6 March 2014 Dependence of optical scattering from Intralipid in gelatin-gel based tissue-mimicking phantoms on mixing temperature and time
Author Affiliations +
Abstract
Intralipid is widely used as an optical scattering agent in tissue-mimicking phantoms. Accurate control when using Intralipid is critical to match the optical diffusivity of phantoms to the prescribed value. Currently, most protocols of Intralipid-based hydrogel phantom fabrication focus on factors such as Intralipid brand and concentration. In this note, for the first time to our knowledge, we explore the dependence of the optical reduced scattering coefficient (at 532 nm optical wavelength) on the temperature and the time of mixing Intralipid with gelatin-water solution. The studied samples contained 1% Intralipid and were measured with oblique-incidence reflectometry. It was found that the reduced scattering coefficient increased when the Intralipid-gelatin-water mixture began to solidify at room temperature. For phantoms that had already solidified completely, the diffusivity was shown to be significantly influenced by the temperature and the duration of the mixing course. The dependence of the measured diffusivity on the mixing conditions was confirmed by experimental observations. Moreover, the mechanism behind the dependence behavior is discussed.

1.

Introduction

Tissue-mimicking phantoms are widely used in biomedical optics research for system validation, optimization, calibration, stability evaluation, and quantitative studies.1 They can be easily fabricated with controlled geometries, homogeneities, and physical properties. Depending on specific applications, phantoms use various types of materials as the bulk matrix, such as aqueous suspensions,2,3 gelatin/agar/polyacrylamide-based hydrogels,4,5 polyester/epoxy/polyurethane-based resins,68 and room temperature vulcanizing silicone.9 Absorbing (e.g., India ink, whole blood, and Evans blue) and scattering (e.g., Intralipid, polystyrene microspheres, and white metal oxide powders) agents can then be mixed into the phantom matrix to emulate the optical properties (such as the absorption coefficient μa and the reduced scattering coefficient μs) of real tissues.1

Intralipid is a scattering agent broadly used in phantoms to reproduce the optical scattering coefficient μs and the anisotropy coefficient g of real tissues. Intralipid is biologically similar to the bilipid membrane of cells and organelles1,10—the major constituents causing optical scattering in tissue. Its main scatterers, lipids, have a particle size on the order of 100 nm.1,3,11,12 When Intralipid is added to the bulk matrix, the resulting scattering coefficient μs and the reduced scattering coefficient μs [μs=μs(1g)] of the phantom have a linear dependency on the Intralipid concentration.13,14 For example, at 751-nm wavelength, μs is measured to be around 100, 200, and 300cm1 for 10%, 20%, and 30% Intralipid solutions, respectively.14 In addition, Intralipid has negligible absorption in the visible spectral region,14,15 and a refractive index close to that of soft tissue.1,16 Further, the optical properties of Intralipid from different production batches or vendors have little variation.4,14,15,17 These features, together with its low cost and wide availability, have made Intralipid the chosen diffusive reference standard for many phantom studies.14,15 As a common practice, Intralipid-based phantoms can be made to simulate the optical properties of different types of tissues by varying the concentrations of the scatterer—Intralipid and the absorber, such as India ink.1,11

Intralipid phantoms exist as either aqueous solutions or hydrogels. Many of the studies characterizing Intralipid’s optical properties were performed on its aqueous solution, in which case Intralipid can be easily mixed well with the matrix to form a stable emulsion having the designed scattering and absorption properties. In circumstances where a static matrix with limited water mobility or with complex sample structures is desired, hydrogels, such as gelatin gel, are preferred.1823 Fabrication of such hydrogel phantoms with Intralipid additive is nontrivial—the main variables being the mixing temperature and time. To mix Intralipid thoroughly with hydrogel in the liquid phase, e.g., a gelatin-water solution, the mixture has to be kept warm to prevent the hydrogel matrix from solidifying.20,24 For a complex phantom with multiple parts,6,19 its fabrication usually takes several steps, during which the Intralipid and the matrix must remain liquid. Therefore, an Intralipid-gelatin-water mixture could stay on a hot/stirring plate for up to 1 to 2 h when a phantom is being made.

In previous studies, the optical properties of Intralipid hydrogel phantoms were assumed to be no different from those of aqueous solution phantoms. This presumption, however, contradicts our experimental observation: when the mixing temperature is high and/or the mixing time is long, lipid droplets start to amalgamate near the Intralipid-gelatin-water mixture surface, indicating a separation of the optical scatters (lipids) from the matrix. This separation is irreversible, even when the temperature is lowered. As a result, the top surface of the solidified gel phantom may appear more opaque than the body beneath. In contrast, in an Intralipid-water solution, some lipid droplets will separate from water if the mixture temperature becomes too high, but when the temperature drops, they will disperse evenly to form a homogeneous emulsion again.1,25 In this paper, for the first time to our knowledge, we explore how the temperature and time of mixing Intralipid with gelatin-water solution influence the resulting hydrogel’s optical scattering property.

2.

Materials and Methods

2.1.

Reagents

Gelatin from porcine skin (#G2500-1kG; gel strength 300, Type A) was purchased from Sigma–Aldrich, St. Louis, Missouri. Intralipid® 20% was manufactured by Fresenius Kabi, Uppsala, Sweden, and was purchased from VWR International, Batavia, Illinois (#68100-276). Sodium azide (#S2002-100G) was purchased from Sigma–Aldrich, St. Louis, Missouri. Ultrapure, deionized, and degassed water were used for all solutions.

2.2.

Phantom Preparation

Two types of Intralipid phantoms (Table 1) were studied. The first type (Groups 1 to 4) was Intralipid hydrogel phantoms made from Intralipid (1% by weight), gelatin (10%), and water (89%). The second type (Groups 5 to 7) was Intralipid aqueous solution phantoms made from Intralipid (1%) and water (99%). The concentration of 1% of Intralipid was chosen to obtain a μs of 10cm1 (i.e., a transport mean free path ltr1/μs of 1mm),3,26 which reasonably mimics that of the whole milk (ltr=1.0±0.1mm) and the pork breast (ltr=1.2±0.2mm).11

Table 1

Preparation conditions for samples used in this study.

TypeGroupNo. of samples examinedMixing temperature (°C)Mixing time (min)
I Intralipid+gelatin+water113830
2634 to 9530
394010 to 150
493710 to 150
II Intralipid+water5634 to 9530
694010 to 150
793710 to 150

To fabricate the Type I samples, we adapted the protocol from Ref. 27 with improved temperature and time control:

  • 1. Add porcine gelatin to ultrapure, deionized, and degassed water at the specified concentration. Place the mixture in a glass beaker on a hot/stirring plate (Barnstead Thermolyne, Model SP18425).

  • 2. Heat the mixture of gelatin and water to a temperature (37 to 50°C, corresponding to a heater setting of 3 to 5) above the gelatin’s melting point of 35°C. Throughout heating, stir with a magnetic stirring bar at a constant speed (a stirrer setting of 4 to 5) until the gelatin dissolves completely and the solution turns transparent [Fig. 1(a)].

  • 3. Add sodium azide to the solution at a concentration of 15 mM to preserve the hydrogel.20,28

  • 4. Add an appropriate amount of Intralipid into the gelatin-water solution, water-bathed in a larger beaker [Fig. 1(b)], and stir at the conditions specified in Table 1. Monitor the water bath temperature with a thermometer. To minimize vapor loss during the mixing process, keep the beakers covered with an aluminum foil.24

  • 5. Fill cylindrical molds with the final solution, and let the hydrogel solidify at room temperature (20°C).

Fig. 1

(a) Illustration of the preparation of gelatin-water solution. (b) Illustration of mixing Intralipid with the prepared gelatin-water solution, with controlled temperature and duration.

JBO_19_3_035002_f001.png

The procedure to make the Type II samples was similar, except that no gelatin or sodium azide were added, and the mixtures did not gel.

After fabrication, the reduced optical scattering coefficient μs of the Type I samples was measured using oblique-incidence reflectometry (OIR).29 For the Group 1 sample, the measurement was taken during the gelling process at room temperature (20°C). Other samples, after they solidified at room temperature, were stored in sealed zip bags in a refrigerator (temperature set at 4°C) for at least 12 h before measurements. These storage conditions ensured complete crosslinking (gelling).24

2.3.

Optical Scattering Measurement Using Oblique-Incidence Reflectometry

The cylindrical molds containing the Intralipid hydrogel samples were sufficiently large (6 cm in diameter and 6 cm in depth) to justify the approximation of treating the phantoms as semi-infinite turbid media. Furthermore, the optical absorption coefficient μa of the phantoms was much smaller than the reduced scattering coefficients μs. Therefore, OIR29 could be used to the measure μs of the samples accurately.

Figure 2 shows a schematic of the OIR experimental setup. An attenuated 1-mm diameter beam from a 532 nm laser source (Coherent, Verdi V-5, Santa Clare, California) illuminated the phantom surface at an angle αi. A CCD camera (Andor Technology, iStar 734, Belfast, United Kingdom) was mounted above the phantom to image the diffuse reflectance of the incident beam. An example of the two-dimensional diffuse reflectance pattern captured by the CCD is given in Fig. 3(a). In Fig. 3(b), the light intensity distribution along the dashed line in Fig. 3(a) is represented as curve M. Curve C is the center of the symmetrical reflectance profile and Δx is the horizontal distance between C and the point of incidence. According to Refs. 29 and 30,

Eq. (1)

Δx=sinαinsample(0.35μa+μs),
where nsample, the refractive index of the Intralipid hydrogel, was approximated to be 1.33, which could be measured with a clear gelatin gel containing no Intralipid. This approximation is justified since both gelatin and Intralipid have refractive indexes close to that of water (nwater=1.33).4,16 The angle αi was measured to be 37.9 deg, and μa of a clear gelatin gel was measured to be 0.12cm1 by a spectrophotometer (Varian, Cary 50, Palo Alto, California). Note that the absorption due to 1% Intralipid is negligible.31 Therefore, once Δx was determined from the sample’s reflectance pattern, as shown in Fig. 3, the reduced scattering coefficient can be calculated by

Eq. (2)

μs=sinαinwaterΔx0.35μa.

Fig. 2

Schematic of the oblique-incidence reflectometer used to measure the reduced scattering coefficient of the prepared phantoms. XYZ are system coordinates.

JBO_19_3_035002_f002.png

Fig. 3

(a) An example of the oblique incidence diffuse reflectance pattern. The length of the scale bar is 5 mm. (b) Curve M is the light intensity (photon count) distribution profile along the dashed line in (a), and curve C is composed of the midpoints of the left and right sides of M for specific photon counts. Δx is the horizontally shifted distance between the vertical portion of curve C and the point of incidence.

JBO_19_3_035002_f003.png

3.

Results

For the Group 1 sample, right after the Intralipid-gelatin-water mixture was poured into the cylindrical mold, we started to record the OIR pattern until the phantom solidified completely. The values of μs calculated from the reflectance patterns are shown in Fig. 4 as a function of time during the gelling process at room temperature. As can be seen, μs was around 8.0cm1 or slightly higher in the beginning, and increased to 9.5 to 11cm1 when the hydrogel began to solidify after 20min.

Fig. 4

Measured reduced scattering coefficient as a function of time for the Group 1 sample during the cooling process at room temperature (20°C). The dashed lines are to highlight two different phases before and after solidifications.

JBO_19_3_035002_f004.png

In Groups 2 to 4, samples were completely solidified and taken out of their molds before the OIR measurements, enabling us to examine both the top and the bottom surfaces of the phantoms. To reduce the measurement error caused by sample inhomogeneity, we recorded the reflectance patterns from five different spots on each surface.

In Fig. 5, the calculated μs based on the reflectance patterns of the Group 2 samples is shown as a function of mixing temperature. As we can see μs10cm1 when the mixing temperature was 34°C, and dropped to 5.0cm1 when the temperature was 41°C. At these temperatures, the OIR measurements performed at different spots of both top and bottom surfaces yielded consistent μs values, indicating optical homogeneity of the phantoms. This homogeneity, however, started to break down at higher mixing temperatures. μs measured from the bottom surface decreased monotonically from 2.5cm1 at 52°C to 0.65cm1 at 95°C, while μs measured from the top surface increased monotonically from 3.2cm1 at 52°C to >100cm1 at 95°C. Moreover, measurements from the top at higher temperatures had much larger error bars, suggesting increased optical inhomogeneity.

Fig. 5

Measured reduced scattering coefficients for Group 2 samples, which were prepared with different mixing temperatures but the same duration (30 min). For each phantom, both the top (red circles) and bottom (blue squares) surfaces were examined. On each surface, five measurements were performed at five different locations, from which a mean value was obtained as the mean μs for that surface. The standard deviation is represented by the error bars.

JBO_19_3_035002_f005.png

In Groups 3 and 4, samples were prepared with the same mixing temperatures (40°C from Group 3 and 37°C from Group 4), but different mixing times. μs was measured at the samples’ bottom surfaces, and is shown in Fig. 6 as a function of mixing time. At the 40°C mixing temperature, μs10cm1 when the mixing time was <20min. As the mixing time increased, μs became inversely related to the duration. However, at a mixing temperature lower than 40°C, μs remained relatively constant around 10cm1 over a long period, until the mixing time exceeded 150 min, in which case μs decreased to 8.5cm1.

Fig. 6

Measured reduced scattering coefficients for Groups 2 and 3 samples, prepared with different mixing durations but the same temperatures, at 40°C (red circles) and 37°C (blue squares), respectively. Note that results from only the bottom surfaces are shown here.

JBO_19_3_035002_f006.png

4.

Discussion

The objective of this study was to explore the influence of the mixing temperature and duration on the resulting optical diffusivity when the Intralipid was used as the scatterer in gelatin-based tissue-mimicking hydrogel phantoms.

In Fig. 4, we showed μs as a function of time for the Group 1 sample. The investigated period covered the cooling process from the initial 38°C to the final 20°C, and the formation/reinforcement of crosslinking (solidification) after 20min. The dependence of μs on the measured time, and consequentially the measured temperature, is potentially related to two factors. First, the reduction in mixture volume—measured to be <2%—following the solidification increases the scatterer concentration and results in a higher μs.32 Second, the refractive index of water changes with the measured temperature,33 which alters the mismatch of refractive index between the Intralipid particles and the water, and introduces variations in μs.31

In Fig. 5, we presented the measured μs from the solidified Group 2 samples at room temperature. Apparently, the samples’ diffusivity was affected by the temperature when mixing Intralipid with the hydrogel. Moreover, samples made with relatively high mixing temperatures showed increased divergence between measurements implemented at the top and bottom surfaces. As seen, the measured μs at the bottom surface decreased as a function of mixing temperature, but those measured at the top decreased first, then later increased dramatically with high mixing temperatures. Such variations of μs primarily arise from the change of particle concentration of the lipid droplets, the main constituents that contribute scattering in fat emulsions. Usually, these fat emulsions are quite stable when mixed with water, but the addition of gelatin (a type of collagen) destabilizes the emulsions.34 As a result, some of the lipid droplets, initially evenly distributed throughout the scattering mixture, may flocculate, aggregate, and coalesce.35 Eventually, some lipid droplets can be deemulsified, forming two separate layers: the dispersed droplets float on the top, leading to enlarged but inhomogeneous μs; the continuous (homogenous) phase that has a reduced concentration of lipid droplets remains in the lower portion remains, and has a correspondingly lower μs.

The above hypothesis can be further verified if we look at the photos of two samples from Group 2 (Fig. 7), processed with mixing temperatures of 34 and 95°C, respectively. The first sample (mixed at 34°C) is relatively uniform [Fig. 7(a)], but the second (mixed at 95°C) has a top surface different from the other fractions of the volume [Fig. 7(b)]. The lower portion of the second sample is more transparent than that of the first one, suggesting a lower concentration of lipid droplets. On the other hand, the top surface of the second sample appears to be whiter but more uneven [Fig. 7(d)] than that of the first sample [Fig. 7(c)], indicating increased diffusivity and inhomogeneity. All these observations are consistent with the measurements in Fig. 5 and the hypothesis.

Fig. 7

Photographs of two solidified phantoms from Group 2, with the same mixing time length of 30 min but different mixing temperatures (34°C and 95°C, respectively), before they were removed from the sample molds. (a) and (b) are viewed from the side, while (c) and (d) are from the top.

JBO_19_3_035002_f007.png

It is further shown, in Fig. 6, that the occurrence and degree of phase separation are also affected by the duration of the mixing process, even when the mixing temperature is moderate (e.g., 40°C). However, if the mixing temperature is further reduced, say, down to human body temperature (37°C) or even lower, the effect of phase separation can be minimized and a relatively constant μs can be maintained within a mixing course of 2 h. This finding is particularly important when preparing a time-consuming complex sample requiring uniform diffusivity among different compartments.

However, μs variations shown in Figs. 5 and 6 might also be associated with changes of lipid droplet size—if the changes indeed happened. To confirm this possibility, we also measured the diffusivity and the droplet diameter of Type II samples (Groups 5 to 7) that were prepared under the conditions specified in Table 1. In our study, the measurement of particle diameter was performed with dynamic light scattering (DLS) equipment (Malvern Instruments, Zetasizer Nano ZS, Worcestershire, UK), which handles only aqueous solution. Therefore, no gelatin was added to samples in Groups 5 to 7. Right before measurements of both diffusivity and particle size, these samples were remixed to ensure sample homogeneity. The measured μs from all these samples was between 9.0 and 11cm1, confirming the previous reports25,31 on the diffusivity of Intralipid-water samples. Figure 8(a) is an example of a lipid droplet size distribution histogram measured by the DLS method. In Fig. 8(b), the measured diameters from all Type II samples are shown and sorted according to their groups. Clearly, the diameters of lipid droplets are fairly constant (most are between 200 and 400 nm) from sample to sample, regardless of the differences in mixing conditions. This behavior is in good agreement with the literature,1,3,11,12,25 confirming that in Intralipid-water solution, even deemulsification occurs, when temperature drops, lipid droplets could disperse evenly again to form a homogenous aqueous solution. This reversibility, however, is not seen in gelatin hydrogels (Figs. 57). Since the DLS equipment can only characterize aqueous solutions, we were not able to directly measure the Intralipid particle sizes at different mixing conditions for gelatin gel-based samples. Nevertheless, the μs changes in samples in Groups 2 to 4 were mainly related to the particle concentration variation caused by phase separation, as discussed earlier.

Fig. 8

(a) An example of lipid droplet size distribution measured by the dynamic light scattering method. (b) Measured lipid droplet size of samples in Groups 5 to 7. Each point is the mean diameter of particle size of a measured sample (red squares for Group 5 samples, blue circles for Group 6 samples, and green triangles for Group 7 samples). The error bar is the standard deviation of the size distribution shown in (a).

JBO_19_3_035002_f008.png

It needs to be pointed out that two minor factors in the current study potentially cause inaccuracies in the measurements of μs. First, in our calculation [Eqs. (1) and (2)], the refractive index of water at 20°C and 1 bar was used to represent the refractive index of the whole sample. Because the refractive indexes of gelatin and Intralipid do not differ much from that of water, this is a fair approximation when water is the dominant constituent. The other potential inaccuracy emerges when the μs of the measured sample is very high: x (the horizontal shift between the center of symmetrical reflectance profile and the point of incidence) then might be too small to be determined precisely. However, the overall measurement accuracy based on the OIR method is reasonably accurate.29 More importantly, in the current study, we are more interested in the relative changes, rather than the absolute magnitudes, of the resulting μs. Thus, the two sources of inaccuracy mentioned above do not severely degrade the rigor of the investigation or the derived conclusions.

5.

Conclusion

Our work was spurred by the experimentally observed irreversible inhomogenization of optical diffusivity in the fabrication of Intralipid-gelatin-water phantoms. We investigated the effect of mixing temperature and duration on the phantom’s reduced optical scattering coefficient μs quantitatively, using OIR. It was found that high temperature and prolonged mixing time increasingly caused phase separation in the Intralipid-gelatin-water emulsion, resulting in a decreasing μs. On the other hand, the optical diffusivity of the Intralipid-water emulsion remains constant regardless of the mixing temperature or mixing time, as confirmed by the measurement of μs and the optical scatterer sizes using OIR and DLS. We, therefore, suspect that the phase separation in the Intralipid-gelatin-water emulsion was caused by irreversible changes of the Intralipid–gelatin structure at a higher temperature or a longer mixing time. Further investigation can clarify this matter, but is outside the scope of this paper. To fabricate Intralipid-gelatin-water phantoms having optical diffusivity that is consistent with that of Intralipid-water counterparts,25,31 we found it is crucial to maintain the mixing temperature below 37°C and use mixing times not exceeding 2 h. We recommend our modified protocol for making Intralipid-gelatin-water phantoms for future optical experiments.

Acknowledgments

The authors thank the Nano Research Facility of Washington University in St. Louis for providing the dynamic light scattering equipment to measure the lipid droplet diameters. We also thank Professor James Ballard for editing the paper. This work was sponsored in part by National Institutes of Health Grants DP1 EB016986 (NIH Director’s Pioneer Award) and R01 CA186567 (NIH Director’s Transformative Research Award) as well as National Academies Keck Futures Initiative Grant IS 13.

References

1. 

B. W. PogueM. S. Patterson, “Review of tissue simulating phantom for optical spectroscopy, imaging and dosimetry,” J. Biomed. Opt., 11 (4), 041102 (2006). http://dx.doi.org/10.1117/1.2335429 JBOPFO 1083-3668 Google Scholar

2. 

J. Linfordet al., “Development of a tissue-equivalent phantom for diaphanography,” Med. Phys., 13 (6), 869 –875 (1986). http://dx.doi.org/10.1118/1.595948 MPHYA6 0094-2405 Google Scholar

3. 

H. J. van Staverenet al., “Light scattering in Intralipid-10% in the wavelength range of 400–1100 nm,” Appl. Opt., 30 (31), 4507 –4514 (1991). http://dx.doi.org/10.1364/AO.30.004507 APOPAI 0003-6935 Google Scholar

4. 

A. J. DurkinS. JaikumarR. Richards-Kortum, “Optically dilute, absorbing, and turbid phantoms for fluorescence spectroscopy of homogeneous and inhomogeneous samples,” Appl. Spectrosc., 47 (12), 2114 –2121 (1993). http://dx.doi.org/10.1366/0003702934066244 APSPA4 0003-7028 Google Scholar

5. 

M. Market al., “Multi-modality tissue-mimicking phantom for thermal therapy,” Phys. Med. Biol., 49 (13), 2767 –2778 (2004). http://dx.doi.org/10.1088/0031-9155/49/13/001 PHMBA7 0031-9155 Google Scholar

6. 

U. Sukowskiet al., “Preparation of solid phantoms with defined scattering and absorption properties for optical tomography,” Phys. Med. Biol., 41 (9), 1823 –1844 (1996). http://dx.doi.org/10.1088/0031-9155/41/9/017 PHMBA7 0031-9155 Google Scholar

7. 

M. FirbankM. OdaD. T. Delpy, “An improved design for a stable and reproducible phantom material for use in near-infrared spectroscopy and imaging,” Phys. Med. Biol., 40 (5), 955 –961 (1995). http://dx.doi.org/10.1088/0031-9155/40/5/016 PHMBA7 0031-9155 Google Scholar

8. 

M. L. Vernonet al., “Fabrication and characterization of a solid polyurethane phantom for optical imaging through scattering media,” Appl. Opt., 38 (19), 4247 –4251 (1999). http://dx.doi.org/10.1364/AO.38.004247 APOPAI 0003-6935 Google Scholar

9. 

R. Bayset al., “Three-dimensional optical phantom and its application in photodynamic therapy,” Lasers Surg. Med., 21 (3), 227 –234 (1997). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

10. 

Tissue Optics: Light Scattering Methods and Instruments for Medical Diagnosis, SPIE Press, Bellingham, Washington (2007). Google Scholar

11. 

L. Wang, Picosecond Kerr Gated Imaging of Phantoms in Turbid Media, City University of New York, New York (1995). Google Scholar

12. 

I. Driveret al., “The optical properties of aqueous suspensions of Intralipid, a fat emulsion,” Phys. Med. Biol., 34 (12), 1927 –1930 (1989). http://dx.doi.org/10.1088/0031-9155/34/12/015 PHMBA7 0031-9155 Google Scholar

13. 

P. Di NinniF. MartelliG. Zaccanti, “Effect of dependent scattering on the optical properties of Intralipid tissue phantoms,” Biomed. Opt. Express, 2 (8), 2265 –2278 (2011). http://dx.doi.org/10.1364/BOE.2.002265 BOEICL 2156-7085 Google Scholar

14. 

P. D. Ninniet al., “Fat emulsions as diffusive reference standards for tissue simulating phantoms?,” Appl. Opt., 51 (30), 7176 –7182 (2012). http://dx.doi.org/10.1364/AO.51.007176 APOPAI 0003-6935 Google Scholar

15. 

P. D. NinniF. MartelliG. Zaccanti, “Intralipid: towards a diffusive reference standard for optical tissue phantoms,” Phys. Med. Biol., 56 (2), N21 –N28 (2011). http://dx.doi.org/10.1088/0031-9155/56/2/N01 PHMBA7 0031-9155 Google Scholar

16. 

H. Dinget al., “Determination of refractive indices of porcine skin tissues and Intralipid at eight wavelengths between 325 and 1557 nm,” J. Opt. Soc. Am. A, 22 (6), 1151 –1157 (2005). http://dx.doi.org/10.1364/JOSAA.22.001151 JOAOD6 0740-3232 Google Scholar

17. 

J. Allardiceet al., “Standardization of intralipid for light scattering in clinical photodynamic therapy,” Lasers Med. Sci., 7 (1), 461 –465 (1992). http://dx.doi.org/10.1007/BF02594089 LMSCEZ 1435-604X Google Scholar

18. 

X. XuH. LiuL. V. Wang, “Time-reversed ultrasonically encoded optical focusing into scattering media,” Nat. Photonics, 5 (3), 154 –157 (2011). http://dx.doi.org/10.1038/nphoton.2010.306 1749-4885 Google Scholar

19. 

P. Laiet al., “Reflection-mode time-reversed ultrasonically encoded (TRUE) optical focusing into turbid media,” J. Biomed. Opt., 16 (8), 080505 (2011). http://dx.doi.org/10.1117/1.3609001 JBOPFO 1083-3668 Google Scholar

20. 

A. M. De Grandet al., “Tissue-like phantoms for near-infrared fluorescence imaging system assessment and the training of surgeons,” J. Biomed. Opt., 11 (1), 014007 (2006). http://dx.doi.org/10.1117/1.2170579 JBOPFO 1083-3668 Google Scholar

21. 

P. LaiX. XuL. V. Wang, “Ultrasound-modulated optical tomography at new depth,” J. Biomed. Opt., 17 (6), 066006 (2012). http://dx.doi.org/10.1117/1.JBO.17.6.066006 JBOPFO 1083-3668 Google Scholar

22. 

P. Laiet al., “Focused fluorescence excitation with time-reversed ultrasonically encoded light and imaging in thick scattering media,” Laser Phys. Lett., 10 (7), 075604 (2013). http://dx.doi.org/10.1088/1612-2011/10/7/075604 1612-2011 Google Scholar

23. 

Y. Suzukiet al., “High-sensitivity ultrasound-modulated optical tomography with a photorefractive polymer,” Opt. Lett., 38 (6), 899 –901 (2013). http://dx.doi.org/10.1364/OL.38.000899 OPLEDP 0146-9592 Google Scholar

24. 

J. R. CookR. R. BouchardS. Y. Emelianov, “Tissue-mimicking phantoms for photoacoustic and ultrasonic imaging,” Biomed. Opt. Express, 2 (11), 3193 –3206 (2011). http://dx.doi.org/10.1364/BOE.2.003193 BOEICL 2156-7085 Google Scholar

25. 

R. MichelsF. FoschumA. Kienle, “Optical properties of fat emulsions,” Opt. Express, 16 (8), 5907 –5925 (2008). http://dx.doi.org/10.1364/OE.16.005907 OPEXFF 1094-4087 Google Scholar

26. 

S. T. Flocket al., “Optical properties of intralipid: a phantom medium for light propagation studies,” Lasers Surg. Med., 12 (5), 510 –519 (1992). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

27. 

C. Kimet al., “Optical phantoms for ultrasound-modulated optical tomography,” Proc. SPIE, 6870 68700M (2008). http://dx.doi.org/10.1117/12.766773 PSISDG 0277-786X Google Scholar

28. 

G. Wagniereset al., “Design and characterization of a phantom that simultaneously simulates tissue optical properties between 400 and 650 nm,” Proc. SPIE, 2926 94 –103 (1996). http://dx.doi.org/10.1117/12.260824 PSISDG 0277-786X Google Scholar

29. 

L. WangS. L. Jacques, “Use of a laser beam with an oblique angle of incidence to measure the reduced scattering coefficient of a turbid medium,” Appl. Opt., 34 (13), 2362 –2366 (1995). http://dx.doi.org/10.1364/AO.34.002362 APOPAI 0003-6935 Google Scholar

30. 

S.-P. Linet al., “Measurement of tissue optical properties by the use of oblique-incidence optical fiber reflectometry,” Appl. Opt., 36 (1), 136 –143 (1997). http://dx.doi.org/10.1364/AO.36.000136 APOPAI 0003-6935 Google Scholar

31. 

B. Cletuset al., “Temperature-dependent optical properties of Intralipid® measured with frequency-domain photon-migration spectroscopy,” J. Biomed. Opt., 15 (1), 017003 (2010). http://dx.doi.org/10.1117/1.3290820 JBOPFO 1083-3668 Google Scholar

32. 

V. A. McGloneet al., “Measuring optical temperature coefficients of Intralipid®,” Phys. Med. Biol., 52 (9), 2367 –2378 (2007). http://dx.doi.org/10.1088/0031-9155/52/9/003 PHMBA7 0031-9155 Google Scholar

33. 

“Index of refraction of water,” Handbook of Chemistry and Physics (2000–2001), CRC Press, Boca Raton, FL (2000). Google Scholar

34. 

M. L. CarassoW. N. RowlandsR. A. Kennedy, “Electroacoustic determination of droplet size and zeta potential in concentrated intravenous fat emulsions,” J. Colloid Interface Sci., 174 (2), 405 –413 (1995). http://dx.doi.org/10.1006/jcis.1995.1408 JCISA5 0021-9797 Google Scholar

35. 

E. D. Doktorgrades, Influence of Lecithin on Structure and Stability of Parenteral Fat Emulsions, Friedrich-Alexander-Universität Erlangen-Nürnberg, Germany (1998). Google Scholar

Biographies of the authors are not available.

© 2014 Society of Photo-Optical Instrumentation Engineers (SPIE) 0091-3286/2014/$25.00 © 2014 SPIE
Puxiang Lai, Xiao Xu, and Lihong V. Wang "Dependence of optical scattering from Intralipid in gelatin-gel based tissue-mimicking phantoms on mixing temperature and time," Journal of Biomedical Optics 19(3), 035002 (6 March 2014). https://doi.org/10.1117/1.JBO.19.3.035002
Published: 6 March 2014
Lens.org Logo
CITATIONS
Cited by 73 scholarly publications and 7 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scattering

Tissue optics

Temperature metrology

Water

Biomedical optics

Particles

Refractive index

RELATED CONTENT


Back to Top