14 July 2015 Ballistic imaging of biological media with collimated illumination and focal plane detection
Barak Brezner, Sarah Cahen, Ziv Glasser, Shmuel Sternklar, Er’el Granot
Author Affiliations +
Abstract
A simple, affordable method for imaging through biological tissue is investigated. The method consists of (1) imaging with a wavelength that has a relatively small scattering coefficient (1310 nm in this case) and (2) collimated illumination together with (3) focal plane detection to enhance the detection of the ballistic photons relative to the diffusive light. We demonstrate ballistic detection of an object immersed in a 1-cm-thick cuvette filled with 4% Intralipid, which is equivalent to ∼1 to 2 cm of skin tissue. With the same technology, a ballistic image of a 1-mm-wide object in 10-mm-thick chicken breast is also presented.

1.

Introduction

Optical imaging, in general, and optical mammography in particular, are holy-grails of clinical imaging. Visible and infrared (IR) light is safe (as opposed to x rays) and also gives significantly better spatial resolution than ultrasound. Moreover, different types of tissues interact differently with visible light. One of the most useful features is the difference in the absorption coefficient at 700 nm of oxidized and nonoxidized hemoglobin, which can be utilized not only in identifying activated metabolism but also as a tumor detector (see, for example, Refs. 1 and 2). In addition, the scattering coefficient is sensitively dependent upon the cells’ organelles, such as the nucleus and mitochondria. Therefore, visible and near-IR light are extremely useful for diagnosing different anomalies in biological tissue.

The main problem with visible and near-IR radiation is that they undergo significant scattering. In fact, for visible light, the mean free path is <0.1mm, and therefore, beyond 2 mm, the ballistic component of the light is negligible.35 For this reason, most optical imaging methods focus on diffusive imaging techniques, such as photon density waves and inverse-scattering solutions of the diffusion equation.58 Methods that are based on ballistic imaging are only effective for thin scattering media. An important example is optical coherence tomography,911 which is a leading technology to image skin layers and the retina with micrometer resolution. This technology works in the reflection configuration; therefore, it is useless beyond 1 to 1.5 mm of tissue, where the number of ballistic photons is negligible and the ballistic image is buried in diffusive light.

In 1990, it was demonstrated by Yoo and Alfano12 that time-resolved streak-camera techniques can be utilized to differentiate between the ballistic and diffusive components of light in diffusive medium. A year later,13 a similar technique based on an ultrafast optical shutter was used to create a high-resolution image through a 3.5-mm-thick human tissue. Since then, this technology was improved and became an important tool in the research of diffusive media (see Refs. 1415.16.17.18.19.20.21.22.23.24.25).

It has recently been shown26 that by applying two simple and well-known modifications, it is possible to substantially extend the ballistic imaging depth of a diffusive medium. This method was based on previous research,27 which demonstrated that the transition point between the ballistic and diffusive regimes depends not only on the scattering coefficient, but on the collection angle of the detection as well.

It is the objective of this work to apply this technology in constructing a ballistic image that can resolve a 1-mm-wide object immersed in 1 to 2 cm (depending on the specific type) of a biological tissue medium.

2.

Theory

Photons experience both absorption and scattering while propagating in any medium. However, in biological tissue, scattering is the dominant mechanism for both visible and near-IR light. In many types of tissue, the scattering coefficient can be more than three orders of magnitude larger than the absorption coefficient. In fact, without scattering, most tissues would be almost completely transparent. In such a nonscattering world, all absorbing/blocking structures, such as veins and tumors, would clearly appear in any optical image exactly like bones do in x-ray images. However, since the small cellular organelles that biological tissues consist of have micrometer dimensions, their scattering cross-section (in the visible and near-IR domain) is huge.28,29 As a consequence, the amount of ballistic, i.e., nonscattered, light rapidly decreases exponentially. In fact, according to Beer’s law, the dependence of the ballistic light on the longitudinal coordinate z is3,5

Eq. (1)

IB(z)=I0exp[(μS+μA)z],
where the subscript B denotes ballistic, I0 is the light intensity entering the medium, and μS and μAμS are the scattering and absorption coefficients, respectively.

The scattered photons, on the other hand, experience a slower decay. However, due to the fact that their optical trajectory is considerably larger than the ballistic path, they suffer more from absorption. In fact, they obey the diffusion equation 3,5

Eq. (2)

D2IcμAI=It,
where Dc/3(μS+μA) is the diffusion coefficient, μSμS(1g) is the reduced scattering coefficient, and gcosθ is the mean cosine of the scattering angle.

In the stationary case, the equation is reduced to

Eq. (3)

2IμeffI=0,
where μeff3(μS+μA)μA is an effective attenuation coefficient.

In a quasi-one-dimensional model, where both x and y directions are degenerate, the solution is approximately

Eq. (4)

ID(z)=I0exp(μeffz),
where the subscript D denotes diffusive.

Since in biological tissue μeffμS, it follows that diffusive photons penetrate the medium over much longer distances. However, since all the photons are initially ballistic, it is clear that their dominancy changes (quite dramatically) inside the medium, so that an undistorted ballistic image is transformed into a blurred one during propagation.

Previous work23,3032 indicated that the critical width of the diffusive medium, beyond which the number of diffusive photons exceeds the photon number of ballistic light, depends on the measurement’s collection angle, or numerical aperture.26,27 It was demonstrated that the total intensity can be written as a superposition of the diffusive and the ballistic components,27 i.e.,

Eq. (5)

I(z)=I0[exp(μSz)+δΩ4πexp(μeffz)],
where δΩ is the collection angle.

A direct consequence of Eq. (5) is that the transition width zc, beyond which the diffusive portion of the light becomes dominant, can easily be extracted.27

Eq. (6)

zc1μSμeffln(4πδΩ).

A ballistic image of the medium can only be extracted for a medium thickness smaller than zc. In Eqs. (5) and (6), it was assumed that the absorption coefficient is negligible in comparison to the scattering coefficient (μSμA).

If no special measures are taken, then zcμs1, which, for visible light, is approximately zcμs10.1mm. It is clear why the naked eye (or camera) cannot image veins that are >1mm deep. In fact, with the naked eye, one cannot clearly discern any object below the skin layer.

It was demonstrated in Ref. 27 that Eq. (6) gives a roadmap for improving the ballistic image. Any decrease in the scattering coefficient μs and/or in the collection angle δΩ will increase the transition width .

Reference 27 suggested to increase the ballistic penetration depth by imaging with 1310 nm light, which has a much larger absorption coefficient than 700 nm; however, it has a smaller scattering coefficient.33 The ballistic photon attenuation is weaker, and therefore is larger. However, it was further suggested to decrease the collection angle by using an additional collimator for focal-plane detection. Collimators can easily reduce the collection angle to δΩ/4π2×105. With these two simple changes, it was shown that the ballistic regime can be increased by at least two orders of magnitude,26 and the goal of optical ballistic imaging turned from unrealistic to tenable. These suggestions are consistent with other work (see Refs. 34 and 35 for the collimator imaging and Refs. 3637.38 for the wavelength selection).

Equation (5) suggests that the image of the medium can be approximated as a superposition of the ballistic image MB(x) and the diffusive one MD(x), where x is the transversal coordinate. Whereas MB(x) is a shadow image of the medium, is the solution of the diffusion equation at the end of the diffusive medium.

Therefore, when the object is scanned with a detection angle δΩ, the image can be written as a superposition of weighted images, i.e.,

Eq. (7)

I(z,x)=I0[MB(x)exp(μsz)+MD(x)δΩ4πexp(μeffz)].

In this work, we evaluate and measure the image of an opaque wire of diameter D0=1mm embedded within a scattering medium. The ballistic image MB(x) can be described as

Eq. (8)

MB(x)[1rectD0(x)]*C(2x/Δ),
where rectD0(x){1|x|<D0/20else is the rectangle function, the asterisk stands for spatial convolution, and the output collimator response, which is a consequence of the collimator’s circular shape, is C(ξ)={(2/π)1ξ2|ξ|10|ξ|>1, where Δ is the diameter of the collimator. Due to the response function C(ξ), the image resembles a Gaussian function.

If the longitudinal thickness of the medium is w, and the object is centered within the medium, then the diffusive image of the object can be approximated as

Eq. (9)

MD(x)1exp[2μeff(w/2)2+x2]1+(2x/w)2.

This expression can be verified by calculating the product of the probabilities (which are proportional to the solution of the diffusion equation) of migrating from z=0,x=0 to z=w,x=0 via the point z=w/2 for any x.

As a consequence, while in the ballistic image the object’s transversal width is approximately ΔxD01mm; in the diffuse image, it is approximately [the FWHM of Eq. (9) when the width of the object is taken into account]:

Eq. (10)

Δx2(ln22μeff+wD02)(wD02)2+D0.

We therefore expect a very sharp transition (of 0.1mm) from a ballistic image to a diffusive one.

3.

Experiment

In the following experiments, a 1-cm-thick emulsion of Intralipid and water is the turbid medium in which the wire is submerged. Light from a 10 mW, 1310 nm fiber-coupled diode laser is collimated and illuminates the sample in both setups. In the first experiment (Fig. 1), an output lens (an additional collimator) focuses the light into a detector located at the focal plane to reduce the collection angle, whereas in the second setup (Fig. 2), the fiber end is located 18 mm from the medium. In both experiments, we gradually change the density of the Intralipid while monitoring the image (the thickness remains constant at 1 cm).

Fig. 1

System schematic.

JBO_20_7_076006_f001.png

Fig. 2

Same system as in Fig. 1 with the exception that the detection fiber end is 18 mm from the medium without the additional collimator.

JBO_20_7_076006_f002.png

It is convenient to write the scattering coefficient in terms of the Intralipid density ρ measured in percent of the emulsion, i.e.,

Eq. (11)

μsσ0ρ[%],
where ρ[%] is measured in percent and σ0 is the normalized cross-section in (%cm)1.

In this case, the critical density, beyond which the image is mainly diffusive, is [from Eqs. (6) and (11)]

Eq. (12)

ρc1σ0[2wD0ln(4πδΩ)+μeff],
such that for focal plane detection, this leads to

Eq. (13)

ρc1σ0[4wD0ln(2NA)+μeff],
where NA is the numerical aperture of the collimator.

In the second case, where the bare fiber is used at the detection end, Eq. (7) cannot be applied since the fiber’s core is considerably smaller than the cross-section of the illuminating beam. Instead, the ballistic component is reduced by the ratio between the fiber’s core and the collimator cross-section, i.e.,

I(z,x)=I0AfiberAcoll[MB(x)exp(μsz)+MD(x)AcollAfiberδΩ4πexp(μeffz)],
where Afiber and Acoll are the collecting fiber and collimator cross-sections, respectively.

Hence,

Eq. (14)

ρc1σ0[4wD0ln(4ΔzΔ)+μeff],
where Δz is the distance between the medium’s boundary and the fiber, and Δ is the illuminating collimator (i.e., the first collimator) diameter. Note that the fiber’s diameter is not a parameter in this case.

For w=1cm, we measured σ07.2(%cm)1 such that μeff2.4ρ[%]2.

We therefore expect that the width of the image will have the following approximate behavior:

Eq. (15)

Δx(ρ){D0ρρc2(ln22μeff+wD02)(wD02)2+D0ρ>ρc.

In the second experiment, the distance from the medium to the collecting fiber was Δz18mm, and since the illuminating collimator core is approximately Δ3.4mm, from Eq. (14), we have ρc2.2%.

Therefore,

Eq. (16)

Δx(ρ){1mmρ<2.2%4.3mmρ>2.2%.

For the first experiment with a numerical aperture NA=8.9×103, according to Eq. (13), ρc3.75%, and therefore, the transition in the object width should correspond to

Eq. (17)

Δx(ρ){1mmρ<3.75%4.3mmρ>3.75%.

In Fig. 3, the normalized power (normalized to the minimum value) as a function of x is plotted for the two configurations and for two different concentrations: below the concentration ρc=2.2cm1 and above it. This critical concentration represents many types of human tissue. It is clearly seen that below ρ<ρc, there is no significant difference between the two configurations. With both systems, the width of the image is approximately Δx1mm as should be expected for the ballistic image. However, above ρ>ρc, in the two-collimator configuration, there is almost no difference in the width of the image compared to the low-concentration case, i.e., the image remains basically ballistic, while in the one-collimator configuration, the image width is considerably wider.

Fig. 3

The normalized power as a function of the transversal coordinate in both configurations is plotted for two concentrations. The dashed line with the crosses is for the one-collimator configuration, whereas the solid line with the circles is for the two-collimator experiment.

JBO_20_7_076006_f003.png

As can be seen from Fig. 4, these evaluations are consistent with the experimental results.

Fig. 4

The width of the detected image as a function of the Intralipid concentration ρ. The circles stand for the one-collimator configuration, and the solid red line stands for the theoretical prediction. The crosses stand for the two-collimator configuration, and the dashed line represents the theoretical prediction.

JBO_20_7_076006_f004.png

Since it is known that an Intralipid emulsion of 2 to 4% simulates several types of tissue,39,40 these results suggest that this simple technique can be utilized in imaging 1 to 2 cm of biological tissue.

In principle, the penetration depth in the one-collimator configuration can be improved by increasing the distance between the medium and the detector; however, to reach ρc3.75%, the distance should be increased 12-fold to 22cm, which, in practice, would decrease the signal to an unacceptably low level. The clear advantage of focal plane detection can be realized from Fig. 5. In this figure, the ratio between the power measured in the two-collimator configuration p(2col) and the one-collimator one p(1col) is plotted as a function of the transverse coordinate. Since the measurement locations are different in the two experiments, the ratio was calculated between the interpolations of the experimental data (this is the reason that the plots are continuous). In the ballistic regime (i.e., ρ<2.2%), the sensitivity of the two-collimator method is at least an order of magnitude better; beyond the ballistic regime, there is more than three orders of magnitude improvement.

Fig. 5

The ratio between the measured power in the two-collimator configuration (p(2col)) and the one-collimator one (p(1col)) for the two concentrations.

JBO_20_7_076006_f005.png

To illustrate the potential of this technology, the Intralipid emulsion was replaced with a 1-cm-wide cuvette filled with chicken breast. In Fig. 6, a transverse scan of the medium is plotted. The dashed line is the ideal ballistic image of the object, while the solid line represents the measured power (in pW). As can be seen from this plot, not only can the object be detected even in a 1-cm-thick tissue, but the fact that its width is still 1mm indicates that it is indeed a ballistic image.

Fig. 6

Transversal scan of the chicken breast medium with the concealed object.

JBO_20_7_076006_f006.png

Since these experiments were carried out using relatively simple equipment, it seems more than reasonable to conclude that by optimizing the wavelength, light power, detection sensitivity, and collecting angle, an extension of this technique to thicker or denser tissue would be feasible.

4.

Conclusions

A simple and affordable method was implemented to view an opaque object in a tissue phantom and in chicken breast. The method is based on the premise that to create a ballistic image of a diffusive medium, two requirements must hold: (1) the ballistic portion of the light should be separated from the main diffusive portion, and (2) the ballistic part component should be sufficiently strong to be detected.

To obtain the first requirement, the collecting angle is reduced substantially. The second requirement is achieved by illuminating the medium with a wavelength that has a relatively small scattering coefficient in tissue. In our experiment, the angle reduction was carried out by using focal plane detection, and the scattering coefficient reduction was reached by using 1310-nm light. With these simple modifications, it was possible to detect a ballistic image of a concealed opaque object in 1 cm of 4% Intralipid emulsion, which is equivalent to 1 to 2 cm of human tissue. Moreover, the method was implemented in detecting a concealed object in 1 cm of chicken breast with a resolution of 1mm. One application would be in bone age assessment, where an image of the hand bones is necessary.

References

1. 

P. Vaupel et al., “Oxygenation of human tumors: evaluation of tissue oxygen distribution in breast cancers by computerized O2 tension measurements,” Cancer Res., 51 (12), 3316 –3322 (1991). Google Scholar

2. 

P. Vaupel et al., “Current status of knowledge and critical issues in tumor oxygenation. Results from 25 years research in tumor pathophysiology,” Adv. Exp. Med. Biol., 454 591 –602 (1998). AEMBAP 0065-2598 Google Scholar

4. 

A. N. Bashkatov, E. A. Genina and V. V. Tuchin, “Optical properties of skin, subcutaneous, and muscle tissues: a review,” J. Innov. Opt. Health Sci., 4 9 –38 (2011). http://dx.doi.org/10.1142/S1793545811001319 Google Scholar

5. 

V. Tuchin, Tissue Optics, SPIE Press, Bellingham, Washington (2000). Google Scholar

6. 

A. Yodh and B. Chance, “Spectroscopy and imaging with diffusing light,” Phys. Today, 48 34 –40 (1995). http://dx.doi.org/10.1063/1.881445 PHTOAD 0031-9228 Google Scholar

7. 

H. Jiang et al., “Optical image reconstruction using frequency-domain data: simulations and experiments,” J. Opt. Soc. Am. A, 13 253 –266 (1996). http://dx.doi.org/10.1364/JOSAA.13.000253 JOAOD6 0740-3232 Google Scholar

8. 

T. O. McBride et al., “Initial studies of in vivo absorption and scattering heterogeneity in near-infrared tomographic breast imaging,” Opt. Lett., 26 822 –824 (2001). http://dx.doi.org/10.1364/OL.26.000822 OPLEDP 0146-9592 Google Scholar

9. 

A. F. Fercher et al., “Optical coherence tomography—principles and applications,” Rep. Prog. Phys., 66 239 –303 (2003). http://dx.doi.org/10.1088/0034-4885/66/2/204 RPPHAG 0034-4885 Google Scholar

10. 

U. Morgner et al., “Spectroscopic optical coherence tomography,” Opt. Lett., 25 111 –113 (2000). http://dx.doi.org/10.1364/OL.25.000111 OPLEDP 0146-9592 Google Scholar

11. 

W. Drexler et al., “In vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett., 24 1221 –1223 (1999). http://dx.doi.org/10.1364/OL.24.001221 OPLEDP 0146-9592 Google Scholar

12. 

K. M. Yoo and R. R. Alfano, “Time-resolved coherent and incoherent components of forward light scattering in random media,” Opt. Lett., 15 320 –322 (1990). http://dx.doi.org/10.1364/OL.15.000320 OPLEDP 0146-9592 Google Scholar

13. 

L. Wang et al., “Ballistic 2D imaging through scattering walls using an ultrafast optical Kerr gate,” Science, 253 769 –771 (1991). http://dx.doi.org/10.1126/science.253.5021.769 SCIEAS 0036-8075 Google Scholar

14. 

B. B Das, F. Liu and R. R. Alfano, “Time-resolved fluorescence and photon migration studies in biomedical and model random media,” Rep. Prog. Phys., 60 227 –292 (1997). http://dx.doi.org/10.1088/0034-4885/60/2/002 RPPHAG 0034-4885 Google Scholar

15. 

E. N. Leith et al., “Realization of time gating by use of spatial filtering,” Appl. Opt., 38 1370 –1376 (1999). http://dx.doi.org/10.1364/AO.38.001370 APOPAI 0003-6935 Google Scholar

16. 

A. Kuditcher et al., “Ultrafast cross-correlated harmonic imaging through scattering media,” Appl. Opt., 40 45 –51 (2001). http://dx.doi.org/10.1364/AO.40.000045 APOPAI 0003-6935 Google Scholar

17. 

J. C. Hebden, “Evaluating the spatial resolution performance of a time-resolved optical imaging system,” Med. Phys., 19 1081 –1087 (1992). http://dx.doi.org/10.1118/1.596826 MPHYA6 0094-2405 Google Scholar

18. 

Q. Z. Wang et al., “Fourier spatial filter acts as a temporal gate for light propagation through a turbid medium,” Opt. Lett., 20 1498 –1500 (1995). http://dx.doi.org/10.1364/OL.20.001498 OPLEDP 0146-9592 Google Scholar

19. 

S. G. Demos and R. R. Alfano, “Temporal gating in highly scattering media by the degree of optical polarization,” Opt. Lett., 21 161 –163 (1996). http://dx.doi.org/10.1364/OL.21.000161 OPLEDP 0146-9592 Google Scholar

20. 

S. G. Demos and R. R. Alfano, “Optical polarization imaging,” Appl. Opt., 36 150 –155 (1997). http://dx.doi.org/10.1364/AO.36.000150 APOPAI 0003-6935 Google Scholar

21. 

B. B. Das, K. M. Yoo and R. R. Alfano, “Ultrafast time-gated imaging in thick tissues: a step toward optical mammography,” Opt. Lett., 18 1092 –1094 (1993). http://dx.doi.org/10.1364/OL.18.001092 OPLEDP 0146-9592 Google Scholar

22. 

K. M. Yoo, B. B. Das and R. R. Alfano, “Imaging of a translucent object hidden in a highly scattering medium from the early portion of the diffuse component of a transmitted ultrafast laser pulse,” Opt. Lett., 17 958 –960 (1992). http://dx.doi.org/10.1364/OL.17.000958 OPLEDP 0146-9592 Google Scholar

23. 

K. M. Yoo, F. Liu and R. R. Alfano, “When does the diffusion approximation fail to describe photon transport in random media?,” Phys. Rev. Lett., 64 2647 (1990). http://dx.doi.org/10.1103/PhysRevLett.64.2647 PRLTAO 0031-9007 Google Scholar

24. 

M. Kempe et al., “Ballistic and diffuse light detection in confocal and heterodyne imaging systems,” J. Opt. Soc. Am. A, 14 216 –223 (1997). http://dx.doi.org/10.1364/JOSAA.14.000216 JOAOD6 0740-3232 Google Scholar

25. 

R. R. Alfano et al., “Light propagation in highly scattering turbid media: concepts, techniques, and biomedical applications,” Photonics, Scientific Foundations, Technology and Applications, IV 367 John Wiley & Sons Inc., New York (2015). Google Scholar

26. 

Z. Glasser et al., “Effect of measurement on the ballistic-diffusive transition in turbid media,” J. Biomed. Opt., 18 106006 (2013). http://dx.doi.org/10.1117/1.JBO.18.10.106006 JBOPFO 1083-3668 Google Scholar

27. 

A. Yaroshevsky et al., “Transition from the ballistic to the diffusive regime in a turbid medium,” Opt. Lett., 36 1395 –1397 (2011). http://dx.doi.org/10.1364/OL.36.001395 OPLEDP 0146-9592 Google Scholar

28. 

J. R. Mourant et al., “Mechanisms of light scattering from biological cells relevant to noninvasive optical-tissue diagnosis,” Appl. Opt., 37 3586 –3593 (1998). http://dx.doi.org/10.1364/AO.37.003586 APOPAI 0003-6935 Google Scholar

29. 

S. Thomsen and D. Tatman, “Physiological and pathological factors of human breast disease that can influence optical diagnosis,” Ann. N.Y. Acad. Sci., 838 171 –193 (1998). http://dx.doi.org/10.1111/j.1749-6632.1998.tb08197.x ANYAA9 0077-8923 Google Scholar

30. 

R. H. J. Kop et al., “Observation of anomalous transport of strongly multiple scattered light in thin disordered slabs,” Phys. Rev. Lett., 79 4369 (1997). http://dx.doi.org/10.1103/PhysRevLett.79.4369 PRLTAO 0031-9007 Google Scholar

31. 

R. Elaloufi, R. Carminati and J.-J. Greffet, “Diffusive-to-ballistic transition in dynamic light transmission through thin scattering slabs: a radiative transfer approach,” J. Opt. Soc. Am. A, 21 1430 (2004). http://dx.doi.org/10.1364/JOSAA.21.001430 JOAOD6 0740-3232 Google Scholar

32. 

Z. Q. Zhang et al., “Wave transport in random media: the ballistic to diffusive transition,” Phys. Rev. E, 60 4843 (1999). http://dx.doi.org/10.1103/PhysRevE.60.4843 Google Scholar

33. 

C. Chen et al., “A primary method for determination of optical parameters of turbid samples and application to intralipid between 550 and 1630 nm,” Opt. Express, 14 7420 –7435 (2006). http://dx.doi.org/10.1364/OE.14.007420 OPEXFF 1094-4087 Google Scholar

34. 

G. Jarry et al., “Imaging mammalian tissues and organs using laser collimated transillumination,” J. Biomed. Eng., 6 70 –74 (1984). http://dx.doi.org/10.1016/0141-5425(84)90013-X JBIEDR 0141-5425 Google Scholar

35. 

A. O. Wist, P. P. Fatouros and S. L. Herr, “Increased spatial resolution in transillumination using collimated light,” IEEE Trans. Med. Imaging, 12 751 –757 (1993). http://dx.doi.org/10.1109/42.251126 ITMID4 0278-0062 Google Scholar

36. 

A. W. Sainter, T. A. King and M. R. Dickinson, “Effect of target biological tissue and choice of light source on penetration depth and resolution in optical coherence tomography,” J. Biomed. Opt., 9 193 –199 (2004). http://dx.doi.org/10.1117/1.1628243 Google Scholar

37. 

U. Sharma, E. W. Chang and S. H. Yun, “Long-wavelength optical coherence tomography at 1.7 μm for enhanced imaging depth,” Opt. Express, 16 (24), 19712 –19723 (2008). http://dx.doi.org/10.1364/OE.16.019712 OPEXFF 1094-4087 Google Scholar

38. 

V. M. Kodach et al., “Quantitative comparison of the OCT imaging depth at 1300 nm and 1600 nm,” Biomed. Opt. Express, 1 176 –185 (2010). http://dx.doi.org/10.1364/BOE.1.000176 Google Scholar

39. 

H. G. van Staveren et al., “Light scattering in Intralipid-10% in the wavelength range of 400-1100 nanometers,” Appl. Opt., 30 4507 –4514 (1991). http://dx.doi.org/10.1364/AO.30.004507 Google Scholar

40. 

T. L. Troy and S. N. Thennadil, “Optical properties of human skin in the near infrared wavelength range of 1000 to 2200 nm,” J. Biomed. Opt., 6 167 –176 (2001). http://dx.doi.org/10.1117/1.1344191 JBOPFO 1083-3668 Google Scholar

Biographies for the authors are not available.

© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2015/$25.00 © 2015 SPIE
Barak Brezner, Sarah Cahen, Ziv Glasser, Shmuel Sternklar, and Er’el Granot "Ballistic imaging of biological media with collimated illumination and focal plane detection," Journal of Biomedical Optics 20(7), 076006 (14 July 2015). https://doi.org/10.1117/1.JBO.20.7.076006
Published: 14 July 2015
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scattering

Collimators

Light scattering

Tissues

Tissue optics

Collimation

Photons

RELATED CONTENT

Quasi-cw tissue transillumination at 1064 nm
Proceedings of SPIE (August 18 1997)
Diffusing temporal light correlation for burn diagnosis
Proceedings of SPIE (August 18 1997)
Analysis Of Light Distribution In Tissue
Proceedings of SPIE (June 03 1988)

Back to Top