Open Access
4 May 2015 Multidistance diffuse correlation spectroscopy for simultaneous estimation of blood flow index and optical properties
Author Affiliations +
Abstract
Traditionally, diffuse correlation spectroscopy (DCS) measures microvascular blood flow by fitting a physical model to the measurement of the intensity autocorrelation function from a single source-detector pair. This analysis relies on the accurate knowledge of the optical properties, absorption, and reduced scattering coefficients of the medium. Therefore, DCS is often deployed together with diffuse optical spectroscopy. We present an algorithm that employs multidistance DCS (MD-DCS) for simultaneous measurement of blood flow index, as well as an estimate of the optical properties of the tissue. The algorithm has been validated through noise-free and noise-added simulated data and phantom measurements. A longitudinal in vivo measurement of a mouse tumor is also shown. MD-DCS is introduced as a stand-alone system for small source-detector separations (<2  cm) for noninvasive measurement of microvascular blood flow.

1.

Introduction

Diffuse correlation spectroscopy (DCS) is a relatively new technique based on temporal speckle fluctuations of the diffuse light for blood flow measurements in the microvasculature.1,2 DCS has been widely used in different medical applications such as neurology and oncology.35 In traditional DCS analysis, the optical properties of the medium, absorption (μa), and reduced scattering (μs) coefficients should be known and the errors in the estimation of these parameters influence the calculated blood flow index (BFI). Most significant errors come from errors in estimating the scattering coefficient, whereas the results are less sensitive to the absorption coefficient.6 DCS is often applied in tandem with a near-infrared diffuse optical spectroscopy (NIRS/DOS) system in order to get more physiological information and also because μa and μs change in time.2

Previously, it has been shown that simultaneously fitting μs and/or μa with BFI was impractical from a single intensity autocorrelation measurement, i.e., from a single source-detector pair,7 due to strong coupling between the parameters. We have hypothesized and observed that the effect of this coupling is dependent on the source-detector separation which has led us to explore if there is a regime where multiple source-detector separations in an analogous fashion to continuous-wave or frequency-domain NIRS/DOS can be used to completely or partially decouple these parameters.

In this work, we describe a multidistance DCS (MD-DCS) algorithm for simultaneous measurement of μa, μs, and BFI in small source-detector separations using noise-added simulated data, phantom measurements, and on a longitudinal in vivo study of a mouse tumor.

2.

Theory, Algorithms, and Experimental Methods

Information about the dynamics of the scatterers, in most cases about the motion of red blood cells, i.e., blood flow, can be derived from the measurement of the intensity autocorrelation of a single speckle generated by diffuse light in tissues.2 For DCS analysis, we converted measured intensity autocorrelation g2 to field autocorrelation function using the Siegert relation g1(τ)={[g2(τ)1]/β}, where β is a constant determined primarily by the collection optics of the experiment. We note that in our experiments, we use nonpolarized light sources and single-mode fibers without polarizers to collect the diffuse light. Therefore, in our experiments, β0.5.

The electric field autocorrelation function (G1), Eq. (1), and the normalized field autocorrelation function (g1) can then be derived and a correlation diffusion model can be used to fit the data and extract the blood flow information.2 The Green’s function solution of the correlation diffusion equation for semi-infinite boundary conditions is given by

Eq. (1)

G1(ρ,τ)=3μs4π{exp[K(τ)r1]r1exp[K(τ)rb]rb}.

Here, τ is the delay time, r1=(1/μs)2+ρ2, rb=(2zb+1/μs)2+(1/μs)2+ρ2, and K(τ)=3μaμs+6μs2κ2ατBFI. α is the fraction of photon scattering events that occur from moving particles in the medium and κ is the wave-number of light in the medium.

Here, we posit that the source-detector separation dependence of the g1 allows us to decouple the contribution of static (μa and μs) and dynamic, BFI, properties of the medium under certain conditions. Therefore, we fit for all desired variables (μa, μs, and BFI) using a single penalty function, χ2, defined as

Eq. (2)

χ2=i=1Nρj=1Nτ[g1theory(ρi,τj,BFI,μs,μa)g1measured(ρi,τj,BFI,μs,μa)]2,
where ρi is the i’th source-detector separation and Nρ is the number of source-detector separations. τj is the j’th delay time and Nτ is the number of delay times. In our implementation, the Nelder–Mead derivative-free simplex method (“fminsearch”) implemented in MATLAB® (Mathworks) was used to minimize this function.

There is more information in the MD-DCS measurements since we are recording the average detected intensity in an analogous fashion to continuous-wave NIRS/DOS. This measured intensity at each detector position from the continuous-wave DCS laser can also be utilized to obtain the effective attenuation coefficient (μeff=3μaμs) to improve the robustness of the algorithm by constraining the relationship between absorption and scattering. We will demonstrate the utility of these in experimental results.

There is one important caveat that we should mention here: DCS uses the intensity autocorrelation functions and, as such, it does not suffer from the so-called source-detector coupling problems. NIRS/DOS, on the other hand, uses the source-detector separation dependence of the intensity alone to measure μeff which requires that these coefficients are calibrated. For the purposes of this work, we did not construct a new probe and our existing probe configuration does not allow the measurement of μeff through the DCS fibers due to this calibration issue.8,9 However, it contains a set of self-calibrating NIRS/DOS fibers that can be used for the measurement of μeff and we utilized them in some of the experimental data. It is beyond the scope of this paper to overcome this limitation.

2.1.

Numerical Data Generation

We have simulated 125 media with five different absorption (μa from 0.04 to 0.16 with steps of 0.03cm1), five different scattering (μs from 4 to 16 with steps of 3cm1), and five different flow (BFI from 4×109 to 16×109 with steps of 3×109cm2/s).

For the main results, the same source-detector separations (2.5 to 5.0 mm) as our existing probe, which is used for phantom and in vivo measurements,8 were simulated. We have also simulated intermediate (10 to 20 mm) and large (20 to 34 mm) optode distances that are shown in Appendices A and B. In a single distance fitting for each separation category (small, intermediate, and large), one separation in the middle is chosen as a representative.

Realistic noise is added to the simulated autocorrelation curves using the previously published DCS noise model.10 In a DCS experiment, the normalized field autocorrelation function decays exponentially [g1(τ)=eΓτ]. The optical/mechanical properties of medium and experimental conditions define the value of Γ. The standard deviation [σ(τ), noise] of the measured intensity autocorrelation, g2(τ), at each delay time (τ) is estimated to be

Eq. (3)

σ(τ)=T/t[β2(1+e2ΓT)(1+e2Γτ)+2m(1e2ΓT)e2Γτ1e2ΓT]12+T/t[2n1β(1+e2Γτ)+n2(1+βeΓτ)]12.

Here, T is the correlator bin time interval, and m is the bin index corresponding to the delay time τ. In our correlator, the bin time interval is T=200ns for the first 16 channels and is doubled every eight channels afterward. t is the total averaging time which is assumed to be 1 min in the following simulations. The average number of photons within bin time T is n, i.e., n=IT, where I is the detected photon count rate. The maximum photon count rate of 1000 kHz is assigned to the smallest source-detector separations. This is the upper limit of the combination of the avalanche photodiodes and digital correlators that are used in the experimental setup. The photon count rate for each optode is calculated from Green’s function solution for photon fluence in semi-infinite boundary conditions.2

Furthermore, we have simulated 16 acquisitions in each optode distance which could be experimentally acquired either by having multiple fibers for each detector11,12 and/or by repeated measurements. We note that these simulations provide a usable signal-to-noise ratio even at the largest source-detector separations which is often not available for real-life DCS experiments. This was chosen to avoid the simulation results being dominated simply by signal-to-noise levels and a realistic phantom and in vivo data were utilized to validate the algorithms.

2.2.

Quantification of the Error of the Fitted Blood Flow Index

To compare the efficacy of MD-DCS fitting with and without providing μeff in different separation types (small, intermediate, and large), we have calculated the error in fitted BFI (εBFI). For each of the simulated BFI, there are five different absorption (μa) and scattering (μs). For each BFI, the estimation error is calculated as the root mean square of the relative difference between fitted and expected values of BFI as

Eq. (4)

εBFI=1Nμsμa(BFIfittedBFIexpectedBFIexpected)2,
where N is the total number of different sets of μa and μs for a given BFI (in this case: 5×5=25).

2.3.

Evaluation of the Penalty Function

For a simulated DCS measurement with μa=0.1cm1, μs=10cm1, and BFI=108cm2/s, we have evaluated the penalty function, χ2, for discrete values of 0.09μa0.11cm1, 9μs11cm1, and 0.9×108BFI1.1×108cm2/s. The χ2 is calculated for two cases: multidistance DCS (MD-DCS) and single-distance DCS (SD-DCS) fitting. MD-DCS fits for eight source-detector separations (2.5 to 5.0 mm) and in SD-DCS, the fourth separation is chosen as a representative (3.6 mm).

2.4.

Tissue Simulating Liquid Phantoms

As mentioned above, the systematic error due to the coupling of BFI with the optical properties is dependent strongly on μs and relatively weakly on μa. To that end, MD-DCS was then validated on liquid phantoms where μs was titrated and μa and BFI were kept constant. In these MD-DCS measurements, a handheld probe with eight source-detector separations from 2.5 to 5.0 mm (2.5, 2.9, 3.2, 3.6, 3.9, 4.3, 4.6, and 5.0 mm) is utilized based on our previous murine tumor measurements.9 Eleven sets of data were acquired with an averaging time of 1 s per acquisition, and the probe was removed then replaced and this procedure was repeated 10 times. In other words, a total of 110 measurements were obtained for each source-detector position. This procedure mimics our in vivo measurement protocols.

Seven liquid phantoms starting with 160 ml of tap water with stepwise 5 ml increments (except 10 ml for the last step) of Lipofundin MCT 20% (BBRAUN, Spain) were measured. The results are shown with box-plots for each phantom representing the variability of each measurement. Prior to this study, μs of 1% Lipofundin was measured by a time-resolved spectroscopy instrument and, assuming the linear relation between μs and the Lipofundin concentration, the expected μs for each phantom was calculated.

2.5.

In Vivo Measurements on a Mouse Tumor

The data were acquired with a hybrid system using a broadband, multidistance NIRS/DOS device to measure the optical properties. The source and detector fibers for both NIRS/DOS and DCS systems were combined on a single handheld probe (six separations from 1.9 to 5.0 mm).9 The fitted values of μa and μs from the MD-DCS method are then compared with these NIRS/DOS results. The fitting of μa and μs from broadband continuous-wave spectroscopy data was previously demonstrated.9,13 This also allows us to compare the MD-DCS fitted BFI values with those obtained by single source-detector pair DCS measurements with NIRS/DOS input—presumably providing the best estimate of the bulk optical properties.

This mouse was part of a study where an antiangiogenic therapy was administered on an implanted renal cell carcinoma model.9 The mouse was measured just before the start of therapy (day 0), received the therapy twice per week, and was measured before each drug injection. In every measurement, in order to have better statistics of the optical properties of tumor, we repositioned the probe 10 times and took 11 data points at each position with an averaging time of 1 s per acquisition. In total, 110 measurements were obtained for each source-detector position. The results are represented as means with error bars showing the standard deviation at each day of the measurement.

To compare whether the measured BFI by MD-DCS and hybrid device are in agreement, we have utilized a two-tailed student test (with 95% confidence interval). We did not repeat this analysis for other mice in the study since the biostatistical analysis of the therapy effects is beyond the scope of this paper.

3.

Results

3.1.

Penalty Function in Single-Distance and Multidistance Diffuse Correlation Spectroscopy Fitting

As described above, a simulated DCS measurement with μa=0.1cm1, μs=10cm1, and BFI=108cm2/s was used to calculate the penalty function χ2 defined in Eq. (2) for a range of values. The χ2 for the expected μa plane, i.e., the plane where μa=0.1cm1, is shown in Fig. 1 for SD-DCS (a) and MD-DCS (b).

Fig. 1

The penalty function of (a) SD-DCS and (b) MD-DCS. χ2 is plotted for the expected μa plane, i.e., the plane where μa=0.1cm1.

JBO_20_5_055001_f001.png

In the single distance fitting, the minimum of the penalty function occurs over a wide range of values of optical parameters and BFI as previously reported.7 Adding more measurements from multiple source-detector separations, i.e., the MD-DCS approach, confines χ2 to a smaller area which leads to better convergence of the algorithm.

3.2.

Multidistance and Single-Distance Diffuse Correlation Spectroscopy Fitting on the Simulated Data

3.2.1.

Noise-free simulations

We have applied SD-DCS and MD-DCS to noise-free data for eight source-detector separations from 2.5 to 5.0 mm (2.5, 2.9, 3.2, 3.6, 3.9, 4.3, 4.6, and 5.0 mm). Figure 2 shows that both SD-DCS and MD-DCS can accurately retrieve all parameters except for one point with lowest μs/BFI and highest μa that SD-DCS fails to fit. We have performed a similar fitting for intermediate (10 to 20 mm) and large (20 to 34 mm) separations, and the results are shown in Appendix A. While MD-DCS fits all the parameters (μa, μs, and BFI) without any error, SD-DCS has considerable error even in noise-free data and the error rises by increasing the source-detector separation (Figs. 7 and 8).

Fig. 2

The values of fitted μs, μa, and BFI versus the expected values in noise-free, small source-detector separations (2.5 to 5.0 mm). The top three plots are the results of SD-DCS and the bottom ones are from MD-DCS fitting. The slope (expected slope is 1) and intercept (expected intercept is 0) of the fitted values versus expected ones for each parameter are displayed in the inset. Each color represents one μs (red: μs=4, green: μs=7, blue: μs=10, cyan: μs=13, and black: μs=16) and each μa is shown by a different marker (star: μa=0.04, circle: μa=0.07, square: μa=0.10, cross: μa=0.13, and triangle: μa=0.16). Dodging is used (moving the points slightly across the x-axis) for better visualization of BFI values where for each μs or μa, BFI increases from left to right. (a) SD-DCS and (b) MD-DCS retrieve all three parameters μs (left), μa (middle), and BFI (right) accurately except for one point with the lowest μs/BFI and highest μa that SD-DCS fails to fit.

JBO_20_5_055001_f002.png

3.2.2.

Noise-added simulations

We have applied SD-DCS and MD-DCS to noise-added simulated data in small optode distances (2.5 to 5 mm), and the results are presented in Fig. 3. It shows that SD-DCS has a considerable error when dealing with realistic experimental noise, whereas MD-DCS fits all parameters with a high accuracy (less than 5% error).

Fig. 3

The values of fitted μs, μa, and BFI by (a) SD-DCS and (b) MD-DCS versus the expected values in small source-detector separations (2.5 to 5.0 mm). The slope (expected slope is 1) and intercept (expected intercept is 0) of the fitted values versus the expected ones for each parameter are displayed in the inset. (a) SD-DCS has a fairly large error in fitting μa, μs, and BFI. (b) In MD-DCS, the fitted line for all parameters has a slope of 1 and 0 intercept, which indicate the complete agreement of the fitted and expected values. Each color represents one μs (red: μs=4, green: μs=7, blue: μs=10, cyan: μs=13, and black: μs=16) and each μa is shown by a different marker (star: μa=0.04, circle: μa=0.07, square: μa=0.10, cross: μa=0.13, and triangle: μa=0.16). Dodging is used (moving the points slightly across the x-axis) for better visualization of BFI values where for each μs or μa, BFI increases from left to right.

JBO_20_5_055001_f003.png

The results of fitting for intermediate and large separations are shown in Appendix B. In these simulations, it was observed that the accuracy of the fitted parameters by MD-DCS decreases when the source-detector separation is larger than 2cm. Therefore, from this point on, we focus on small source-detector separations for both phantom and in vivo validations.

3.3.

Multidistance Diffuse Correlation Spectroscopy Fitting on the Liquid Phantom Data

As explained above, to validate the algorithm in experimental conditions, we have measured homogenous liquid phantoms with both MD-DCS and NIRS. Figure 4(a) shows the measured (fitted by MD-DCS) μs and the agreement with the expected values (slope=0.9, intercept=0.4). The absorption coefficient of all phantoms is expected to be the same. As shown in Fig. 4(b), the fitted absorption coefficient of all phantoms (0.0282±0.0007cm1) is the same and is in the range of the expected μa of water (μa785nm=0.02 to 0.03cm1). Since the changes in viscosity due to Lipofundin titration are negligible, it is expected that BFI also does not change by Lipofundin titration. Figure 4(c) shows that the measured BFI (1.83×108±0.06×108cm2/s) is constant over all phantoms. We note that the relatively large error bars are due to the protocol that was chosen which, as explained above, involves the lifting and replacement of the probe as it is done for the in vivo measurements.

Fig. 4

The measured μa, μs, and BFI by MD-DCS versus expected values in the liquid phantom measurement. On each box, the central mark is the median and the edges of the box are the 25’th and 75’th percentiles over all 10 relocating measurements. (a) The measured (fitted by MD-DCS) μs versus expected values and their increase in each step of titration. The linear fitting equation of fitted μs versus expected is presented (slope=0.9, intercept=0.4). (b) The measured (fitted by MD-DCS) absorption coefficient of all phantoms is the same and in the range of the expected μa for water (μa785nm=0.02 to 0.03cm1). (c) The measured (fitted by MD-DCS fitting) BFI is constant over all phantoms.

JBO_20_5_055001_f004.png

3.4.

Multidistance Diffuse Correlation Spectroscopy Fitting on the Longitudinal In Vivo Data from a Mouse Tumor

To validate the MD-DCS technique in a more realistic scenario where the heterogeneity of the medium optical properties may pose difficulties to a multidistance measurement, we have applied the method on a mouse tumor. The results of fitting values with both methods (MD-DCS and NIRS/DOS) for all three parameters over 24 days are presented in Fig. 5. Although the fitted μa and μs from NIRS/DOS and MD-DCS deviate slightly and MD-DCS has a significantly higher standard deviation, the fitted BFI from both fittings shows a fair agreement.

Fig. 5

The fitted values from both methods (MD-DCS and NIRS/DOS) for μs, μa, and BFI over 24 days on a mouse tumor. The central markers represent the mean values and the error bars show the standard deviation at each day of the measurement. (a) The fitted μs from NIRS/DOS and MD-DCS. (b) The fitted μa from NIRS/DOS and MD-DCS. (c) The fitted BFI from hybrid device and standalone MD-DCS. The blue stars show where the null hypothesis of equal means is rejected.

JBO_20_5_055001_f005.png

We have examined the null hypothesis that the BFI values measured by standalone MD-DCS and hybrid device have equal means. The null hypothesis was rejected for only 3 days which are shown by blue stars in Fig. 5.

4.

Discussion

In this study, we have introduced an algorithm that utilizes MD-DCS to simultaneously estimate the absolute values of static (reduced scattering and absorption coefficients) and dynamic (blood flow) properties of the tissue. The main goal of this algorithm is to be able to obtain a robust estimate of the BFI even without an independent measurement of the optical properties.

As it was previously shown7 in traditional DCS fitting (SD-DCS), the cross-talk between optical properties (μs and μa) and BFI does not allow the simultaneous fitting of all three parameters. Here, we show that the source-detector separation dependence of g1 allows us to decouple completely (in case of noise-free data) or partially (under the influence of noise) the contribution of μa, μs, and BFI by fitting for multiple separations aiming to minimize a single cost function. Figure 1 illustrates this point in an objective manner where it is observed that in the case of the single distance fitting, the minimum of penalty function (χ2) occurs over a wide range of values which can be translated as the high cross-talk between μs and BFI in SD-DCS. By considering multiple separations, MD-DCS, χ2 is confined to a smaller area which leads to a better convergence of the algorithm.

To demonstrate that this can be utilized in a fitting algorithm, we have applied SD-DCS and MD-DCS to noise-free data for small, intermediate, and large source-detector separations (Figs. 2, 7, and 8) showing that MD-DCS can accurately retrieve all parameters while SD-DCS has considerable error even in noise-free data and the error increases by increasing the source-detector separation. Since the data are noise-free and the signal-to-noise ratio is not affected by the optode distance, these results demonstrate that, as expected from the cost function, SD-DCS is not a reliable method to simultaneously fit multiple parameters. These results are in agreement with previous work.7

Furthermore, SD-DCS has a considerable error when dealing with realistic noise in the simulated data as shown in Figs. 3, 9, and 10. On the other hand, in the noise-added simulated data, we have observed that in the small source-detector regime, MD-DCS fits all parameters with high accuracy (less than 5% error), but by increasing the source-detector separations, the accuracy of the fitted parameters decreases.

We attribute this to the interplay of several factors: for the larger separations the signal-to-noise ratio is worse, the autocorrelation curve decays faster, and the bin size for each delay time is finite. As shown in Eq. (1), the decay of the autocorrelation function depends on the optical properties and BFI via K(τ) and the source-detector separation ρ. The couplings of the optical properties, in particular μs, and BFI are evident in this equation but the strength of the coupling depends on both ρ and the relative values of the τ dependent and independent parts of K. As the signal-to-noise ratio gets worse and the curve decays faster, the sensitivity to different parameters decreases, which becomes even worse by the finite bin size. In the future, a more advanced theoretical analysis may allow for better algorithms to decouple the different parameters.

Our next step was to accept these findings, limit ourselves to smaller source-detector separations, and validate MD-DCS on liquid phantoms with μs titration while μa and BFI were kept constant. As was shown in Fig. 4, the measured (fitted by MD-DCS) μa, μs, and BFI agreed with the expected values. We once more stress that this measurement has validated the algorithm in challenging conditions which resemble the in vivo measurements. The liquid phantoms were measured by a handheld probe and for each measurement the probe was removed and then replaced 10 times.

We have then tested the algorithm on our previous measurements9 from a mouse tumor undergone a longitudinal study after receiving a cancer therapy. In Fig. 5, we have compared the fitted μa and μs by NIRS/DOS and MD-DCS, which has demonstrated that the results of MD-DCS alone are noisier than a dedicated NIRS/DOS measurement. This was expected since the NIRS/DOS system used a multidistance, broadband dataset with optimized detector sizes for an improved signal-to-noise ratio and a great deal of redundancy. MD-DCS, on the other hand, used small (6μm) single-mode fibers which are not optimized to collect the maximum intensity. However, the BFI results that were shown in Fig. 5 were in good agreement over the majority of the days with a comparable variance. We argue that these results are satisfactory since the goal of this algorithm is to obtain reliable BFI measurements rather than to replace NIRS/DOS.

Finally, we note that for the in vivo results we have used estimates of μeff to constrain the optical properties. We have argued that the continuous-wave DCS laser and the mean intensity could also be used if the source-detector coupling coefficients are calibrated. This illustrates both a possible shortcoming, i.e., the need to constrain the values, and also a potential solution, i.e., the use of the MD-DCS data to estimate that value. To explore this further, in noise-added simulated data, we have tested all the source-detector separation ranges (small, intermediate, and large) and have evaluated the error in fitted BFI (εBFI) deploying Eq. (4) with and without the knowledge of μeff as a prior. The estimated error for noise-added data is presented in Fig. 6. It shows that in simulated homogenous media, providing μeff does not affect the accuracy of MD-DCS considerably when the conditions are near-ideal. However, MD-DCS is highly sensitive to the source-detector separation regime (small, intermediate, and large) and its accuracy dramatically decreases in large optode distances. This means there is more room for further improvement in the experimental implementation of this algorithm.

Fig. 6

The error in fitted BFI by MD-DCS on noise-added simulated data for different source-detector separation regimes (small, intermediate, and large). For each source-detector separation range, the estimated error with and without providing μeff is presented.

JBO_20_5_055001_f006.png

5.

Conclusion

We have investigated the possibility of simultaneous extraction of absorption coefficient (μa), reduced scattering coefficient (μs), and BFI of the tissue using MD-DCS. Numerical simulations demonstrate that the fitting is more robust in relatively smaller source-detector separations (from 2 to 20 mm). MD-DCS is then validated on tissue simulating phantoms. Finally, it is applied for in vivo measurements on a mouse tumor. The results suggest that MD-DCS can be used to decouple the absorption and scattering coefficients from BFI in small source-detector separations to allow simpler, DCS-only approaches in cases where BFI is the primary parameter of interest.

Appendices

Appendix A:

Noise-Free Simulated Data of Intermediate and Large Optode Distances

In this section, the results of SD-DCS and MD-DCS fitting in noise-free data are presented. Figure 7 shows the values of fitted μs, μa, and BFI fitted by (a) SD-DCS and (b) MD-DCS versus the expected values for intermediate source-detector separations (10 to 20 mm).

Fig. 7

The values of fitted μs, μa, and BFI by (a) SD-DCS and (b) MD-DCS versus the expected values in noise-free data of intermediate source-detector separations (10 to 20 mm). The slope (expected slope is 1) and intercept (expected intercept is 0) of the fitted values versus expected ones for each parameter are displayed in the inset. (a) SD-DCS has error in estimation of μa, μs, and BFI. (b) In MD-DCS, the fitted line for all parameters has slope of 1 and 0 intercept, which indicate the complete agreement of fitted and expected values. Each color represents one μs (red: μs=4, green: μs=7, blue: μs=10, cyan: μs=13, and black: μs=16) and each μa is shown by a different marker (star: μa=0.04, circle: μa=0.07, square: μa=0.10, cross: μa=0.13, and triangle: μa=0.16). Dodging is used (moving the points slightly across the x-axis) for a better visualization of BFI values where for each μs or μa, BFI increases from left to right.

JBO_20_5_055001_f007.png

In Fig. 8, the results of SD-DCS and MD-DCS on noise-free data in large source-detector separations (20 to 34 mm) are presented. It can be observed that while MD-DCS fits all parameters perfectly without any error, SD-DCS has a considerable error in fitting μa, μs, and BFI and the error in fitted values rises by increasing the source-detector separations.

Fig. 8

The values of fitted μs, μa, and BFI by (a) SD-DCS and (b) MD-DCS versus the expected values in noise-free data in large source-detector separations (20 to 34 mm). The slope (expected slope is 1) and intercept (expected intercept is 0) of the fitted values versus expected ones for each parameter are displayed in the inset. (a) SD-DCS has a considerable error in estimation of μa, μs, and BFI. (b) In MD-DCS, the fitted line for all parameters has a slope of 1 and 0 intercept, which indicate the complete agreement of fitted and expected values. Each color represents one μs (red: μs=4, green: μs=7, blue: μs=10, cyan: μs=13, and black: μs=16) and each μa is shown by a different marker (star: μa=0.04, circle: μa=0.07, square: μa=0.10, cross: μa=0.13, and triangle: μa=0.16). Dodging is used (moving the points slightly across the x-axis) for better visualization of BFI values where for each μs or μa, BFI increases from left to right.

JBO_20_5_055001_f008.png

Appendix B:

Noise-Added Simulated Data of Intermediate and Large Optode Distances

In this section, the results of SD-DCS and MD-DCS fitting in noise-added simulated data are presented. Figure 9 shows the values of fitted μs, μa, and BFI fitted by (a) SD-DCS and (b) MD-DCS versus the expected values intermediate source-detector separations (10 to 20 mm). It can be observed that while MD-DCS fits all parameters with high accuracy, SD-DCS fails to fit in many cases.

Fig. 9

The values of fitted μs, μa, and BFI by (a) SD-DCS and (b) MD-DCS versus the expected values in noise-added data of intermediate source-detector separations (10 to 20 mm). The slope (expected slope is 1) and intercept (expected intercept is 0) of the fitted values versus expected ones for each parameter are displayed in the inset. (a) SD-DCS fails to fit μa, μs, and BFI. Some of the points are not presented since they are out of the y-axis range of these plots. (b) In MD-DCS, the fitted values have a good agreement with the expected values. Each color represents one μs (red: μs=4, green: μs=7, blue: μs=10, cyan: μs=13, and black: μs=16) and each μa is shown by a different marker (star: μa=0.04, circle: μa=0.07, square: μa=0.10, cross: μa=0.13, and triangle: μa=0.16). Dodging is used (moving the points slightly across the x-axis) for better visualization of BFI values where for each μs or μa, BFI increases from left to right.

JBO_20_5_055001_f009.png

In large source-detector separations (20 to 34 mm) (Fig. 10), the results of SD-DCS and MD-DCS on noise-added data are presented. Similar to intermediate optode distances, SD-DCS fails to fit μs, μa, and BFI in most points. In MD-DCS, the fitted values have a fair agreement with the expected values in the points where μs and μa are small, but it fails to fit in high μs and μa. It can be observed that the accuracy of the fitted parameters by MD-DCS decreases when the source-detector separation is larger than 2cm.

Fig. 10

The values of fitted μs, μa, and BFI by (a) SD-DCS and (b) MD-DCS versus the expected values in noise-added data in large source-detector separations (20 to 34 mm). The slope (expected slope is 1) and intercept (expected intercept is 0) of the fitted values versus expected ones for each parameter are displayed in the inset. (a) SD-DCS fails to fit μs, μa, and BFI. Some of the points are not presented since they are out of y-axis range of this plots. (b) In MD-DCS, the fitted values have a fair agreement with the expected values in the points where μs and μa are small but it fails to fit in high μs and μa values. Each color represents one μs (red: μs=4, green: μs=7, blue: μs=10, cyan: μs=13, and black: μs=16) and each μa is shown by a different marker (star: μa=0.04, circle: μa=0.07, square: μa=0.10, cross: μa=0.13, and triangle: μa=0.16). Dodging is used (moving the points slightly across the x-axis) for better visualization of BFI values where for each μs or μa, BFI increases from left to right.

JBO_20_5_055001_f010.png

Acknowledgments

The project was funded by Fundació Cellex Barcelona, Ministerio de Economía y Competitividad (PHOTOSTROKE), l’Obra Social La Caixa (MedLlumBCN), and LASERLAB Europe III (Bioptichal). We acknowledge useful discussions with Johannes Johansson, Miguel Mireles, and Behrooz Hashemian.

References

1. 

D. Boas, L. Campbell and A. Yodh, “Scattering and imaging with diffusing temporal field correlations,” Phys. Rev. Lett., 75 1855 –1858 (1995). http://dx.doi.org/10.1103/PhysRevLett.75.1855 PRLTAO 0031-9007 Google Scholar

2. 

T. Durduran et al., “Diffuse optics for tissue monitoring and tomography,” Rep. Prog. Phys., 73 076701 (2010). http://dx.doi.org/10.1088/0034-4885/73/7/076701 RPPHAG 0034-4885 Google Scholar

3. 

T. Durduran and A. G. Yodh, “Diffuse correlation spectroscopy for non-invasive, micro-vascular cerebral blood flow measurement,” NeuroImage, 85 (Pt 1), 51 –63 (2014). http://dx.doi.org/10.1016/j.neuroimage.2013.06.017 NEIMEF 1053-8119 Google Scholar

4. 

E. M. Buckley et al., “Diffuse correlation spectroscopy for measurement of cerebral blood flow: future prospects,” Neurophotonics, 1 011009 (2014). http://dx.doi.org/10.1117/1.NPh.1.1.011009 NEUROW 2329-423X Google Scholar

5. 

G. Yu, “Near-infrared diffuse correlation spectroscopy in cancer diagnosis and therapy monitoring,” J. Biomed. Opt., 17 010901 (2012). http://dx.doi.org/10.1117/1.JBO.17.1.010901 JBOPFO 1083-3668 Google Scholar

6. 

D. Irwin et al., “Influences of tissue absorption and scattering on diffuse correlation spectroscopy blood flow measurements,” Biomed. Opt. Express, 2 1969 –1985 (2011). http://dx.doi.org/10.1364/BOE.2.001969 BOEICL 2156-7085 Google Scholar

7. 

L. Dong et al., “Simultaneously extracting multiple parameters via fitting one single autocorrelation function curve in diffuse correlation spectroscopy,” IEEE Trans. Biomed. Eng., 60 361 –368 (2013). http://dx.doi.org/10.1109/TBME.2012.2226885 IEBEAX 0018-9294 Google Scholar

8. 

P. Farzam and T. Durduran, “Design of a broadband near infrared spectroscopy (NIRS) and diffuse correlation spectroscopy (DCS) device with a self-calibrated probe for experimental oncology,” in European Conf. Biomedical Optics, (2011). Google Scholar

9. 

P. Farzam, “Hybrid diffuse optics for monitoring of tissue hemodynamics with applications in oncology,” ICFO-The Institute of Photonic Sciences, (2014). Google Scholar

10. 

C. Zhou et al., “Diffuse optical correlation tomography of cerebral blood flow during cortical spreading depression in rat brain,” Opt. Express, 14 1125 –1144 (2006). http://dx.doi.org/10.1364/OE.14.001125 OPEXFF 1094-4087 Google Scholar

11. 

L. Koban et al., “Processing of emotional words measured simultaneously with steady-state visually evoked potentials and near-infrared diffusing-wave spectroscopy,” BMC Neurosci., 11 85 (2010). http://dx.doi.org/10.1186/1471-2202-11-85 BNMEA6 1471-2202 Google Scholar

12. 

P. Farzam et al., “Noninvasive characterization of the healthy human manubrium using diffuse optical spectroscopies,” Physiol. Meas., 35 1469 –1491 (2014). http://dx.doi.org/10.1088/0967-3334/35/7/1469 PMEAE3 0967-3334 Google Scholar

13. 

H.-W. Wang et al., “Broadband reflectance measurements of light penetration, blood oxygenation, hemoglobin concentration, and drug concentration in human intraperitoneal tissues before and after photodynamic therapy,” J. Biomed. Opt., 10 (1), 014004 (2005). http://dx.doi.org/10.1117/1.1854679 JBOPFO 1083-3668 Google Scholar

Biography

Parisa Farzam received her PhD in biomedical photonics from ICFO-The Institute of Photonics Sciences, Spain. Her PhD research was focused on developing hybrid diffuse optics for monitoring of tissue hemodynamics with applications in oncology. Currently, she is a postdoctoral researcher at the Institute of Applied Physics at the University of Bern, Switzerland, working on quantitative measurement of preterm-neonate cerebral hemodynamics.

Turgut Durduran was trained at the University of Pennsylvania. In 2009, he moved to ICFO-The Institute of Photonic Sciences, Spain, where he leads the medical optics group. His research interests revolve around the use of diffuse light to noninvasively probe tissue function. The group develops new technologies and algorithms and routinely translates them to preclinical, clinical, and industrial applications.

© 2015 Society of Photo-Optical Instrumentation Engineers (SPIE) 1083-3668/2015/$25.00 © 2015 SPIE
Parisa Farzam and Turgut Durduran "Multidistance diffuse correlation spectroscopy for simultaneous estimation of blood flow index and optical properties," Journal of Biomedical Optics 20(5), 055001 (4 May 2015). https://doi.org/10.1117/1.JBO.20.5.055001
Published: 4 May 2015
Lens.org Logo
CITATIONS
Cited by 23 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical properties

Spectroscopy

Blood circulation

Absorption

Biological research

In vivo imaging

Scattering

Back to Top