate0.935611.00690
13Low or intermediate1.030570.69261
15Low or intermediate0.783710.86216

Table 4

Normalized ratios (R2) of emission intensity peaks from RM2=I340/I460 over RN2=I340/I460.

Patient number (n)GradeExcitation 280 nmExcitation 300 nm
1High5.525915.63311
2High2.979661.64052
3High2.576684.71759
5High4.069144.29812
6High1.393721.68522
7High1.333701.35229
8High1.076241.61274
10High5.111034.64725
12High8.657705.08158
14High6.294565.90878
4Low or intermediate0.884741.06531
9Low or intermediate0.556380.44834
11Low or intermediate1.188211.15523
13Low or intermediate1.204070.52290
15Low or intermediate0.884290.85510

Similar results can be seen for R2, with a higher overall R2 for patients with high-grade cancer and a lower R2 for patients with intermediate-/low-grade cancer. R2 for high-grade tumors, from 280 and 300 nm excitations, had average values of 3.90183 (with standard deviation of 2.49327 and p-value of 0.00484<0.05) and 3.65772 (with a standard deviation of 1.759722 and p-value of 0.00081<0.05). The averages of ratios R1 and R2 were 3.31149 for high-grade tumors and 0.88436 for low-grade tumors. Higher ratios correlated strongly with grade 3 (high) histologic grade, while tumors that were grade 2 or 1 (intermediate/low) demonstrated low ratios.

Large tumor size also correlated with high ratios. This is not surprising as tumor size frequently correlates with the degree of tumor grade.1 Exceptions to the association of high ratios and large tumor sizes occurred where the tumor was lower grade (n=9 and 15) and the histologic grade could be regarded as a confounder. Tables 5 and 6 compare the tumor size with the ratios R1 and R2.

Table 5

Correlation between tumor size and ratio (R1=RM1/RN1) of emission intensity peaks from RM1=I340/I440 over RN1=I340/I440.

Patient number (n)Tumor size (cm)Excitation 280 nmExcitation 300 nm
1Above 4.53.555594.24868
5Above 4.51.778441.45143
6Above 4.52.972602.62358
22.5 to 4.52.979661.81906
32.5 to 4.51.942093.68094
72.5 to 4.51.067600.99885
82.5 to 4.51.144531.50294
9a2.5 to 4.50.516360.37816
122.5 to 4.55.580904.30446
142.5 to 4.53.694365.50504
15a2.5 to 4.50.783710.86216
40 to <2.50.961241.03272
110 to <2.50.935611.10069
130 to <2.51.030570.69261

aLow-grade tumor, grade acts as a confounder.

Table 6

Correlation between tumor size and ratio (R2=RM2/RN2) of emission intensity peaks from RM2=I340/I460 over RN2=I340/I460.

Patient number (n)Tumor size (cm)Excitation 280 nmExcitation 300 nm
1Above 4.53.555594.24868
5Above 4.54.069144.29812
6Above 4.51.393721.68522
22.5 to 4.52.979661.64052
32.5 to 4.52.576684.71759
72.5 to 4.51.333701.35229
82.5 to 4.51.076241.61274
9a2.5 to 4.50.556380.44834
122.5 to 4.58.657708.08158
142.5 to 4.56.294565.90878
15a2.5 to 4.50.884290.85510
40 to <2.50.884741.06531
110 to <2.51.188211.15523
130 to <2.51.204070.52290

aLow-grade tumor, grade acts as a confounder.

With the exception of n=9 and 15, the average of R1 and R2, from 280 and 300 nm excitations at different tumor sizes, was 2.99007 with a standard deviation of 1.16402 (above 4.5 cm), 3.41884 with a standard deviation of 2.32900 (2.5 to 4.5 cm), and 0.98116 with a standard deviation of 0.20324 (0 to <2.5cm).

Although histologic grade and tumor size (to a lesser degree) correlate with high ratios, the number of axillary lymph node metastases, a major determinant of staging, was found not to correlate with the ratios, with tumors from patients who had a small number or zero positive nodes demonstrating high ratios if high grade, and tumors from patients with a large number of positive nodes showing low ratios if they were lower grade. For example, analysis of the samples from patient n=14 showed high R1 and R2 because the cancer was high grade, despite the fact that this patient had zero positive lymph nodes. The sample from patient n=12 demonstrated a high ratio presumably because it was high grade, despite the fact that only one lymph node was positive for cancer. On the other hand, analysis of samples n=4, 7, and 11 gave low ratios presumably because their cancers were lower grade, despite the fact that each of these patients had more than 10 positive lymph nodes. No correlation of increased ratios could be found with estrogen receptor, progesterone receptor, or HER-2-Neu receptor statuses.

4.

Conclusion

A number of researchers have shown that cancerous tissues have an increase in their tryptophan content compared with normal tissues and cells.917 The emission spectra show similar peak maxima at 340nm using an excitation of 280 and 300 nm and suggest that the fluorescence is dominated by tryptophan residues. We note possible energy transfer between tryptophan (donor) and NADH (acceptor) in breast tissue samples with excitation wavelengths of 280 and 300 nm. Further studies are needed in order to determine the tryptophan residues involved in this event and the efficiency of such an energy transfer. We also report that the increase in tryptophan content is correlated with the degree of tumor histologic grade. Indeed, the difference in tryptophan content between higher and lower grade tumors that we observed in our study was marked. The averages of ratios R1 and R2 were 3.31149 for high-grade tumors and 0.88436 for low-grade tumors. Tumor size also correlates with increased tryptophan content, but this appears to be due to the known association of large tumor size with high histologic grade.1 As histologic grade is becoming recognized as an increasingly important measure of prognosis for patients with breast carcinoma, the measurement of tryptophan ratios may also be useful in the assessment of these cancers. Further studies will be needed to determine whether very high breast cancer tryptophan levels in those patients with high-grade tumors might carry a worse prognosis, and thus necessitate more aggressive treatment than would be done in patients with only slightly elevated levels.

References

1. 

E. A. Rakhaet al., “Breast cancer prognostic classification in the molecular era: the role of histological grade,” Breast Cancer Res., 12 (4), 207 (2010). http://dx.doi.org/10.1186/bcr2607 BCTRD6 Google Scholar

2. 

G. J. R. Porteret al., “Patterns of metastatic breast carcinoma: influence of tumor histological grade,” Clin. Radiol., 59 (12), 1094 –1098 (2004). http://dx.doi.org/10.1016/j.crad.2004.05.001 CLRAAG 0009-9260 Google Scholar

3. 

I. Soerjomataramet al., “An overview of prognostic factors for long-term survivors of breast cancer,” Breast Cancer Res. Treat., 107 309 –330 (2008). http://dx.doi.org/10.1007/s10549-007-9556-1 BCTRD6 0167-6806 Google Scholar

4. 

S. B. Edgeet al., AJCC Cancer Staging Manual, 347 –376 7th ed.Springer, New York (2010). Google Scholar

5. 

D. E. Hensonet al., “Relationship among outcome, stage of disease, and histologic grade for 22,616 cases of breast cancer. The basis for a prognostic index,” Cancer, 68 (10), 2142 –2149 (1991). http://dx.doi.org/10.1002/(ISSN)1097-0142 0008-543X Google Scholar

6. 

E. A. Rakkaet al., “Prognostic significance of Nottingham histologic grade in invasive breast carcinoma,” J. Clin. Oncol., 26 3153 –3158 (2008). http://dx.doi.org/10.1200/JCO.2007.15.5986 JCONDN 0732-183X Google Scholar

7. 

R. R. Alfanoet al., “Laser induced fluorescence spectroscopy from native cancerous and normal tissue,” IEEE J. Quantum Electron., 20 1507 –1511 (1984). http://dx.doi.org/10.1109/JQE.1984.1072322 IEJQA7 0018-9197 Google Scholar

8. 

R. R. Alfanoet al., “Light sheds light on cancer,” Bull. N. Y. Acad. Med., 67 143 –150 (1991). BNYMAM 0028-7091 Google Scholar

9. 

A. Tankiewiczet al., “Tryptophan and its metabolites in patients with oral squamous cell carcinoma: preliminary study,” Adv. Med. Sci., 51 (Suppl 1), 221 –224 (2006). AMSDCH 1896-1126 Google Scholar

10. 

F. V. DeGeorgeR. R. Brown, “Differences in tryptophan metabolism between breast cancer patients with and without cancer at other sites,” Cancer, 26 767 –770 (1970). http://dx.doi.org/10.1002/(ISSN)1097-0142 0008-543X Google Scholar

11. 

B. Banerjeeet al., “Tryptophan autoflorescence imaging of neoplasms of the human colon,” J. Biomed. Opt., 17 (1), 016003 (2012). http://dx.doi.org/10.1117/1.JBO.17.1.016003 JBOPFO 1083-3668 Google Scholar

12. 

Y. Yanget al., “Optical spectroscopy of benign and malignant breast tissues,” Lasers Life Sci., 7 115 –127 (1996). LLSCES 0886-0467 Google Scholar

13. 

P. K. GuptaS. K. MajumderA. Uppal, “Breast cancer diagnosis using N2 laser excited autoflorescence spectroscopy,” Laser Surg. Med., 21 (5), 417 –422 (1997). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

14. 

Y. Puet al., “Changes of collagen and nicotinamide adenine dinucleotide in human cancerous and normal prostate tissues studied using native fluorescence spectroscopy with selective excitation wavelength,” J. Biomed. Opt., 15 (4), 047008 (2010). http://dx.doi.org/10.1117/1.3463479 JBOPFO 1083-3668 Google Scholar

15. 

L. A. Sordilloet al., “Optical spectral fingerprints of tissues from patients with different breast cancer histologies using a novel fluorescence spectroscopic device,” Technol. Cancer Res. Treat., 12 455 –461 (2013). http://dx.doi.org/10.7785/tcrt.2012.500330 TCRTBS 1533-0346 Google Scholar

16. 

L. A. Sordilloet al., “Compact Stokes shift and fluorescence spectroscopic diagnostics LED ratiometer unit with no moving parts for cancer detection,” Proc. SPIE, 8220 82200I (2012). http://dx.doi.org/10.1117/12.906487 PSISDG 0277-786X Google Scholar

17. 

L. Zhanget al., “Tryptophan as the fingerprint for distinguishing aggressiveness among breast cancer cell lines using native fluorescence spectroscopy,” J. Biomed. Opt., 19 (3), 037005 (2014). http://dx.doi.org/10.1117/1.JBO.19.3.037005 JBOPFO 1083-3668 Google Scholar

18. 

J. R. Lakowicz, Principles of Fluorescence Spectroscopy, Kluwer Academics/Plenum Publisher, New York (1999). Google Scholar

19. 

T. Torikataet al., “Lifetimes and NADH quenching of tryptophan fluorescence in pig heart lactate dehydrogenase,” Biochemistry, 18 (2), 385 –390 (1979). http://dx.doi.org/10.1021/bi00569a024 0006-2960 Google Scholar

20. 

T. Torikataet al., “Lifetimes and NADH quenching of tryptophan fluorescence in pig heart cytoplasmic malate dehydrogenase,” J. Biol. Chem., 254 (9), 3516 –3520 (1979). JBCHA3 0021-9258 Google Scholar

21. 

C. C. FieldsW. T. BirdsongR. H. Goodman, “Differential binding of NAD+ and NADH allows the transcriptional corepressor carboxyl-terminal binding protein to serve as a metabolic sensor,” Proc. Natl. Acad. Sci. U. S. A., 100 (16), 9202 –9207 (2003). http://dx.doi.org/10.1073/pnas.1633591100 PNASA6 0027-8424 Google Scholar

22. 

M. J. KronmanL. G. Holmes, “The fluorescence of native, denatured and reduced-denatured proteins,” Photochem. Photobiol., 14 (2), 113 –134 (1971). http://dx.doi.org/10.1111/php.1971.14.issue-2 PHCBAP 0031-8655 Google Scholar

23. 

I. Z. Steinberg, “Long-range nonradiative transfer of electronic excitation energy in proteins and polypeptides,” Annu. Rev. Biochem., 40 83 –11 (1971). http://dx.doi.org/10.1146/annurev.bi.40.070171.000503 ARBOAW 0066-4154 Google Scholar

Biography

Laura A. Sordillo is a researcher at the Institute for Ultrafast Spectroscopy and Lasers in the Department of Physics at The City College of CUNY. She has developed a novel portal device for the assessment of cancerous lesions, studied the use of an octreotate indocyanine dye to target breast cancer, and investigated the use of the three NIR optical windows to image microfractures of human tibial bone. She is a recipient of the CCNY-MSKCC Partnership Graduate Award.

Peter P. Sordillo is a physician and cancer researcher, whose specialty is the treatment of extremely rare cancers. He also holds graduate degrees in philosophy and in physics. He is currently an attending hematologist and oncologist at Lenox Hill Hospital and research consultant at the Institute for Ultrafast Spectroscopy and Lasers in the Department of Physics at The City College of CUNY.

Yang Pu obtained his PhD degree in electrical engineering at the City University of New York. He is a multidisciplinary researcher in the fields of biomedical optics. His research is concentrated on breaking two limits of optics: 1) enhancing the resolution of microscope to break the limitation of diffraction, and 2) imaging deep organs from large animal and human using optical techniques.

Robert R. Alfano is a distinguished professor of Science and Engineering at The City College of CUNY. He has pioneered many applications of light and photonics technologies to the study of biological, biomedical, and condensed matter systems and invented and used in his research supercontinuum and novel tunable solid state lasers. He received his PhD degree in physics from New York University and is a Fellow of the American Physical Society, Optical Society of America, and IEEE.

Yury Budansky: Biography is not available.

© 2014 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2014/$25.00 © 2014 SPIE
Laura A. Sordillo, Peter P. Sordillo M.D., Yury Budansky, Yang Pu, and Robert R. Alfano "Differences in fluorescence profiles from breast cancer tissues due to changes in relative tryptophan content via energy transfer: tryptophan content correlates with histologic grade and tumor size but not with lymph node metastases," Journal of Biomedical Optics 19(12), 125002 (18 December 2014). https://doi.org/10.1117/1.JBO.19.12.125002
Published: 18 December 2014
Lens.org Logo
CITATIONS
Cited by 36 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Tumors

Tissues

Breast cancer

Luminescence

Lymphatic system

Cancer

Breast

Back to Top